AED-delivery Using Drones in Suspected OHCA

An interventional study of Automated flying Drone carrying an Automated external defibrillator (AED) in Out-Of-Hospital Cardiac Arrest, sponsored by Karolinska Institutet. Completed at 1 site in Sweden. Open to participants aged 8 Years and older. Per ClinicalTrials.gov, last updated 2022-06-06.

Sponsored by Karolinska Institutet · Not applicable, Interventional, and Treatment

Phase
Not applicable
Study type
Interventional
Enrollment
59
Allocation
Not applicable
Ages
8 Years and older
Sex
All
01

Study summary

Time to defibrillation is the most important predictor of survival in cardiac arrest. Traditional emergency medical system response is often to slow. The overall aim of this pilot study is to evaluate if drones can deliver Automated external defibrillators (AEDs) to the scene of suspected out-of-hospital cardiac arrest (OHCA) prior to ambulance arrival with clinically relevant time benefits.

Read the detailed description

This is an observational, non-randomized single group assignment study with the primary purpose of evaluating proportion of AED-delivery prior to ambulance arrival and time benefits.

Totally, five AED-equipped drones will be placed in hangars at five different sites covering a geographic area of about 200 km2 and approximately 125,000 inhabitants for dispatch to suspected cases of OHCA as a complement to standard care i.e., ambulance. By coordinating the operations with local Air Traffic Control (ATC), the risk of airspace conflicts, or in worst case collisions, is mitigated to an acceptable level.

The investigators estimate to enroll some 80 alerts and 60 participants during a study period of up to 14 months. The investigators will closely follow and study the potential of delivering AEDs prior ambulance arrival and timesaving effects as compared to ambulance during a period of up to 14 months in which the investigators will follow and study the potential of delivering AEDs prior ambulance arrival and timesaving effects as compared to ambulance.

The drone operator, Everdrone, has integrated all software and hardware for the concept on the DJI Matrice 600 Pro drone. Examples of added features: an integrated Parachute Rescue System (validated by independent third party using the ASTM F3322 - 18 standard), a high precision sense and avoid system and a winching device to safely lower the AED to ground at delivery. Furthermore, Everdrone AB has fully integrated their system and operational routines with the dispatch central, emergency medical services and aviation organizations. As a result from the previous study (version 1.0) a number of hardware and software improvements have been made to the system (version 2.0):

Five hangars will be setup at optimal locations in each of the five administrative areas. Hangars are large isolated tents 4x4 meters in where the drone is placed, remotely operated and surveiled by camera. Hangars are within fenced areas so that the public has no access. They are equipped with weather stations and an automated gate which automatically will open on alert.

When an emergency 112-call is answered it is indexed by the dispatcher depending on the patients condition. If an OHCA is suspected, the dispatcher indexes the event as a)"ongoing cardio-pulmonary resuscitation (CPR)" or b) "Unconscious with abnormal or absent breathing". If an emergency call is identified within administrative areas AND indexed as "a" or "b" an automated (not manual as with ambulance alerts) alert is sent via radio-alert to the drone operator. This triggers:

  • boot-up of the drone
  • drone hangar gate opening
  • route-planning of the mission
  • weather condition confirmation
  • remote pilot acceptance of the mission
  • remote pilot calling ATC for takeoff approval

The drone system receives an alert from the dispatch center and the drone takes off from the hangar after approval from the drone pilot. For safety reasons all beyond visual line of sight (BVLOS) flights are done within the controlled airspace of Säve airport approval from ATC granted for every single flight. Each flight will be carried out at 60-80 meters of altitude, have a preliminary maximum flight speed of 70 km/h, and a maximum range of 6 km one-way. The drone system is remotely deployed and surveiled from the drone operating center at Everdrone Säve Airport, the drone is dispatched alongside ambulance as a complement to standard care. The drone systems are designed to operate in dry conditions with median winds not exceeding 8 m/s and primarily during daylight conditions.

The AED used is an ultra-light AED (FRED Easyport from Schiller), weighting totally 800 grams with basket and siren. Upon arrival onsite the basket with the AED will be winched down from about 30 m altitude and delivered within 10 meters from the object i.e., the front door or onsite in public areas. Local AED drone support after completed mission includes electrode change battery test, potentially pic-up up the drone after dispatch by the Everdrone personnel.

Dispatch of the drone in suspected OHCA cases will be executed as a complement to standard care i.e., ambulance between 21 April to 1 June 2022, totally up to 14 months of full-scale real-life flights to suspected OHCA victims. Based on historical data of OHCA from the Swedish register for cardiopulmonary resuscitation (SRCR) and first ever dispatch of AED-equipped drones during 2020 the drone systems will be alerted in an estimated 80 cases (about 3.5 flights/ week) of suspected OHCA over the study period. In about 25% of cases, predefined exclusion criteria (rain, median wind > 8 m/s, high rise buildings >20m, pre-flight checklists not OK) will presumably prohibit flights, thus making an estimated 60 cases eligible for inclusion (i.e., drone takes off to deliver AEDs in suspected OHCA.)

For more details about the methods please see the research protocol.

02

Conditions studied

  • Out-Of-Hospital Cardiac Arrest

Keywords

  • Drone
  • AED
  • OHCA
  • EMS
  • Dispatch centre
  • UAV
03

In context

Heart Arrest

966 studies on the registry are indexed under Heart Arrest; 226 are open to participants now.

This study's enrollment of 59 is below the median of 100 across 557 interventional studies indexed under Heart Arrest.

Browse Heart Arrest studies →

Lead sponsor

Karolinska Institutet is the lead sponsor of 1,113 studies on the registry; 267 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
8 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Inclusion criteria

  • All suspected OHCA during 112-calls including drowning
  • Within prespecified administrative areas (excluding no delivery- and no-fly zones)
  • Hours of operation: 08:00-22:00 (daylight conditions)
  • Drone system online

Exclusion criteria

Exclusion Criteria:

Pre-alert:

  • Children \<8 years
  • Trauma
  • EMS-witnessed cases

Post alert:

  • ATC non-approval of flight
  • Rain, winds exceeding 8m/s (median)
05

Study design

Phase
Not applicable
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
59 participants (actual)

Study arms

  • Experimental
    Automated flying Drone carrying an Automated external defibrillator (AED)

    Five drone systems are setup to be deployed in suspected OHCA cases as a complement to EMS. This is a single-arm intervention evaluating time benefit of drone delivery of AEDs in suspected OHCA.

    Device: Automated flying Drone carrying an Automated external defibrillator (AED)

Interventions

  • DeviceAutomated flying Drone carrying an Automated external defibrillator (AED)

    Totally n=5 drones are equipped with automated external defibrillators (AEDs). These drones are deployed by the dispatch centre to cases of suspected out-of-hospital cardiac (OHCA) as a complement to standard care (ambulance/EMS) over up to 14 months during daytime Monday to Sunday 08:00-22:00. The bystander onsite receives instructions from the dispatcher to retrieve the AED outside the house when it has been delivered by the drone. The bystander attaches the AED to the patients chest to facilitate early defibrillation.

06

What researchers measure

Primary outcomes

  1. Proportion (%) of AED-drone arrival prior to ambulance.

    In all alerts of suspected OHCA where a drone took off, comparisons will be done on proportion (%) of AED-drones arriving first on scene vs the proportion (%) of ambulances arriving first on scene. Accurate completion of mission is defined as delivery of an AED onsite on the ground in close connection from the OHCA and drone operator reporting arrival status "F" back to the dispatch centre. AEDs shall be accessible to the bystander through T-CPR instructions given by the dispatcher.

    Time frame: Up to 14 months

Secondary outcomes

  1. Median time difference (min:sec) between AED-drone and ambulance.

    In all alerts of suspected OHCA where a drone took off, analysis of time benefit will be done in all cases when the AED-drone arrived first to the scene of a suspected OHCA. Median time difference (min:sec) between AED-drone and ambulance will be measured. Accurate completion of mission is defined as delivery of an AED onsite on the ground in close connection from the OHCA and drone operator reporting arrival status "F" back to the dispatch centre. Time benefit will be presented overall (% \> 3minutes) as well as in continuous data intervals i.e. proportions in: 0-1 min, 1-2 min, 2-3 min, 3-4 min.

    Time frame: Up to 14 months

  2. c) Proportion (%) of attached drone delivered AEDs prior the arrival of EMS. Percentages.

    In all alerts to suspected OHCA where a drone took off, a bystander retrieves the drone delivered AED as instructed in T-CPR protocol by the calltaker and attaches it to the patient before ambulance arrival. Percentages.

    Time frame: Up to 14 months

  3. d) Proportion (%) of cases defibrillated when a drone delivers an AED before ambulance arrival (real OHCA)

    In all alerts to suspected OHCA where a drone took off, a bystander attached the AED to the patient as instructed in T-CPR protocol by the calltaker and administers a shock before ambulance arrival. Percentages.

    Time frame: Up to 14 months

07

Study locations

1 site
  • Center for Resuscitation science
    Stockholm, Sweden
08

References and documents

Individual participant data

Plan to share: No

No publications or documents are linked to this record.

09

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Jun 6, 2022, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT04723368
Lead sponsor
Karolinska Institutet
Collaborators
SOS Alarm Sverige AB, Everdrone AB, Vastra Gotaland Region, Aviation capacity resources AB (ACR), Castellum AB
Responsible party
Andreas Claesson (RN, Associate professor, Karolinska Institutet) — Principal investigator
First posted
Jan 25, 2021
Start date
Apr 21, 2021
Primary completion
Jun 1, 2022
Completion
Jun 1, 2022
Last update
Jun 6, 2022

Oversight

FDA-regulated drug
No
FDA-regulated device
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Jun 2022. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion