An interventional study of Nelfilcon A digital contact lenses and Nelfilcon A contact lenses in Refractive Errors, sponsored by Alcon Research. Completed at 5 sites in United States. Open to participants aged 18 Years to 35 Years. Per ClinicalTrials.gov, last updated 2019-06-18.
Sponsored by Alcon Research (part of Alcon) · Not applicable, Interventional, and Treatment
The purpose of this study is to evaluate the visual performance of DAILIES® AquaComfort Plus® (DACP) Digital lenses by assessing distance visual acuity (VA) as compared to DACP sphere contact lenses after one week of wear.
427 studies on the registry are indexed under Refractive Errors; 62 are open to participants now.
This study's enrollment of 60 is below the median of 81 across 307 interventional studies indexed under Refractive Errors.
Browse Refractive Errors studies →Alcon Research is the lead sponsor of 624 studies on the registry; 9 are open to participants now.
Of its 111 completed or terminated interventional studies of FDA-regulated products, 100 (90%) have results posted.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
Nelfilcon A digital contact lenses worn first, followed by nelfilcon A contact lenses. Each product worn bilaterally (in both eyes) for 7 days in a daily disposable modality.
Device: Nelfilcon A digital contact lenses · Device: Nelfilcon A contact lenses
Nelfilcon A contact lenses worn first, followed by nelfilcon A digital contact lenses. Each product worn bilaterally for 7 days in a daily disposable modality.
Device: Nelfilcon A digital contact lenses · Device: Nelfilcon A contact lenses
Silicone hydrogel digital contact lenses
Also known as: DAILIES® AquaComfort Plus® Digital, DACP Digital
Silicone hydrogel spherical contact lenses
Also known as: DAILIES® AquaComfort Plus®, DACP
Distance Visual Acuity (VA) (logMAR, OU)
VA was tested under photopic (well-lit) conditions using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart positioned 6 meters from the subject. VA was collected bilaterally (OU) and measured in logMAR (logarithm of the minimum angle of resolution), with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on the ETDRS chart. A lower numeric value represents better visual acuity.
Time frame: Day 7, each product
Overall Vision
Subjective rating of overall vision on a scale of 1 (Poor) to 10 (Excellent). Both eyes contributed to the analysis.
Time frame: Day 7, each product
Subjects were recruited from 5 investigative sites located in the United States.
| Milestone | DACP Digital Then DACP | DACP Then DACP Digital |
|---|---|---|
| Started | 29 | 31 |
| Completed | 29 | 31 |
| Not completed | 0 | 0 |
| Milestone | DACP Digital Then DACP | DACP Then DACP Digital |
|---|---|---|
| Started | 29 | 31 |
| Completed | 29 | 31 |
| Not completed | 0 | 0 |
VA was tested under photopic (well-lit) conditions using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart positioned 6 meters from the subject. VA was collected bilaterally (OU) and measured in logMAR (logarithm of the minimum angle of resolution), with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on the ETDRS chart. A lower numeric value represents better visual acuity.
| logMAR | DACP Digital | DACP |
|---|---|---|
| Distance Visual Acuity (VA) (logMAR, OU) | -0.11 ± 0.10 | -0.12 ± 0.09 |
Subjective rating of overall vision on a scale of 1 (Poor) to 10 (Excellent). Both eyes contributed to the analysis.
| units on a scale | DACP Digital | DACP |
|---|---|---|
| Overall Vision | 8.8 ± 1.0 | 8.9 ± 1.0 |
Collected over Dispense through study completion, an average of 14 days. Non-serious events are listed at a 5% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| DACP Digital | 0/60 (0%) | 0/60 (0%) | 0/60 (0%) |
| DACP | 0/60 (0%) | 0/60 (0%) | 0/60 (0%) |
This analysis population includes all randomized subjects who were exposed to any study lenses evaluated in this study (Full Analysis Set).
| Age, Continuous(years) | Overall |
|---|---|
| Mean | 25.2 ± 5.3 |
| Sex: Female, Male(Participants) | Overall |
|---|---|
| Female | 36 |
| Male | 24 |
| Race/Ethnicity, Customized(Participants) | Overall |
|---|---|
| White | 56 |
| Black or African American | 3 |
| American Indian or Alaska Native | 0 |
| Asian | 0 |
| Native Hawaiian or Other Pacific Islander | 0 |
| Other | 1 |
Documents are hosted by the registry — open the source record to download them.
Plan to share: No
Ran from Jun 2018 to primary completion in Aug 2018, 2 months.
Results were first posted in Jun 2019, 9 months after primary completion.
Of 122 industry-sponsored interventional studies with no phase under Refractive Errors, half reached primary completion within 4 months of starting (the median), counting those still running; a quarter within 1 and three quarters within 10. This study's 2 months is below that.
Counting every sponsor, half of 174 studies reached it within 7 months; a quarter within 2 and three quarters within 23.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in May 2019. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.