Evaluation of Daily Disposable Toric Soft Lenses Manufactured With an Alternative Hydration Process

An interventional study of Investigational daily disposable, soft, toric lenses manufactured on 4th Generation Technology (4GT) lines with extended mini-stack hydration using propylene glycol (PG) as the hydration solvent. and ACUVUE® Oasys 1-Day for Astigmatism (AO1DfA) soft contact lenses manufactured on 3rd Generation Technology (3GT) legacy lines using isopropyl alcohol (IPA) as the hydration solvent. in Astigmatism, sponsored by Johnson & Johnson Vision Care, Inc.. Completed at 20 sites in United States. Open to participants aged 18 Years to 39 Years, including healthy volunteers. Per ClinicalTrials.gov, last updated 2026-11-04.

Sponsored by Johnson & Johnson Vision Care, Inc. (part of Johnson & Johnson) · Not applicable, Interventional, and Treatment

Phase
Not applicable
Study type
Interventional
Enrollment
281
Allocation
Randomized
Ages
18 Years to 39 Years
Sex
All
01

Study summary

This is a prospective, multi-site, dispensing, randomized, controlled, double-masked, bilateral wear, 2x2 crossover study to evaluate the safety and effectiveness of the test lens.

02

Conditions studied

  • Astigmatism

Browse trials for

03

In context

Astigmatism

417 studies on the registry are indexed under Astigmatism; 50 are open to participants now.

This study's enrollment of 281 is above the median of 55 across 325 interventional studies indexed under Astigmatism.

Browse Astigmatism studies →

Lead sponsor

Johnson & Johnson Vision Care, Inc. is the lead sponsor of 221 studies on the registry; 1 is open to participants now.

Of its 111 completed or terminated interventional studies of FDA-regulated products, 103 (93%) have results posted.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years to 39 Years
Sexes eligible
All
Accepts healthy volunteers
Yes

Eligibility criteria

Potential subjects must satisfy all of the following criteria to be enrolled in the study:

  1. Read, understand, and sign the STATEMENT OF and receive a fully executed copy of the form.
  2. Appear able and willing to adhere to the instructions set forth in this clinical protocol.
  3. Be between 18 and 39 (inclusive) years of age at the time of screening.
  4. By self-report, habitually wear soft contact lenses in both eyes in a daily reusable or daily disposable wear modality (i.e. not extended wear modality). Habitual wear is defined as a minimum of 6 hours of wear per day, for a minimum of 2 days per week during last 30 days.
  5. Possess a wearable pair of spectacles that provide correction for distance vision.
  6. In both eyes, have astigmatic refractive error suitable for correction with the toric contact lens powers available in this study:

    1. Sphere powers (DS) -1.50 through -4.00 (inclusive) in 0.25 steps
    2. Cylinder powers (DC) -0.75 and -1.25
    3. Axes (°) 170, 180, 10, 80, 90, 100
  7. Have best corrected monocular distance visual acuity of 20/30 or better in each eye.

Potential subjects who meet any of the following criteria will be excluded from participating in the study:

  1. Be currently pregnant or lactating.
  2. Be diabetic.
  3. Be currently using any ocular medications or have an ocular infection of any type.
  4. By self-report, have any ocular or systemic disease, allergies, infection, or use of medication that might contraindicate or interfere with contact lens wear, or otherwise compromise study endpoints, including infectious disease (e.g., hepatitis, tuberculosis), contagious disease (e.g., [HIV]), disease (e.g. rheumatoid arthritis, Sjögren's syndrome), or history of serious mental illness or seizures. See section 9.1 for additional details regarding excluded systemic medications.
  5. Have habitually worn rigid gas permeable (RGP) lenses, orthokeratology lenses, or hybrid lenses (e.g. SynergEyes, SoftPerm) within the past 6 months.
  6. Be currently wearing monovision or multifocal contact lenses.
  7. Be currently wearing lenses in an extended wear modality.
  8. Have a history of strabismus or amblyopia.
  9. Be an employee (e.g., Investigator, Coordinator, Technician) or immediate family member of an employee (including partner, child, parent, grandparent, grandchild or sibling of the employee or their spouse) of the clinical site.
  10. Have participated in a contact lens or lens care product clinical trial within 7 days prior to study enrollment.
  11. Have (grade 3 or higher on the FDA grading scale) slit lamp findings (e.g., corneal edema, neovascularization or staining, tarsal abnormalities or bulbar injection) or other corneal or ocular disease or abnormalities that contraindicate contact lens wear or may otherwise compromise study endpoints (including entropion, ectropion, chalazia, recurrent styes, glaucoma, history of recurrent corneal erosions, aphakia, moderate or above corneal distortion, herpetic keratitis).
  12. Have fluctuations in vision due to clinically significant dry eye or other ocular conditions.
  13. Have had or have planned (within the study period) any ocular or intraocular surgery (e.g., radial keratotomy, Photorefractive Keratectomy [PRK], Laser-Assisted in Situ Keratomileusis [LASIK], iridotomy, retinal laser photocoagulation, etc.).
Terms in these criteria, in plain language(5)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorised to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

Immunosuppression

A weakened immune system. It can come from an illness, such as HIV or some blood cancers, or from medicines that turn the immune system down, for example after an organ transplant or for an autoimmune disease; steroids at higher doses count.

Read whether the line is about a condition or about medicines, and which medicines and doses it means.

HIV

The human immunodeficiency virus, which weakens the immune system.

Many studies now accept people whose HIV is controlled on treatment, judged by the viral load and the CD4 count. Check the limits this study sets.

Autoimmune disease

A condition in which the immune system attacks the body’s own tissues, such as rheumatoid arthritis, lupus, psoriasis, type 1 diabetes or inflammatory bowel disease.

Immunotherapy studies often leave out active autoimmune disease, and some name exceptions, such as thyroid disease on replacement hormone. Check the exceptions this study lists.

Clinically significant

Judged by the study doctor to matter for your health or for the study. It is a doctor’s decision, so ask the study team if you are unsure whether something counts.

05

Study design

Phase
Not applicable
Primary purpose
Treatment
Allocation
Randomized
Intervention model
Crossover assignment
Masking
Double (Participant, Investigator)
Enrollment
281 participants (actual)

Study arms

  • Experimental
    Test/Control

    Eligible subjects will be randomized to the Test/Control sequence, to wear the Test Lens followed by the Control Lens bilaterally for approximately 1-week each in a daily disposable modality with a washout period of approximately 1-week between wear periods.

    Device: Investigational daily disposable, soft, toric lenses manufactured on 4th Generation Technology (4GT) lines with extended mini-stack hydration using propylene glycol (PG) as the hydration solvent. · Device: ACUVUE® Oasys 1-Day for Astigmatism (AO1DfA) soft contact lenses manufactured on 3rd Generation Technology (3GT) legacy lines using isopropyl alcohol (IPA) as the hydration solvent.

  • Experimental
    Control/Test

    Eligible subjects will be randomized to the Control/Test sequence, to wear the Control Lens followed by the Test Lens bilaterally for approximately 1-week each in a daily disposable modality with a washout period of approximately 1-week between wear periods.

    Device: Investigational daily disposable, soft, toric lenses manufactured on 4th Generation Technology (4GT) lines with extended mini-stack hydration using propylene glycol (PG) as the hydration solvent. · Device: ACUVUE® Oasys 1-Day for Astigmatism (AO1DfA) soft contact lenses manufactured on 3rd Generation Technology (3GT) legacy lines using isopropyl alcohol (IPA) as the hydration solvent.

Interventions

  • DeviceInvestigational daily disposable, soft, toric lenses manufactured on 4th Generation Technology (4GT) lines with extended mini-stack hydration using propylene glycol (PG) as the hydration solvent.

    Test Lens

  • DeviceACUVUE® Oasys 1-Day for Astigmatism (AO1DfA) soft contact lenses manufactured on 3rd Generation Technology (3GT) legacy lines using isopropyl alcohol (IPA) as the hydration solvent.

    Control Lens

06

What researchers measure

Primary outcomes

  1. Distance Monocular logMAR Visual Acuity

    Distance Monocular visual acuity was measured on a logMAR (Logarithm of Minimal Angle of Resolution) scale under high luminance high contrast condition. Visual acuity is assessed using ETDRS (Early Treatment Diabetic Retinopathy Study) charts. Letter-by-letter results calculated the visual performance score for each chart read. LogMAR scores closer to zero, or below zero, indicate better visual acuity. A logMAR visual performance score of 0.0 is equivalent to Snellen visual acuity of 20/20.

    Time frame: 1-Week Follow-up

  2. Proportion of Eyes With Absolute Rotation Less Than or Equal to 10 Degrees

    Absolute rotation was assessed for each subject eye using a slit lamp 15 minutes after lens insertion. Acceptable absolute rotation was dichotomized into a binary response where Y=1 if absolute rotation was less than or equal to 10 degrees and Y=0 otherwise (absolute rotation greater than 10 degrees). The proportion of eyes with absolute rotation less than or equal to 10 degrees was reported for each lens.

    Time frame: 15-Minutes after Lens Insertion

  3. Proportion of Eyes With Rotational Stability Less Than or Equal to 5 Degrees

    Rotational Stability was assessed for each subject eye using a slit lamp 15 minutes after lens insertion. Rotational Stability was dichotomized into a binary response where Y=1 if Rotational Stability was less than or equal to 5 degrees and Y=0 otherwise (Rotational Stability greater than 5 degrees). The proportion of eyes with Rotational Stability less than or equal to 5 degrees was reported for each lens.

    Time frame: 15-minutes after lens insertion

  4. Proportion of Eyes With Grade 3 or Higher Slit Lamp Findings

    Slit Lamp Findings (SLF) were assessed using a biomicroscope and were graded using the FDA grading scale (Grade: 0, 1,2, 3 and 4) with grade 0 represents the absence of findings and 1 to 4 representing successively worse findings (i.e. Grade 1 = trace, Grade 2 = Mild, Grade 3 = moderate and Grade 4 = severe). This was performed on each subject eye at every study visit (baseline, unscheduled visits and 1-week follow-up). The data was then dichotomized into two groups. Those with grade 3 or higher and those with grade 2 or lower. The proportion of eyes with SLF with grade 3 or higher was reported for each lens.

    Time frame: Up to 1-Week Follow-up

  5. Proportion of Eyes With Acceptable Lens Fitting

    Contact lens fitting acceptance was assessed for each subject eye using a biomicroscope post lens insertion at the 1-week follow-up. Lens fit was a binary variable where acceptable lens fit=1 and unacceptable lens fit=0. The proportion of eyes with acceptable lens fit was reported for each lens.

    Time frame: Up to 1-Week Follow-up

Secondary outcomes

  1. CLUE Vision Score

    Subjective vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.

    Time frame: 1-Week Follow-up

  2. CLUE Comfort Score

    Subjective comfort was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.

    Time frame: 1-Week Follow-up

  3. CLUE Handling Score

    Subjective handling was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.

    Time frame: 1-Week Follow-up

07

Results

Posted Sep 15, 2026

Participant flow

Period 1
Participant flow — Period 1
MilestoneTest (Senofilcon A)/Control (Senofilcon A)Control (Senofilcon A)/ Test (Senofilcon A)
Started135139
Completed135138
Not completed01
Withdrew: Subject no longer meets eligibility criteria01
Period 2
Participant flow — Period 2
MilestoneTest (Senofilcon A)/Control (Senofilcon A)Control (Senofilcon A)/ Test (Senofilcon A)
Started135138
Completed134138
Not completed10
Withdrew: Protocol deviation wrong left lens cylinder power dispensed10

Outcome measures

PrimaryDistance Monocular logMAR Visual Acuity

Distance Monocular visual acuity was measured on a logMAR (Logarithm of Minimal Angle of Resolution) scale under high luminance high contrast condition. Visual acuity is assessed using ETDRS (Early Treatment Diabetic Retinopathy Study) charts. Letter-by-letter results calculated the visual performance score for each chart read. LogMAR scores closer to zero, or below zero, indicate better visual acuity. A logMAR visual performance score of 0.0 is equivalent to Snellen visual acuity of 20/20.

Time frame:
1-Week Follow-up
Reported as:
Mean · logMAR
Distance Monocular logMAR Visual Acuity
logMARTest (Senofilcon A)Control (Senofilcon A)
Distance Monocular logMAR Visual Acuity-0.099 ± 0.0724-0.099 ± 0.0719
Statistical analysis
  • Test (Senofilcon A) vs Control (Senofilcon A) · Linear Mixed Model · p = <0.0001 (The 1-sided p-value was reported. a one-sided alpha level of 0.025 was used.) · Least square mean: 0.0001 · 95% CI -0.0047 to 0.0050The between within method was used for the calculation of the denominator degrees of freedom.
PrimaryProportion of Eyes With Absolute Rotation Less Than or Equal to 10 Degrees

Absolute rotation was assessed for each subject eye using a slit lamp 15 minutes after lens insertion. Acceptable absolute rotation was dichotomized into a binary response where Y=1 if absolute rotation was less than or equal to 10 degrees and Y=0 otherwise (absolute rotation greater than 10 degrees). The proportion of eyes with absolute rotation less than or equal to 10 degrees was reported for each lens.

Time frame:
15-Minutes after Lens Insertion
Reported as:
Number · Proportion of eyes
Proportion of Eyes With Absolute Rotation Less Than or Equal to 10 Degrees
Proportion of eyesTest (Senofilcon A)Control (Senofilcon A)
Proportion of Eyes With Absolute Rotation Less Than or Equal to 10 Degrees0.98710.9852
Statistical analysis
  • Test (Senofilcon A) vs Control (Senofilcon A) · Bayesian Beta-binomial Model · Mean central posterior proportion: 0.0016 · 95% CI -0.0165 to 0.0195
PrimaryProportion of Eyes With Rotational Stability Less Than or Equal to 5 Degrees

Rotational Stability was assessed for each subject eye using a slit lamp 15 minutes after lens insertion. Rotational Stability was dichotomized into a binary response where Y=1 if Rotational Stability was less than or equal to 5 degrees and Y=0 otherwise (Rotational Stability greater than 5 degrees). The proportion of eyes with Rotational Stability less than or equal to 5 degrees was reported for each lens.

Time frame:
15-minutes after lens insertion
Reported as:
Number · Proportion of eyes
Proportion of Eyes With Rotational Stability Less Than or Equal to 5 Degrees
Proportion of eyesTest (Senofilcon A)Control (Senofilcon A)
Proportion of Eyes With Rotational Stability Less Than or Equal to 5 Degrees11
Statistical analysis
  • Test (Senofilcon A) vs Control (Senofilcon A) · Bayesian Beta-binomial Model · Mean central posterior proportion: -0.0001 · 95% CI -0.0106 to 0.0102
PrimaryProportion of Eyes With Grade 3 or Higher Slit Lamp Findings

Slit Lamp Findings (SLF) were assessed using a biomicroscope and were graded using the FDA grading scale (Grade: 0, 1,2, 3 and 4) with grade 0 represents the absence of findings and 1 to 4 representing successively worse findings (i.e. Grade 1 = trace, Grade 2 = Mild, Grade 3 = moderate and Grade 4 = severe). This was performed on each subject eye at every study visit (baseline, unscheduled visits and 1-week follow-up). The data was then dichotomized into two groups. Those with grade 3 or higher and those with grade 2 or lower. The proportion of eyes with SLF with grade 3 or higher was reported for each lens.

Time frame:
Up to 1-Week Follow-up
Reported as:
Number · Proportion of Eyes
Proportion of Eyes With Grade 3 or Higher Slit Lamp Findings
Proportion of EyesTest (Senofilcon A)Control (Senofilcon A)
Proportion of Eyes With Grade 3 or Higher Slit Lamp Findings00.002
Statistical analysis
  • Test (Senofilcon A) · Bayesian Beta-binomial Model · Mean central posterior proportion: 0.0012 · 95% CI 0.0000 to 0.0061
PrimaryProportion of Eyes With Acceptable Lens Fitting

Contact lens fitting acceptance was assessed for each subject eye using a biomicroscope post lens insertion at the 1-week follow-up. Lens fit was a binary variable where acceptable lens fit=1 and unacceptable lens fit=0. The proportion of eyes with acceptable lens fit was reported for each lens.

Time frame:
Up to 1-Week Follow-up
Reported as:
Number · Proportion of Eyes
Proportion of Eyes With Acceptable Lens Fitting
Proportion of EyesTest (Senofilcon A)Control (Senofilcon A)
Proportion of Eyes With Acceptable Lens Fitting11
Statistical analysis
  • Test (Senofilcon A) · Bayesian Beta-binomial Model · Mean central posterior proportion: 0.9924 · 95% CI 0.9840 to 0.9979
SecondaryCLUE Vision Score

Subjective vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.

Time frame:
1-Week Follow-up
Reported as:
Mean · Units on a scale
CLUE Vision Score
Units on a scaleTest (Senofilcon A)Control (Senofilcon A)
CLUE Vision Score67.57 ± 19.84469.76 ± 18.858
SecondaryCLUE Comfort Score

Subjective comfort was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.

Time frame:
1-Week Follow-up
Reported as:
Mean · Units of a scale
CLUE Comfort Score
Units of a scaleTest (Senofilcon A)Control (Senofilcon A)
CLUE Comfort Score61.78 ± 23.80865.33 ± 22.450
SecondaryCLUE Handling Score

Subjective handling was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.

Time frame:
1-Week Follow-up
Reported as:
Mean · Units of a scale
CLUE Handling Score
Units of a scaleTest (Senofilcon A)Control (Senofilcon A)
CLUE Handling Score69.08 ± 22.19370.45 ± 20.367

Adverse events

Collected over Throughout the duration of the study; The study duration for each subject was approximately 21 days.. Non-serious events are listed at a 5% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Test (Senofilcon A)0/273 (0%)0/273 (0%)0/273 (0%)
Control (Senofilcon A)0/272 (0%)0/272 (0%)0/272 (0%)

Baseline characteristics

All subjects dispensed at least one study lens.

Age, Continuous
Age, Continuous(Years)Overall Reporting Group
Mean29.6 ± 6.16
Sex: Female, Male
Sex: Female, Male(Participants)Overall Reporting Group
Female171
Male103
Race (NIH/OMB)
Race (NIH/OMB)(Participants)Overall Reporting Group
American Indian or Alaska Native3
Asian15
Native Hawaiian or Other Pacific Islander3
Black or African American17
White231
More than one race5
Unknown or Not Reported0
Region of Enrollment
Region of Enrollment(Participants)Overall Reporting Group
United States274
08

Study locations

20 sites
  • James R. Dugue, Optometrist
    Mission Viejo, California 92691, United States
  • Scripps Poway Eyecare & Optometry
    San Diego, California 92131, United States
  • Stam & Associates Eye Care
    Jacksonville, Florida 32256, United States
  • Pearson Research Center PA
    Longwood, Florida 32779, United States
  • Maitland Vision Centers - North Orlando Ave
    Maitland, Florida 32751, United States
  • Mid-State Eye
    Clinton, Illinois 61727, United States
  • Franklin Park Eye Center, P.C.
    Franklin Park, Illinois 60131, United States
  • Kannarr Eye Care - 101 North Broadway
    Pittsburg, Kansas 66762, United States
  • Complete Eye Care of Medina
    Medina, Minnesota 55340, United States
  • Advanced Eyecare - Raytown, MO
    Raytown, Missouri 64133, United States
  • The Koetting Associates
    St Louis, Missouri 63144, United States
  • ABQ Eye Care
    Albuquerque, New Mexico 87109, United States
  • Spectrum Eye Care
    Jamestown, New York 14750, United States
  • Sacco Eye Group
    Vestal, New York 13850, United States
  • ProCare Vision Centers
    Granville, Ohio 43023, United States
  • Professional Vision Care Inc. - Westerville
    Westerville, Ohio 43081, United States
  • Optometry Group, PLLC
    Memphis, Tennessee 38111, United States
  • Total Eye Care
    Memphis, Tennessee 38119, United States
  • Tyler Eye Associates
    Tyler, Texas 75703, United States
  • Clarke EyeCare Center
    Wichita Falls, Texas 76308, United States
09

References and documents

Study documents

PDFs hosted by the registry, ClinicalTrials.gov.

Individual participant data

Plan to share: Yes — Johnson \& Johnson Medical Device Companies have an agreement with the Yale Open Data Access (YODA) to serve as the independent review panel for evaluation of requests for clinical study reports and participant level data from investigators and physicians for scientific research that will advance medical knowledge and public health. Requests for access to the study data can be submitted through the YODA Project site at http://yoda.yale.edu

Data sharing details ↗ · yoda.yale.edu

10

Timeline

2 registry updates since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Ran from Apr 2025 to primary completion in Jul 2025, 2 months.

Results were first posted in Sep 2026, 1 year 2 months after primary completion.

Of 106 industry-sponsored interventional studies with no phase under Astigmatism, half reached primary completion within 7 months of starting (the median), counting those still running; a quarter within 2 and three quarters within 19. This study's 2 months is below that.

Counting every sponsor, half of 183 studies reached it within 12 months; a quarter within 4 and three quarters within 26.

How these figures are worked out, and which studies are left out of them.

Registered
Apr 22, 2025
First posted
May 13, 2025
StartActual
Apr 25, 2025
Primary completionActual
Jul 14, 2025
CompletionActual
Jul 14, 2025
Results first posted
Sep 15, 2026
Last update
Nov 4, 2026
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
Minor edits
Nothing that changes what the study is or who can join. Edited: verification date
2 updates, last Nov 4, 2026
Show all 2 updates
  1. Nov 4, 2026
    Minor edits only
    + 1 other change: verification date
  2. Sep 28, 2026
    Minor edits only
    + 1 other change: verification date

From the registry record's own update history. This site started tracking changes on Sep 25, 2026; for anything earlier, see the record history on ClinicalTrials.gov ↗

11

Registry details

Key details

Study ID
NCT06967129
Responsible party
Sponsor
First posted
May 13, 2025
Start date
Apr 25, 2025
Primary completion
Jul 14, 2025
Completion
Jul 14, 2025
Results posted
Sep 15, 2026
Last update
Nov 4, 2026

Study contacts

Johnson & Johnson Vision Care, Inc. Clinical Trial
study director · Johnson & Johnson Vision Care, Inc.

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Nov 2026. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow