An observational study in Sudden Cardiac Death, Left Ventricular Dysfunction and Cardiac Event, sponsored by Zoll Medical Corporation. Completed at 2 sites in France. Per ClinicalTrials.gov, last updated 2024-12-09.
Sponsored by Zoll Medical Corporation · Observational
This post-market study is a prospective observational study evaluating the efficacy and safety of the LifeVest in real-life settings.
LifeVest has obtained coverage for 4 indications: (i) implantable cardiac defibrillator (ICD) removal due to cardiac device infections, (ii) a bridge to heart transplantation, (iii) in the early post-myocardial infarction period with left ventricular (LV) dysfunction (LVEF \<30%), (iv) and a recent coronary revascularization with LV dysfunction (LVEF \< 30%). There is no current comparator to LifeVest. In France, patients at high risk for sudden cardiac arrest (SCD) and waiting for heart transplant or ICD are hospitalized or discharged home without any particular ECG monitoring. LifeVest has been demonstrated to automatically detect and terminate rapid ventricular arrhythmias. Introduction of LifeVest into the current patient management would definitely improve the clinical outcomes and benefit the patients with temporary risk of SCD. The use of LifeVest was shown to be effective in protecting patients against SCD in the United States. However, the clinical impact of using LifeVest in France remains unassessed. Therefore we are planning to conduct a post-market study in France to provide efficacy and safety data for the reimbursement dossier renewal before April 2017.
363 studies on the registry are indexed under Death, Sudden, Cardiac; 84 are open to participants now.
This study's enrollment of 1,164 is above the median of 449 across 202 observational studies indexed under Death, Sudden, Cardiac.
Browse Death, Sudden, Cardiac studies →Zoll Medical Corporation is the lead sponsor of 36 studies on the registry; 6 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Residents of France All patient who were prescribed a LifeVest were approached for consent. In addition to those prospectively enrolled, patients who had completed LifeVest use prior to the study start were eligible for retrospective enrollment.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Patients receiving a LifeVest prescription in clinical routine for the following indications:
Exclusion Criteria:
Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
The share of blood the heart’s main pumping chamber (the left ventricle) pushes out with each beat, measured by an echocardiogram (heart ultrasound) or a MUGA scan. About 50–70% is normal.
LVEF (left ventricular ejection fraction)
The share of blood the heart’s main pumping chamber (the left ventricle) pushes out with each beat, measured by an echocardiogram (heart ultrasound) or a MUGA scan. About 50–70% is normal.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Patients who were prescribed and used a LifeVest in France
Device: Wearable Cardioverter Defibrillator
A non-invasive wearable cardioverter defibrillator detects and treats VT/VF without the need for bystander intervention
Also known as: LifeVest
Appropriate Shock Events Per Patient
Appropriate shock events were defined as at least one WCD therapy shock delivered during sustained (at least 30 seconds) ventricular tachycardia (VT) or ventricular fibrillation (VF). All shocks delivered within 24 hours of the first shock were considered one event. Adjudication of cardiac rhythms recorded during WCD therapy shocks was performed by 3 cardiologists familiar with device-recorded ECG tracings. For assessing incidence per 100 patient years, all appropriate shock events were considered.
Time frame: start to end of WCD use (3 months typical prescription)
Risk of Not Receiving Appropriate Shocks When Necessary
While wearing a WCD, how often does a shock not occur when 1) VF or a sustained (\> 30 seconds duration) VT occurs and 2) the response buttons are not used. Response button use is a surrogate for either a conscious patient or bystander presence.
Time frame: from start to end of WCD use (3 months typical prescription)
Inappropriate Shock Events Per Patient
Inappropriate shock events were defined as having at least one WCD therapy shock delivered when neither sustained (at least 30 seconds in duration) ventricular tachycardia (VT) or ventricular fibrillation (VF) was not present. All shocks delivered within 24 hours of the first shock were considered one event. Adjudication of cardiac rhythms recorded during WCD therapy shocks was performed by 3 cardiologists familiar with device-recorded ECG tracings. For assessing incidence per 100 patient years, all inappropriate shock events were considered.
Time frame: from start to end of WCD use (3 months typical prescription)
Survival During WCD Use
The number of patients that were alive at the end of the use of the Wearable Cardioverter Defibrillator (WCD) compared to the total population of patients prescribed the LifeVest
Time frame: from start to end of WCD use (3 months typical prescription)
Quality of Life Relative to Health State
Measured through the administration of a patient quality of life survey. This survey has been used previously in a published study of patient quality of life. Agreement score according to 5-point Likert scale (lower is better): 1. = Strongly Agree 2. = Agree 3. = Neither Agree nor Disagree 4. = Disagree 5. = Strongly Disagree
Time frame: Time of enrollment
Causes of Mortality
Mortality was adjudicated by 3 cardiologists into these categories: Cardiac, Vascular, or Non-cardiovascular.
Time frame: Duration of prescribed device use, up to 3 months.
Deaths While Wearing the WCD
Deaths during device use (i.e., physically wearing the WCD) were adjudicated by 3 cardiologists as initiated by asystole, initiated by ventricular fibrillation (VF), or having non-arrhythmic causes.
Time frame: Terminal event may have occurred any time during device use, up to 3 months. All data available up to 24 hours prior to the terminal event was used for adjudication.
All LifeVest users completing use between May 2014 and December 2016 were approached for retrospective enrollment. All patients prescribed a LifeVest January 2017 through March 2018 were approached for prospective enrollment. Recruitment took place in France at 88 cardiology centers.
| Milestone | WCD Users |
|---|---|
| Started | 1164 |
| Completed | 1157 |
| Not completed | 7 |
| Withdrew: Data not available. | 7 |
Appropriate shock events were defined as at least one WCD therapy shock delivered during sustained (at least 30 seconds) ventricular tachycardia (VT) or ventricular fibrillation (VF). All shocks delivered within 24 hours of the first shock were considered one event. Adjudication of cardiac rhythms recorded during WCD therapy shocks was performed by 3 cardiologists familiar with device-recorded ECG tracings. For assessing incidence per 100 patient years, all appropriate shock events were considered.
| Participants | All LifeVest Users |
|---|---|
| Patients not shocked for VT or VF | 1139 |
| Patients with 1 appropriate shock event | 17 |
| Patients with 2 appropriate shock events | 1 |
While wearing a WCD, how often does a shock not occur when 1) VF or a sustained (\> 30 seconds duration) VT occurs and 2) the response buttons are not used. Response button use is a surrogate for either a conscious patient or bystander presence.
| Recorded instances of VT/VF | Subjects With VT/VF |
|---|---|
| Appropriate shock for VT/VF | 19 |
| Response buttons used for tolerated VT/VF | 15 |
| Sustained VT, self-terminated before shock | 8 |
Inappropriate shock events were defined as having at least one WCD therapy shock delivered when neither sustained (at least 30 seconds in duration) ventricular tachycardia (VT) or ventricular fibrillation (VF) was not present. All shocks delivered within 24 hours of the first shock were considered one event. Adjudication of cardiac rhythms recorded during WCD therapy shocks was performed by 3 cardiologists familiar with device-recorded ECG tracings. For assessing incidence per 100 patient years, all inappropriate shock events were considered.
| Participants | All LifeVest Users |
|---|---|
| Patients without inappropriate shock events | 1149 |
| Patients with 1 inappropriate shock event | 8 |
The number of patients that were alive at the end of the use of the Wearable Cardioverter Defibrillator (WCD) compared to the total population of patients prescribed the LifeVest
| Participants | All LifeVest Users |
|---|---|
| Survival to end of prescribed WCD use (WCD wearing period) | 1132 |
| Death during WCD wearing period | 25 |
| Death after WCD wearing period | 1 |
Measured through the administration of a patient quality of life survey. This survey has been used previously in a published study of patient quality of life. Agreement score according to 5-point Likert scale (lower is better): 1. = Strongly Agree 2. = Agree 3. = Neither Agree nor Disagree 4. = Disagree 5. = Strongly Disagree
| units on a scale | Prospective Subjects |
|---|---|
| The LifeVest gives me peace of mind | 1.96 ± 0.9 |
| I don't worry as much because I know the LifeVest is protecting me | 2.17 ± 1 |
| I sleep significantly better knowing I am protected by the LifeVest | 2.51 ± 1.12 |
| I feel more confident returning to my normal daily activities when wearing the LifeVest | 2.11 ± 0.95 |
| LifeVest has given me the confidence to perform exercise or cardiac rehabilitation | 2.35 ± 1.03 |
| Wearing the LifeVest makes me take my condition seriously | 1.66 ± 0.9 |
| I take significantly better care of myself since being prescribed the LifeVest | 2.14 ± 1.06 |
| I follow lifestyle modification recommendations from my physician | 1.56 ± 0.68 |
| I would recommend LifeVest to family or friends with a similar medical condition | 1.67 ± 0.9 |
Mortality was adjudicated by 3 cardiologists into these categories: Cardiac, Vascular, or Non-cardiovascular.
| Participants | WCD Users |
|---|---|
| Cardiac cause | 17 |
| Vascular cause (includes stroke) | 1 |
| Non-cardiovascular cause | 6 |
| Unknown (no information available) | 1 |
Deaths during device use (i.e., physically wearing the WCD) were adjudicated by 3 cardiologists as initiated by asystole, initiated by ventricular fibrillation (VF), or having non-arrhythmic causes.
| Participants | Deaths While Wearing the WCD |
|---|---|
| Arrhythmic cause: primary asystole | 0 |
| Arrhythmic cause: VF (includes electrical storm) | 2 |
| Non-arrhythmic cause | 7 |
Collected over Data was collected until the end of the prescribed WCD use time (typically 3 months).. Non-serious events are listed at a 0% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| All LifeVest Users | 26/1,157 (2.2%) | 6/1,157 (0.5%) | 0/1,157 (0%) |
| Event | All LifeVest Users |
|---|---|
| Inappropriate shocksProduct Issues | 4/1157 |
| AllergySkin and subcutaneous tissue disorders | 1/1157 |
| Dyspnea with device alertsRespiratory, thoracic and mediastinal disorders | 1/1157 |
| Age, Continuous(years) | All LifeVest Users |
|---|---|
| Mean | 60.0 ± 11.5 |
| Sex: Female, Male(Participants) | All LifeVest Users |
|---|---|
| Female | 183 |
| Male | 974 |
| Race and Ethnicity Not Collected(Participants) | All LifeVest Users |
|---|
| Enrollment type(Participants) | All LifeVest Users |
|---|---|
| Prospective enrollment | 529 |
| Retrospective enrollment | 628 |
| BMI(kilograms per meter squared) | All LifeVest Users |
|---|---|
| Mean | 26.2 ± 5.0 |
| Resting HR(beats per minute) | All LifeVest Users |
|---|---|
| Mean | 80.0 ± 17.6 |
| Indication for use(Participants) | All LifeVest Users |
|---|---|
| ICD extraction due to infection | 119 |
| Bridge to heart transplant | 88 |
| Early after myocardial infarction with LVEF 30% or less | 647 |
| Early after coronary revascularization with LVEF 30% or less | 303 |
| LVEF (4 groups)(Participants) | All LifeVest Users |
|---|---|
| >45% | 41 |
| >35% and ≤45% | 48 |
| >25% and ≤35% | 483 |
| ≤25% | 577 |
2 further baseline measures are reported on the registry.
PDFs hosted by the registry, ClinicalTrials.gov.
This study is completed, as verified in Oct 2024. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.