CClinicalTrials.gg
CompletedNCT03475992Updated Jun 7, 2021Results posted

Pilot Clinical Study on a Low-power Electromagnetic Wave Breast Imaging Device for Cancer Screening Purposes.

An interventional study of Low-power microwave breast imaging system in Breast Cancer and Breast Cyst, sponsored by MVG Industries SAS. Completed at 1 site in Ireland. Open to female participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2021-06-07.

Sponsored by MVG Industries SAS · Not applicable, Interventional, and Diagnostic

Phase
Not applicable
Study type
Interventional
Enrollment
25
Allocation
Non-randomized
Ages
18 Years and older
Sex
Female
01

Study summary

The trialed investigational medical imaging device is a low-power microwave breast imaging system for cancer screening purposes. It is an active device which uses non-ionizing radiation.

Microwave imaging is an emerging imaging modality for the early detection of the breast cancer. The physical basis of microwave imaging is the dielectric contrast between healthy and cancerous breast tissues at microwave frequencies. Microwave imaging can potentially be used for monitoring neoadjuvant chemotherapy treatment, breast health monitoring, and for routine screening and diagnosis of the breast cancer at the early-stage. The non-invasive and the non-ionizing characteristics of microwaves should allow for frequent scans of the breast using microwave imaging, unlike X-ray mammography. In addition to safety, microwave imaging does not require uncomfortable breast compression and it is potentially a lower-cost modality.

This is a first-in-human clinical test of the investigational device, which has been so far tested only with experimental phantoms modelling the human female breast.

The clinical data that will be collected in the context of this study is intended to provide early safety information for the investigational medical imaging device. In addition, this exploratory data will guide the refinement of the device hardware and the imaging algorithm design, before decision to proceed (or not) with further clinical tests.

Furthermore, this study will be used to guide sample size calculation for a subsequent study designed to evaluate efficacy should that appear warranted once this study is completed.

Read the detailed description

This is a single-site early-phase pilot clinical study, taking place at Galway University Hospital/HRB Clinical Research Facility, Galway.

Patients will have a conventional history and breast examinations (Mammogram and / or Ultrasound and Clinical assessment) performed by the physician, as per normal practice in the Symptomatic Breast Unit of University Hospital Galway.

If patients are interested in the study, provide written informed consent and are deemed eligible, they will attend the Clinical Research Facility Galway for a Microwave Breast Imaging (MBI) scan. The data from the scan will be collected and stored securely. Patients will be followed up 7-21 days after the microwave breast investigation or before surgery, whichever comes first. Patients will be assessed for their experience of the microwave breast investigation, and for any adverse events.

The investigational medical imaging device consists of two subsystems, both performing a non-invasive examination:

  • the Microwave Breast Imaging subsystem (MBI subsystem);
  • the Optical Breast Contour Detection subsystem (OBCD subsystem);

The MBI subsystem is an active device which uses non-ionizing radiation. It illuminates the breast with low-power electromagnetic waves in the microwave frequency spectrum, which penetrate the breast under examination. The subsystem collects the scattered electromagnetic waves and recovers useful information on the breast tissue consistency, given the dielectric contrast of these tissues.

Multi-static radar detection technology is employed in the MBI subsystem for breast image formation.The well-established Microwave Vision Group (MVG) technology for fast antenna measurement, using multiple sensors in a vertical arch configuration, has been transposed to a horizontal arch of sensors. In addition, vertical translation of the horizontal arch has been enabled, such that 3D multi-static short-range radar imaging is possible.

The sensors are in contact with a cylindrical container filled with a liquid; the so-called coupling liquid has been designed to have electromagnetic (EM) properties appropriately selected such that the EM wave penetration in the breast is maximized.

During the MBI scan, the patient is lying in a face down position on a special bed, integrated with the MBI subsystem. The breast under examination is immersed in the coupling liquid, through a dedicated circular opening of the bed. The breast is then scanned. This vertical scan takes approximately 10-15 minutes to complete, depending on the size of each breast.

In order to compute the required volume of coupling liquid, such that the container of the MBI subsystem is optimally filled after immersion of the breast, a simple process for optical assessment of the total volume of the breast takes place just before starting the MBI scan.

The Optical Breast Contour Detection (OBCD) subsystem serves to provide the total volume of the breast, and also its external contour, as a priori information to the MBI subsystem. The OBCD subsystem consists of a 3D optical camera placed below the examination table, at a distance of several tens of centimetres below the breast. An azimuthal scan of the 3D camera permits to reconstruct the external surface of the breast.

In order for the breast contour to be a useful a priori information for the MBI subsystem, it is important that the patient is lying in the same face down position during both the MBI and OBCD scans. Thus, an identical examination table, as the one integrated with the MBI subsystem, is also integrated with the OBCD subsystem. During the OBCD scan, the patient is lying on the examination table, with her breast under examination inserted in the circular opening of the examination table. For this scan, there is no coupling liquid; the breast is in the air, hanging below the examination table.

Both the breast bearing the palpable lump and the contralateral breast will be scanned. The contralateral breast scan will serve as control in the data analysis.

The MBI scan duration of each breast should not exceed 15 minutes. The recorded data will be checked and in the case of non-optimal positioning of the patient, the scan may be repeated.

It is expected that the duration of the microwave breast investigation (including OBCD scan of both breasts for breast volume assessment, preparation of the MBI subsystem, MBI scan of both breasts, and data verification) should not exceed 1 hour. If the Data Quality Check suggests repetition of the MBI scan, and upon agreement by the patient, the total duration of the microwave breast investigation will be extended.

The written radiology reports from conventional imaging will be acquired and used to evaluate the performance of the MBI subsystem in terms of detecting and estimating the size and consistency of the breast lump.

MBI data analysis will be performed off-line, at MVG premises (France), and is expected to be completed few months after the end-date of data collection. Following the assessment of the MBI results by the medical experts involved in the study, a decision will be made regarding the potential of this emerging imaging modality and the interest in proceeding with clinical studies involving larger sample sizes. Based on these results, MVG will decide and plan actions for refinement of the system, towards an upgraded version to be used for potential future tests with a more significant impact.

02

Conditions studied

  • Breast Cancer
  • Breast Cyst

Browse trials for

03

Who can participate

Ages eligible
18 Years and older
Sexes eligible
Female
Accepts healthy volunteers
No

Inclusion criteria

  1. Subjects must be able and willing to give written informed consent and to comply with the requirements of this study protocol
  2. Subject must have attended the Symptomatic Breast Unit with a palpable breast lump
  3. Subjects must have had a mammogram in the clinical assessment period (≤ 6 weeks before the microwave breast investigation)
  4. Subjects must be able to comfortably lie reasonably still in a prone position for up to 15 minutes
  5. Subjects with bra size larger than 32B and cup size larger or equal to B.
  6. Subjects whose breast size is adapted to the cylindrical container of the MBI system with sufficient margin to allow the presence of transition liquid around the breast. Final decision to be taken by the physicians based on their judgement.

Exclusion criteria

Exclusion Criteria:

  1. Subjects unable to provide written informed consent
  2. Subjects who are pregnant or breast-feeding
  3. Subjects who have had previous surgery to the breast
  4. Subjects who have previously received chemotherapy or radiotherapy to the breast
  5. Subjects who have had a breast biopsy less than two weeks prior to imaging
  6. Subjects with any active or metallic implant (e.g. cardiac pacemaker, stents, internal cardiac defibrillator, cardiac resynchronisation device, nerve stimulator...), or subjects bearing any non-removable metallic object (e.g. piercing) on their torso
  7. Post-biopsy patients whose breast tissue is not healed sufficiently for the imaging procedure, in the opinion of the investigator
  8. Patients who have had or plan to have a breast cyst aspiration before MBI.
  9. Subjects with significant co-morbidities which, in the opinion of the investigator, may influence the result of the study
  10. Subjects with prior or concurrent malignancy
  11. Subjects under the age of 18 years old
  12. Subjects with evidence of inflammation and/or erythema of the breast as well as any subjects who have a break in the skin which would be in contact with the coupling fluid
  13. Subjects who would be unsuitable for an MBI scan, unlikely to attend a follow up visit, or would otherwise be unsuitable for such an investigation, in the opinion of the Investigator
04

Study design

Phase
Not applicable
Primary purpose
Diagnostic
Allocation
Non-randomized
Intervention model
Parallel assignment
Masking
None (open label)
Enrollment
25 participants (actual)

Study arms

  • Experimental
    Pre-diagnosed breast cancer

    Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation

    Device: Low-power microwave breast imaging system

  • Experimental
    Pre-diagnosed breast cyst

    Low-power microwave breast imaging system. No prior biopsy

    Device: Low-power microwave breast imaging system

  • Experimental
    Pre-diagnosed benign lesion

    Low-power microwave breast imaging system. Core needle biopsy performed ≥14 days before the microwave breast investigation

    Device: Low-power microwave breast imaging system

Interventions

  • DeviceLow-power microwave breast imaging system

    Investigate the capacity of microwave imaging to detect and characterise diagnosed palpable breast lump

05

What researchers measure

Primary outcomes

  1. Number of Serious Adverse Events, Directly Related to Normal Functioning of the Device

    To provide early safety information for the proposed investigational medical imaging device.

    Time frame: Through study completion, up to 21 days

Other outcomes

  1. Capacity of the Low-power Microwave Breast Imaging (MBI) System to Detect Pre-diagnosed Palpable Breast Lesions. Detection Rate for Malignant and Benign Breast Lesions.

    Percentage of total number of biopsy-confirmed malignant breast lesions being detected with MBI; Percentage of total number of pre-diagnosed breast cysts being detected with MBI; Percentage of total number of biopsy-confirmed benign breast lesions being detected with MBI.

    Time frame: Duration of study - 19 months

06

Results

Posted Jun 7, 2021
Limitations and caveats
First in Human, exploratory, open clinical study was conducted in order to gather clinical data to support refinement of Prototype # 1 of the low-power MBI system as well as initial safety information

Participant flow

Participant flow — Overall Study
MilestonePre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy Confirmed
Started1195
Completed1195
Not completed000

Outcome measures

PrimaryNumber of Serious Adverse Events, Directly Related to Normal Functioning of the Device

To provide early safety information for the proposed investigational medical imaging device.

Time frame:
Through study completion, up to 21 days
Reported as:
Number · Serious Adverse Events
Number of Serious Adverse Events, Directly Related to Normal Functioning of the Device
Serious Adverse EventsPre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy Confirmed
Number of Serious Adverse Events, Directly Related to Normal Functioning of the Device000
Other pre-specifiedCapacity of the Low-power Microwave Breast Imaging (MBI) System to Detect Pre-diagnosed Palpable Breast Lesions. Detection Rate for Malignant and Benign Breast Lesions.

Percentage of total number of biopsy-confirmed malignant breast lesions being detected with MBI; Percentage of total number of pre-diagnosed breast cysts being detected with MBI; Percentage of total number of biopsy-confirmed benign breast lesions being detected with MBI.

Time frame:
Duration of study - 19 months
Reported as:
Number · Percentage(%) of lesions being detected
Capacity of the Low-power Microwave Breast Imaging (MBI) System to Detect Pre-diagnosed Palpable Breast Lesions. Detection Rate for Malignant and Benign Breast Lesions.
Percentage(%) of lesions being detectedPre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy Confirmed
Capacity of the Low-power Microwave Breast Imaging (MBI) System to Detect Pre-diagnosed Palpable Breast Lesions. Detection Rate for Malignant and Benign Breast Lesions.8210080

Adverse events

Collected over Safety data collected over 15 months. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Pre-diagnosed Breast Cancer - Biopsy Confirmed0/11 (0%)0/11 (0%)0/11 (0%)
Pre-diagnosed Breast Cyst0/9 (0%)0/9 (0%)1/9 (11.1%)
Pre-diagnosed Benign Lesion - Biopsy Confirmed0/5 (0%)0/5 (0%)0/5 (0%)
Most frequent other events
Most frequent other events
EventPre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy Confirmed
Eyelid infection-styeEye disorders0/111/90/5

Baseline characteristics

Overall, there have been twenty-five (25) patients scanned with the low-power Microwave Breast Imaging (MBI) system. Out of the twenty-five (25) patients, twenty-four (24) were included in the final analysis. The patient who was excluded from the final analysis was a patient who presented with a palpable lump which was subsequently determined to be normal breast tissue

Age, Continuous
Age, Continuous(years)Pre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy ConfirmedTotal
Mean64.6 (42 to 83)38.3 (35 to 49)39.4 (36 to 47)50.5 (35 to 83)
Sex: Female, Male
Sex: Female, Male(Participants)Pre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy ConfirmedTotal
Female118524
Male0000
Race/Ethnicity, Customized
Race/Ethnicity, Customized(Participants)Pre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy ConfirmedTotal
Race — Irish107522
Race — Brazilian1102
Region of Enrollment
Region of Enrollment(participants)Pre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy ConfirmedTotal
Ireland118524
Volumetric Breast Density
Volumetric Breast Density(Percentage(%) of dense tissue volume)Pre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy ConfirmedTotal
Mean8.01 (4.00 to 15.40)18.04 (8.40 to 27.40)13.22 (5.80 to 25.45)12.56 (4.00 to 25.45)
Breast Volume
Breast Volume(milliliters (mL))Pre-diagnosed Breast Cancer - Biopsy ConfirmedPre-diagnosed Breast CystPre-diagnosed Benign Lesion - Biopsy ConfirmedTotal
Mean757.3 (293 to 1369)647.8 (150 to 1338)646.5 (222.7 to 1105.6)696.8 (150 to 1369)
07

Study locations

1 site
  • HRB Clinical Research Facility Galway
    Galway, Ireland
08

References and documents

Study documents

  • Protocol and statistical analysis plan · Jan 19, 2021

Documents are hosted by the registry — open the source record to download them.

Individual participant data

Plan to share: No

09

Registry details

Key details

Study ID
NCT03475992
Lead sponsor
MVG Industries SAS
Collaborators
HRB Clinical Research Facility Galway
Responsible party
Sponsor
First posted
Mar 23, 2018
Start date
Oct 10, 2018
Primary completion
May 7, 2020
Completion
May 7, 2020
Results posted
Jun 7, 2021
Last update
Jun 7, 2021

Study contacts

Michael Kerin, Prof.
principal investigator · Galway University Hospital

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in May 2021. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion