ClinicalTrials.gg
CompletedNCT02310230Updated Jun 25, 2024Results posted

An Evaluation of the Utility of the ExSpiron Respiratory Variation Monitor During Upper GI Endoscopy

An interventional study of ExSpiron Respiratory Variation Monitor in Biliary Tract Diseases, Stomach Neoplasms and Carcinoma, Pancreatic Ductal, sponsored by Donald Mathews. Completed at 1 site in United States. Open to participants aged 21 Years and older. Per ClinicalTrials.gov, last updated 2024-06-25.

Sponsored by Donald Mathews · Not applicable, Interventional, and Diagnostic

How we know

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Phase
Not applicable
Study type
Interventional
Enrollment
100
Allocation
Randomized
Ages
21 Years and older
Sex
All
01

Where this study stands

Counted across the records on this site · 2026-11-24

Among 15 studies in Biliary Tract Diseases with a site in United States — none open to participants now, none under way, none of them phase 3. Biliary Tract Diseases · United States.

Enrollment
Largest of 12
100 · median 78
Started
2014
with 1 other
No phase
3 of 15
none open now
Registered
4th of 15
by date
Lead sponsor
1 of 15
Donald Mathews

Among the 15 studies in Biliary Tract Diseases with a site in United States, this one has the largest enrollment of the 12 interventional ones that give a count, is one of 2 that started in 2014, is one of 3 with no phase, and was the 4th registered.

Biliary Tract Diseases

59 studies on the registry are indexed under Biliary Tract Diseases; 12 are open to participants now.

Browse Biliary Tract Diseases studies →

Open to participants now

None of the 15 studies in Biliary Tract Diseases with a site in United States is open now. The open studies under Biliary Tract Diseases, anywhere →

Who these studies take

Of the 2 Biliary Tract Diseases studies recruiting now, none set no upper age limit and none stop at 75 or younger; none take children under 18; 1 takes healthy volunteers.

This study: 21 Years and older · all sexes · no healthy volunteers.

The full table on the hub →

This tab as a report to print or save as PDF → With an account.

Was this useful?
Download CSVNarrow it in search →With an account: the 15 studies in Biliary Tract Diseases with a site in United States as a file.
02

Study summary

Purpose: To assess the utility of a new medical device that monitors a patient's breathing during medical procedures in which a patient is sedated, but not mechanically ventilated. In minor procedures, such as endoscopy (where the doctor examines a patient's digestive tract by a TV camera inserted through the mouth), patients do not require general anesthesia, in which a machine would take over their breathing while they are unconscious for surgery. However, during endoscopic procedures it is sometimes difficult for the anesthesiologist to monitor the patient's breathing-specifically, to monitor changes in breathing patterns and the adequacy of breathing. In endoscopy procedures, the room is darkened, and the patient's mouth is generally occupied by the endoscope. While the anesthesiologist can listen to the patient's breathing sounds with a stethoscope, this type of monitoring can only be done periodically, and there is limited ability to gauge the adequacy of ventilation. This study will use the ExSpiron Respiratory Volume Monitor (RVM), which measures non-invasive minute ventilation (MV), tidal volume (TV) and respiratory rate (RR), in patients undergoing an endoscopic procedure to provide additional information regarding the effects of clinical interventions such as drug administrations or airway maneuvers on the patient's respiratory status.

For patients who give informed consent, study participation means that they will have a PadSet consisting of 3 electrodes applied to the chest. Another component, a nasal cannula (a thin clear plastic tube that goes under the nose) will give patients supplemental oxygen, and is standard of care for endoscopy at UVM Medical Center. Patients will then be asked to breathe in and out of a portable spirometer (breath meter) for 30 seconds up to five times. This data will be compared to data recorded by the monitor to confirm that the monitor is recording accurately. The procedure will then go forward in the normal fashion. Patients will be randomly placed into one of two groups. In the first group during the procedure, the anesthesiologist will not be able to see the numbers (MV, TV, and RR) displayed screen of the monitor, so the data will not be used to guide the patient's clinical care. In the second group, the anesthesiologist will be able to see the RVM measurements of MV, TV, and RR to evaluate the effect of the interventions. Monitoring for both groups will continue in the recovery room, until discharge.

Read the detailed description

The purpose of this study is to assess the utility of the ExSpiron Respiratory Variation Monitor in patients undergoing an interventional procedure with anesthesia. The primary outcome measure will be the average minute ventilation of the patient during the procedure. This study will also examine the correlation between clinical interventions such as drug administrations or airway maneuvers with data from the monitor. Additionally the study may provide information about the ability of this new monitoring system to prevent hypoxemia during these procedures and to consider its utility, compared with capnography, to detect hypopnea.

Background Assessing the adequacy of ventilation during Monitored Anesthesia Care (MAC) and sedation is difficult. Agents used to provide procedural sedation can depress ventilatory drive and can interfere with airway patency, making it much more likely that the patient will hypoventilate and experience partial or complete obstruction to ventilation. [It is actually clinically easier to administer general anesthesia and insert an airway management device such as an endotracheal tube or a laryngeal mask airway , however given the superior recovery profile following sedation vs. general anesthesia, it is usually preferable to administer sedation, if the procedure can be performed with its use.] The American Society of Anesthesiologists maintains a database of closed malpractice claims which allows analysis of patterns of injury. A study considering claims from MAC/sedation cases found that 40% of the MAC claims resulted in death or brain damage, that 25% of the claims that were associated with oversedation/hypoventilation occurred in the endoscopy suite, and that 44% of the oversedation cases could have been prevented by better monitoring.1 A study comparing claims from remote locations (such as endoscopy) with operating rooms, found that 50% of remote location claims involving monitored anesthesia care, the proportion of claims for death was increased in remote location claims [54 vs. 29% (operating room claims)], respiratory damaging events were more common in remote location claims (44 vs. 20%), with inadequate oxygenation/ventilation the most common specific event (21 vs. 3% in operating room claims), and that remote location claims were more often judged as being preventable by better monitoring (32 vs. 8% for operating room claims).2

Upper endoscopy procedures, such as upper GI endoscopy and endoscopic retrograde cholangiopancreatography (ERCP), are particularly challenging where the anesthesiologist has limited access to the patient's head and is working in darkened rooms. Indeed investigators have found it very difficult to find a combination of propofol and short acting opioid that allows instrumentation of the esophagus without "intolerable ventilatory depression".3 Trying to assess ventilation by observing chest excursion or listening to ventilation with a stethoscope under these conditions can be difficult and can be done only periodically. Side-stream nasal cannula end-tidal capnography offers a possible solution, but the cannula can become dislodged and the proceduralist's endoscope often interferes with its function. End-tidal capnography in the non-intubated patient is very insensitive to changes in minute ventilation. At best, this technology can usually assess only the presence or absence of ventilation: there is limited ability to judge the adequacy of ventilation or to quantify it.4

The Study Device The ExSpiron Respiratory Volume Monitor (Respiratory Motion, Inc., Waltham MA) has FDA 501(k) clearance to be used for patient care. It consists of a proprietary PadSet that is placed on the chest in the mid-axillary line and connected to a free standing monitor. This monitor measures and analyzes the changes in thoracic electrical impedance that occur during the respiratory duty cycle. It is able to quantify respiratory rate and also determine tidal volume and minute ventilation as well as graphically display changes in lung volumes. It has been shown to accurately measure the effect of interventions (medications, airway maneuvers) in endoscopy patients.5 Preliminary work has determined optimal lead placement demonstrated excellent correlation to spirometry with an R value of 0.94-0.98 and excellent accuracy and precision.6

Preliminary work cited below has shown that the monitor provides information about the effect of medications and airway maneuvers on breathing patterns that is not available with routine monitoring. This study seeks to extend that knowledge by investigating whether using this information in real-time results in different respiratory measurements compared to patients with routine monitoring.

03

Conditions studied

  • Biliary Tract Diseases
  • Stomach Neoplasms
  • Carcinoma, Pancreatic Ductal
  • Barrett Esophagus

Keywords

  • Endoscopy, Upper
  • Minute Ventilation
  • Tidal Volume
  • Respiratory Rate
04

Who can participate

Ages eligible
21 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Subjects undergoing upper gastrointestinal endoscopy

Exclusion criteria

Exclusion Criteria:

  • Patients with a history of thoracotomy with resection of lung tissue
  • Patients with a history of severe (defined as an inability to climb a flight of stairs or /VC of less than 30% of predicted)
  • greater than 43.
Terms in these criteria, in plain language(3)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

COPD

Chronic obstructive pulmonary disease: long-term lung disease that makes it hard to breathe out, including emphysema and chronic bronchitis.

FEV1

From a breathing test (spirometry): how much air you can blow out in the first second. Often given as a percentage of what is expected for your age, height and sex.

BMI (body mass index)

Weight in kilograms divided by height in metres squared. Example: 70 kg and 1.75 m gives 70 ÷ (1.75 × 1.75) ≈ 22.9.

What counts as overweight or obesity differs between countries (some use lower cut-offs), so go by the range the study states.

05

Study design

Phase
Not applicable
Primary purpose
Diagnostic
Allocation
Randomized
Intervention model
Parallel assignment
Masking
Single (Care provider)
Enrollment
100 participants (actual)

Study arms

  • Other
    Blinded Group

    Data from the ExSpiron Respiratory Variation Monitor (minute ventilation, tidal volume, and respiratory rate) will not be displayed. The anesthesia provider will care for the patient in the usual manner.

    Device: ExSpiron Respiratory Variation Monitor

  • Experimental
    Monitor Group

    The ExSpiron Respiratory Variation Monitor will display continuous real-time measurements of minute ventilation, tidal volume, and respiratory rate, and the anesthesia provider will be instructed to utilize this information in the care of the patient as they deem appropriate.

    Device: ExSpiron Respiratory Variation Monitor

Interventions

  • DeviceExSpiron Respiratory Variation Monitor

    Non-invasive monitoring of minute ventilation, tidal volume, and respiratory rate.

06

What researchers measure

Primary outcomes

  1. Average Minute Ventilation in Patients Cared for Using the ExSpiron Respiratory Variation Monitor (RVM) Compared to Patients With Routine Monitoring in Patients Undergoing Upper Gastrointestinal Endoscopy

    Minute ventilation (or respiratory minute volume) is defined as the volume of air that can be inhaled or exhaled during one minute. The ExSpiron Respiratory Variation Monitor is able to quantify respiratory rate and also determine tidal volume and minute ventilation. The monitor measures a patient's baseline minute ventilation (MV) at the beginning of the case, and then compares subsequent minute volumes to that baseline, measured as a percentage of the baseline. In that respect, the minute ventilation (MV) values can often exceed 100 percent.

    Time frame: Duration of procedure: Approximately 60 minutes

Secondary outcomes

  1. Incidence of Airway Maneuvers Required to Maintain Ventilation in Each Group

    Time frame: Duration of procedure: Approximately 60 minutes

  2. Time With Low Minute Ventilation

    Percentage of procedure time spent with Minute Ventilation less than 40% of the Minute Ventilation Baseline. Minute ventilation (or respiratory minute volume) is defined as the volume of air that can be inhaled or exhaled during one minute. The ExSpiron Respiratory Variation Monitor is able to quantify respiratory rate and also determine tidal volume and minute ventilation.

    Time frame: Duration of Procedure: Approximately 60 Minutes

07

Results

Posted Jun 25, 2024

Participant flow

Participant flow — Overall Study
MilestoneBlinded GroupMonitor Group
Started5050
Completed4132
Not completed918
Withdrew: Physician decision13
Withdrew: Monitor baseline taken after induction27
Withdrew: Poor signal quality58
Withdrew: Case cancelled10

Outcome measures

PrimaryAverage Minute Ventilation in Patients Cared for Using the ExSpiron Respiratory Variation Monitor (RVM) Compared to Patients With Routine Monitoring in Patients Undergoing Upper Gastrointestinal Endoscopy

Minute ventilation (or respiratory minute volume) is defined as the volume of air that can be inhaled or exhaled during one minute. The ExSpiron Respiratory Variation Monitor is able to quantify respiratory rate and also determine tidal volume and minute ventilation. The monitor measures a patient's baseline minute ventilation (MV) at the beginning of the case, and then compares subsequent minute volumes to that baseline, measured as a percentage of the baseline. In that respect, the minute ventilation (MV) values can often exceed 100 percent.

Time frame:
Duration of procedure: Approximately 60 minutes
Reported as:
Mean · % of minute ventilation baseline
Average Minute Ventilation in Patients Cared for Using the ExSpiron Respiratory Variation Monitor (RVM) Compared to Patients With Routine Monitoring in Patients Undergoing Upper Gastrointestinal Endoscopy
% of minute ventilation baselineBlinded GroupMonitor Group
Average Minute Ventilation in Patients Cared for Using the ExSpiron Respiratory Variation Monitor (RVM) Compared to Patients With Routine Monitoring in Patients Undergoing Upper Gastrointestinal Endoscopy86.7 ± 5.9109.8 ± 9.3
SecondaryIncidence of Airway Maneuvers Required to Maintain Ventilation in Each Group
Time frame:
Duration of procedure: Approximately 60 minutes
Reported as:
Mean · one airway maneuver
Incidence of Airway Maneuvers Required to Maintain Ventilation in Each Group
one airway maneuverBlinded GroupMonitor Group
Incidence of Airway Maneuvers Required to Maintain Ventilation in Each Group2.2 (0 to 7)2.3 (0 to 8)
SecondaryTime With Low Minute Ventilation

Percentage of procedure time spent with Minute Ventilation less than 40% of the Minute Ventilation Baseline. Minute ventilation (or respiratory minute volume) is defined as the volume of air that can be inhaled or exhaled during one minute. The ExSpiron Respiratory Variation Monitor is able to quantify respiratory rate and also determine tidal volume and minute ventilation.

Time frame:
Duration of Procedure: Approximately 60 Minutes
Reported as:
Mean · percentage of procedure duration
Time With Low Minute Ventilation
percentage of procedure durationBlinded GroupMonitor Group
Time With Low Minute Ventilation15.3 ± 2.87.1 ± 1.4

Adverse events

Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Blinded Group—0/50 (0%)0/50 (0%)
Monitor Group—0/50 (0%)0/50 (0%)

Baseline characteristics

Age, Continuous
Age, Continuous(years)Blinded GroupMonitor GroupTotal
Mean52.9 ± 17.060.5 ± 11.056.2 ± 15.2
Sex: Female, Male
Sex: Female, Male(Participants)Blinded GroupMonitor GroupTotal
Female231639
Male181634
Height
Height(centimeters)Blinded GroupMonitor GroupTotal
Mean169.5 ± 9.7169.3 ± 12.0169.4 ± 10.8
Weight
Weight(kilograms)Blinded GroupMonitor GroupTotal
Mean82.1 ± 23.875.3 ± 17.179.1 ± 21.4
BMI
BMI(kg/m^2)Blinded GroupMonitor GroupTotal
Mean28.3 ± 6.926.2 ± 5.027.4 ± 6.2
Length of Procedure
Length of Procedure(minutes)Blinded GroupMonitor GroupTotal
Mean30.1 ± 1.937.2 ± 3.233.2 ± 1.8
Intra-Op Propofol
Intra-Op Propofol(mcg/kg/min)Blinded GroupMonitor GroupTotal
Mean239.7 ± 108.2241.4 ± 112.6240.4 ± 110.2
08

Study locations

1 site
  • University of Vermont Medical Center
    Burlington, Vermont 05401, United States
09

References and documents

Publications

  • Bhananker SM, Posner KL, Cheney FW, Caplan RA, Lee LA, Domino KB. Injury and liability associated with monitored anesthesia care: a closed claims analysis. Anesthesiology. 2006 Feb;104(2):228-34. doi: 10.1097/00000542-200602000-00005. PubMed 16436839 ↗
  • Metzner J, Posner KL, Domino KB. The risk and safety of anesthesia at remote locations: the US closed claims analysis. Curr Opin Anaesthesiol. 2009 Aug;22(4):502-8. doi: 10.1097/ACO.0b013e32832dba50. PubMed 19506473 ↗
  • LaPierre CD, Johnson KB, Randall BR, White JL, Egan TD. An exploration of remifentanil-propofol combinations that lead to a loss of response to esophageal instrumentation, a loss of responsiveness, and/or onset of intolerable ventilatory depression. Anesth Analg. 2011 Sep;113(3):490-9. doi: 10.1213/ANE.0b013e318210fc45. Epub 2011 Mar 17. PubMed 21415430 ↗
  • Holley K, Mathews D, Freeman J, Brayanov J, Schapiro H. The Use of a Respiratory Volume Monitor to Assess Ventilation Before & After Airway Maneuvers during Upper Endoscopy. Poster Presentation, American Society of Anesthesiologists, October 2013, San Francisco, CA
  • Voscopoulos C, Brayanov J, Ladd D, Lalli M, Panasyuk A, Freeman J. Special article: evaluation of a novel noninvasive respiration monitor providing continuous measurement of minute ventilation in ambulatory subjects in a variety of clinical scenarios. Anesth Analg. 2013 Jul;117(1):91-100. doi: 10.1213/ANE.0b013e3182918098. Epub 2013 Jun 3. PubMed 23733842 ↗
  • Freeman J, Lalli M, Yocum N, Panasyuk A, Panasyuk S, and Lew R, 328: Non-invasive Monitoring of Tidal Volume and Minute Ventilation in Non-intubated Patients. Critical Care Medicine, 2011. 39(12): p. 88 10.1097/01.ccm.0000408627.24229.88.

Individual participant data

Plan to share: No

10

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reached

Ran from Sep 2014 to primary completion in May 2016, 19 months.

Results were first posted in Jun 2024, 8 years 1 month after primary completion.

How these figures are worked out, and which studies are left out of them.

Registered
Sep 18, 2014
First posted
Dec 8, 2014
StartActual
Sep 22, 2014
Primary completionActual
May 9, 2016
CompletionActual
May 9, 2016
Results first posted
Jun 25, 2024
Last update
Jun 25, 2024
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Jun 25, 2024, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT02310230
Lead sponsor
Donald Mathews
Collaborators
Respiratory Motion, Inc.
Responsible party
Donald Mathews (Attending Anesthesiologist and Professor of Anesthesiology, University of Vermont) — Sponsor-investigator
First posted
Dec 8, 2014
Start date
Sep 22, 2014
Primary completion
May 9, 2016
Completion
May 9, 2016
Results posted
Jun 25, 2024
Last update
Jun 25, 2024

Study contacts

Donald M Mathews, M.D.
principal investigator · University of Vermont

Oversight

Data monitoring committee
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in May 2024. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion