ClinicalTrials.gg
CompletedNCT06190808Updated Sep 29, 2026Results posted

Continuous Glucose Monitoring in the Intensive Care Unit

An interventional study of Dexcom G7 Continuous Glucose Monitor in Diabetes Mellitus, sponsored by University of Massachusetts, Worcester. Completed at 1 site in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2026-09-29.

Sponsored by University of Massachusetts, Worcester · Not applicable, Interventional, and Other

How we know

Updated Sep 29, 2026Results postedStudy documents posted+1 moreGo to Updates ↓

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Phase
Not applicable
Study type
Interventional
Enrollment
16
Allocation
Not applicable
Ages
18 Years and older
Sex
All
01

Where this study stands

Counted across the records on this site · 2026-11-23

Among 3,023 studies in Diabetes Mellitus with a site in United States — 289 open to participants now, 388 under way, 12 of them phase 3. Diabetes Mellitus · United States.

Enrollment
2,126th of 2,502
16 · median 82
Started
2024
with 87 others
No phase
1,418 of 3,023
159 open now
Registered
2,767th of 3,023
by date

Among the 3,023 studies in Diabetes Mellitus with a site in United States, this one is the 2,126th largest by enrollment of the 2,502 interventional ones that give a count, is one of 88 that started in 2024, is one of 1,418 with no phase, and was the 2,767th registered.

Diabetes Mellitus

8,295 studies on the registry are indexed under Diabetes Mellitus; 1,071 are open to participants now.

Browse Diabetes Mellitus studies →

Open to participants now

289 of the 3,023 studies in Diabetes Mellitus with a site in United States are open now. Recruiting first, the newest first:

284 more — sign in →

All 289 open studies →

A study can be open while its sites in United States have not opened yet or have stopped. Only those recruiting at a site in United States →

Who these studies take

Of the 649 Diabetes Mellitus studies recruiting now, 358 (55%) set no upper age limit and 194 (30%) stop at 75 or younger; 89 take children under 18; 181 take healthy volunteers.

This study: 18 Years and older · all sexes · no healthy volunteers.

The full table on the hub →

This tab as a report to print or save as PDF → With an account.

Was this useful?
Get alerts for this landscapeDownload CSVNarrow it in search →With an account: a weekly mail of what changed among 3,023 studies in Diabetes Mellitus with a site in United States, or all of them as a file.
02

Study summary

The aim of the proposed clinical trial is to study the use of the Dexcom G7 continuous glucose monitor (CGM) within the intensive care unit (ICU) environment. The overall goal is to validate the accuracy and practicality of the Dexcom G7 when used to assess glycemia in critically ill patients.

Read the detailed description

The specific goals are 1) to work with intensive care unit (ICU) staff, in a respiratory ICU setting, to determine how best to place and maintain Dexcom G7 continuous glucose monitors (CGMs) on patients and how to place recording devices and chargers in that environment, 2) download CGM readings to the Dexcom Dashboard in an analyzable format, and 3) to compare simultaneous CGM readings with glucose measurements obtained in the course of routine care as recorded in the electronic medical record. The primary analytical goal is the correlation of CGM glucose measurements with those obtained in the course of routine care. A secondary goal is to determine the frequency of device malfunctions that occur either as a result of device failure (e.g. it falls off) or adverse events (e.g. removal due to skin irritation).

Study subjects will be patients admitted to the 6-ICU UMass Memorial Medical Center (UMMMC) University Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized. All enrolled volunteers will have a Dexcom G7 CGM place on the skin according to manufacturer's specifications. The devices will remain in place for 10 days or until discharge from the ICU

03

Conditions studied

  • Diabetes Mellitus

Keywords

  • ICU
  • CGM
  • Accuracy
  • Diabetes
  • Hypoglycemia
  • Diabetes mellitus
  • Continuous glucose monitor
  • Intensive care
04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

Campus medical intensive care unit for any reason who have known diabetes or who are hyperglycemic (blood glucose concentration ≥200 mg/dL) with diabetes that was previously undiagnosed or unrecognized.

  • Persons with either type 1 or type 2 diabetes.
  • Persons of either sex.
  • Persons of any age over 18.
  • Persons able to give or cognitively impaired adults whose legally authorized representative (LAR) can give informed consent.
  • English or Spanish speaking adults.

Exclusion criteria

Exclusion Criteria:

  • Persons unable to provide .
  • Cognitively impaired persons whose LAR declines to provide informed consent.
  • Patients receiving hydroxyurea, which interferes with sensor function.
  • Patients receiving high dose acetaminophen (> 4 gm/day), which interferes with sensor function.
  • Anasarca affecting preferred areas for device placement (arm, upper buttocks, or thigh), which may interfere with sensor accuracy.
  • Severe cachexia with absence of subcutaneous fat at preferred sensor placement sites, which might interfere with sensor accuracy.
  • Persons who are pregnant.
  • Prisoners.
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

05

Study design

Phase
Not applicable
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
16 participants (actual)

Study arms

  • Experimental
    Single Arm

    All members of the arm will have a CGM place and glucose concentration downloaded.

    Diagnostic Test: Dexcom G7 Continuous Glucose Monitor

Interventions

  • Diagnostic testDexcom G7 Continuous Glucose Monitor

    Continuous glucose monitor

06

What researchers measure

Primary outcomes

  1. Mean Absolute Relative Difference (MARD).

    Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Mean Absolute Relative Difference (MARD). The MARD is the average percentage difference between the continuous glucose monitor readings measured over 6 months and glucose concentration results obtained by near simultaneous (+/- 5 minutes) standard test methods was used to gauge the accuracy of the CGM in the ICU setting. Typical MARD values reported for CGMs in the outpatient setting average approximately 8%. Typical MARD values reported for CGMs other than the one used in this study n the ICU setting average approximately 12%.

    Time frame: 6 months

  2. Clarke Error Grid

    Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Clarke Error Grid. The Clarke Error Grid plots of reference glucose values obtained by point-of-care fingerstick or laboratory determination (x-axis) against CGM values (y-axis). The grid is divided into 5 zones (A-E) that indicate clinical implications of CGM glucose measurements when compared to a "gold standard" Region A: Clinically accurate. Region B: Benign measurement error. Region C: Overcorrection error region CGM values suggest unnecessary treatment. Region D: Serious hypo- or hyperglycemia detection error region. Region E: Erroneous treatment region. CGM fails to detect potentially dangerous hypo- or hyperglycemia Ideally, all values are in Regions A and B. We report the percentage of correlated values that fall in each region.

    Time frame: 6 months

Secondary outcomes

  1. Unplanned Event Frequency

    Number of occurrences per patient of device failure or or unforeseen device removal

    Time frame: 6 months

07

Results

Posted Sep 29, 2026
Limitations and caveats
A limitation of the study was the small number of subjects, due principally to the difficulty of obtaining informed consent from either the critically ill patient or their Legally Authorized Representative, many of whom were unavailable in a timely manner. Another limitation is that all subjects were enrolled in respiratory ICUs at a single medical center. CGM use in other kinds of ICUs needs to be investigated separately.

Participant flow

Patients with hyperglycemia admitted to a respiratory intensive care unit (ICU) or their legal representative were approached to participate

Participant flow — Overall Study
MilestoneSingle Arm
Started16
Completed16
Not completed0

Outcome measures

PrimaryMean Absolute Relative Difference (MARD).

Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Mean Absolute Relative Difference (MARD). The MARD is the average percentage difference between the continuous glucose monitor readings measured over 6 months and glucose concentration results obtained by near simultaneous (+/- 5 minutes) standard test methods was used to gauge the accuracy of the CGM in the ICU setting. Typical MARD values reported for CGMs in the outpatient setting average approximately 8%. Typical MARD values reported for CGMs other than the one used in this study n the ICU setting average approximately 12%.

Time frame:
6 months
Reported as:
Mean · Percentages
Mean Absolute Relative Difference (MARD).
PercentagesSingle Arm
Mean Absolute Relative Difference (MARD).13.81 ± 12.40
PrimaryClarke Error Grid

Concordance of glucose concentration results obtained by standard test methods with those obtained nearly simultaneously by a continuous glucose monitoring device was assessed by Clarke Error Grid. The Clarke Error Grid plots of reference glucose values obtained by point-of-care fingerstick or laboratory determination (x-axis) against CGM values (y-axis). The grid is divided into 5 zones (A-E) that indicate clinical implications of CGM glucose measurements when compared to a "gold standard" Region A: Clinically accurate. Region B: Benign measurement error. Region C: Overcorrection error region CGM values suggest unnecessary treatment. Region D: Serious hypo- or hyperglycemia detection error region. Region E: Erroneous treatment region. CGM fails to detect potentially dangerous hypo- or hyperglycemia Ideally, all values are in Regions A and B. We report the percentage of correlated values that fall in each region.

Time frame:
6 months
Reported as:
Number · Percent In Clarke Regions
Clarke Error Grid
Percent In Clarke RegionsSingle Arm
Percent of Readings in Clarke Region A77.2
Percent of Readings in Clarke Region B22.5
Percent of Readings in Clarke Regions C or D0.3
SecondaryUnplanned Event Frequency

Number of occurrences per patient of device failure or or unforeseen device removal

Time frame:
6 months
Reported as:
Number · Device events
Unplanned Event Frequency
Device eventsSingle Arm
Unplanned Event Frequency0

Adverse events

Collected over Nine Months. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Single Arm0/16 (0%)0/16 (0%)0/16 (0%)

Baseline characteristics

Patients with hyperglycemia (\>200 mg/dL) admitted to the ICU

Age, Categorical
Age, Categorical(Participants)Single Arm
<=18 years0
Between 18 and 65 years16
>=65 years0
Sex: Female, Male
Sex: Female, Male(Participants)Single Arm
Female4
Male12
Race and Ethnicity Not Collected
Race and Ethnicity Not Collected(Participants)Single Arm
08

Study locations

1 site
  • UMass Memorial Medical Center
    Worcester, Massachusetts 01605, United States
09

References and documents

Study documents

PDFs hosted by the registry, ClinicalTrials.gov.

Individual participant data

Plan to share: Yes — Study design and deidentified aggregate glucose concentration data will be shared. No individual patient data will be shared.

Supporting information: Study protocol, Statistical analysis plan, Informed consent form

Time frame: One year

Access criteria: Publication

10

Timeline

1 registry update since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Ran from Feb 2024 to primary completion in Jul 2025, 17 months.

Results were first posted in Sep 2026, 1 year 2 months after primary completion.

Of 1,590 non-industry interventional studies with no phase under Diabetes Mellitus, half reached primary completion within 22 months of starting (the median), counting those still running; a quarter within 10 and three quarters within 43. This study's 17 months is below that.

Counting every sponsor, half of 1,750 studies reached it within 21 months; a quarter within 9 and three quarters within 42.

How these figures are worked out, and which studies are left out of them.

Registered
Dec 7, 2023
First posted
Jan 5, 2024
StartActual
Feb 20, 2024
Primary completionActual
Jul 25, 2025
CompletionActual
Jul 25, 2025
Results first posted
Sep 29, 2026
Last update
Sep 29, 2026
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
Results
Results posted
posted Sep 29, 2026
Primary outcomes
2 description edited
What changed in the outcomes
  • DescriptionClarke Error Grid: how it is measured was reworded
  • DescriptionMean Absolute Relative Difference (MARD).: how it is measured was reworded
Sep 29, 2026
Secondary outcomes
1 edited, 1 removed
What changed in the outcomes
  • EditedUnplanned event frequencyEvent FrequencyTime frame: 6 months
  • RemovedAdverse event nature and frequencyTime frame: 6 months
Sep 29, 2026
Show all 1 update
  1. Sep 29, 2026
    Results posted
    Primary outcomes 2 description edited
    What changed in the outcomes
    • DescriptionClarke Error Grid: how it is measured was reworded
    • DescriptionMean Absolute Relative Difference (MARD).: how it is measured was reworded
    Secondary outcomes 1 edited, 1 removed
    What changed in the outcomes
    • EditedUnplanned event frequencyEvent FrequencyTime frame: 6 months
    • RemovedAdverse event nature and frequencyTime frame: 6 months
    + 2 other changes: verification date and index terms

From the registry record's own update history. This site started tracking changes on Sep 25, 2026; for anything earlier, see the record history on ClinicalTrials.gov ↗

11

Registry details

Key details

Study ID
NCT06190808
Responsible party
John P Mordes, MD (Professor Emeritus of Medicine, University of Massachusetts, Worcester) — Principal investigator
First posted
Jan 5, 2024
Start date
Feb 20, 2024
Primary completion
Jul 25, 2025
Completion
Jul 25, 2025
Results posted
Sep 29, 2026
Last update
Sep 29, 2026

Study contacts

John P Mordes, MD
principal investigator · UMass Chan Medical School

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Sep 2026. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion