Comparison of Two OCT Device RNFL and GCIPL Measurements in Mild, Moderate, and Severe Glaucoma

An observational study in Glaucoma, sponsored by Twin Cities Eye Consultants. Completed at 1 site in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2026-01-20.

Sponsored by Twin Cities Eye Consultants · Observational

How we know

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Study type
Observational
Model
Case-only
Time perspective
Prospective
Enrollment
141
Ages
18 Years and older
Sex
All
01

Where this study stands

Counted across the records on this site · 2026-11-22

Among 490 studies in Glaucoma with a site in United States — 75 open to participants now, 94 under way, 5 of them phase 3. Glaucoma · United States.

Enrollment
40th of 110
141 · median 100
Started
2024
with 20 others
Observational
110 of 490
18 open now
Registered
467th of 490
by date

Among the 490 studies in Glaucoma with a site in United States, this one is the 40th largest by enrollment of the 110 observational ones that give a count, is one of 21 that started in 2024, is one of 110 observational ones, and was the 467th registered.

Glaucoma

1,421 studies on the registry are indexed under Glaucoma; 221 are open to participants now.

Browse Glaucoma studies →

Open to participants now

75 of the 490 studies in Glaucoma with a site in United States are open now. Recruiting first, the newest first:

70 more — sign in →

All 75 open studies →

A study can be open while its sites in United States have not opened yet or have stopped. Only those recruiting at a site in United States →

Who these studies take

Of the 154 Glaucoma studies recruiting now, 112 (73%) set no upper age limit and 9 (6%) stop at 75 or younger; 13 take children under 18; 44 take healthy volunteers.

This study: 18 Years and older · all sexes · no healthy volunteers.

The full table on the hub →

This tab as a report to print or save as PDF → With an account.

Was this useful?
Get alerts for this landscapeDownload CSVNarrow it in search →With an account: a weekly mail of what changed among 490 studies in Glaucoma with a site in United States, or all of them as a file.
02

Study summary

This research study was designed to evaluate how consistently two different imaging devices measure the thickness of retinal layers in patients with glaucoma. The two machines, the Topcon Maestro2 and the Zeiss Cirrus 5000, are commonly used in eye clinics to take high-resolution scans of the retina and optic nerve. These scans help doctors monitor glaucoma and detect disease progression.

The goal of this study was to compare the results from both machines when scanning the same patients, and to determine how closely their measurements align. Researchers also developed mathematical models that allow doctors to convert results from one machine to the other, which could help ensure consistency when a patient receives scans on different devices over time or at different clinics.

A total of 141 eyes from patients with varying levels of glaucoma severity were included in the study. Scans were performed using standard imaging protocols on both machines. Researchers analyzed the average thickness of two important retinal layers: the retinal nerve fiber layer (RNFL) and the ganglion cell-inner plexiform layer (GCIPL).

This study helps improve our understanding of how different imaging devices compare and provides tools for standardizing results across platforms. The findings may help eye care providers make more confident decisions when reviewing images from different machines, improving the care and monitoring of patients with glaucoma.

This was an investigator-initiated study sponsored by Twin Cities Eye Consultants with funding support from Topcon Healthcare. No new medications, treatments, or procedures were tested as part of this research. The study only involved non-invasive eye imaging scans that are already widely used in clinical practice.

Read the detailed description

This investigator-initiated, cross-sectional study was designed to assess the level of agreement between optical coherence tomography (OCT) measurements captured by two commercially available imaging platforms: the Topcon Maestro2 and the Zeiss Cirrus 5000. These devices are widely used in the clinical management of glaucoma to quantify structural biomarkers such as retinal nerve fiber layer (RNFL) thickness and ganglion cell-inner plexiform layer (GCIPL) thickness.

The motivation behind this study stems from real-world variability in imaging hardware across clinical settings. Patients may undergo imaging on different machines due to equipment availability, clinic referral patterns, or evolving device infrastructure. As each device uses proprietary software algorithms and scan acquisition protocols, there can be small but clinically significant differences in reported measurements-even when scanning the same eye under similar conditions. Understanding this variability is critical for ensuring continuity of care and accuracy in longitudinal monitoring of glaucoma.

The study enrolled 141 eyes from patients with established glaucoma, screened from a single clinical practice. The first 72 eyes enrolled were designated as the derivation cohort, while the subsequent 69 eyes comprised the validation cohort. Each subject underwent back-to-back imaging on both devices during the same session using manufacturer-recommended scan protocols: the Topcon Wide Scan, RNFL Cube, and Macular Cube protocols; and the Zeiss Optic Disc Cube 200×200 and Macular Cube 512×128 protocols. All imaging was performed by a single, trained technician using factory-default acquisition settings and standard dilation (phenylephrine 2.5% and tropicamide 1%), with eye-tracking (FastTrac) enabled where available.

Average RNFL thickness and GCIPL thickness were extracted from each scan and compared between devices. Measurements were further analyzed by quadrant (RNFL) and sector (GCIPL) to evaluate regional agreement. Generalized linear models and Deming regression were used to evaluate the level of agreement and construct conversion equations between devices. Results were evaluated based on clinical thresholds of ±3 microns-a commonly accepted tolerance for scan-to-scan variability on a single device.

Repeatability of Topcon scans was also assessed using three consecutive scans on the same visit. Intra-device variability was analyzed using mean differences and 95% limits of agreement to determine reliability across protocols. Exploratory subgroup analyses were performed by glaucoma severity, gender, race/ethnicity, and visual acuity to explore potential drivers of scan variability or poor agreement.

The study found that while mean RNFL and GCIPL measurements from the two devices were highly correlated, statistically significant differences in average values were observed. This reinforces the importance of accounting for device-specific variability in both research and clinical practice. Derived conversion equations demonstrated strong performance in the validation cohort, particularly for global RNFL and GCIPL metrics. Sector-level agreement was more variable, particularly in the temporal RNFL and superotemporal GCIPL sectors, indicating that clinicians should use caution when interpreting regional values across platforms.

Importantly, scan repeatability on the Maestro2 was excellent, particularly for macular GCIPL metrics, highlighting its reliability for longitudinal monitoring. This supports the growing role of widefield and integrated OCT imaging in clinical glaucoma care.

The study was approved by Sterling IRB and conducted at Twin Cities Eye Consultants (Minnesota). It involved no investigational treatments or interventions beyond standard OCT imaging. Funding support was provided by Topcon Healthcare, which had no role in study design, data analysis, or manuscript preparation. Results from this study are intended to inform clinicians, researchers, and device manufacturers regarding the interoperability and clinical translation of OCT imaging biomarkers in glaucoma.

03

Conditions studied

  • Glaucoma

Browse trials for

Keywords

  • Glaucoma
  • OCT
  • Optical Coherence Tomography
04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
No
Sampling method
Non-probability sample

Study population

The study population consisted of adult patients aged 18 years or older with established or suspected glaucoma. Subjects were consecutively enrolled from a single tertiary ophthalmology practice in Minnesota. The final cohort included a range of glaucoma severity (mild, moderate, and severe) and was predominantly composed of White (59.6%) and female (59.6%) individuals. All participants underwent same-day, back-to-back OCT imaging on both the Topcon Maestro2 and Zeiss Cirrus 5000 systems.

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Male or female patients at least 18 years of age
  • A diagnosis of mild, moderate or severe glaucoma.
  • Optic nerve appearance characteristic of glaucoma.
  • \< +/- 6D Spherical equivalent Refractive Error
  • BCVA ≥ 20/40
  • Previous Zeiss OCT RNFL with signal strength of ≥ 7/10
  • The subject is able to understand the requirements of the study and provide written .
  • Subject is willing to follow study instructions, agrees to comply with all study procedures and testing.

Exclusion criteria

Exclusion Criteria:

  • If pseudophakic, PCO must be ≤ trace
  • If phakic, the cataract must be ≤ 1+
  • No diagnosis of optic nerve disease other than glaucoma, and no macular disease.
  • Current participation or participation in any drug or device clinical trial (which includes the fellow eye) within the past 30 calendar days.
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

05

Study design

Observational model
Case-only
Time perspective
Prospective
Enrollment
141 participants (actual)
Target follow-up
1 Day
Patient registry
Yes

Groups and cohorts

  • Derivation

    The derivation cohort consisted of the first 72 eyes enrolled in the study. These participants had established glaucoma and were selected consecutively from a single clinical practice. Each eye underwent back-to-back OCT imaging using both the Topcon Maestro2 and Zeiss Cirrus 5000 devices during the same visit. Data from this cohort were used to develop statistical conversion models assessing agreement and potential equivalence in RNFL and GCIPL thickness measurements across platforms. Non-interventional imaging procedures were performed using standard manufacturer-recommended scan protocols. These included the Topcon Wide Scan, RNFL Cube, and Macular Cube protocols, as well as the Zeiss Optic Disc Cube 200×200 and Macular Cube 512×128 protocols. All scans were conducted by a single trained technician under controlled clinical conditions.

    Diagnostic Test: Optical Coherence Tomography

  • Validation

    The validation cohort included the subsequent 69 eyes enrolled in the study, following the derivation cohort. Imaging procedures and clinical characteristics mirrored those of the derivation group. This cohort was used to test and validate the accuracy and reliability of the conversion models derived from the initial group. Subjects underwent the same non-invasive imaging protocols on both the Topcon Maestro2 and Zeiss Cirrus 5000 devices during a single clinic session. Scan protocols, technician, and environmental conditions were consistent with those of the derivation cohort to preserve standardization.

    Diagnostic Test: Optical Coherence Tomography

Interventions

  • Diagnostic testOptical Coherence Tomography

    Subjects underwent bilateral Optical Coherence Tomography (OCT) imaging using both the Topcon Maestro2 and Zeiss Cirrus 5000 devices during the same clinical session. Manufacturer-recommended protocols were used: Wide Scan, RNFL Cube, and Macular Cube (Topcon), and Optic Disc Cube 200×200 and Macular Cube 512×128 (Zeiss). Imaging was performed back-to-back using factory-default settings and standardized pupil dilation to assess cross-platform agreement and repeatability of RNFL and GCIPL thickness measurements in glaucomatous eyes.

    Also known as: OCT, Maestro2, Cirrus 5000

06

What researchers measure

Primary outcomes

  1. Agreement Between Topcon Maestro2 and Zeiss Cirrus 5000 OCT Devices

    This primary outcome evaluates the cross-platform agreement of retinal nerve fiber layer (RNFL) and ganglion cell-inner plexiform layer (GCIPL) thickness measurements between the Topcon Maestro2 and Zeiss Cirrus 5000 devices in glaucoma and glaucoma-suspect eyes. The analysis will assess absolute differences, standard deviation, 95% limits of agreement, and proportion of eyes within a ±3 micron threshold for global and sectoral measurements.

    Time frame: Day of single study visit

Secondary outcomes

  1. Repeatability of Topcon Maestro2 OCT Measurements

    Evaluate the test-retest repeatability of RNFL and GCIPL thickness measurements acquired with the Topcon Maestro2 Wide, RNFL Cube, and Macular Cube scan protocols. Repeatability will be assessed using standard deviation of repeated measures, coefficient of variation (CV), and intraclass correlation coefficients (ICC) when applicable.

    Time frame: Day of single study visit

  2. Impact of Glaucoma Severity on Device Agreement

    Assess whether glaucoma severity (mild, moderate, or severe) influences agreement between RNFL and GCIPL thickness values measured by the Topcon Maestro2 and Zeiss Cirrus 5000. Stratified analyses will be conducted for both derivation and validation cohorts.

    Time frame: Day of single study visit

  3. Cross-Device Agreement by Quadrant and Sectoral Analysis

    Evaluate device agreement on a quadrant-by-quadrant and sectoral basis for both RNFL and GCIPL layers. The analysis will include superior, inferior, nasal, and temporal quadrants for RNFL, and superotemporal, superonasal, inferotemporal, and inferonasal sectors for GCIPL.

    Time frame: Day of single study visit

  4. Predictive Modeling to Convert Measurements Between Devices

    Develop linear regression models using the derivation cohort to predict Zeiss Cirrus RNFL and GCIPL values based on Topcon Maestro2 outputs. Validate model performance in a separate validation cohort within ±3 microns.

    Time frame: Day of single study visit

07

Study locations

1 site
  • Twin Cities Eye Consultants
    Coon Rapids, Minnesota 55433, United States
08

References and documents

Publications

  • Tan O, Chopra V, Lu AT, Schuman JS, Ishikawa H, Wollstein G, Varma R, Huang D. Detection of macular ganglion cell loss in glaucoma by Fourier-domain optical coherence tomography. Ophthalmology. 2009 Dec;116(12):2305-14.e1-2. doi: 10.1016/j.ophtha.2009.05.025. Epub 2009 Sep 10. PubMed 19744726 ↗
  • Wong E, Yoshioka N, Kalloniatis M, Zangerl B. Cirrus HD-OCT short-term repeatability of clinical retinal nerve fiber layer measurements. Optom Vis Sci. 2015 Jan;92(1):83-8. doi: 10.1097/OPX.0000000000000452. PubMed 25479451 ↗
  • Kotowski J, Wollstein G, Folio LS, Ishikawa H, Schuman JS. Clinical use of OCT in assessing glaucoma progression. Ophthalmic Surg Lasers Imaging. 2011 Jul;42 Suppl(0):S6-S14. doi: 10.3928/15428877-20110627-01. PubMed 21790113 ↗
  • Vazirani J, Kaushik S, Pandav SS, Gupta P. Reproducibility of retinal nerve fiber layer measurements across the glaucoma spectrum using optical coherence tomography. Indian J Ophthalmol. 2015 Apr;63(4):300-5. doi: 10.4103/0301-4738.158064. PubMed 26044467 ↗
  • Mwanza JC, Oakley JD, Budenz DL, Anderson DR; Cirrus Optical Coherence Tomography Normative Database Study Group. Ability of cirrus HD-OCT optic nerve head parameters to discriminate normal from glaucomatous eyes. Ophthalmology. 2011 Feb;118(2):241-8.e1. doi: 10.1016/j.ophtha.2010.06.036. Epub 2010 Oct 28. PubMed 20920824 ↗

Individual participant data

Plan to share: No

09

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Ran from Oct 2024 to primary completion in Feb 2025, 3 months.

Of 111 non-industry observational studies under Glaucoma, half reached primary completion within 47 months of starting (the median), counting those still running; a quarter within 12 and three quarters within 208. This study's 3 months is below that.

Counting every sponsor, half of 135 studies reached it within 31 months; a quarter within 11 and three quarters within 208.

How these figures are worked out, and which studies are left out of them.

Registered
Dec 29, 2025
First posted
Jan 20, 2026
StartActual
Oct 28, 2024
Primary completionActual
Feb 14, 2025
CompletionActual
Feb 14, 2025
Last update
Jan 20, 2026
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Jan 20, 2026, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT07350304
Lead sponsor
Twin Cities Eye Consultants
Responsible party
Sponsor
First posted
Jan 20, 2026
Start date
Oct 28, 2024
Primary completion
Feb 14, 2025
Completion
Feb 14, 2025
Last update
Jan 20, 2026

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Dec 2025. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow