ClinicalTrials.gg
CompletedNCT02111993ULVUpdated May 22, 2019Results posted

Evaluating Myocardial Injury for Defibrillation Threshold Testing Methods for ICD Implantation

An interventional study of Standard Defibrillation Testing and Upper Limit of Vulnerability Testing in Cardiomyopathy, Heart Failure and Cardiac Arrhythmia, sponsored by University of California, San Diego. Completed at 1 site in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2019-05-22.

Sponsored by University of California, San Diego · Not applicable, Interventional, and Other

How we know

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Phase
Not applicable
Study type
Interventional
Enrollment
60
Allocation
Randomized
Ages
18 Years and older
Sex
All
01

Where this study stands

Counted across the records on this site · 2026-11-26

Among 231 studies in Cardiomyopathies with a site in United States — 56 open to participants now, 69 under way, 6 of them phase 3. Cardiomyopathies · United States.

Enrollment
51st of 132
60 · median 42
Started
2011
with 2 others
No phase
51 of 231
6 open now
Registered
90th of 231
by date

Among the 231 studies in Cardiomyopathies with a site in United States, this one is the 51st largest by enrollment of the 132 interventional ones that give a count, is one of 3 that started in 2011, is one of 51 with no phase, and was the 90th registered.

Cardiomyopathies

556 studies on the registry are indexed under Cardiomyopathies; 153 are open to participants now.

Browse Cardiomyopathies studies →

Open to participants now

56 of the 231 studies in Cardiomyopathies with a site in United States are open now. Recruiting first, the newest first:

51 more — sign in →

All 56 open studies →

A study can be open while its sites in United States have not opened yet or have stopped. Only those recruiting at a site in United States →

Who these studies take

Of the 116 Cardiomyopathies studies recruiting now, 71 (61%) set no upper age limit and 18 (16%) stop at 75 or younger; 36 take children under 18; 22 take healthy volunteers.

This study: 18 Years and older · all sexes · no healthy volunteers.

The full table on the hub →

This tab as a report to print or save as PDF → With an account.

Was this useful?
Get alerts for this landscapeDownload CSVNarrow it in search →With an account: a weekly mail of what changed among 231 studies in Cardiomyopathies with a site in United States, or all of them as a file.
02

Study summary

The objective of this study is to evaluate myocardial injury, if any, as quantified by cardiac markers (Troponin-T) in defibrillation threshold (DFT) testing during implantation of implantable cardioverter defibrillators (ICDs) using the upper limit of vulnerability (ULV) method vs. standard defibrillation threshold method.

Read the detailed description

This is a prospective, randomized (1:1 DFT vs ULV testing), controlled clinical research trial. Patients undergoing ICD implant or device change out will be candidates for this study. Medical records for these patients will be screened prior to enrollment to ensure eligibility for the study. Each patient who is a candidate and who agrees to participate in the study will undergo the standard of care pre-procedure evaluation including a complete medical history evaluation, a complete physical examination and routine laboratory testing including chemistry panel, hematology panel, coagulation panel, a 12-lead electrocardiogram, and a urine pregnancy test (all females of child-bearing age). A candidate patient will be randomly assigned to either the standard DFT testing method group or the ULV testing method group.

For the standard DFT testing group, per routine standard of care, the patient will be admitted to the hospital. The study requires a pre-procedure blood test of 3 ml for Troponin-T (TnT) which will be drawn prior to any DFT testing. The blood sample will be sent to the University of California, San Diego (UCSD) central laboratory for processing. The patient will then undergo the scheduled ICD implant procedure in a standard fashion. The standard defibrillation threshold testing will be performed using a low energy shock delivered on T-wave to induce ventricular fibrillation (VF). If VF is induced, the first shock is 25 Joules (J). If success, we will wait 5 minutes and repeat the process. If the 2nd shock yields success, the DFT testing is complete and a 10J safety margin achieved (device shock energy is set at 35J or as clinically indicated determined by treating Cardiac Electrophysiologist). A back-up external defibrillator will always be immediately available in case transvenous shocks fail. Blood pressure will be monitored before and immediately after each shock. After 4 hours, 8 hours and 16 hours post DFT testing, blood samples (3 ml for Troponin-T) will be drawn and sent to the UCSD central laboratory for processing.

For the ULV testing group, per routine standard of care, the patient will be admitted to the hospital. The study requires a pre-procedure blood test (3 ml for Troponin-T.) which will be drawn prior to any ULV testing. The blood sample will be sent to the UCSD central laboratory for processing. The patient will then undergo the scheduled ICD implant procedure as is standard clinical care. The ULV testing is performed as follows: for single coil device, a 18J shock will be delivered at 0msec, +20 msec, -20msec and -40msec measured from the peak of the T-wave; for dual coil device, a 18J shock will be delivered at 0msec, -20msec, +20msec, and +40msec measured from the peak of the T-wave. If VF is not induced with any of the above mentioned shocks, the ULV will be considered as 18J and the ICD will be implanted with the first shock strength set at 30 joules or as clinically indicated determined by treating Cardiac Electrophysiologist. If VF is induced, a 25J rescue shock will be delivered as is standard and if successful, the device will be programmed to the maximum output of 35 joules or as clinically indicated determined by treating Cardiac Electrophysiologist. Blood pressure will be monitored before and immediately after each shock. After 4 hours, 8 hours and 16 hours post ULV testing, blood samples (3 ml for Troponin-T) will be drawn and sent to the UCSD central laboratory for processing.

Other than the blood tests for Troponin-T, there is no change in routine post-operative management. Patients will be followed every three months, as is standard management for patients with ICDs.

The TnT blood draws at 4-hour, 8-hour and 16-hour post implantation will not require longer hospital stay than is standard for patients undergoing ICD implantation. The results of Troponin-T measurement for the study patients will be reviewed by the cardiac electrophysiology (EP) research staff. If myocardial injury is noted (any TnT value >0.1 as per current assay at the UCSD laboratory), the patient will be monitored as is deemed clinically necessary by treating Cardiac Electrophysiologist, including possible repeat TnT measurements for evaluation of myocardial infarction. We will follow standard of care for management of myocardial injury as recommended by the American Heart Association and the American College of Cardiology. Elevated TnT measurements may delay hospitalization, if determined to be clinically necessary. If the repeat TnT measurement is \<0.1, then myocardial injury is most unlikely and the positive TnT value is likely due to DFT shocks.

03

Conditions studied

  • Cardiomyopathy
  • Heart Failure
  • Cardiac Arrhythmia

Keywords

  • Defibrillation testing
  • Upper Limit of Vulnerability
  • ULV
  • DFT
  • ICD
  • implantable cardioverter defibrillator
  • cardiomyopathy
  • cardiac arrhythmia
  • cardiac device testing
04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Patients eligible to receive an ICD implant or change out.
  • Patients with a single or dual coil right ventricle (RV) defibrillation lead located in the RV apex.
  • Patients greater than 18 years of age.
  • Patients willing and able to give .
  • Patients who have insurance which may cover additional hospital stay if myocardial injury resulted from testing

Exclusion criteria

Exclusion Criteria:

  • Patients with medical conditions that would limit study participation.
  • Patients that are pregnant
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

05

Study design

Phase
Not applicable
Primary purpose
Other
Allocation
Randomized
Intervention model
Parallel assignment
Masking
None (open label)
Enrollment
60 participants (actual)

Study arms

  • Active comparator
    Standard Defibrillation Testing

    Standard Defibrillation Threshold Testing is a procedure in which a low energy shock will be delivered to the heart to induce ventricular fibrillation (VF) followed by a rescue shock to restore sinus rhythm. Process is repeated after 5 minutes.

    Device: Standard Defibrillation Testing

  • Active comparator
    Upper Limit of VulnerabilityTesting

    Upper Limit of Vulnerability Testing is a procedure in which four 18J shocks will be delivered at specified intervals. If VF is induced, then a 25 J rescue shock will be delivered.

    Device: Upper Limit of Vulnerability Testing

Interventions

  • DeviceStandard Defibrillation Testing

    Once the ICD is implanted, Standard Defibrillation Threshold (DFT) testing will begin. A low energy shock will be delivered to the heart at the peak of the T wave to induce ventricular fibrillation (VF). Once VF is induced, a rescue shock of 25J will be delivered to restore sinus rhythm. This process will be repeated once more after 5 minutes.

    Also known as: ICD implantation

  • DeviceUpper Limit of Vulnerability Testing

    Once the ICD is implanted, Upper Limit of Vulnerability (ULV) testing will commence. An 18 J shock will be delivered will at different points from the t-wave. If VF is not induced with any of these shocks, then the ULV will be considered to be 18J. If VF is induced, then a 25 J rescue shock will be delivered.

    Also known as: ICD implantation

06

What researchers measure

Primary outcomes

  1. Myocardial Damage

    Myocardial damage will be evaluated by cardiac troponin (cTNT) lab collection before and then after ULV or DFT testing at 4 hours, 8 hours, and 20 hours.

    Time frame: 20 hours (hr)

Secondary outcomes

  1. Defibrillation Testings (DFT) vs. Upper Limit of Vulnerability (ULV) (VF Induced)

    the DFT group cardiac troponin (cTNT) values were compared to the subjects within the ULV group that required VF induction.

    Time frame: 20 hours (hr)

  2. Defibrillation Testing (DFT) Versus Upper Limit of Vulnerability (ULV) (Non-induced)

    The DFT group was compared to the ULV subgroup that did not require ventricular fibrillation (VF) induction. Cardiac troponin (cTNT) levels were analyzed.

    Time frame: 20 hours (hr)

07

Results

Posted May 22, 2019

Participant flow

Participant flow — Overall Study
MilestoneStandard Defibrillation TestingUpper Limit of VulnerabilityTesting
Started3030
Completed2526
Not completed54
Withdrew: Physician decision54

Outcome measures

PrimaryMyocardial Damage

Myocardial damage will be evaluated by cardiac troponin (cTNT) lab collection before and then after ULV or DFT testing at 4 hours, 8 hours, and 20 hours.

Time frame:
20 hours (hr)
Reported as:
Mean · ng/mL
Myocardial Damage
ng/mLStandard Defibrillation TestingUpper Limit of VulnerabilityTesting
Δ4 hr cTnT0.001 ± 0.0230.011 ± 0.066
Δ8 hr cTnT0.002 ± 0.0300.012 ± 0.072
Δ20 hr cTnT0.002 ± 0.0320.001 ± 0.048
Statistical analysis
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting · t-test, 2 sided · p = 0.02 (This p-value correlates to Δ4 hr cTnT)
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting · t-test, 2 sided · p = 0.03 (This p-value correlates to Δ8 hr cTnT.)
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting · t-test, 2 sided · p = 0.16 (This p-value correlates to Δ20 hr cTnT)
SecondaryDefibrillation Testings (DFT) vs. Upper Limit of Vulnerability (ULV) (VF Induced)

the DFT group cardiac troponin (cTNT) values were compared to the subjects within the ULV group that required VF induction.

Time frame:
20 hours (hr)
Reported as:
Mean · ng/mL
Defibrillation Testings (DFT) vs. Upper Limit of Vulnerability (ULV) (VF Induced)
ng/mLStandard Defibrillation TestingUpper Limit of VulnerabilityTesting (VF Induced)
Δ4 hr cTnT0.001 ± 0.0230.075 ± 0.086
Δ8 hr cTnT0.002 ± 0.0300.078 ± 0.093
Δ20 hr cTnT0.002 ± 0.0320.042 ± 0.057
Statistical analysis
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting (VF Induced) · t-test, 2 sided · p = 0.04 (This p-value correlates to Δ4 hr cTnT)
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting (VF Induced) · t-test, 2 sided · p = 0.06 (This p-value correlates to Δ8 hr cTnT)
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting (VF Induced) · t-test, 2 sided · p = 0.16 (This p-value correlates to Δ20 hr cTnT)
SecondaryDefibrillation Testing (DFT) Versus Upper Limit of Vulnerability (ULV) (Non-induced)

The DFT group was compared to the ULV subgroup that did not require ventricular fibrillation (VF) induction. Cardiac troponin (cTNT) levels were analyzed.

Time frame:
20 hours (hr)
Reported as:
Mean · ng/mL
Defibrillation Testing (DFT) Versus Upper Limit of Vulnerability (ULV) (Non-induced)
ng/mLStandard Defibrillation TestingUpper Limit of VulnerabilityTesting (No VF Induction)
Δ4 hr cTnT0.001 ± 0.0230.031 ± 0.031
Δ8 hr cTnT0.002 ± 0.0300.038 ± 0.038
Δ20 hr cTnT0.002 ± 0.0320.031 ± 0.031
Statistical analysis
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting (No VF Induction) · t-test, 2 sided · p = 0.17 (This p-value correlates to Δ4 hr cTnT)
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting (No VF Induction) · t-test, 2 sided · p = 0.19 (This p-value correlates to Δ8 hr cTnT)
  • Standard Defibrillation Testing vs Upper Limit of VulnerabilityTesting (No VF Induction) · t-test, 2 sided · p = 0.38 (This p-value correlates to Δ20 hr cTnT)

Adverse events

Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Standard Defibrillation Testing—0/30 (0%)0/30 (0%)
Upper Limit of VulnerabilityTesting—0/30 (0%)0/30 (0%)

Baseline characteristics

Age, Continuous
Age, Continuous(years)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Mean67 ± 1663 ± 1165 ± 14
Sex: Female, Male
Sex: Female, Male(Participants)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Female6511
Male192140
Region of Enrollment
Region of Enrollment(participants)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
United States252651
Body Mass Index (BMI)
Body Mass Index (BMI)(kg/m^2)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Mean28.2 ± 7.027.4 ± 4.627.8 ± 5.1
History of Hypertension (HTN)
History of Hypertension (HTN)(participants)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Number161228
Presence of Left Ventricular Hypertrophy (LVH)
Presence of Left Ventricular Hypertrophy (LVH)(participants)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Number91726
History of Ventricular Tachycardia (VT)
History of Ventricular Tachycardia (VT)(participants)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Number81018
History of Non-Ischemic Cardiomyopathy (NICM)
History of Non-Ischemic Cardiomyopathy (NICM)(participants)Standard Defibrillation TestingUpper Limit of VulnerabilityTestingTotal
Number268

9 further baseline measures are reported on the registry.

08

Study locations

1 site
  • UCSD Sulpizio Cardiovascular Center
    La Jolla, California 92093, United States
09

References and documents

Publications

  • Swerdlow CD, Shehata M, Chen PS. Using the upper limit of vulnerability to assess defibrillation efficacy at implantation of ICDs. Pacing Clin Electrophysiol. 2007 Feb;30(2):258-70. doi: 10.1111/j.1540-8159.2007.00659.x. PubMed 17338725 ↗
  • Day JD, Doshi RN, Belott P, Birgersdotter-Green U, Behboodikhah M, Ott P, Glatter KA, Tobias S, Frumin H, Lee BK, Merillat J, Wiener I, Wang S, Grogin H, Chun S, Patrawalla R, Crandall B, Osborn JS, Weiss JP, Lappe DL, Neuman S. Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations). Circulation. 2007 May 8;115(18):2382-9. doi: 10.1161/CIRCULATIONAHA.106.663112. Epub 2007 Apr 30. PubMed 17470697 ↗
  • Chen PS, Shibata N, Dixon EG, Martin RO, Ideker RE. Comparison of the defibrillation threshold and the upper limit of ventricular vulnerability. Circulation. 1986 May;73(5):1022-8. doi: 10.1161/01.cir.73.5.1022. PubMed 3698224 ↗
  • Chen PS, Feld GK, Kriett JM, Mower MM, Tarazi RY, Fleck RP, Swerdlow CD, Gang ES, Kass RM. Relation between upper limit of vulnerability and defibrillation threshold in humans. Circulation. 1993 Jul;88(1):186-92. doi: 10.1161/01.cir.88.1.186. PubMed 8319332 ↗
  • Hwang C, Swerdlow CD, Kass RM, Gang ES, Mandel WJ, Peter CT, Chen PS. Upper limit of vulnerability reliably predicts the defibrillation threshold in humans. Circulation. 1994 Nov;90(5):2308-14. doi: 10.1161/01.cir.90.5.2308. PubMed 7955188 ↗
  • Martin DJ, Chen PS, Hwang C, Gang ES, Mandel WJ, Peter CT, Swerdlow CD. Upper limit of vulnerability predicts chronic defibrillation threshold for transvenous implantable defibrillators. J Cardiovasc Electrophysiol. 1997 Mar;8(3):241-8. doi: 10.1111/j.1540-8167.1997.tb00786.x. PubMed 9083873 ↗
  • Birgersdotter-Green U, Undesser K, Fujimura O, Feld GK, Kass RM, Mandel WJ, Peter CT, Chen PS. Correlation of acute and chronic defibrillation threshold with upper limit of vulnerability determined in normal sinus rhythm. J Interv Card Electrophysiol. 1999 Jul;3(2):155-61. doi: 10.1023/a:1009825731592. PubMed 10387143 ↗
  • Swerdlow CD, Davie S, Ahern T, Chen PS. Comparative reproducibility of defibrillation threshold and upper limit of vulnerability. Pacing Clin Electrophysiol. 1996 Dec;19(12 Pt 1):2103-11. doi: 10.1111/j.1540-8159.1996.tb03285.x. PubMed 8994950 ↗
  • Swerdlow CD, Ahern T, Kass RM, Davie S, Mandel WJ, Chen PS. Upper limit of vulnerability is a good estimator of shock strength associated with 90% probability of successful defibrillation in humans with transvenous implantable cardioverter-defibrillators. J Am Coll Cardiol. 1996 Apr;27(5):1112-8. doi: 10.1016/0735-1097(95)00603-6. PubMed 8609329 ↗
  • Swerdlow CD. Implantation of cardioverter defibrillators without induction of ventricular fibrillation. Circulation. 2001 May 1;103(17):2159-64. doi: 10.1161/01.cir.103.17.2159. PubMed 11331256 ↗
  • Green UB, Garg A, Al-Kandari F, Ungab G, Tone L, Feld GK. Successful implantation of cardiac defibrillators without induction of ventricular fibrillation using upper limit of vulnerability testing. J Interv Card Electrophysiol. 2003 Feb;8(1):71-5. doi: 10.1023/a:1022304417889. PubMed 12652181 ↗
  • Swerdlow C, Shivkumar K, Zhang J. Determination of the upper limit of vulnerability using implantable cardioverter-defibrillator electrograms. Circulation. 2003 Jun 24;107(24):3028-33. doi: 10.1161/01.CIR.0000074220.19414.18. Epub 2003 Jun 16. PubMed 12810611 ↗
  • Shehata M, Belk P, Kremers M, Saba S, Cao J, Swerdlow CD. Automatic determination of timing intervals for upper limit of vulnerability using ICD electrograms. Pacing Clin Electrophysiol. 2008 Jun;31(6):691-700. doi: 10.1111/j.1540-8159.2008.01072.x. PubMed 18507541 ↗
10

Timeline

Tracking since Sep 25, 2026
Not marked actual or estimatedDates not marked actual or estimated

The record gives a start of Aug 2011, a primary completion of Apr 2015 and a completion of Jun 2015. It does not say whether these dates are actual or estimated.

Results were first posted in May 2019, 4 years 1 month after primary completion.

How these figures are worked out, and which studies are left out of them.

Registered
Apr 9, 2014
First posted
Apr 11, 2014
Start
Aug 2011
Primary completionActual
Apr 2015
CompletionActual
Jun 2015
Results first posted
May 22, 2019
Last update
May 22, 2019
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on May 22, 2019, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT02111993
Collaborators
Medtronic
Responsible party
Ulrika Birgersdotter-Green (Professor, University of California, San Diego) — Principal investigator
First posted
Apr 11, 2014
Start date
Aug 2011
Primary completion
Apr 2015
Completion
Jun 2015
Results posted
May 22, 2019
Last update
May 22, 2019

Study contacts

Ulrika Birgersdotter-Green, MD
principal investigator · University of California, San Diego

Oversight

Data monitoring committee
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in May 2019. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion