Closed Loop Glucose Control in a Simulated ICU Setting

An interventional study of FUSION closed loop glucose control system in Type 1 Diabetes, Type 2 Diabetes and Stress Induced Hyperglycemia, sponsored by Ideal Medical Technologies. Not yet recruiting. Open to participants aged 18 Years to 70 Years. Per ClinicalTrials.gov, last updated 2026-06-12.

Sponsored by Ideal Medical Technologies · Not applicable, Interventional, and Treatment

Phase
Not applicable
Study type
Interventional
Enrollment
6
Allocation
Not applicable
Ages
18 Years to 70 Years
Sex
All
01

Study summary

This is a proof of concept safety study of a rule-based closed loop glucose control system designed for use in the intensive care unit setting.

The persons with type 1 and type 2 diabetes in this study will have their glucose controlled to a range of 100-140 mg/dL by a fully autonomous novel rule-based closed loop glucose control system for a period of 24 hours. The subjects will consume three standardized meals and receive a continuous intravenous infusion of dextrose (D-glucose) during the 24 hour study period. Meals will not be announced to the glucose control system.

Read the detailed description

Tight glucose control in the intensive care unit (ICU) setting is difficult to achieve. The investigators hypothesize that a closed loop glucose control system that is rule-based will improve upon the glucose control currently achieved by the current open loop manual methods, and that this improved glucose control may improve the outcomes of critically ill patients.

This Earl Feasibility Study will test the ability of a prototype rule-based closed loop glucose control system named FUSION, to provide safe and effective glucose control in subjects with type 1 and type 2 diabetes in a clinical research center (CRC) setting. Subjects with type 1 diabetes have been chosen as safe and effective glucose control is difficult to achieve in these subjects during meal challenges. Subjects with type 2 diabetes have been chosen as they are insulin resistant, which makes their insulin resistance profile similar to that of ICU patients. As this is a safety study of a new medical device, the controlled environment of the CRC is preferable to the less controlled environment of an ICU setting.

The prototype FUSION system to be used in this study will consist to two Dexcom G6 continuous glucose monitors (CGM), the rule-based glucose control software run on an all-in-one medical computer, and two syringe pumps. The prototype system is housed on a medical cart. Based on the average glucose value of the two Dexcom G6 CGM's, and the algorithms of the FUSION systems rule-based glucose control software, the FUSION system will make rate adjustments every 5-10 minutes to the intravenous infusion rates of short acting insulin (NovoLog) and dextrose (D10NS) under its control, in an attempt to keep the subjects glucose in the range of 100-140 mg/dL. The FUSION system only requires entry of the subjects study identification number and weight in kilograms to initiate the system.

For safety reasons, the subjects will have their blood glucose independently measured every 10-60 minutes on the point of care Nova StatStrip system, throughout the 24 hour study period.

The study has halting criteria to avoid recurrent instances of severe hypoglycemia (\< 54 mg/dL).

The average of the two CGM's, that is used by the FUSION system for glucose control, will be used for statistical analysis.

02

Conditions studied

  • Type 1 Diabetes
  • Type 2 Diabetes
  • Stress Induced Hyperglycemia

Keywords

  • Artificial Pancreas
  • Intensive Care Unit
  • ICU
  • Closed Loop Glucose Control
  • Stress Induced Hyperglycemia
  • Type 1 Diabetes
  • Type 2 Diabetes
  • Inpatient
  • Hospital
  • Hypoglycemia
  • Hyperglycemia
  • Glucose Variability
  • Fully Autonomous
03

In context

Diabetes Mellitus, Type 1

3,754 studies on the registry are indexed under Diabetes Mellitus, Type 1; 600 are open to participants now.

This study's planned enrollment of 6 is below the median of 40 across 2,820 interventional studies indexed under Diabetes Mellitus, Type 1.

Browse Diabetes Mellitus, Type 1 studies →

Lead sponsor

Ideal Medical Technologies is the lead sponsor of 3 studies on the registry; 1 is open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years to 70 Years
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Participants with type 1 and type 2 diabetes are eligible to be included in the study only if they meet all the following criteria:

    1. Are 18-70 years of age, inclusive.
    2. Can understand and sign an , communicate with the investigator, and understand and comply with the protocol requirements.
    3. Have had a diagnosis of type 1 or type 2 diabetes for a period of at least 3 months.
    4. Use insulin injections at home for glucose control.
    5. Have a (HbA1c) in the range of 7.0 - 10.0%.
    6. Have a in the range for sex:

      1. Females: 7-15.5 grams/dL.
      2. Males: 7-17.5 grams/dl.
    7. Have adequate venous access sites in upper extremities.
    8. Body weight between 40 - 150 kg.

Exclusion criteria

Exclusion Criteria:

  1. Have participated in an interventional medical, surgical, or pharmaceutical study within 30 days of screening.
  2. Have a known to any of the components of study treatment.
  3. Have skin disease/injury at Dexcom G6 CGM insertion site(s) that would prevent insertion of the CGM.
  4. Currently abuses drugs or alcohol or has a history of abuse that in the investigator's opinion would cause the individual to be noncompliant.
  5. Have a medical condition that in the opinion of the investigator could affect study participation and/or personal well-being.
  6. Have a history or presence of any of the following conditions:

    1. Hepatic failure or has (ALT) greater than 3 times the .
    2. Has an estimated \<30 ml/min/1.73 m2 or End Stage Kidney Disease on renal replacement therapy.
    3. Have congestive heart failure greater than class 1 on the classification system.
    4. Have a history of seizures.
    5. Have a history of cerebrovascular accident.
    6. Have a history of ischemic heart disease.
  7. For female subjects of potential childbearing age (age 18 to 55) they will be excluded if:

    1. Pregnant.
    2. Refuse to agree to a pregnancy test at the time of enrollment.
    3. Have a positive urine pregnancy test at the time of enrollment.
  8. Have a positive Influenza or COVID test within 14 days of visit 3.
  9. Have any viral lower respiratory related symptoms (e.g., cough, chest tightness, respiratory distress) in the 14-day period prior to visit 3.
  10. Have a known unprotected Influenza or COVID exposure in the 14-day period prior to visit 3.
  11. Are currently being treated with Hydroxyurea, a medication which may cause the Dexcom G6 CGM to produce glucose readings that are higher than the actual glucose value.
Terms in these criteria, in plain language(9)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorised to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

HbA1c

A blood test of average blood sugar over the past two to three months, given as a percentage or in mmol/mol. 6.5% (48 mmol/mol) or more is the usual threshold for diabetes, normally confirmed by a second test.

Hemoglobin (Hb)

The protein in red blood cells that carries oxygen. A low level is anaemia, which can cause tiredness and breathlessness. Measured in g/dL or g/L (9 g/dL is 90 g/L).

Hypersensitivity

An allergy, or another reaction of the immune system, to a substance. “Hypersensitivity to the study drug” means a known allergic reaction to it — not every side effect a medicine can have.

Clinically significant

Judged by the study doctor to matter for your health or for the study. It is a doctor’s decision, so ask the study team if you are unsure whether something counts.

ALT and AST (liver enzymes)

Two enzymes measured in a routine blood test of the liver. Higher levels can mean the liver is under strain or damaged. SGPT and SGOT are older names for ALT and AST.

Criteria usually give a limit as a multiple of the upper limit of normal (ULN), often higher when cancer has spread to the liver.

ULN (upper limit of normal)

The top of the normal range for a blood test, as set by the laboratory that runs it. Each lab prints its own range next to your result.

“≤ 2.5 × ULN” means the result can be up to two and a half times that top value.

GFR and eGFR (kidney filtering)

GFR (glomerular filtration rate) is how much blood the kidneys filter each minute. It is usually estimated from a blood test — eGFR, by formulas such as CKD-EPI or MDRD — and some studies ask for it to be measured directly instead. Usually given in mL/min/1.73 m².

Check the method, the units and the minimum the study states. A value of 90 or above is in the normal-or-high range, but kidney damage can show in other tests too.

NYHA class (heart failure)

The New York Heart Association’s four classes of how much heart failure limits physical activity.

I
No limits: ordinary activity causes no undue tiredness or breathlessness.
II
Slight limits: comfortable at rest, ordinary activity causes symptoms.
III
Marked limits: comfortable at rest, less than ordinary activity causes symptoms.
IV
Symptoms even at rest; any activity makes them worse.

Studies set their own limit — many outside heart failure leave out class III–IV, while heart failure studies may ask for a particular class. Check the class this study names.

05

Study design

Phase
Not applicable
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
6 participants (estimated)

Study arms

  • Experimental
    FUSION closed loop glucose control system safety study

    All subjects will be treated with the FUSION closed loop glucose control system for up to 24 hours

    Device: FUSION closed loop glucose control system

Interventions

  • DeviceFUSION closed loop glucose control system

    The FUSION system will be used to control the subjects glucose to a range of 100-140 mg/dL. Data will be collected for up to 24 hours, or upon early termination of the study session.

06

What researchers measure

Primary outcomes

  1. Primary Safety Measure

    The percentage of all glucose values \<70 mg/dL

    Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time up to 24 hours

  2. Primary Efficacy Measure

    The percentage of all glucose values within the glucose range of 70-180 mg/dL

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

Secondary outcomes

  1. Percent of All Glucose Values Within Glucose Range <54 mg/dL

    The percentage of all glucose values that are less than 54 mg/dL

    Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

  2. Percent of All Glucose Values Within Glucose Range 54-69 mg/dL

    The percentage of all glucose values that are within the glucose range of 54-69 mg/dL

    Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of up to 24 hours.

  3. Number of Hypoglycemic Events (<70 mg/dL)

    The total number of hypoglycemic events less than 70 mg/dL as measured by a continuous series of glucose values less than 70 mg/dL for at least 15 minutes duration.

    Time frame: From beginning of use of the FUSION System until end of use of the FUSION System, which will be a period of time of up to 24 hours.

  4. Percent of All Glucose Values Within Glucose Range 70-100 mg/dL

    The percentage of all glucose values within the glucose range of 70-100 mg/dL

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  5. Percent of All Glucose Values Within Glucose Range 70-140 mg/dL

    The percentage of all glucose values within the glucose range of 70-140 mg/dL

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  6. Percent of All Glucose Values Within Glucose Range 100-140 mg/dL

    The percentage of all glucose values within the glucose range of 100-140 mg/dL

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  7. Percent of All Glucose Values Within Glucose Range >140 mg/dL

    The percentage of all glucose values greater than 140 mg/dL

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  8. Percent of All Glucose Values Within Glucose Range >180 mg/dL

    The percentage of all glucose values that are greater than 180 mg/dL

    Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

  9. Percent of All Glucose Values Within Glucose Range >250 mg/dL

    The percentage of all glucose values that are greater than 250 mg/dL

    Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

  10. Average glucose value in mg/dL

    The average of all measured glucose values in mg/dL

    Time frame: From beginning of use of the FUSION System to end of use of the FUSION System, which will be a period of time of up to 24 hours

  11. Measure of dispersion - Coefficient of Variation

    The Coefficient of variation of glucose will be calculated by dividing the standard deviation of the average glucose value by the average glucose value, then multiplying this value by 100.

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  12. Clarke Error Grid Analysis of Continuous Glucose Monitor Data

    The study data will be used to determine the percentage of the two Dexcom G6 CGM's glucose values in each zone using a Clarke error grid analysis. This same calculation will be performed on the average of the two Dexcom G6 CGM systems used by the FUSION system for purposes of glucose control.

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  13. FDA integrated Continuous Glucose Monitor Standards

    The study data will be used to determine if each individual Dexcom G6 CGM system, and their average value meet FDA criteria for integrated Continuous Glucose Monitor (iCGM) accuracy for the glucose zones \<70 mg/dL, 70-180 mg/dL, and \>180 mg/dL.

    Time frame: From beginning of use of FUSION System to end of use of FUSION System, which will be a period of time of up to 24 hours

  14. Hemoglobin A1c Percentage Level

    The subjects hemoglobin A1c measurement from a blood draw taken within 2 weeks of the start of the closed loop glucose control session

    Time frame: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session

  15. C-Peptide Level

    The subjects C-peptide level will be measured prior to the start of the closed loop glucose control session.

    Time frame: As measured from a blood draw within two weeks prior to the start of the 24-hour closed loop glucose control session

  16. Total Daily Dose of Insulin in Units/Kilogram

    The subjects total daily dose of insulin used prior to the start of the closed loop glucose control session was calculated by adding up all of the insulin doses used by the subjects (both short and long acting insulin) in units and dividing this value by the subjects weight in Kilograms

    Time frame: As documented within two weeks prior to the start of the 24-hour closed loop glucose control session

  17. Weight in Kilograms

    Average weight in kilograms of the subjects

    Time frame: As measured just prior to the start of the 24-hour closed loop glucose control session

  18. Body Mass Index

    The body mass index will be calculated by dividing the subjects weight in kilograms by the square of their height in meters

    Time frame: As measured just prior to the start of the 24-hour closed loop glucose control session

07

Study locations

No study locations are listed for this record.

08

References and documents

Publications

  • Sardu C, D'Onofrio N, Balestrieri ML, Barbieri M, Rizzo MR, Messina V, Maggi P, Coppola N, Paolisso G, Marfella R. Outcomes in Patients With Hyperglycemia Affected by COVID-19: Can We Do More on Glycemic Control? Diabetes Care. 2020 Jul;43(7):1408-1415. doi: 10.2337/dc20-0723. Epub 2020 May 19. PubMed 32430456 ↗
  • DeJournett L. Essential elements of the native glucoregulatory system, which, if appreciated, may help improve the function of glucose controllers in the intensive care unit setting. J Diabetes Sci Technol. 2010 Jan 1;4(1):190-8. doi: 10.1177/193229681000400124. PubMed 20167184 ↗
  • DeJournett J, Nekludov M, DeJournett L, Wallin M. Performance of a closed-loop glucose control system, comprising a continuous glucose monitoring system and an AI-based controller in swine during severe hypo- and hyperglycemic provocations. J Clin Monit Comput. 2021 Apr;35(2):317-325. doi: 10.1007/s10877-020-00474-2. Epub 2020 Jan 31. PubMed 32006145 ↗
  • DeJournett J, DeJournett L. Comparative Simulation Study of Glucose Control Methods Designed for Use in the Intensive Care Unit Setting via a Novel Controller Scoring Metric. J Diabetes Sci Technol. 2017 Nov;11(6):1207-1217. doi: 10.1177/1932296817711297. Epub 2017 Jun 22. PubMed 28637358 ↗
  • DeJournett L, DeJournett J. In Silico Testing of an Artificial-Intelligence-Based Artificial Pancreas Designed for Use in the Intensive Care Unit Setting. J Diabetes Sci Technol. 2016 Nov 1;10(6):1360-1371. doi: 10.1177/1932296816653967. Print 2016 Nov. PubMed 27301982 ↗

Individual participant data

Plan to share: No

09

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Start is estimated for Jul 2026; primary completion for Dec 2026, 5 months later.

Of 105 industry-sponsored interventional studies with no phase under Diabetes Mellitus, Type 1, half reached primary completion within 10 months of starting (the median), counting those still running; a quarter within 5 and three quarters within 18. This study's planned 5 months is below that.

Counting every sponsor, half of 949 studies reached it within 18 months; a quarter within 9 and three quarters within 35.

How these figures are worked out, and which studies are left out of them.

Registered
Jun 9, 2026
First posted
Jun 12, 2026
StartEst.
Jul 2026
Primary completionEst.
Dec 2026
CompletionEst.
Dec 2027
Last update
Jun 12, 2026
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Jun 12, 2026, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT07646067
Lead sponsor
Ideal Medical Technologies
Responsible party
Sponsor
First posted
Jun 12, 2026
Start date
Jul 2026 (estimated)
Primary completion
Dec 2026 (estimated)
Completion
Dec 2027 (estimated)
Last update
Jun 12, 2026

Study contacts

Leon M DeJournett, MD
Contact
leondej@idealmedtech.com
1-828-337-9960
Jeremy DeJournett, EE
Contact
jeremy@idealmedtech.com
1-828-545-9110
Leon M DeJournett, MD
study director · Ideal Medical Technologies
Francisco Pasquel, MD, MPH
principal investigator · Emory School of Medicine

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is not yet recruiting, as verified in Jun 2026. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion