An interventional study of Test Lens in Visual Acuity, sponsored by Johnson & Johnson Vision Care, Inc.. Completed at 5 sites in United States. Open to participants aged 40 Years to 70 Years, including healthy volunteers. Per ClinicalTrials.gov, last updated 2023-07-27.
Sponsored by Johnson & Johnson Vision Care, Inc. (part of Johnson & Johnson) · Not applicable, Interventional, and Treatment
The study is a bilateral, single-masked, single-visit, non-dispensing 2x2 crossover study to confirm a finalized design of a prototype contact lens. There will be eight study lens types, however, each subject will only be randomized to receive two study lens types.
Johnson & Johnson Vision Care, Inc. is the lead sponsor of 221 studies on the registry; 2 are open to participants now.
Of its 111 completed or terminated interventional studies of FDA-regulated products, 103 (93%) have results posted.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Potential subjects must satisfy all of the following criteria to be enrolled in the study:
Exclusion Criteria:
Potential subjects who meet any of the following criteria will be excluded from participating in the study:
Eligible subjects that are habitual contact lens wearers will be randomized to the Myope, against-the-rule-astigmatism, (LOW/HIGH DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Myope, against-the-rule-astigmatism, (HIGH/LOW DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Myope, with-the-rule-astigmatism, (LOW/HIGH DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Myope, with-the-rule-astigmatism, (HIGH/LOW DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Hyperope, against-the-rule-astigmatism, (LOW/HIGH DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Hyperope, against-the-rule-astigmatism, (HIGH/LOW DC) sequence and wear two different study lens designs one at time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Hyperope, with-the-rule-astigmatism, (LOW/HIGH DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Eligible subjects that are habitual contact lens wearers will be randomized to the Hyperope, with-the-rule-astigmatism, (HIGH/LOW DC) sequence and wear two different study lens designs one at a time bilaterally over 2 wear periods.
Device: Test Lens
Multifocal Toric Contact Lenses manufactured in Senofilcon A (C3) material with UV/HEV filter
Proportion of Eyes With Absolute Rotation ≤ 10 Degrees
Absolute rotation was assessed for each subject eye using a slit lamp, at 1-, 3-, 7-, 15- and 25-minutes after lens insertion. However, absolute rotation at 15-minutes was the primary endpoint. Acceptable absolute rotation was dichotomized into a binary response where Y=1 if absolute rotation was less than or equal to 10 degrees and Y=0 otherwise (absolute rotation greater than 10 degrees). The proportion of eyes with absolute rotation less than or equal to 10 degrees was report for each lens.
Time frame: 15-minutes post lens insertion
| Milestone | Senofilcon A(C3) Toric Lens UV/HEV Filter(Low DC)/Senofilcon A(C3) Toric Lens UV/HEV Filter(High DC) | Senofilcon A(C3) Toric Lens UV/HEV Filter(High DC)/Senofilcon A(C3) Toric Lens UV/HEV Filter(Low DC) |
|---|---|---|
| Started | 53 | 49 |
| Completed | 53 | 49 |
| Not completed | 0 | 0 |
| Milestone | Senofilcon A(C3) Toric Lens UV/HEV Filter(Low DC)/Senofilcon A(C3) Toric Lens UV/HEV Filter(High DC) | Senofilcon A(C3) Toric Lens UV/HEV Filter(High DC)/Senofilcon A(C3) Toric Lens UV/HEV Filter(Low DC) |
|---|---|---|
| Started | 53 | 49 |
| Completed | 53 | 49 |
| Not completed | 0 | 0 |
Absolute rotation was assessed for each subject eye using a slit lamp, at 1-, 3-, 7-, 15- and 25-minutes after lens insertion. However, absolute rotation at 15-minutes was the primary endpoint. Acceptable absolute rotation was dichotomized into a binary response where Y=1 if absolute rotation was less than or equal to 10 degrees and Y=0 otherwise (absolute rotation greater than 10 degrees). The proportion of eyes with absolute rotation less than or equal to 10 degrees was report for each lens.
| Proportion | Senofilcon A (C3) Multifocal Toric Lens With UV/HEV Filter (Low DC) | Senofilcon A (C3) Multifocal Toric Lens With UV/HEV Filter (High DC) |
|---|---|---|
| Proportion of Eyes With Absolute Rotation ≤ 10 Degrees | 0.9167 | 0.8873 |
Collected over Throughout the duration of the study. Approximately 1-Day per Subject.. Non-serious events are listed at a 5% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| Senofilcon A (C3) Multifocal Toric Lens With UV/HEV Filter (Low C) | 0/102 (0%) | 0/102 (0%) | 0/102 (0%) |
| Senofilcon A (C3) Multifocal Toric Lens With UV/HEV Filter (High DC) | 0/102 (0%) | 0/102 (0%) | 0/102 (0%) |
All subjects dispensed at least one study lens.
| Age, Continuous(years) | Total |
|---|---|
| Mean | 52.3 ± 8.11 |
| Sex: Female, Male(Participants) | Total |
|---|---|
| Female | 81 |
| Male | 21 |
| Race/Ethnicity, Customized(Participants) | Total |
|---|---|
| Asian | 2 |
| Black or African American | 6 |
| White | 93 |
| Other | 1 |
| Region of Enrollment(Participants) | Total |
|---|---|
| United States — participants | 102 |
Documents are hosted by the registry — open the source record to download them.
Plan to share: Yes — Johnson \& Johnson Medical Device Companies have an agreement with the Yale Open Data Access (YODA) to serve as the independent review panel for evaluation of requests for clinical study reports and participant level data from investigators and physicians for scientific research that will advance medical knowledge and public health. Requests for access to the study data can be submitted through the YODA Project site at http://yoda.yale.edu
Data sharing details ↗ · yoda.yale.edu
Ran from Jun 2022 to primary completion in Jul 2022, 0 months.
Results were first posted in Jul 2023, 1 year after primary completion.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in Jun 2023. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Johnson & Johnson Vision Care, Inc. →