An interventional study of Implantation of premium monofocal IOL, LuxSmart (device under investigation) and Implantation of monofocal IOL, LuxGood (control device) in Cataract and Lens Opacities, sponsored by Cutting Edge SAS. Completed at 8 sites in 6 countries. Open to participants aged 22 Years and older, including healthy volunteers. Per ClinicalTrials.gov, last updated 2026-11-25.
Sponsored by Cutting Edge SAS · Not applicable, Interventional, and Treatment
Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →
Among 18 studies in Cataract with a site in Czechia — 1 open to participants now, 1 under way, none of them phase 3. Cataract · Czechia.
Among the 18 studies in Cataract with a site in Czechia, this one is the 2nd largest by enrollment of the 15 interventional ones that give a count, is one of 2 that started in 2023, is one of 10 with no phase, and was the 13th registered.
1,635 studies on the registry are indexed under Cataract; 233 are open to participants now.
Browse Cataract studies →1 of the 18 studies in Cataract with a site in Czechia is open now. Recruiting first, the newest first:
Carl Zeiss Meditec AG · Sites in Czechia, Germany, Spain and 1 more country
Ages 18+
A study can be open while its sites in Czechia have not opened yet or have stopped. Only those recruiting at a site in Czechia →
Of the 142 Cataract studies recruiting now, 87 (61%) set no upper age limit and 8 (6%) stop at 75 or younger; 10 take children under 18; 14 take healthy volunteers.
This study: 22 Years and older · all sexes · healthy volunteers too.
The full table on the hub →This tab as a report to print or save as PDF → With an account.
18 studies; the count open now beside each. This study's row is marked.
First in 2007; this study’s year is marked. 2 started in 2023.
By drug class, each counting the studies that test any of its medicines:
This study's enrollment of 251 is above the median of 79 across 1,096 interventional studies indexed under Cataract.
Median enrollment 135 across the 15 interventional studies, 164 across the 3 observational. Of the 16 that completed or were terminated, 1 (6%) have results posted.
This tab as a report to print or save as PDF → With an account.
14 of 18 have an industry lead sponsor. The 10 most frequent:
Cutting Edge SAS is the lead sponsor of 5 studies on the registry; 2 are open to participants now.
Cutting Edge SAS has 4 studies filed under Cataract — fewer than ten, so there is no comparison to draw.
None of the studies under way reaches primary completion within a year.
Among the 18 studies in Cataract with a site in Czechia, from their registry updates:
This tab as a report to print or save as PDF → With an account.
This study is a multicentric, prospective, randomised, controlled, post-market clinical follow up (PMCF) study to investigate safety, visual outcomes and contrast sensitivity after bilateral implantation of either LuxSmart IOLs (study group) or LuxGood IOLs (control group).
This study is a multicentric, prospective, randomised, controlled, post-market clinical follow up (PMCF) study whereby subjects undergoing routine cataract surgery will have bilateral implantation of either LuxSmart IOLs (study group) or LuxGood IOLs (control group).
The subjects will be randomized in a 1:1 ratio to receive the study or control lenses.
Both study and control IOLs, are CE approved The study and control IOLs, and all devices used for the study examinations, will be used within the intended use specifications from the manufacturer. In addition, no invasive or other burdening examinations will occur for the subject.
The study device (LuxSmart) is a hydrophobic, premium monofocal intraocular lens (IOL) manufactured by the sponsor of this study. The control lens (LuxGood) is the monofocal parent lens sharing the same material and optic design but with slight differences in the optic design. The IOLs will be implanted as part of the routine cataract surgery on subjects suffering from cataract. The targeted study cohort represents the standard subject cohort for cataract surgery.
In total 238 subjects will be enrolled for this clinical study and receive bilateral implantation of the study or control lens based on a 1:1 randomization given by the EDC.
Subjects participating in the trial will attend a total of 6 to 10 study visits (1 preoperative, 1 or 2 operative and 4 - 7 postoperative visits that may be carried out at the same day if requirements allow) over a period of 6 months.
Data analyses will be performed after the last patient finished the 120-180 days postoperative examination.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
In addition to above mentioned in- and exclusion criteria, subjects shall be discontinued when certain conditions are present at the time of surgery, including:
Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.
You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.
Judged by the study doctor to matter for your health or for the study. It is a doctor’s decision, so ask the study team if you are unsure whether something counts.
Ways of preventing pregnancy. “Highly effective” methods fail less than 1% of the time a year when used correctly — for example an IUD or an implant.
Criteria usually say which methods count and how long to keep using them after the last dose, sometimes for male participants too. Some studies accept a partner’s vasectomy if it was confirmed successful and that person is the participant’s only sexual partner.
A reason a treatment, test or procedure should not be given to a particular person, such as an allergy, another illness or a medicine they take.
Informed consent
Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.
You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Clinically significant
Judged by the study doctor to matter for your health or for the study. It is a doctor’s decision, so ask the study team if you are unsure whether something counts.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Contraception
Ways of preventing pregnancy. “Highly effective” methods fail less than 1% of the time a year when used correctly — for example an IUD or an implant.
Criteria usually say which methods count and how long to keep using them after the last dose, sometimes for male participants too. Some studies accept a partner’s vasectomy if it was confirmed successful and that person is the participant’s only sexual partner.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Contraindication
A reason a treatment, test or procedure should not be given to a particular person, such as an allergy, another illness or a medicine they take.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Experimental arm: Premium monofocal intraocular lens LuxSmart
Device: Implantation of premium monofocal IOL, LuxSmart (device under investigation)
Comparator arm: Monofocal intraocular lens LuxGood
Device: Implantation of monofocal IOL, LuxGood (control device)
Patients will be implanted with study IOL in both eyes
Patients will be implanted with Control IOL in both eyes
Primary Safety Endpoint: Mean Corrected Distance Visual Acuity
The primary safety endpoint is to show that the mean monocular Corrected Distance Visual Acuity (CDVA) at at 120-180 days postoperative is statistically non-inferior to outcomes obtained in the control group using a non-inferiority margin of 0.1 logMAR, a 1-sided test and a significance level of 0.05. This analysis will be done for first implanted eyes as well as all implanted eyes per subject.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Monocular Contrast Sensitivity - Between-group mean difference
The co-primary safety endpoint is to compare the outcomes on monocular contrast sensitivity under mesopic light conditions (with and without glare) between the study and the control group at 120-180 days postoperative. This analysis will be done for first implanted eyes per subject. The following Log Contrast Sensitivity Analysis will be performed: \- Between-group mean difference in log contrast sensitivity with 90% non-parametric confidence interval for each spatial frequency.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Monocular Contrast Sensitivity - descriptive statistics
The co-primary safety endpoint is to compare the outcomes on monocular contrast sensitivity under mesopic light conditions (with and without glare) between the study and the control group at 120-180 days postoperative. This analysis will be done for first implanted eyes per subject. The following Log Contrast Sensitivity Analysis will be performed: \- Provide descriptive statistics for the log contrast sensitivity for each group (mean, SD, median, 0th, 25th, 50th 75th, and 100th percentiles) and for each spatial frequency.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Monocular Contrast Sensitivity - frequency with glare
The co-primary safety endpoint is to compare the outcomes on monocular contrast sensitivity under mesopic light conditions with glare between the study and the control group at 120-180 days postoperative. This analysis will be done for first implanted eyes per subject. The following Log Contrast Sensitivity Analysis will be performed: \- The percentage of eyes that can and cannot see the reference pattern for each spatial frequency.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Monocular Contrast Sensitivity - frequency without glare
The co-primary safety endpoint is to compare the outcomes on monocular contrast sensitivity under mesopic light conditions without glare between the study and the control group at 120-180 days postoperative. This analysis will be done for first implanted eyes per subject. The following Log Contrast Sensitivity Analysis will be performed: \- The percentage of eyes that can and cannot see the reference pattern for each spatial frequency.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Best Corrected Distance Visual Acuity compared to historical data
The objective is to compare best corrected distance visual acuities (CDVA) above defined thresholds of the investigational product to the normative data stated in the according ISO norm (EN ISO 11979-7:2018) for posterior chamber intraocular lenses.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Rates of Adverse Events
The objective is to compare the SPE (Safety and Performance Endpoints) rates of the investigational product to reference data stated in the according ISO standard for posterior intraocular lenses (EN ISO 11979-7:2018) based on minimum 100 subjects.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Incidence of all SAE
The incidence of all Serious Adverse Events, including Secondary Surgical Interventions (SSIs) related to the optical properties of the IOL, in first-implanted eyes as well as all implanted eyes will be collected through Post-Operative Visit 4 (120 to 180 days after first and second eye implantation). The proportion of first implanted eyes and all implanted eyes with at least one serious adverse event will be summarized using categorical summary statistics by treatment received. Each eye will be counted only once in the calculation of the rate. There is no statistical hypothesis associated with the proportion of first implanted eyes with at least one serious adverse event.
Time frame: 120-180 days postoperative
Co-Primary Safety Endpoint: Rate of SSI Related to Optical Properties of the IOL
The rate of SSIs related to the optical properties of the IOL for first-implanted eyes as well as all implanted eyes will be reported through 120 to 180 days after first and second eye implantation. Secondary surgical interventions related to the optical properties of the IOL will be defined as IOL explantation, replacement, or repositioning due to subject intolerance of visual symptoms not adequately improved by spectacle correction. Each eye will be classified as either having undergone a secondary surgical intervention related to the optical properties of the IOL or not having undergone such an intervention. Secondary surgical interventions related to the optical properties of the IOL will be summarized categorically (Yes, No) by actual treatment received in a table.
Time frame: 120-180 days postoperative
Primary Performance Endpoint: Mean Distance Corrected Intermediate Visual Acuity
The primary efficacy endpoint is to show that the monocular Distance Corrected Intermediate Visual Acuity (DCIVA) at 120-180 days postoperative is statistically superior to outcomes obtained in the control group (1 sided test using significance of 0.025). To avoid bias, only the first implanted eye per subject will be considered for this calculation.
Time frame: 120-180 days postoperative
Co-Primary Performance Endpoint: Cumulative Distance Corrected Intermediate Visual Acuity
One co-primary efficacy endpoint is to show that the outcomes of the study device on monocular Distance Corrected Intermediate Visual Acuity (DCIVA) at 66 cm at 120-180 days postoperative needs to have at least 50% of eyes achieving 0.2 logMAR or better. To avoid bias, only the first implanted eye per subject will be considered for this calculation.
Time frame: 120-180 days postoperative
Co-Primary Performance Endpoint: Monocular best-corrected distance defocus
Monocular best-corrected distance defocus testing will be performed. The defocus range where the mean visual acuity of 0.2 logMAR or better is achieved will be derived by visual inspection of the mean defocus curve. The monocular depth of focus is defined as the defocus range from zero to the first negative vergence level, where the visual acuity is 0.2 logMAR or less. One co-primary efficacy endpoint is to show that the monocular depth of focus for the eyes implanted with the study device is at least 0.5 D greater than the depth of focus for the eyes implanted with the control device at 0.2 logMAR.
Time frame: 120-180 days postoperative
Secondary Safety Endpoint: Best Corrected Distance Visual Acuity compared to historical data
The objective is to compare best corrected distance visual acuities (CDVA) above defined thresholds of the investigational product to the normative data stated in the according ISO norm (EN ISO 11979-7:2018) for posterior chamber intraocular lenses.
Time frame: 120-180 days postoperative
Secondary Performance Endpoint: Distance Corrected Near Visual Acuity (DCNVA)
Secondary performance endpoint is to show a statistically significant increase between pre-and postoperative findings on monocular Distance Corrected Near Visual Acuity (DCNVA) under photopic light conditions on the first implanted eye.
Time frame: 120-180 days postoperative
Secondary Performance Endpoint: Distance Corrected Intermediate Visual Acuity (DCIVA)
Secondary performance endpoint is to show a statistically significant increase between pre-and postoperative findings on monocular Distance Corrected Intermediate Visual Acuity (DCIVA) under photopic light conditions on the first implanted eye.
Time frame: 120-180 days postoperative
Binocular best-corrected distance defocus
In addition to the co-primary study endpoint on monocular defocus, binocular best-corrected distance defocus testing will be performed. The defocus range where the mean visual acuity of 0.2 logMAR or better is achieved will be derived by visual inspection of the mean defocus curve. The monocular depth of focus for the eyes implanted with the study device needs to be at least 0.5 D greater than the depth of focus for the eyes implanted with the control device at 0.2 logMAR \[4\]. All measured outcomes data will be statistically compared to outcomes obtained in the control group.
Time frame: 120-180 days postoperative
Manifest refraction
The manifest refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2018. This data will also be used to calculate the manifest refractive spherical equivalent (MRSE)
Time frame: Preoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Adjusted Mean Refractive Spherical Equivalent (MRSE)
The predictability of the postoperative manifest refraction will be evaluated using the absolute value of the adjusted MRSE calculated by the following formula: MRSEadjusted = MRSEpostop - MRSEtarget.
Time frame: 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Monocular Uncorrected Distance Visual Acuity (UDVA) under photopic light conditions
Monocular UDVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 4m distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: Preoperative, 1-2 days postoperative; 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Binocular Uncorrected Distance Visual Acuity (UDVA) under photopic light conditions
Binocular UDVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 4m distance according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 120-180 days postoperative
Monocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions
Monocular CDVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: Preoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Monocular Corrected Distance Visual Acuity (low contrast CDVA) using low contrast (10%) optotypes under photopic light conditions
Monocular CDVA using low contrast (10%) optotypes is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions. This test is performed using a low-contrast acuity chart with 10% contrast (using the Weber definition: \[background - optotype\]/background).
Time frame: 120-180 days postoperative
Binocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions
Binocular CDVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 120-180 days postoperative
Monocular Uncorrected Intermediate Visual Acuity (UIVA) under photopic light conditions
Monocular UIVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 66cm distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: 30-60 days postoperative, 120-180 days postoperative
Binocular Uncorrected Intermediate Visual Acuity (UIVA) under photopic light conditions
Binocular UIVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 66cm distance according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 120-180 days postoperative
Binocular Uncorrected Near Visual Acuity (UNVA) under photopic light conditions
Binocular UNVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 40cm distance according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 120-180 days postoperative
Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) under photopic light conditions
Monocular DCIVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 66cm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: 30-60 days postoperative, 120-180 days postoperative
Monocular Distance Corrected Intermediate Visual Acuity (DCIVA) under mesopic light conditions
Monocular DCIVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 66cm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done monocularly under mesopic light conditions.
Time frame: 120-180 days postoperative
Monocular Distance Corrected Intermediate Visual Acuity (low contrast DCIVA) using low contrast (10%) optotypes under photopic light conditions
Monocular DCIVA using low contrast (10%) optotypes is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 66cmm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions. This test is performed using a low-contrast acuity chart with 10% contrast (using the Weber definition: \[background - optotype\]/background).
Time frame: 120-180 days postoperative
Binocular Distance Corrected Intermediate Visual Acuity (DCIVA) under photopic light conditions
Binocular DCIVA is measured using the Clinical Trial Suite (M\&S Technologies, Niles, USA) placed in 66cm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 120-180 days postoperative
Binocular Distance Corrected Near Visual Acuity (DCNVA) under photopic light conditions
Binocular DCNVA is measured with ETDRS charts placed in 40cm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 120-180 days postoperative
Subject Symptom Assessment
Subject ocular and visual symptoms are collected at each postoperative visit through the use of non-directed questions. Responses will be recorded and characterized by severity (mild, moderate, severe) and divided into distinct groups of symptoms (halo, glare, ocular pain, ocular dryness, other). The percentage of eyes with ocular and visual symptoms will be reported divided into the mentioned severity levels and symptom groups. No scoring or scaling is used for this type of symptom assessment. All data obtained in the study group will be statistically compared to outcomes obtained in the control group.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Patient reported outcomes - Quality of Vision questionnaire
To assess the subjective perception of disturbances by photic phenomena, the validated and verified Quality of Vision (QoV) questionnaire will be used. This questionnaire is a linear-scaled 30-item instrument and is designed with 10 symptoms rated in each of three scales (frequency, severity, and bothersome). Each of the three scales provides 4 answer options with the first answer being the best outcome and the 4th answer being the worst outcome: * Frequency answer options: never, occasionally, quite often, very often; * Severity answer options: not at all, mild, moderate, severe; * bothersome answer options: not at all, a little, quite, very. Based on the given answers, a QoV score in terms of symptom frequency, severity, and bothersome is calculated. The score ranges from 0 to 100 with 100 being the worst possible outcome.
Time frame: 30-60 days postoperative, 120-180 days postoperative
Patient reported outcomes - Catquest-9SF 2011 questionnaire
To assess if difficulties in daily life occur due to impaired sight. This questionnaire is a 9-item Rasch-scaled instrument to assess subjective perception of visual impairment after surgery. Each item has 4 answer options ranging from 'Yes, very great difficulty' to 'No, no difficulty'. with first answer being the best outcome and the last answer being the worst outcome. In addition, the subjects can state, that they cannot decide on the answer. The percentage of subjects per item will be reported per answer option. No scoring or scaling is used for this type of symptom assessment. All data obtained in the study group will be statistically compared to outcomes obtained in the control group.
Time frame: 30-60 days postoperative, 120-180 days postoperative
Pupil Size under photopic light conditions
Photopic pupil diameters are measured with a ruler allowing for a precision of at least +/-0.5 mm. In addition to the measurement with a ruler, pupil diameters under photopic light conditions will be recorded from the preoperative Pentacam measurement. For the photopic pupil diameters, eye illumination should be identical to that used for photopic contrast sensitivity testing. Pupil measurements should only be made after the eye has had time to fully adapt to the testing conditions (at least 5 minutes). The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: preoperative, 120-180 days postoperative
Pupil Size under mesopic light conditions
Mesopic pupil diameters are measured with a ruler allowing for a precision of at least +/-0.5 mm. For the mesopic pupil diameters, eye illumination should be identical to that used for mesopic contrast sensitivity testing. Pupil measurements should only be made after the eye has had time to fully adapt to the testing conditions (at least 5 minutes). The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: preoperative, 120-180 days postoperative
Fundus examination with dilated pupil
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Fundus
Time frame: preoperative, 120-180 days postoperative
Slitlamp examination - Corneal status
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Corneal Status
Time frame: Preoperative, 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - Iris status
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: • Iris Status
Time frame: Preoperative, 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - Signs of inflammation
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: * Signs of inflammation * Anterior chamber cells, * Anterior chamber flare, * Cystoid macular oedema, * Hypopyon, and * Endophthalmitis.
Time frame: Preoperative, 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - Pupillary block
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- Pupillary block.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - Retinal detachment
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- Retinal detachment.
Time frame: preoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - Status of anterior and posterior capsule
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- Status of anterior and posterior capsule.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - IOL decentration
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- IOL decentration.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - IOL tilt
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- IOL tilt.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - IOL discoloration
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- IOL discoloration.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Slitlamp examination - IOL opacity
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp: \- IOL opacity.
Time frame: 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Intraocular pressure (IOP) measurement
The IOP will be measured with non-contact tonometer as part of the routine follow up examinations. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative, 1-2 days postoperative, 7-14 days postoperative, 30-60 days postoperative, 120-180 days postoperative
Keratometry
Keratometric measurements are performed to calculate the required IOL power. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
IOL power
The power of the implanted IOL will be recorded.
Time frame: during surgery
Target refraction
The target refraction given by the IOL calculator is recorded. This parameter is needed to calculate the accuracy of achieving the target refraction. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: during surgery
Fundus OCT
An OCT (optical coherence tomography) image will be taken at the preoperative visit to identify AMD suspicious eyes that need to be excluded from this study. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
Biometry
Biometry measurements are performed to calculate the required IOL power. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
Corneal topography
Pentacam device (Oculus, Germany) will be used to perform corneal topography. This examination is needed to exclude patients showing an irregular corneal astigmatism.
Time frame: Preoperative
Corneal aberrometry
Pentacam device (Oculus, Germany) will be used to perform corneal aberrometry measurement. The following values will be evaluated in this study: Spherical aberrations, high order aberrations.
Time frame: Preoperative
Halo and glare scores
A Halo and Glare simulator software (Eyeland Design Network GmbH, Germany) will be used to assess the subjective perception of photic phenomena. The patients will be asked to adjust the halo and glare level in a simulated image to the amount they perceive of such photic phenomena. The scaling ranges from 0 to 100 with 0 indicating no halo or glare and 100 being the worst outcome. All data obtained in the study group will be statistically compared to outcomes obtained in the control group.
Time frame: 30-60 days postoperative
Plan to share: No
Ran from Jun 2023 to primary completion in Feb 2025, 20 months.
Of 190 industry-sponsored interventional studies with no phase under Cataract, half reached primary completion within 16 months of starting (the median), counting those still running; a quarter within 10 and three quarters within 28. This study's 20 months is above that.
Counting every sponsor, half of 496 studies reached it within 17 months; a quarter within 9 and three quarters within 29.
Since we started tracking on Sep 25, 2026, the dates moved once, on Nov 25, 2026: primary completion Apr 2025 → Feb 2025 and completion Apr 2025 → Feb 2025.
Earlier changes are on the record history on ClinicalTrials.gov ↗
How these figures are worked out, and which studies are left out of them.
From the registry record's own update history. This site started tracking changes on Sep 25, 2026; for anything earlier, see the record history on ClinicalTrials.gov ↗
This study is completed, as verified in Nov 2026. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Cutting Edge SAS