Real-World Validation of A Minimally Disruptive Ambulatory Blood Pressure Monitoring (ABPM) Huawei Smartwatch in Malaysia

An observational study in Hypertension, Vascular Disease and Cardiovascular (CV) Risk, sponsored by Sunway University. Enrolling by invitation at 1 site in Malaysia. Open to participants aged 18 Years and older, including healthy volunteers. Per ClinicalTrials.gov, last updated 2026-11-03.

Sponsored by Sunway University · Observational

Updated Nov 3, 2026Newly registeredGo to Updates ↓
Study type
Observational
Model
Ecologic or community
Time perspective
Cross-sectional
Enrollment
500
Ages
18 Years and older
Sex
All
01

Study summary

Background:

Ambulatory Blood Pressure Monitoring (ABPM) is the gold standard for diagnosing high blood pressure (hypertension) because it tracks blood pressure continuously across day and night. However, standard ABPM requires a cuff that inflates every 15 to 30 minutes during the day and every 30 to 60 minutes at night. These frequent inflations often cause discomfort, interrupt sleep, and cause patients to abandon testing.

Purpose:

This study aims to evaluate whether a new cuff-based Huawei smartwatch can provide a less-disturbing ABPM experience. Researchers will test whether a reduced-frequency measurement schedule (e.g., 7 readings: 4 awake, 3 asleep) can provide the same diagnostic accuracy as a conventional, full 24-hour ABPM protocol (typically 48 readings) while improving patient comfort.

Study Design \& Procedures:

Study Population:

Approximately 500 adult participants will be recruited over a 24-month study period, with a targeted subgroup of 50 participants selected for clinical validation. The subgroup will represent normotensive, pre-hypertensive, and hypertensive individuals across diverse age groups and ethnicities.

Monitoring \& Comparison: Participants in the validation group will undergo standard 24-hour ABPM at an accredited clinical facility alongside monitoring with the Huawei smartwatch. Data from the smartwatch will be used to compare a reduced-frequency reading schedule against full 24-hour ABPM recordings.

User Feedback: Participants will complete questionnaires assessing comfort, sleep quality, ease of use, and overall preference between the smartwatch and standard ABPM cuffs.

Goal:

The main goal is to determine if a reduced-frequency wearable monitoring protocol can maintain clinical diagnostic accuracy for blood pressure and sleep-dipping patterns while significantly reducing patient burden and sleep disruption.

02

Conditions studied

  • Hypertension
  • Vascular Disease
  • Cardiovascular (CV) Risk
  • Cardiovascular Disease

Keywords

  • Hypertension
  • High Blood Pressure
  • Ambulatory Blood Pressure Monitoring (ABPM)
  • Minimally-Disruptive ABPM
  • Reduced-Frequency Monitoring
  • Wearable Blood Pressure Devices
  • Cuff-based Smartwatch Blood Pressure
  • Nocturnal Blood Pressure Dipping
  • Sleep Disruption and Patient Comfort
  • Cardiovascular Risk Assessment
03

In context

Hypertension

6,845 studies on the registry are indexed under Hypertension; 976 are open to participants now.

This study's planned enrollment of 500 is above the median of 212 across 1,393 observational studies indexed under Hypertension.

Browse Hypertension studies →

Lead sponsor

Sunway University is the lead sponsor of 11 studies on the registry; 7 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
Yes
Sampling method
Non-probability sample

Study population

The study population comprises adult community-dwelling individuals residing in Malaysia recruited via the Huawei Research Platform. From this general screening pool (target N = 500), a targeted clinical validation subgroup (n = 50) will be prospectively selected to undergo dual ABPM and smartwatch evaluation.

To ensure clinical applicability and statistical validity in alignment with ISO 81060-2 and AAMI validation standards, the validation subgroup will be stratified into three distinct cohorts:

  • Normotensive Cohort (n = 10): Participants with blood pressure \< 135/85 mmHg.
  • Undiagnosed Elevated Blood Pressure Cohort (n = 20): Participants with untreated blood pressure > 135/85 mmHg.
  • Documented Hypertensive Cohort (n = 20): Participants with documented or self-reported history of history of hypertension.

Selection across all cohorts will maintain balanced demographic representation across age groups, biological sex, and major Malaysian ethnic populations (Malay, Chinese, Indian)

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Possesses the specific Huawei wearable required for the study
  • Aged ≥18 years
  • Malaysian citizen/permanent resident
  • Understands English, Malay, or Chinese
  • Able to download the required application on a compatible smartphone
  • Willing to wear devices continuously
  • Able to provide informed consent

Exclusion criteria

Exclusion Criteria:

  • Current pregnancy
  • Presence of implantable cardiac devices (e.g., permanent pacemakers, •implantable cardioverter-defibrillators [ICDs])
  • Uncontrolled cardiac arrhythmias (e.g., active atrial fibrillation, frequent premature ventricular contractions [PVCs])
  • Movement disorders or active tremors that interfere with blood pressure measurement or signal acquisition (e.g., severe Parkinson's disease, essential tremor)
  • Acute medical conditions (e.g., acute coronary syndromes, stroke/TIA within 3 months, acute decompensated heart failure, severe active infections, or hypertensive emergencies)
  • Resting blood pressure >=180/110 mmHg
05

Study design

Observational model
Ecologic or community
Time perspective
Cross-sectional
Enrollment
500 participants (estimated)
Target follow-up
2 Years
Patient registry
Yes

Groups and cohorts

  • Normotensive Cohort

    Adult participants with normal blood pressure (defined as blood pressure \< 135/85 mmHg; n = 10) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring.

  • Undiagnosed Elevated Blood Pressure Cohort

    Adult participants with undiagnosed elevated blood pressure (defined as blood pressure \> 135/85 mmHg; n = 20) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring.

  • Documented Hypertensive Cohort

    Adult participants with documented or self-reported hypertension (n = 20) who undergo simultaneous standard 24-hour ABPM and Huawei smartwatch monitoring.

  • Screening Population

    General adult participants recruited across Malaysia who enroll via the Huawei Research Platform app to provide continuous wearable ambulatory blood pressure and physiological data and complete baseline health assessments over the study period.

06

What researchers measure

Primary outcomes

  1. Mean Difference in Blood Pressure Measured by Reduced-Frequency Smartwatch Schedule Versus Standard 24-Hour Ambulatory Blood Pressure Monitor (ABPM)

    Measure Description: Mean blood pressure difference (bias) for 24-hour, daytime, and night-time periods calculated by subtracting reference 24-hour ABPM values (SpaceLabs/Mobil-O-Graph cuff monitor) from reduced-frequency Huawei smartwatch ABPM values (7-measurement schedule). Unit of Measure: mmHg

    Time frame: Month 13 through Month 18 (during clinical ABPM validation testing).

  2. Mean Absolute Error (MAE) of Blood Pressure Readings Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM

    Measure Description: The average absolute difference between blood pressure readings obtained from the reduced-frequency Huawei smartwatch schedule and the standard 24-hour cuff-based ABPM monitor. Unit of Measure: mmHg

    Time frame: Month 13 through Month 18.

  3. Root Mean Square Error (RMSE) of Blood Pressure Readings Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM

    Measure Description: The root mean square error evaluating agreement between blood pressure values recorded by the reduced-frequency Huawei smartwatch schedule and the standard 24-hour cuff-based ABPM monitor. Unit of Measure: mmHg

    Time frame: Month 13 through Month 18

  4. Intraclass Correlation Coefficient (ICC) of Blood Pressure Measurements Between Reduced-Frequency Smartwatch Schedule and Standard 24-Hour ABPM

    Measure Description: Agreement reliability calculated using absolute-agreement two-way random-effects ICC comparing reduced-frequency Huawei smartwatch blood pressure estimates against full 24-hour ABPM readings. Unit of Measure: Intraclass Correlation Coefficient (score ranging from -1.0 to +1.0)

    Time frame: Month 13 through Month 18

Secondary outcomes

  1. Limits of Agreement (Bland-Altman) for Blood Pressure Readings Between Huawei Smartwatch and Conventional Cuff ABPM

    Measure Description: Upper and lower 95% limits of agreement (mean difference +/- 1.96 x standard deviation) calculated via Bland-Altman analysis comparing simultaneous Huawei smartwatch and conventional cuff-based ABPM measurements. Unit of Measure: mmHg

    Time frame: Month 13 through Month 18

  2. Participant Comfort Score Evaluated via Post-Monitoring User Experience Survey

    Measure Description: Mean total score for participant-reported physical comfort derived from a standardized 5-point Likert scale questionnaire administered immediately after 24-hour monitoring. Unit of Measure: Score on a scale of 1 to 5 (where 1 = Extreme Discomfort and 5 = Complete Comfort)

    Time frame: Immediately following completion of 24-hour monitoring (Month 13 through Month 18)

  3. Number of Participants Indicating Preference for Smartwatch Monitoring Over Conventional Cuff ABPM

    Measure Description: Count of participants who express a preference for smartwatch-based monitoring over traditional arm-cuff ABPM on the post-monitoring user experience survey. Unit of Measure: Participant count

    Time frame: Immediately following completion of 24-hour monitoring (Month 13 through Month 18)

Other outcomes

  1. Cohen's Kappa Coefficient for Nocturnal Dipping Pattern Classification Concordance

    Measure Description: Inter-rater agreement (Cohen's Kappa) for categorizing participants into nocturnal dipping status categories (dipper vs. non-dipper) comparing reduced-frequency smartwatch schedules against standard full-day ABPM. Unit of Measure: Kappa Statistic (score ranging from -1.0 to +1.0)

    Time frame: Month 16 through Month 20.

  2. Minimum Number of Measurements Required to Maintain Diagnostic Accurac

    Measure Description: The minimum threshold count of active blood pressure inflations required per 24-hour period to maintain an ICC \>= 0.80 relative to standard 24-hour ABPM. Unit of Measure: Number of blood pressure inflations

    Time frame: Month 16 through Month 20.

07

Study locations

1 site
08

References and documents

Publications

  • Stergiou GS, Kario K, Aristizabal D, Beheiry H, Cho MC, Colombari DSA, Itoh H, Jafar TH, Jones E, Li Y, Romero CA, Steckelings UM, Williams B, Borghi C, Charchar F, Doku A, Dorobantu M, Flynn J, Khan N, Kollias A, Kreutz R, Mahmud A, Mancia G, Parati G, Poulter N, Reckelhoff J, Renna N, Shehab A, Schlaich M, Schutte A, Teo BW, Tomaszewski M, Touyz R, Wang J, Weber T, Whelton P, Zhang Y. 2026 International Society of Hypertension Global Hypertension Practice Guidelines. J Hypertens. 2026 Oct 20. doi: 10.1097/HJH.0000000000004451. Online ahead of print. PubMed 42909506 ↗
  • Stergiou GS, Menti A, Asmar R, Bilo G, Ntineri A, Palatini P, Parati G, Shennan A, Wang JG, Sierra A, Kollias A, Schutte AE; STRIDE BP and European Society of Hypertension Working Group on Blood Pressure Monitoring.. Updated practical guide for validation studies of automated cuff blood pressure monitors according to the Universal Standard by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018) and Amendments 1:2020 and 2:2024. J Hypertens. 2026 Oct 20. doi: 10.1097/HJH.0000000000004448. Online ahead of print. PubMed 42853222 ↗
  • Stergiou GS, Menti A, Asayama K, Kollias A, Parati G, Schutte AE, Wang J, Aristizabal D, Asmar R, Bilo G, Bothe T, Clark C, Kario K, Li Y, McManus R, Ohkubo T, Palatini P, Picone D, Shimbo D, Tsaganos T, De La Sierra A. Wearable wrist-watch type cuff oscillometric blood pressure monitors: consensus statement by the European Society of Hypertension Working Group on blood pressure monitoring. J Hypertens. 2026 Oct 1;44(10):1721-1731. doi: 10.1097/HJH.0000000000004415. Epub 2026 Aug 26. PubMed 42644439 ↗
  • Chia J, Bhatia KS, Mihailidou AS, Kanagaratnam LB. The Role of Ambulatory Blood Pressure Monitoring in Current Clinical Practice. Heart Lung Circ. 2022 Oct;31(10):1333-1340. doi: 10.1016/j.hlc.2022.06.670. Epub 2022 Aug 4. PubMed 35934633 ↗
  • Kawaja A, Moosa AS, Lee EKP, Phoon IKY, Ang ATW, Chang ZY, Lim ACA, Yap J, Huang W, Ng DX, Sng MY, Loh HY, Ng CJ. Comparison of patients' acceptance of cuff-based vs wearable 24-hour ambulatory blood pressure monitoring devices: A multi-method study. PLoS One. 2025 Nov 17;20(11):e0336961. doi: 10.1371/journal.pone.0336961. eCollection 2025. PubMed 41248137 ↗
  • McEvoy JW, McCarthy CP, Bruno RM, Brouwers S, Canavan MD, Ceconi C, Christodorescu RM, Daskalopoulou SS, Ferro CJ, Gerdts E, Hanssen H, Harris J, Lauder L, McManus RJ, Molloy GJ, Rahimi K, Regitz-Zagrosek V, Rossi GP, Sandset EC, Scheenaerts B, Staessen JA, Uchmanowicz I, Volterrani M, Touyz RM; ESC Scientific Document Group. 2024 ESC Guidelines for the management of elevated blood pressure and hypertension. Eur Heart J. 2024 Oct 7;45(38):3912-4018. doi: 10.1093/eurheartj/ehae178. No abstract available. PubMed 39210715 ↗
  • Stergiou GS, Menti A, Mariglis D, Kollias A. The quest for accurate wearable blood pressure monitors. Hypertens Res. 2026 Mar;49(3):1025-1029. doi: 10.1038/s41440-025-02410-w. Epub 2025 Nov 5. PubMed 41193706 ↗

Individual participant data

Plan to share: Yes — De-identified individual participant data (IPD) collected during this study will be shared with qualified external researchers. Specific shared data elements include: Demographics \& Baseline Characteristics: De-identified participant age, sex, body mass index, and baseline clinical blood pressure measurements. Wearable Sensor Data: Continuous physiological streams captured via the Huawei Research Platform, including raw photoplethysmography (PPG) signals, heart rate, heart rate variability, physical activity metrics, and sleep tracking parameters. Reference ABPM Data: Matched 24-hour, daytime, and night-time blood pressure and heart rate values, along with derived nocturnal dipping indices, obtained from clinical cuff-based ABPM and reduced-frequency Huawei smartwatch schedules. User Experience Data: De-identified survey scores on device comfort, ease of use, sleep disturbance, and monitoring preference.

Supporting information: Study protocol, Statistical analysis plan, Informed consent form

Time frame: Access Start Date: 6 months after the publication of the primary study results in a peer-reviewed journal. Access End Date: 5 years after initial publication.

Access criteria: Method of Access: Data will be made available via a secure, managed data repository (or directly from the Principal Investigator upon formal request). Qualified Requesters: Scientific researchers affiliated with recognized academic, clinical, or research institutions. Request Requirements: Requesters must submit a methodologically sound research proposal and a clear scientific rationale for secondary analysis. Data Use Agreement (DUA): Execution of a formal Data Use Agreement (DUA) is required to ensure data protection, mandate non-re-identification, and enforce secure handling. Compliance: All shared files will be fully de-identified in compliance with the Personal Data Protection Act (PDPA) Malaysia and institutional Ethics Committee guidelines.

09

Updates

1 registry update since Sep 25, 2026
Registered
First appeared on the registry. No changes since
Nov 3, 2026
Show all 1 update
  1. Nov 3, 2026
    First appeared on the registry

From the registry record's own update history. This site started tracking changes on Sep 25, 2026; for anything earlier, see the record history on ClinicalTrials.gov ↗

10

Registry details

Key details

Study ID
NCT07913607
Lead sponsor
Sunway University
Collaborators
Huawei Device Co., Ltd
Responsible party
YookChin Chia (Professor, Sunway University) — Principal investigator
First posted
Nov 3, 2026
Start date
Dec 15, 2026 (estimated)
Primary completion
Nov 30, 2028 (estimated)
Completion
Dec 31, 2028 (estimated)
Last update
Nov 3, 2026

Study contacts

Yook Chin Chia
principal investigator · Sunway University
Yee How Say, BSc PhD
principal investigator · Sunway University

Oversight

Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
View the source record on ClinicalTrials.gov ↗

Interested in this study?

Eligibility is decided by the study team. Share this record with your doctor or contact the team directly.

No contact was published for this record. The registry link below has the sponsor’s details.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion