Validation of Point Partial User Needs With Partial Finger Amputees

An interventional study of Point Partial in Amputation; Traumatic, Hand, sponsored by Point Designs. Completed at 2 sites in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2024-11-08.

Sponsored by Point Designs · Not applicable, Interventional, and Device feasibility

From the registry’s dates

  • Registered 9 months after the study started (first participant enrolled Nov 2020, registered Aug 2021).
Phase
Not applicable
Study type
Interventional
Enrollment
5
Allocation
Not applicable
Ages
18 Years and older
Sex
All
01

Study summary

The objective of the device feasibility study will be to validate the user needs of the Point Partial system. This study will be a single group intervention model where one group of 5 partial finger amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.

02

Conditions studied

  • Amputation; Traumatic, Hand

Browse trials for

03

In context

Amputation, Traumatic

112 studies on the registry are indexed under Amputation, Traumatic; 24 are open to participants now.

This study's enrollment of 5 is below the median of 15 across 75 interventional studies indexed under Amputation, Traumatic.

Browse Amputation, Traumatic studies →

Lead sponsor

Point Designs is the lead sponsor of 8 studies on the registry; 2 are open to participants now.

Of its 5 completed or terminated interventional studies of FDA-regulated products, 5 (100%) have results posted.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Partial finger amputees with partial absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met.
  • Fluent in English
  • Individuals aged 18 or greater
  • Patients may present with either fused or mobile MCP joints.

Exclusion criteria

Exclusion Criteria:

  • Significant cognitive deficits as determined upon clinical evaluation
  • Significant neurological deficits as determined upon clinical evaluation
  • Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation
  • Uncontrolled pain or phantom pain impacting full participation in the study as determined upon OT evaluation
  • Serious uncontrolled medical problems as judged by the project therapist.
05

Study design

Phase
Not applicable
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
5 participants (actual)

Study arms

  • Experimental
    Prosthesis

    Patient is temporarily fit with Point Partial partial finger prosthetic system

    Device: Point Partial

Interventions

  • DevicePoint Partial

    The Point Partial partial finger prosthetic system consists of 1-4 ratcheting mechanical partial digits and mounting brackets

06

What researchers measure

Primary outcomes

  1. VAL-1: System Comfort Validation Test

    User wears Point Partial(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (8 hours)

  2. VAL-2: Unilateral Function Validation Test

    User flexes and extends the Point Partial(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  3. VAL-3.1: System Robustness Validation Test With 25 lb Bag

    User lifts a 25 lb. bag with Point Partial(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  4. VAL-3.2: Robustness Validation Test With Hammer

    User grips a hammer with Point Partial(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  5. VAL-4: Object Release Validation Test

    User releases Point Partial(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  6. VAL-5: Position Function Validation Test

    User positions Point Partial(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  7. VAL-6: Extension Function Validation Test

    User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh throughout day without failed extension (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  8. VAL-7: Grip Functions Validation Test

    User grips a 5 lb. cylindrical smooth object using Point Partial(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  9. VAL-8: Fine Motor Function Validation Test

    User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (10 minutes)

  10. VAL-9: Overall Function & Robustness Validation Test

    User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

    Time frame: Day 1 (8 hours)

07

Results

Posted Nov 8, 2024

Participant flow

Participant flow — Overall Study
MilestoneProsthesis
Started5
Completed5
Not completed0

Outcome measures

PrimaryVAL-1: System Comfort Validation Test

User wears Point Partial(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (8 hours)
Reported as:
Number · % of participants who Passed test
VAL-1: System Comfort Validation Test
% of participants who Passed testProsthesis
VAL-1: System Comfort Validation Test100
PrimaryVAL-2: Unilateral Function Validation Test

User flexes and extends the Point Partial(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-2: Unilateral Function Validation Test
% of participants who Passed testProsthesis
VAL-2: Unilateral Function Validation Test100
PrimaryVAL-3.1: System Robustness Validation Test With 25 lb Bag

User lifts a 25 lb. bag with Point Partial(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-3.1: System Robustness Validation Test With 25 lb Bag
% of participants who Passed testProsthesis
VAL-3.1: System Robustness Validation Test With 25 lb Bag100
PrimaryVAL-3.2: Robustness Validation Test With Hammer

User grips a hammer with Point Partial(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-3.2: Robustness Validation Test With Hammer
% of participants who Passed testProsthesis
VAL-3.2: Robustness Validation Test With Hammer100
PrimaryVAL-4: Object Release Validation Test

User releases Point Partial(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-4: Object Release Validation Test
% of participants who Passed testProsthesis
VAL-4: Object Release Validation Test100
PrimaryVAL-5: Position Function Validation Test

User positions Point Partial(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-5: Position Function Validation Test
% of participants who Passed testProsthesis
VAL-5: Position Function Validation Test100
PrimaryVAL-6: Extension Function Validation Test

User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh throughout day without failed extension (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-6: Extension Function Validation Test
% of participants who Passed testProsthesis
VAL-6: Extension Function Validation Test100
PrimaryVAL-7: Grip Functions Validation Test

User grips a 5 lb. cylindrical smooth object using Point Partial(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-7: Grip Functions Validation Test
% of participants who Passed testProsthesis
VAL-7: Grip Functions Validation Test100
PrimaryVAL-8: Fine Motor Function Validation Test

User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (10 minutes)
Reported as:
Number · % of participants who Passed test
VAL-8: Fine Motor Function Validation Test
% of participants who Passed testProsthesis
VAL-8: Fine Motor Function Validation Test100
PrimaryVAL-9: Overall Function & Robustness Validation Test

User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Time frame:
Day 1 (8 hours)
Reported as:
Number · % of participants who Passed test
VAL-9: Overall Function & Robustness Validation Test
% of participants who Passed testProsthesis
VAL-9: Overall Function & Robustness Validation Test100

Adverse events

Collected over an average of 1 day. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Prosthesis0/5 (0%)0/5 (0%)0/5 (0%)

Baseline characteristics

Age, Categorical
Age, Categorical(Participants)Prosthesis
<=18 years0
Between 18 and 65 years5
>=65 years0
Sex: Female, Male
Sex: Female, Male(Participants)Prosthesis
Female1
Male4
Ethnicity (NIH/OMB)
Ethnicity (NIH/OMB)(Participants)Prosthesis
Hispanic or Latino0
Not Hispanic or Latino2
Unknown or Not Reported3
Race (NIH/OMB)
Race (NIH/OMB)(Participants)Prosthesis
American Indian or Alaska Native0
Asian0
Native Hawaiian or Other Pacific Islander0
Black or African American0
White2
More than one race0
Unknown or Not Reported3
08

Study locations

2 sites
  • University of Colorado Denver | Anschutz Medical Campus
    Aurora, Colorado 80045, United States
  • Point Designs
    Lafayette, Colorado 80026, United States
09

References and documents

Study documents

PDFs hosted by the registry, ClinicalTrials.gov.

Individual participant data

Plan to share: No

10

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Ran from Nov 2020 to primary completion in Dec 2022, 25 months.

Results were first posted in Nov 2024, 1 year 10 months after primary completion.

Of 46 interventional studies with no phase under Amputation, Traumatic, half reached primary completion within 27 months of starting (the median), counting those still running; a quarter within 14 and three quarters within 51. This study's 25 months is close to that.

How these figures are worked out, and which studies are left out of them.

Registered
Aug 12, 2021
First posted
Aug 19, 2021
StartActual
Nov 3, 2020
Primary completionActual
Dec 31, 2022
CompletionActual
Dec 31, 2022
Results first posted
Nov 8, 2024
Last update
Nov 8, 2024
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Nov 8, 2024, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT05012657
Lead sponsor
Point Designs
Responsible party
Sponsor
First posted
Aug 19, 2021
Start date
Nov 3, 2020
Primary completion
Dec 31, 2022
Completion
Dec 31, 2022
Results posted
Nov 8, 2024
Last update
Nov 8, 2024

Oversight

FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Aug 2024. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion