To perform a comparative study of multi-site left ventricular pacing and cardiac resynchronization therapy effects on ventriculoarterial coupling and energy efficiency of the failing heart
Given that the main function of the cardiovascular system is to provide sufficient blood supply to tissues in order to ensure their normal and effective function, combined with the most efficient possible use of energy produced by ATP degradation, there has been a keen interest in elucidating the interplay between heart and vessels, critically affecting both.
In order to achieve these goals, it is thought that in healthy humans the cardiovascular system as a unity operates at a unique combination of parameters (arterial elastance, heart rate and left ventricular end systolic elastance) so as to:
Ventriculoarterial coupling (VAC) is a composite parameter, defined as the ratio of arterial elastance (Ea) to end systolic left ventricular elastance (Ees). Thus: VAC=Ea/Ees . It is a fundamental property of the cardiovascular system, integrating and assessing the interaction of all individual parameters of the ventricle (pump) and the arterial tree (afterload). Furthermore, VAC may assess both whether SW produced is maximal for a given contractility of the left ventricle (condition for maximization: VAC=1) and whether mechanical efficiency of the ventricle is optimal (optimization condition: VAC=0.5-0.7). Consequently, simultaneous optimization is not possible, and the cardiovascular system operates either at maximal output (as in healthy individuals at rest) or at optimal efficiency (healthy individuals at exercise). Multi-site pacing (MSP) of the left ventricle is a recently introduced technique with excellent studies' findings concerning echocardiographic parameters of ventricular function. Recently, the MultiPoint Pacing (MPP) IDE study showed that a specific choice of electrical dipole for the first left ventricular pulse and a close to simultaneous application of the two left ventricular pulses achieves a very high percent of clinical response (87%), with excellent patient safety. Subsequent studies confirmed these findings, reporting even higher NYHA response rates (95% vs 78% for conventional cardiac resynchronization therapy - CRT).
The underlying rationale lies in the better approximation of the normal sequence of left ventricular activation, through use of two, instead of a single, pulses. According to trial results, one can achieve, compared to conventional CRT, improved coordination between left ventricular segments, improved cardiac output and, possibly, tissue perfusion, and potentially reduction of arrhythmia propensity (mechanism similar to that of CRT). Thus, it would be interesting to study whether these can be independently confirmed by changes in VAC values. In heart failure, VAC values increase considerably due to increases in Ea as a result of the feedback loop regarding pressure (but not volume) maintenance. As a consequence, any reduction would move them closed to both 1 and the 0.5-0.7 area, yielding improvement in both SW maximization and efficiency optimization.
However, there are objective difficulties in achieving lege artis MSP (according to MPP-IDE study standards) given that two prerequisites must be met: 1. Interpolar distance for the first left ventricular pulse >30mm (i.e. non-sequential poles used), 2. Nearly simultaneous (Δt=5msec) left ventricular pulses and 3. Threshold of ≤3.5V@0.5msec.
Moreover, the first pulse should, ideally, be directed to the most delayed, compared to the normal activation sequence, viable myocardial segment, a feat not always possible due to electrode placing constraints. Obviously, presence of scar could alter the course and shape of the activation front and thus diminish its effects (similar to issues already discussed in the case of CRT).
Objective:
To perform a comparative study of multi-site left ventricular pacing and cardiac resynchronization therapy effects on ventriculoarterial coupling and energy efficiency of the failing heart
Hypothesis:
VAC values are improved (shift closer to unity/0.5-0.7 area) and work/efficiency increase with patients on MSP as compared to CRT pacing.
5,701 studies on the registry are indexed under Heart Failure; 1,220 are open to participants now.
This study's enrollment of 80 is below the median of 200 across 1,679 observational studies indexed under Heart Failure.
Browse Heart Failure studies →National and Kapodistrian University of Athens is the lead sponsor of 168 studies on the registry; 50 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Systolic heart failure with an ejection fraction of 35% or less, at NYHA functional class of II or worse, and a QRS displaying either LBBB morphology and >120msec or non-LBBB morphology and >150msec.
Exclusion Criteria:
Adult, consenting patients with any cardiomyopathy type and an existing I/IIa indication for a CRT-D device will receive a device with multi-site pacing capability. Initially, for 6 months, optimal, conventional, resynchronization therapy will be delivered. Following this, all patients will crossover to optimized multi-site pacing, and receive this therapy for 6 more months. Optimization of therapy will be determined based on maximization of cardiac output, i.e. maximization of left ventricular outflow tract velocity-time integral. Baseline measurements of serum creatinine and ventriculoarterial coupling will also be acquired.
Device: Activation of multi-site pacing capability on CRT devices
Instead of administering a single LV pulse at the most (electrically) delayed segment of the ventricle, multi-site pacing allows for a more detailed "sculpting" of the LV activation sequence. Based on the MPP-IDE study results, activation of the antero-lateral wall, or at least its most delayed segments, closely followed by a pulse to the apex and then by a right ventricular one will yield favorable results in terms of hemodynamics and clinical parameters. Programming features: * Interpolar distance for the first left ventricular pulse \>30mm (i.e. non-sequential poles used) * Nearly simultaneous (Δt=5msec) second left ventricular pulse and * Threshold of ≤3.5V@0.5msec
Improvement of ventriculoarterial coupling
Ventriculoarterial coupling value shifts closer to 1
Time frame: 6 months for each intervention (conventional CRT - MPP)
Improvement of energy efficiency
Energy efficiency improvement will be assessed by means of ventriculoarterial coupling value shifts closer to 0.7
Time frame: 6 months for each intervention (conventional CRT - MPP)
Improvement in renal function
Creatinine clearance (Cockcroft-Gault formula) increases
Time frame: 6 months for each intervention (conventional CRT - MPP)
Improvement in percent maximal stroke work
Calculated through use of ventriculoarterial coupling
Time frame: 6 months for each intervention (conventional CRT - MPP)
Plan to share: No
This study is completed, as verified in Sep 2020. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
National and Kapodistrian University of Athens