Efficacy and Safety of Pulmonary Function Analysis With Dynamic Digital Radiography Total Respiratory Cycle Photography

An interventional study of Dynamic digital radiography and Pneumotachograph in Healthy Volunteers, sponsored by Shenzhen People's Hospital. Status unknown. Open to participants aged 18 Years and older, including healthy volunteers. Per ClinicalTrials.gov, last updated 2022-10-04.

Sponsored by Shenzhen People's Hospital · Not applicable, Interventional, and Other

The sponsor has not verified this record recently (last verified Sep 2022), so the status shown — last known as Not yet recruiting — may be out of date.
Phase
Not applicable
Study type
Interventional
Enrollment
20
Allocation
Non-randomized
Ages
18 Years and older
Sex
All
01

Study summary

The digital X-ray system produced by Shenzhen Angell Technology Co., Ltd was selected as the tested equipment, and the powerlab biological signal acquisition and analysis system equipped with a pneumotachograph by ADInstruments was selected as the reference equipment. During the study, the lung volume measurements obtained by the lung function analysis method based on the dynamic digital radiography full respiratory cycle photography of the tested equipment was compared with the lung volume measurements obtained by the reference device pneumotachograph. The consistency of the lung volume measurements was used as the main validity evaluation index to observe the distribution of the differences between the measured values of the test device and the reference device, and to verify the validity of the declared product according to the main validity evaluation index.

Read the detailed description

Pulmonary function test is a test to explore the functional status of human respiratory system with the knowledge of exercise respiratory physiology and modern examination technology. It is an important test content for clinical diagnosis, severity assessment, treatment and prognosis assessment of chest and lung diseases, which has been widely used in the fields of respiratory medicine, surgery and anesthesiology. Compared with chest X-rays and CT examinations, which reflect the static function of the chest and lungs, pulmonary function tests can better reflect the dynamic functional changes of the body.

However, pulmonary function test is not well performed to assess the functional status in patients with poor cooperation (e.g., poor lung function, inability to understand cooperation). Dynamic digital X-ray whole respiratory cycle photography uses advanced dynamic radiography and acquisition technology to obtain high-definition dynamic lung images during the whole respiratory cycle of the patient, and uses the industry's first rotating foot pedal to rotate the patient to obtain weighted 3D scan data and single-angle or multi-angle lung data. The patient's lung function is analyzed by algorithmically extracting the patient's full respiratory cycle airway nodes and branches and data statistics. For patients who cannot cooperate with the lung function, it can provid information on the dynamic function of the patient's chest and lungs for doctors to further improve clinical diagnosis and treatment.

At present, the digital X-ray system (DTP580B-3) and the patient position fixation frame MRW-P1 manufactured by Shenzhen Angell Technology Co., Ltd. have passed the inspection by the inspection agency certified by the State Drug Administration of China (NMPA). Clinical trials were conducted to further evaluate the clinical effectiveness and safety of the products.

The lung volume was measured using a pneumotachograph (with a physiological signal recorder) and a digital X-ray system (UC dynamic DR) in the same healthy subject, and the lung volume values measured by each of the reference device were analyzed for consistency to evaluate the validity of the lung volume measurement by the test device. And To evaluate the safety of the lung volume measurement function of the test device in clinical use based on the occurrence of device malfunctions, adverse events and serious adverse events in the trial.

02

Conditions studied

  • Healthy Volunteers

Keywords

  • dynamic digital X-ray
  • pulmonary function analysis
  • healthy volunteers
03

In context

Lead sponsor

Shenzhen People's Hospital is the lead sponsor of 62 studies on the registry; 7 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
Yes

Inclusion criteria

① age ≥ 18 years; ② subjects willing to undergo chest radiography; ③ normal pulmonary ventilation function (forced vital capacity (FVC) ≥ 80% of the expected value; forced expiratory volume in first second (FEV1) ≥ 80% of the expected value; 1 second rate (FEV1/FVC) ≥ 70%); ④ can perform tidal breathing, deep inspiration and deep expiration; ⑤ no history of smoking; ⑥ no history of chronic respiratory disease; ⑦ no acute respiratory disease within 4 weeks; ⑧ subjects willing to provide written informed consent

Exclusion criteria

Exclusion Criteria:

① Pregnant and lactating women; ② Inability to stand alone; ③ Severe vertigo and other adverse reactions to rotation in place.

05

Study design

Phase
Not applicable
Primary purpose
Other
Allocation
Non-randomized
Intervention model
Parallel assignment
Masking
None (open label)
Enrollment
20 participants (estimated)

Study arms

  • Active comparator
    Pneumotachograph

    The subject stood on the turntable of the instrument and grasped the postural fixator with both hands up. The subjects pinched their nose and breathed through the disposable filter mouthpiece connected to the pneumotachograph, taking care not to leak air at the corners of the mouth. The subjects were instructed to perform tidal breathing, deep inhalation, deep exhalation, etc., and to monitor lung volume by pneumotachograph.

    Device: Pneumotachograph

  • Experimental
    Dynamic digital radiography

    The subject stood on the turntable of the instrument and grasped the postural fixator with both hands up. The subjects pinched their nose and breathed through the disposable filter mouthpiece connected to the pneumotachograph, taking care not to leak air at the corners of the mouth. The subjects were instructed to perform tidal breathing, deep inhalation, deep exhalation, etc., and to monitor lung volume by dynamic digital radiography analysis.

    Device: Dynamic digital radiography

Interventions

  • DeviceDynamic digital radiography

    The use of advanced dynamic X-ray photography and acquisition technology can obtain high-definition dynamic lung images of patients throughout the respiratory cycle for analysis.

  • DevicePneumotachograph

    The pneumotachograph monitors the flow rate, and the lung volume is calculated by integrating.

06

What researchers measure

Primary outcomes

  1. Lung volume changes

    The volume of gas inhaled/exhaled during each respiratory session of the subject during the monitoring period.

    Time frame: 1 day (The start time is when the subject is positioned and begins to breathe on command, and the end time is when the subject safely exits the x-ray system.)

07

Study locations

No study locations are listed for this record.

08

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Oct 4, 2022, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
09

Registry details

Key details

Study ID
NCT05565027
Lead sponsor
Shenzhen People's Hospital
Responsible party
Sponsor
First posted
Oct 4, 2022
Start date
Oct 20, 2022 (estimated)
Primary completion
Jul 20, 2023 (estimated)
Completion
Oct 20, 2023 (estimated)
Last update
Oct 4, 2022

Study contacts

Jianyi Niu, MD
Contact
niujianyi001@163.com
+8617825846046
Lili Guan, PhD
Contact
dr_nickguan@163.com
+8617825846046
Rongchang Chen, PhD
study chair · Shenzhen People's Hospital

Oversight

FDA-regulated drug
No
FDA-regulated device
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is status unknown, as verified in Sep 2022. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion