SpO2 Accuracy In Vivo Testing for Neonates & Infants

An interventional study of MEDLINE RENEWAL PULSE OXIMETRY SENSORS and Reference CO-oximetry sensor in Pulse Oximetry, sponsored by Medline Industries. Withdrawn at 1 site in United States. Open to participants aged Up to 28 Days, including healthy volunteers. Per ClinicalTrials.gov, last updated 2024-08-28.

Sponsored by Medline Industries · Not applicable, Interventional, and Basic science

Why this study was withdrawn
Business decision
Phase
Not applicable
Study type
Interventional
Enrollment
0
Allocation
Non-randomized
Ages
Up to 28 Days
Sex
All
01

Study summary

The purpose of this clinical study is to validate the oxygen saturation (SpO2) accuracy of various pulse oximetry systems and sensors (new designs and/or reprocessed) during non-motion conditions as compared to arterial blood samples, drawn in the normal course of patient care, assessed by CO-Oximetry in neonates and infants. The primary end goal is to provide supporting documentation for the SpO2 accuracy of these pulse oximetry systems. The secondary end goal is to provide data for evaluation of clinical impact of measured to calculated SaO2 and related issues.

Read the detailed description

FDA Guidance for pulse oximeters (March 4, 2013), section 4.1.2, In vivo testing for SpO2 accuracy for neonates, calls for the devices to be first tested in adults during hypoxia tests with arterial samples compared to measured oxygen saturation. This has been previously completed on the devices that will be tested in this protocol. In order to verify safe clinical performance in the neonate, the FDA then requires testing using neonatal arterial samples that are collected in the normal course of care; which is the primary purpose of this protocol. As a secondary purpose the data will be utilized to evaluate the impact of calculated oxygen saturation versus measured oxygen saturation and related issues.

Neonates/infants who will have an arterial line or sheath placed in the normal course of care will be included in this study. The pulse oximeter sensor must be placed on the Neonates/infant's foot or hand. Neonates who either have repaired PDA but are having the placement of the sensor on the foot (which will be arterial blood only) will be allowed as long as other inclusion/exclusion criteria are met.

02

Conditions studied

  • Pulse Oximetry
03

In context

Lead sponsor

Medline Industries is the lead sponsor of 39 studies on the registry; none are open to participants now.

Of its 15 completed or terminated interventional studies of FDA-regulated products, 14 (93%) have results posted.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
Up to 28 Days
Sexes eligible
All
Accepts healthy volunteers
Yes

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Neonates/Infants up to 28 days old
  • Neonates/Infants less than 5kg.

And one of the following:

Without PDA With a PDA with R-L shunting that will be repaired before arterial samples are taken

Exclusion criteria

Exclusion Criteria:

  • Physical malformation of feet or toes or other sensor sites which would limit the ability to place the sensors for the study.
  • Neonates/Infants older than 28 days old.
  • Neonates/Infants outside the weight range for the sensor (Greater than 5kg.)
  • Neonates/Infants with unrepaired PDA. with PDA with R-L shunting
  • Subjects who are undergoing light therapy while samples are being obtained.
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Bilirubin

A yellow substance made when red blood cells are broken down, which the liver clears. A high level can mean a problem with the liver or the bile ducts.

Criteria often allow more for people with Gilbert’s syndrome, a harmless inherited condition that raises bilirubin.

05

Study design

Phase
Not applicable
Primary purpose
Basic science
Allocation
Non-randomized
Intervention model
Parallel assignment
Masking
None (open label)
Enrollment
0 participants (actual)

Study arms

  • Experimental
    MEDLINE RENEWAL PULSE OXIMETRY SENSORS

    Reference sensors from the reprocessed oximeter device will be placed on each subject to evaluate the SpO2 accuracy and performance.

    Device: MEDLINE RENEWAL PULSE OXIMETRY SENSORS

  • Sham comparator
    Reference CO-oximetry sensor

    A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.

    Device: Reference CO-oximetry sensor

Interventions

  • DeviceMEDLINE RENEWAL PULSE OXIMETRY SENSORS

    An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. Because an arterial sample of blood is not required to make the measurement, the pulse oximeter can provide non-invasive real time information.

  • DeviceReference CO-oximetry sensor

    A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.

06

What researchers measure

Primary outcomes

  1. Validation of SpO2 accuracy of pulse oximetry sensors

    To validate SpO2 accuracy performance of Sponsor sensors by an Accuracy Root Mean Square (Arms) value. Reference CO-Oximetry will be used as the basis for comparison for the investigational device SpO2 readings.

    Time frame: 4 hours

Secondary outcomes

  1. Calculated oxygen saturation versus measured oxygen saturation

    The arterial sample will always be measured in the reference co-oximeter. The sample may also be measured in the currently used AVOX 1000E.

    Time frame: 4 hours

07

Study locations

1 site
  • Children's Hospital Colorado
    Aurora, Colorado 80045, United States
08

References and documents

Individual participant data

Plan to share: Undecided

No publications or documents are linked to this record.

09

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Aug 28, 2024, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT03843489
Lead sponsor
Medline Industries
Collaborators
Clinimark, LLC
Responsible party
Sponsor
First posted
Feb 18, 2019
Start date
Mar 19, 2019 (estimated)
Primary completion
Sep 20, 2021 (estimated)
Completion
Sep 20, 2021 (estimated)
Last update
Aug 28, 2024

Study contacts

Robin Deterding, MD
principal investigator · Children's Hospital Colorado

Oversight

Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is withdrawn, as verified in Aug 2024. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion