RF Surgical Sponge-Detecting System on the Function of Pacemakers and Implantable Cardioverter Defibrillators

An interventional study of RF Assure Scanning in Cardiomyopathy, Cardiac Arrhythmia and Heart Failure, sponsored by University of California, San Diego. Completed at 1 site in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2019-12-24.

Sponsored by University of California, San Diego · Not applicable, Interventional, and Other

How we know

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Phase
Not applicable
Study type
Interventional
Enrollment
50
Allocation
Not applicable
Ages
18 Years and older
Sex
All
01

Where this study stands

Counted across the records on this site · 2026-11-26

Among 231 studies in Cardiomyopathies with a site in United States — 56 open to participants now, 69 under way, 6 of them phase 3. Cardiomyopathies · United States.

Enrollment
56th of 132
50 · median 42
Started
2014
with 16 others
No phase
51 of 231
6 open now
Registered
90th of 231
by date

Among the 231 studies in Cardiomyopathies with a site in United States, this one is the 56th largest by enrollment of the 132 interventional ones that give a count, is one of 17 that started in 2014, is one of 51 with no phase, and was the 90th registered.

Cardiomyopathies

556 studies on the registry are indexed under Cardiomyopathies; 153 are open to participants now.

Browse Cardiomyopathies studies →

Open to participants now

56 of the 231 studies in Cardiomyopathies with a site in United States are open now. Recruiting first, the newest first:

51 more — sign in →

All 56 open studies →

A study can be open while its sites in United States have not opened yet or have stopped. Only those recruiting at a site in United States →

Who these studies take

Of the 116 Cardiomyopathies studies recruiting now, 71 (61%) set no upper age limit and 18 (16%) stop at 75 or younger; 36 take children under 18; 22 take healthy volunteers.

This study: 18 Years and older · all sexes · no healthy volunteers.

The full table on the hub →

This tab as a report to print or save as PDF → With an account.

Was this useful?
Get alerts for this landscapeDownload CSVNarrow it in search →With an account: a weekly mail of what changed among 231 studies in Cardiomyopathies with a site in United States, or all of them as a file.
02

Study summary

If a surgical sponge is mistakenly left inside a patient's body after a surgical procedure, it can cause a serious infection. To prevent this from happening, a new device has been developed that uses radiofrequency (RF) signals to detect the presence of surgical sponges inside the body. The device is now being used routinely to make sure that no sponges are left inside a patient at the end of an operation. However, the RF device has not been implemented in procedures for patients with cardiac implantable electronic devices (CIEDs). While the device is FDA approved for use, there is a theoretical concern that the radiofrequency signals used to detect the sponges will change the settings on the pacemaker or the defibrillator. Changing the settings on a pacemaker might make it pace the heart too quickly or too slowly, while changing the settings on a defibrillator might cause unnecessary shocks or prevent it from shocking the heart if the patient were to have cardiac arrest.

The purpose of this study is to test whether the radiofrequency device used to detect sponges can cause a clinically significant change to the settings on pacemakers and defibrillators. To minimize potential risk, the device will be tested only on patients who are having the pacemaker or defibrillator removed or replaced as part of their regular medical care, either because it is infected or because the battery has worn out. Before the pacemaker or defibrillator is removed, the settings will be carefully and completely recorded and the radiofrequency device will be used to scan the body for sponges as it would be done during normal operation.

After the pacemaker or defibrillator is taken out, the settings will again be recorded and compared to the settings before the scan. In a standard device removal procedure, no clinically significant change in CIED settings would be expected. If a new pacemaker or defibrillator is implanted in the patient, it will not be exposed to the detection device at all. We will also test whether the RF device has any effect on temporary pacemakers that patients may receive after open heart surgery. We plan to perform testing in a total of 50 patients, 40 with permanent pacemakers or defibrillators and 10 with temporary pacemakers.

Read the detailed description

Retained surgical items (e.g., sponges, needles, and instruments) are among the most frequently reported medical errors and occur in an estimated 1 in 5000 operations. A retained surgical item frequently causes a serious negative outcome, including re-operation (70-80% of cases), readmission or prolonged hospital stay (30-60% of cases), sepsis (43% of cases), or death (2% of cases). Retained surgical items are classed as "never events" by the Centers for Medicare and Medicaid Services (CMS), and CMS does not reimburse for additional medical care related to the treatment for such a complication.

Sponges are by far the most common retained items and are estimated to account for 50-70% of cases. The standard method to ensure that no sponge is left inside the body is manual counting. Manual counting, however, carries a sensitivity and specificity of only about 77% and 99%, respectively. In one study 62% of retained items were detected after the count was reported as normal. If the manual count suggests that a sponge has been retained, the standard method to detect it is by taking a radiograph. Radiographs, however, are time-consuming and in one large retrospective study failed to detect retained items in 33% of cases. The standard methods of detecting retained surgical sponges are clearly suboptimal.

Radiofrequency (RF) technology has recently been employed to improve the detection of retained surgical sponges. A radiofrequency chip is sewn into the fabric of the sponge, and a circular wand that emits a radiofrequency signal is passed over the patient. The wand also serves as an antenna that detects a return signal from the chip in the sponge. If a chip is detected it triggers an audio and a visual alarm on a console attached to the wand. In a study involving 210 patients, RF detection systems had 100% sensitivity and specificity for detecting retained sponges, even in morbidly obese patients. To minimize human error during the scanning process, a new system has been developed where the patient lies on a mat that emits the RF signal and serves as the antenna. In a separate study involving 203 patients, the system using the RF mat had a sensitivity and specificity of 98.5% and 100% respectively. To carry out these studies, sponges were placed underneath patients in a blinded manner. Although it would seem that RF detection systems are superior to manual counting, it should be emphasized that an RF detection system is not a substitute for manual counting, but rather an adjunct to manual counting. The RF Assure ® Detection System manufactured by RF Surgical System Inc. is now routinely used to detect retained sponges in the operating rooms at the University of California, San Diego (UCSD) and at \~200 other hospitals.

There has been some concern that the radio frequency signals emitted by the RF Assure® device might reprogram the settings on CIED such as pacemakers and internal cardiac defibrillators (ICDs), since CIEDs are commonly programmed remotely using RF signals. For this reason, the technology is not commonly used in surgeries with patients with CIEDs. In vitro testing was done in a controlled setting to assess the compatibility of the RF Surgical Detection Technology with permanent pacemakers and ICDs as well as temporary pacemakers, specifically the Boston Scientific Cognis, St. Jude Medical Promote, and Medtronic Virtuoso devices, finding no interference between the RF Assure® system and these devices. Although anecdotal experience and in vitro testing has indicated that the RF Assure® system does not affect CIEDs, it has not been established whether the RF Assure® system is suitable to use in patients with CIEDs in a clinical setting. The purpose of this study is to determine whether the RF Assure® device actually can reprogram CIEDs or cause them to malfunction when used in routine clinical practice. This question is of particular importance because of the rising number of surgical patients, especially cardiac patients, that have an implanted CIED.

03

Conditions studied

  • Cardiomyopathy
  • Cardiac Arrhythmia
  • Heart Failure
  • Chronic Thromboembolic Pulmonary Hypertension

Keywords

  • CIED
  • pacemaker
  • implantable cardioverter defibrillator
  • ICD
  • radiofrequency
  • cardiac device programming
  • RF sponge detection
  • RF assure
  • cardiac internal electronic devices
  • temporary pacing
  • pulmonary thromboendarterectomy
  • pacing interference
  • Arrhythmia
04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Patients must be at least 18 years of age
  • Patients must be willing and able to provide consent
  • Patients must be eligible for pacemaker/ICD removal or temporary pacemaker

Exclusion criteria

Exclusion Criteria:

  • Pregnancy
  • Inability or unwillingness to comply with the protocol
  • Medical condition that would limit study participation
05

Study design

Phase
Not applicable
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
50 participants (actual)

Study arms

  • Experimental
    RF Assure Scanning

    The patient's CIED will be interrogated prior to the study to obtain a baseline reading. The patient will be asked to lie down on the RF Assure® Detection Mat with a sponge placed underneath his or her shoulder. The RF Assure® mat and wand will be activated to detect the sponge. The sponge will be removed from underneath the patient's shoulder, and the RF system will be re-activated to obtain a clear reading. The patient's CIED will be re-interrogated to determine if the RF Assure system caused any changes to the CIED parameters or function.

    Device: RF Assure Scanning

Interventions

  • DeviceRF Assure Scanning

    CIED and temporary pacemaker patients will be scanned with RF surgical sponge detection wand and mat as well as with and without the RF sponge. The scanning will only be conducted on the device scheduled for removal. Patients will have device interrogated before and after scanning (for CIEDs) to determine if there have been any clinically significant changes in programming or settings due to scanning with RF technology. Vitals will be monitored and recorded before, during, and after scanning.

06

What researchers measure

Primary outcomes

  1. Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The Pacing Mode on the CIED was measured prior to RF scanning (with sponge) and after the scanning (after sponge removal) in order to evaluate if any significant changes in the pacing mode setting resulted. The following CIED modes were evaluated: DDD (dual chamber pacing, sensing, triggered and inhibited mode), VVI (ventricular pacing, sensing, and inhibited mode), DDI (dual pacing, sensing, and inhibited mode), AAI (atrial pacing, sensing, and inhibited mode). The number of patients' device mode switched between these settings was tabulated and is shown in the below table. Multiple post-scan assessments were not made.

    Time frame: Baseline and 15 minutes

  2. Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The patient's pacing polarity was measured prior to scanning with sponge and after scanning with sponge via device interrogation. The following pacing polarity measurements were evaluated: right atrium/right ventricle (RA/RV) bipolar polarity, left ventricle (LV) bipolar polarity, and left ventricle (LV) unipolar polarity. Note that not all study patients had LV leads implanted. Multiple post-scan assessments were not made.

    Time frame: Baseline and 15 minutes

  3. Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure

    The base rate on patients' devices was measured before scanning with sponge and after RF scanning with sponge. The median was determined and is presented below with standard deviation for both times. Multiple post-scan assessments were not made.

    Time frame: Baseline and 15 minutes

  4. Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The max tracking rate on the CIEDs was measured prior to scanning with the sponge and post scanning with sponge. The max tracking rate is the maximum atrial rate at which a pacemaker will deliver a ventricular pacing stimulus following each sensed atrial beat. Below, the median and standard deviation are presented. Please note that multiple post-scan assessments were not made for any patient.

    Time frame: Baseline and 15 minutes

  5. Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The atrio-ventricular delay was measured prior to scanning with sponge and post RF scanning with sponge. Please note that multiple post-scan assessments were not made for any of the patients presented here.

    Time frame: Baseline and 15 minutes

  6. Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    Of the patients enrolled, battery capacity changes were measured prior to and post RF scanning with sponge. Multiple post-scan assessments were not made for any of the patients presented here.

    Time frame: Baseline and 15 minutes

  7. Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The impedance values of right ventricle (RV), right atrium (RA), and left ventricle (LV) leads was measured prior to RF scanning with sponge and post RF scanning with sponge. Multiple post-scan assessments were not made for any patient represented here.

    Time frame: Baseline and 15 minutes

  8. Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The shock impedance changes were measured (right ventricle (RV) coil and superior vena cava (SVC) coil) prior to and post RF scanning with sponge. This was measured by performing a device interrogation in the electrophysiology (EP) lab. Multiple post-scan measurements were not taken for any of the patients presented here.

    Time frame: Baseline and 15 minutes

  9. Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    The P and R waves were measured via device interrogation prior to and post RF scanning with sponge. Multiple post-scan measurements were not made for any of the participants represented here.

    Time frame: Baseline and 15 minutes

  10. Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

    Via device interrogation, capture threshold changes (the minimum amount of electricity that the box has to emit to pace the heart) were measured prior to and post RF Scanning with sponge. Multiple post-scan measurements were not taken for any participants presented here.

    Time frame: Baseline and 15 minutes

07

Results

Posted Dec 24, 2019

Participant flow

Patients that required a pacemaker/ICD generator change or device implant were approached for participation in this study in both the Sulpizio Cardiovascular Center Clinic and Pre-procedural Unit. Patients undergoing PTE Surgery were also approached in these settings.

Participant flow — Overall Study
MilestoneRF Assure Scanning
Started50
Completed50
Not completed0

Outcome measures

PrimaryPacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The Pacing Mode on the CIED was measured prior to RF scanning (with sponge) and after the scanning (after sponge removal) in order to evaluate if any significant changes in the pacing mode setting resulted. The following CIED modes were evaluated: DDD (dual chamber pacing, sensing, triggered and inhibited mode), VVI (ventricular pacing, sensing, and inhibited mode), DDI (dual pacing, sensing, and inhibited mode), AAI (atrial pacing, sensing, and inhibited mode). The number of patients' device mode switched between these settings was tabulated and is shown in the below table. Multiple post-scan assessments were not made.

Time frame:
Baseline and 15 minutes
Reported as:
Number · participants
Pacing Mode Changes Between Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
participantsPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
Pacing Mode - DDD2929
Pacing Mode - VVI88
Pacing Mode - DDI22
Pacing Mode - AAI11
Statistical analysis
  • Pre-Scan Measures for PPM/ICDs Scanned With RF Assure vs Post-Scan Measures for PPM/ICDs Scanned With RF Assure · t-test, 2 sided · p = 1.0
PrimaryPacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The patient's pacing polarity was measured prior to scanning with sponge and after scanning with sponge via device interrogation. The following pacing polarity measurements were evaluated: right atrium/right ventricle (RA/RV) bipolar polarity, left ventricle (LV) bipolar polarity, and left ventricle (LV) unipolar polarity. Note that not all study patients had LV leads implanted. Multiple post-scan assessments were not made.

Time frame:
Baseline and 15 minutes
Reported as:
Number · participants
Pacing Polarity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
participantsPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
RA/RV (bipolar)4040
LV (bipolar)55
LV (unipolar)33
PrimaryBase Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure

The base rate on patients' devices was measured before scanning with sponge and after RF scanning with sponge. The median was determined and is presented below with standard deviation for both times. Multiple post-scan assessments were not made.

Time frame:
Baseline and 15 minutes
Reported as:
Median · bpm
Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure
bpmPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
Base Rate Measurement Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned w/ RF Assure60 ± 10.2760 ± 10.27
PrimaryMax Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The max tracking rate on the CIEDs was measured prior to scanning with the sponge and post scanning with sponge. The max tracking rate is the maximum atrial rate at which a pacemaker will deliver a ventricular pacing stimulus following each sensed atrial beat. Below, the median and standard deviation are presented. Please note that multiple post-scan assessments were not made for any patient.

Time frame:
Baseline and 15 minutes
Reported as:
Median · bpm
Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
bpmPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
Max Tracking Rate Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure130 ± 14.51130 ± 14.51
PrimaryAtrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The atrio-ventricular delay was measured prior to scanning with sponge and post RF scanning with sponge. Please note that multiple post-scan assessments were not made for any of the patients presented here.

Time frame:
Baseline and 15 minutes
Reported as:
Median · ms
Atrio-ventricular (AV) Delay Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
msPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
Sensed AV Delay230 ± 69.3230 ± 69.3
Paced AV Delay250 ± 68.3250 ± 68.3
PrimaryBattery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

Of the patients enrolled, battery capacity changes were measured prior to and post RF scanning with sponge. Multiple post-scan assessments were not made for any of the patients presented here.

Time frame:
Baseline and 15 minutes
Reported as:
Median · mV
Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
mVPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
Battery Capacity Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure2.745 ± 0.232.744 ± 0.23
PrimaryPacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The impedance values of right ventricle (RV), right atrium (RA), and left ventricle (LV) leads was measured prior to RF scanning with sponge and post RF scanning with sponge. Multiple post-scan assessments were not made for any patient represented here.

Time frame:
Baseline and 15 minutes
Reported as:
Median · ohms
Pacing Impedance Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
ohmsPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
RA Impedance500.1 ± 161.18495.1 ± 161.48
RV Impedance531.9 ± 135.96530.6 ± 135.08
LV Impedance592.8 ± 255.19610.1 ± 236.89
PrimaryShock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The shock impedance changes were measured (right ventricle (RV) coil and superior vena cava (SVC) coil) prior to and post RF scanning with sponge. This was measured by performing a device interrogation in the electrophysiology (EP) lab. Multiple post-scan measurements were not taken for any of the patients presented here.

Time frame:
Baseline and 15 minutes
Reported as:
Mean · ohms
Shock Impedance Changes With Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
ohmsPre-Scan Measures for ICDs Scanned With RF AssurePost-Scan Measures for ICDs Scanned With RF Assure
RV Coil51.30 ± 18.3651.5 ± 18.14
SVC Coil53.45 ± 17.5353.70 ± 17.13
PrimaryChanges in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

The P and R waves were measured via device interrogation prior to and post RF scanning with sponge. Multiple post-scan measurements were not made for any of the participants represented here.

Time frame:
Baseline and 15 minutes
Reported as:
Median · mV
Changes in P & R Wave Measurements in Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
mVPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
P Wave2.5 ± 1.382.469 ± 1.37
R Wave10.73 ± 5.2510.76 ± 5.29
PrimaryCapture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure

Via device interrogation, capture threshold changes (the minimum amount of electricity that the box has to emit to pace the heart) were measured prior to and post RF Scanning with sponge. Multiple post-scan measurements were not taken for any participants presented here.

Time frame:
Baseline and 15 minutes
Reported as:
Median · V
Capture Threshold Changes With Permanent Pacemakers (PPMs)/Implantable Cardiac Defibrillators (ICDs) Scanned With RF Assure
VPre-Scan Measures for PPM/ICDs Scanned With RF AssurePost-Scan Measures for PPM/ICDs Scanned With RF Assure
RA Threshold0.806 ± 0.350.819 ± 0.36
RV Threshold0.955 ± 0.490.972 ± 0.50
LV Threshold2.616 ± 1.162.320 ± 1.16

Adverse events

Collected over Patients were evaluated for adverse events for up to 2 days after their procedure. This was not a follow-up study, therefore, any events that may have occurred after patients were discharged from the hospital were not assessed for this study.. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
RF Assure Scanning—0/50 (0%)0/50 (0%)

Baseline characteristics

50 patients were enrolled in this trial. Of this population, 20 had pacemakers, 20 had ICDs, and 10 were undergoing temporary pacing during Pulmonary Thromboendarterectomy (PTE) surgery.

Age, Continuous
Age, Continuous(years)RF Assure Scanning
Pacemaker74.5 ± 17
ICD64 ± 17.4
PTE54.9 ± 13.1
Sex: Female, Male
Sex: Female, Male(Participants)RF Assure Scanning
Female16
Male34
Region of Enrollment
Region of Enrollment(participants)RF Assure Scanning
United States50
Body Mass Index (BMI)
Body Mass Index (BMI)(kg/m^2)RF Assure Scanning
Pacemaker26.3 ± 3.6
ICD29.8 ± 6.1
PTE30.8 ± 3.9
History of Coronary Artery Disease (CAD)
History of Coronary Artery Disease (CAD)(participants)RF Assure Scanning
Pacemaker6
ICD6
PTE1
History of Atrial Fibrillation (AF)
History of Atrial Fibrillation (AF)(participants)RF Assure Scanning
Pacemaker14
ICD5
PTE2
History of Diabetes Mellitus (DM)
History of Diabetes Mellitus (DM)(participants)RF Assure Scanning
Pacemaker3
ICD6
PTE3
History of Hypertension (HTN)
History of Hypertension (HTN)(participants)RF Assure Scanning
Pacemaker15
ICD10
PTE5

1 further baseline measures are reported on the registry.

08

Study locations

1 site
  • UCSD Sulpizio Cardiovascular Center
    La Jolla, California 92093, United States
09

References and documents

Publications

  • Cima RR, Kollengode A, Garnatz J, Storsveen A, Weisbrod C, Deschamps C. Incidence and characteristics of potential and actual retained foreign object events in surgical patients. J Am Coll Surg. 2008 Jul;207(1):80-7. doi: 10.1016/j.jamcollsurg.2007.12.047. Epub 2008 May 23. PubMed 18589366 ↗
  • Egorova NN, Moskowitz A, Gelijns A, Weinberg A, Curty J, Rabin-Fastman B, Kaplan H, Cooper M, Fowler D, Emond JC, Greco G. Managing the prevention of retained surgical instruments: what is the value of counting? Ann Surg. 2008 Jan;247(1):13-8. doi: 10.1097/SLA.0b013e3180f633be. PubMed 18156916 ↗
  • Gawande AA, Studdert DM, Orav EJ, Brennan TA, Zinner MJ. Risk factors for retained instruments and sponges after surgery. N Engl J Med. 2003 Jan 16;348(3):229-35. doi: 10.1056/NEJMsa021721. PubMed 12529464 ↗
  • Steelman VM, Alasagheirin MH. Assessment of radiofrequency device sensitivity for the detection of retained surgical sponges in patients with morbid obesity. Arch Surg. 2012 Oct;147(10):955-60. doi: 10.1001/archsurg.2012.1556. PubMed 23070411 ↗
  • Steelman VM. Sensitivity of detection of radiofrequency surgical sponges: a prospective, cross-over study. Am J Surg. 2011 Feb;201(2):233-7. doi: 10.1016/j.amjsurg.2010.05.001. PubMed 21266216 ↗
10

Timeline

Tracking since Sep 25, 2026
Not marked actual or estimatedDates not marked actual or estimated

The record gives a start of Mar 2014, a primary completion of Mar 2015 and a completion of May 2015. It does not say whether these dates are actual or estimated.

Results were first posted in Dec 2019, 4 years 9 months after primary completion.

How these figures are worked out, and which studies are left out of them.

Registered
Apr 9, 2014
First posted
Apr 11, 2014
Start
Mar 2014
Primary completionActual
Mar 2015
CompletionActual
May 2015
Results first posted
Dec 24, 2019
Last update
Dec 24, 2019
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Dec 24, 2019, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT02111980
Collaborators
RF Surgical Systems, Inc.
Responsible party
Ulrika Birgersdotter-Green (Professor of Medicine, Cardiac Electrophysiology, University of California, San Diego) — Principal investigator
First posted
Apr 11, 2014
Start date
Mar 2014
Primary completion
Mar 2015
Completion
May 2015
Results posted
Dec 24, 2019
Last update
Dec 24, 2019

Study contacts

Ulrika Birgersdotter-Green, MD
principal investigator · University of California, San Diego

Oversight

Data monitoring committee
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Dec 2019. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion