Bedside Resources to Gauge Intravascular Volume Status

An interventional study of CardioQ-EDM and CipherOx-CRI in Hypovolemia and Craniosynostoses, sponsored by University of Colorado, Denver. Completed at 1 site in United States. Open to participants aged 3 Months to 2 Years. Per ClinicalTrials.gov, last updated 2023-06-29.

Sponsored by University of Colorado, Denver · Not applicable, Interventional, and Diagnostic

Phase
Not applicable
Study type
Interventional
Enrollment
23
Allocation
Not applicable
Ages
3 Months to 2 Years
Sex
All
01

Study summary

The goal if this study is to employ the CardioQ-Esophageal Aortic Doppler probe to define fluid responders from non-responders among infants undergoing cranial vault reconstruction for craniosynostosis. After defining these two groups in this single arm prospective trial, the investigators will compare the predictive utility of non-invasive devices such as the CipherOx-Compensatory Reserve Index (CipherOx-CRI) and Inferior Vena Cava Collapsibility Index (IVC CI) to currently employed indices (heart rate, systolic blood pressure, urine output and pulse pressure variability) to gauge the need for additional fluid and ongoing resuscitation. If the CipherOx-CRI or IVC CI proved to be as predictive or better at predicting fluid responders, the investigators hope to replace invasive arterial lines with non-invasive tools to guide resuscitation.

Read the detailed description

Predicting fluid responsiveness in the operating room is essential to guide balanced resuscitation. Aggressive resuscitation may lead to significant morbidities, such as intra-abdominal hypertension, pulmonary edema, difficulty with ventilator liberalization, and consequently increased mortality. Alternatively, under resuscitation may lead to mal-perfusion and end-organ dysfunction.

A plethora of indices and tools have been studied and marketed to assess intravascular volume status with only a few proven reliable with reproducible results. Based on pre-fluid challenge values, several of these tools have been used to predict who may benefit from additional fluid (fluid responders). Alternatively, some of these tools have been used to distinguish fluid responders from non-responders based on changes in pre- and post-fluid challenge values. Among these tools, the pulmonary artery catheter provides measurements of both left and right heart pressures which can be applied to calculate the cardiac output (CO) and stroke volume (SV). Changes in these values (e.g. an increase in the stroke volume by 10%) between pre- and post-fluid challenge have been used to define fluid responders. This devise, however, is invasive with several significant risks, and therefore is rarely used in children. Echocardiography, on the other hand, is a non-invasive bedside study also used to assess CO and SV but is expensive and requires trained echosonagraphers for application. Further, because a transthoracic probe is required to obtain the images, application in the operating room is difficult as the chest is often in the operating field limiting access to the echosonagrapher. Lastly, the esophageal aortic blood flow device (CardioQ-Esophageal Dopler Monitor (Cardio-EDM), Deltex Medical, Chichester, UK) has been found in multiple adult and pediatric studies to reliably distinguish fluid responders from non-responders intensive care unit (ICU) and operating room. Much like an orogastric tube, this device is simply placed by a provider in the patient's esophagus and uses Doppler waveforms to measure aortic blood flow velocities. Variations in the amplitude of peak velocities has been shown to corelate with intravascular volume status. Specifically, a change in the peak velocity by greater than 10% between pre- and post-fluid challenge values has been shown to accurately distinguishes those who are fluid responsive from those who are not with similar accuracy to echocardiography and pulmonary artery catheter readings.

In recent years with continued technological advancements, there has been enthusiasm about less invasive, and in some cases, non-invasive, tools to gauge volume status. Among these, bedside ultrasonography (performed by providers rather than sonographers) is a common tool used to evaluate the inferior vena cava (IVC) collapsibility index (CI) has been shown to be a reliable tool in adults. Another non-invasive device uses a photoplethysmoraphic probe (CipherOx-CRI) placed on a digit to calculate the compensatory reserve index (CRI), a marker of proximity to hemodynamic collapse. Both IVC CI and CRI have been shown in multiple adult studies to predictive the need for volume expansion, but their utility in the pediatric population is unknown.

The goal if this proposed study is to employ the CardioQ-EDM probe to define fluid responders from non-responders among infants undergoing cranial vault reconstruction for craniosynostosis. After defining these two groups in this single arm prospective trial, the investigators will compare the predictive utility of non-invasive devices such as the CipherOx-CRI and IVC CI to currently employed indices (heart rate, systolic blood pressure, urine output and pulse pressure variability) to gauge the need for additional fluid and ongoing resuscitation. If the CipherOx-CRI or IVC CI proved to be as predictive or better at predicting fluid responders, the investigators hope to replace invasive arterial lines with non-invasive tools to guide resuscitation.

The investigators chose this population for several reasons. First, the investigators institution performs approximately 50-70 of these cases a year making them a relatively accessible group. Second, these children are generally healthy which will minimize physiologic confounders. Additionally, the subjects are paralyzed, have normal respiratory compliance, and providers maintain normothermia, all of which will minimizing confounders. Another unique benefit to this population is that these infants have been nil per os for several hours prior to surgery, putting them at risk for hypovolemia, and after induction, independent of the provider's assessment of intravascular volume status, all children receive a bolus of crystalloid (10mL/kg). This baseline data should provide sufficient data for analysis; but because these procedures are associated with significant blood loss and hypovolemia requiring aggressive resuscitation in the form of fluid or blood boluses, the investigators plan to continue to collect pre- and post- bolus data with the hope to further validate the benefit of non-invasive tools such as the CipherOx-CRI and IVC CI in the setting of ongoing blood loss.

As intravascular volume status is often difficult to assess clinically, the investigators aim to determine the predictability of non-invasive devices to guide resuscitation. In this prospective observational study, the investigators hope to identify:

  1. The proportion of children within the cohort who are fluid responsive based on CardioQ-EDM aortic blood flow velocity changes pre- and post-bolus,
  2. The positive predictive value, negative predictive value, sensitivity, specificity, and optimal threshold for CRI, IVC CI, pulse pressure variability, stroke volume variability, heart rate, systolic blood pressure, and mean arterial pressures in predicting fluid responders as determined by CardioQ-EDM, and
  3. Assess confounding variables that may influence the predictive utility of such devices
02

Conditions studied

  • Hypovolemia
  • Craniosynostoses
03

In context

Hypovolemia

122 studies on the registry are indexed under Hypovolemia; 14 are open to participants now.

This study's enrollment of 23 is below the median of 40 across 61 interventional studies indexed under Hypovolemia.

Browse Hypovolemia studies →

Lead sponsor

University of Colorado, Denver is the lead sponsor of 1,543 studies on the registry; 317 are open to participants now.

Of its 147 completed or terminated interventional studies of FDA-regulated products, 92 (63%) have results posted.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
3 Months to 2 Years
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Children with craniosynostosis undergoing cranial vault reconstruction

Exclusion criteria

Exclusion Criteria:

  • Children with known underlying cardiac anomalies or cardiac arrhythmias
  • Weight less than 3 kg
  • Children who have vasopressors adjusted during a fluid bolus
05

Study design

Phase
Not applicable
Primary purpose
Diagnostic
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
23 participants (actual)

Study arms

  • Other
    Fluid Challenge

    After defining fluid responders from non-responders in this single arm prospective trial, we will compare the predictive utility of non-invasive devices such as the CipherOx-CRI and IVC CI to currently employed indices (heart rate, systolic blood pressure, urine output and pulse pressure variability) to gauge the need for additional fluid and ongoing resuscitation.

    Device: CardioQ-EDM and CipherOx-CRI

Interventions

  • DeviceCardioQ-EDM and CipherOx-CRI

    A CardioQ-EDM probe will be placed on the day of surgery after induction of general anesthesia. The anesthesiologist will inform the investigator of plans to provide a fluid or blood bolus per clinical judgement in addition to the protocolized 10 ml/kg bolus provided after induction. While the anesthesiologist is preparing to administer volume expansion, a co-investigator will collect pre-fluid bolus data. Measurements will be recorded for data analysis at the completion of the trial. Additionally, a CipherOx-CRI probe will be placed on the patient's index finger (recorded data will be interpreted post hoc) and a bedside ultrasound will be performed by either the principal investigator (PI) or one of two co-investigators to measure the IVC CI. Ultrasound cine-loops will be recorded, and CI will be calculated post-hoc. Data will be recorded on the Data Collection Form for each fluid bolus administered. The PI and co-investigators will manage all aspects of investigational devices.

06

What researchers measure

Primary outcomes

  1. Utility of Compensatory Reserve Index (CRI) Which Ranges From 0-1 in Order to Predict Fluid Responders From Non-responders

    Using a delta peak aortic velocity threshold of 10% (measured from CardioQ-EDM) before and after a bolus to define fluid responders (=/\>10%) from non-responders (\<10%), we will determine the performance of pre-bolus CRI reading which is an index between 0 and 1 (0=poor reserve and 1=excellent reserve) in order to predict fluid responders from non-responders. Measurements will be recorded three times with one minute between measurements and then averaged. Vital signs analyzed included heart rate, systolic blood pressure, mean arterial pressure, shock index (heart rate/systolic blood pressure), pulse pressure variability, and end-tidal carbon dioxide level. Infants were also monitored with a Compensatory Reserve Index (CRI) monitor, which provides a continuous, individual-specific, beat-to-beat estimate of central volume status, from normovolemia (CRI=1) to decompensation (CRI=0). Each variable's performance was compared using area under the receiver operator curves (AUC).

    Time frame: Through study completion (3-4 hours)

07

Results

Posted Jun 29, 2023

Participant flow

Participant flow — Overall Study
MilestoneFluid Challenge RespondersFluid Challenge Non-responders
Started716
Completed716
Not completed00

Outcome measures

PrimaryUtility of Compensatory Reserve Index (CRI) Which Ranges From 0-1 in Order to Predict Fluid Responders From Non-responders

Using a delta peak aortic velocity threshold of 10% (measured from CardioQ-EDM) before and after a bolus to define fluid responders (=/\>10%) from non-responders (\<10%), we will determine the performance of pre-bolus CRI reading which is an index between 0 and 1 (0=poor reserve and 1=excellent reserve) in order to predict fluid responders from non-responders. Measurements will be recorded three times with one minute between measurements and then averaged. Vital signs analyzed included heart rate, systolic blood pressure, mean arterial pressure, shock index (heart rate/systolic blood pressure), pulse pressure variability, and end-tidal carbon dioxide level. Infants were also monitored with a Compensatory Reserve Index (CRI) monitor, which provides a continuous, individual-specific, beat-to-beat estimate of central volume status, from normovolemia (CRI=1) to decompensation (CRI=0). Each variable's performance was compared using area under the receiver operator curves (AUC).

Time frame:
Through study completion (3-4 hours)
Reported as:
Mean · units on a scale
Utility of Compensatory Reserve Index (CRI) Which Ranges From 0-1 in Order to Predict Fluid Responders From Non-responders
units on a scaleFluid Challenge RespondersFluid Challenge Non-responders
Utility of Compensatory Reserve Index (CRI) Which Ranges From 0-1 in Order to Predict Fluid Responders From Non-responders0.06 ± 0.030.05 ± 0.02

Adverse events

Collected over Data was collected from the time of probe placement in the operating room until discharge (0-3 days). Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Fluid Challenge Responders0/7 (0%)0/7 (0%)0/7 (0%)
Fluid Challenge Non-responders0/16 (0%)0/16 (0%)0/16 (0%)

Baseline characteristics

Age, Categorical
Age, Categorical(Participants)Fluid Challenge RespondersFluid Challenge Non-respondersTotal
<=18 years71623
Between 18 and 65 years000
>=65 years000
Sex: Female, Male
Sex: Female, Male(Participants)Fluid Challenge RespondersFluid Challenge Non-respondersTotal
Female347
Male41216
Ethnicity (NIH/OMB)
Ethnicity (NIH/OMB)(Participants)Fluid Challenge RespondersFluid Challenge Non-respondersTotal
Hispanic or Latino112
Not Hispanic or Latino51419
Unknown or Not Reported112
Race (NIH/OMB)
Race (NIH/OMB)(Participants)Fluid Challenge RespondersFluid Challenge Non-respondersTotal
American Indian or Alaska Native000
Asian000
Native Hawaiian or Other Pacific Islander000
Black or African American011
White71320
More than one race000
Unknown or Not Reported022
Region of Enrollment
Region of Enrollment(participants)Fluid Challenge RespondersFluid Challenge Non-respondersTotal
United States71623
08

Study locations

1 site
  • Children's Hospital Colorado
    Aurora, Colorado 80045, United States
09

References and documents

Publications

  • Malbrain ML, Marik PE, Witters I, Cordemans C, Kirkpatrick AW, Roberts DJ, Van Regenmortel N. Fluid overload, de-resuscitation, and outcomes in critically ill or injured patients: a systematic review with suggestions for clinical practice. Anaesthesiol Intensive Ther. 2014 Nov-Dec;46(5):361-80. doi: 10.5603/AIT.2014.0060. PubMed 25432556 ↗
  • Feissel M, Michard F, Mangin I, Ruyer O, Faller JP, Teboul JL. Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock. Chest. 2001 Mar;119(3):867-73. doi: 10.1378/chest.119.3.867. PubMed 11243970 ↗
  • Monnet X, Rienzo M, Osman D, Anguel N, Richard C, Pinsky MR, Teboul JL. Esophageal Doppler monitoring predicts fluid responsiveness in critically ill ventilated patients. Intensive Care Med. 2005 Sep;31(9):1195-201. doi: 10.1007/s00134-005-2731-0. Epub 2005 Jul 30. PubMed 16059723 ↗
  • Durand P, Chevret L, Essouri S, Haas V, Devictor D. Respiratory variations in aortic blood flow predict fluid responsiveness in ventilated children. Intensive Care Med. 2008 May;34(5):888-94. doi: 10.1007/s00134-008-1021-z. Epub 2008 Feb 8. PubMed 18259726 ↗
  • Absi MA, Lutterman J, Wetzel GT. Noninvasive cardiac output monitoring in the pediatric cardiac Intensive Care Unit. Curr Opin Cardiol. 2010 Mar;25(2):77-9. doi: 10.1097/HCO.0b013e3283362452. PubMed 20104174 ↗
  • Weber T, Wagner T, Neumann K, Deusch E. Low predictability of three different noninvasive methods to determine fluid responsiveness in critically ill children. Pediatr Crit Care Med. 2015 Mar;16(3):e89-94. doi: 10.1097/PCC.0000000000000364. PubMed 25647238 ↗
  • Pereira de Souza Neto E, Grousson S, Duflo F, Ducreux C, Joly H, Convert J, Mottolese C, Dailler F, Cannesson M. Predicting fluid responsiveness in mechanically ventilated children under general anaesthesia using dynamic parameters and transthoracic echocardiography. Br J Anaesth. 2011 Jun;106(6):856-64. doi: 10.1093/bja/aer090. Epub 2011 Apr 26. PubMed 21525016 ↗
  • Lanspa MJ, Grissom CK, Hirshberg EL, Jones JP, Brown SM. Applying dynamic parameters to predict hemodynamic response to volume expansion in spontaneously breathing patients with septic shock. Shock. 2013 Feb;39(2):155-60. doi: 10.1097/SHK.0b013e31827f1c6a. PubMed 23324885 ↗
  • Janak JC, Howard JT, Goei KA, Weber R, Muniz GW, Hinojosa-Laborde C, Convertino VA. Predictors of the Onset of Hemodynamic Decompensation During Progressive Central Hypovolemia: Comparison of the Peripheral Perfusion Index, Pulse Pressure Variability, and Compensatory Reserve Index. Shock. 2015 Dec;44(6):548-53. doi: 10.1097/SHK.0000000000000480. PubMed 26529655 ↗
  • Czerwinski M, Hopper RA, Gruss J, Fearon JA. Major morbidity and mortality rates in craniofacial surgery: an analysis of 8101 major procedures. Plast Reconstr Surg. 2010 Jul;126(1):181-186. doi: 10.1097/PRS.0b013e3181da87df. PubMed 20220557 ↗
  • Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D. Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med. 1970 Aug 27;283(9):447-51. doi: 10.1056/NEJM197008272830902. No abstract available. PubMed 5434111 ↗
  • National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network; Wheeler AP, Bernard GR, Thompson BT, Schoenfeld D, Wiedemann HP, deBoisblanc B, Connors AF Jr, Hite RD, Harabin AL. Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. N Engl J Med. 2006 May 25;354(21):2213-24. doi: 10.1056/NEJMoa061895. Epub 2006 May 21. PubMed 16714768 ↗
  • Shah MR, Hasselblad V, Stevenson LW, Binanay C, O'Connor CM, Sopko G, Califf RM. Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA. 2005 Oct 5;294(13):1664-70. doi: 10.1001/jama.294.13.1664. PubMed 16204666 ↗
  • Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M; Canadian Critical Care Clinical Trials Group. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003 Jan 2;348(1):5-14. doi: 10.1056/NEJMoa021108. PubMed 12510037 ↗
  • Westphal GA, Goncalves AR, Bedin A, Steglich RB, Silva E, Poli-de-Figueiredo LF. Vasodilation increases pulse pressure variation, mimicking hypovolemic status in rabbits. Clinics (Sao Paulo). 2010 Feb;65(2):189-94. doi: 10.1590/S1807-59322010000200011. PubMed 20186303 ↗
  • Monnet X, Marik PE, Teboul JL. Prediction of fluid responsiveness: an update. Ann Intensive Care. 2016 Dec;6(1):111. doi: 10.1186/s13613-016-0216-7. Epub 2016 Nov 17. PubMed 27858374 ↗
  • Chytra I, Pradl R, Bosman R, Pelnar P, Kasal E, Zidkova A. Esophageal Doppler-guided fluid management decreases blood lactate levels in multiple-trauma patients: a randomized controlled trial. Crit Care. 2007;11(1):R24. doi: 10.1186/cc5703. PubMed 17313691 ↗
  • Desgranges FP, Desebbe O, Pereira de Souza Neto E, Raphael D, Chassard D. Respiratory variation in aortic blood flow peak velocity to predict fluid responsiveness in mechanically ventilated children: a systematic review and meta-analysis. Paediatr Anaesth. 2016 Jan;26(1):37-47. doi: 10.1111/pan.12803. Epub 2015 Nov 6. PubMed 26545173 ↗
  • Zhao J, Wang G. Inferior Vena Cava Collapsibility Index is a Valuable and Non-Invasive Index for Elevated General Heart End-Diastolic Volume Index Estimation in Septic Shock Patients. Med Sci Monit. 2016 Oct 20;22:3843-3848. doi: 10.12659/msm.897406. PubMed 27762259 ↗
  • Preau S, Bortolotti P, Colling D, Dewavrin F, Colas V, Voisin B, Onimus T, Drumez E, Durocher A, Redheuil A, Saulnier F. Diagnostic Accuracy of the Inferior Vena Cava Collapsibility to Predict Fluid Responsiveness in Spontaneously Breathing Patients With Sepsis and Acute Circulatory Failure. Crit Care Med. 2017 Mar;45(3):e290-e297. doi: 10.1097/CCM.0000000000002090. PubMed 27749318 ↗
  • Gan H, Cannesson M, Chandler JR, Ansermino JM. Predicting fluid responsiveness in children: a systematic review. Anesth Analg. 2013 Dec;117(6):1380-92. doi: 10.1213/ANE.0b013e3182a9557e. PubMed 24257389 ↗
  • Muniz GW, Wampler DA, Manifold CA, Grudic GZ, Mulligan J, Moulton S, Gerhardt RT, Convertino VA. Promoting early diagnosis of hemodynamic instability during simulated hemorrhage with the use of a real-time decision-assist algorithm. J Trauma Acute Care Surg. 2013 Aug;75(2 Suppl 2):S184-9. doi: 10.1097/TA.0b013e31829b01db. PubMed 23883906 ↗
  • Stewart CL, Mulligan J, Grudic GZ, Convertino VA, Moulton SL. Detection of low-volume blood loss: compensatory reserve versus traditional vital signs. J Trauma Acute Care Surg. 2014 Dec;77(6):892-7; discussion 897-8. doi: 10.1097/TA.0000000000000423. PubMed 25423536 ↗
  • Monnet X, Bleibtreu A, Ferre A, Dres M, Gharbi R, Richard C, Teboul JL. Passive leg-raising and end-expiratory occlusion tests perform better than pulse pressure variation in patients with low respiratory system compliance. Crit Care Med. 2012 Jan;40(1):152-7. doi: 10.1097/CCM.0b013e31822f08d7. PubMed 21926581 ↗
  • Teboul JL, Monnet X, Chemla D, Michard F. Arterial Pulse Pressure Variation with Mechanical Ventilation. Am J Respir Crit Care Med. 2019 Jan 1;199(1):22-31. doi: 10.1164/rccm.201801-0088CI. PubMed 30138573 ↗
  • Airapetian N, Maizel J, Alyamani O, Mahjoub Y, Lorne E, Levrard M, Ammenouche N, Seydi A, Tinturier F, Lobjoie E, Dupont H, Slama M. Does inferior vena cava respiratory variability predict fluid responsiveness in spontaneously breathing patients? Crit Care. 2015 Nov 13;19:400. doi: 10.1186/s13054-015-1100-9. PubMed 26563768 ↗
  • Tibby SM, Hatherill M, Murdoch IA. Use of transesophageal Doppler ultrasonography in ventilated pediatric patients: derivation of cardiac output. Crit Care Med. 2000 Jun;28(6):2045-50. doi: 10.1097/00003246-200006000-00061. PubMed 10890662 ↗

Study documents

PDFs hosted by the registry, ClinicalTrials.gov.

Individual participant data

Plan to share: No — No plan to share individual participant data with other researchers is planned

10

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Ran from Apr 2019 to primary completion in Mar 2020, 11 months.

Results were first posted in Jun 2023, 3 years 3 months after primary completion.

Of 19 interventional studies with no phase under Hypovolemia, half reached primary completion within 12 months of starting (the median); a quarter within 8 and three quarters within 20. This study's 11 months is close to that.

How these figures are worked out, and which studies are left out of them.

Registered
Apr 8, 2019
First posted
Apr 16, 2019
StartActual
Apr 8, 2019
Primary completionActual
Mar 12, 2020
CompletionActual
Mar 12, 2020
Results first posted
Jun 29, 2023
Last update
Jun 29, 2023
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Jun 29, 2023, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT03915587
Responsible party
Sponsor
First posted
Apr 16, 2019
Start date
Apr 8, 2019
Primary completion
Mar 12, 2020
Completion
Mar 12, 2020
Results posted
Jun 29, 2023
Last update
Jun 29, 2023

Study contacts

Sarkis Derderian, MD
principal investigator · Children's Hospital Colorado

Oversight

Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Jun 2023. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion