Repeatability, Reproducibility and Comparison of Cirrus OCT, RTVue OCT, MS-39 OCT, and Insight 100 VHFDU

An interventional study of ArcScan Insight 100 very high-frequency digital ultrasound and Carl Zeiss Meditec Cirrus HD OCT 5000 in Corneal Thickness Measurement, Epithelial Thickness Measurement and LASIK Flap Thickness Measurement, sponsored by London Vision Clinic. Suspended at 1 site in United Kingdom. Open to participants aged 21 Years to 70 Years, including healthy volunteers. Per ClinicalTrials.gov, last updated 2022-04-26.

Sponsored by London Vision Clinic · Not applicable, Interventional, and Diagnostic

Why this study was suspended
Devices currently awaiting software updates
Phase
Not applicable
Study type
Interventional
Enrollment
242
Allocation
Non-randomized
Ages
21 Years to 70 Years
Sex
All
01

Study summary

The purpose of this study is to evaluate the repeatability and reproducibility of corneal, epithelial, and LASIK flap thickness using the Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT, and ArcScan Insight 100 very high-frequency digital ultrasound arc-scanner.

Read the detailed description

Very high-frequency (VHF) digital ultrasound was first used in 1993 to obtain images and measure the thickness of the cornea (the window at the front of the eye) and the layers within the cornea, such as the epithelium (the layer of skin on the surface of the cornea). A commercially available instrument, known as the Artemis, has been used in routine clinical practice since 2001. The ArcScan Insight 100 VHF digital ultrasound scanner is an updated version, which obtained CE mark in 2016.

Optical coherence tomography (OCT) is another method of measuring the thickness of the cornea and the corneal epithelium. OCT has been used for measuring corneal thickness since 1997 and the technology had evolved to also measure epithelial thickness since 2012. There are currently three OCT scanners capable of measuring epithelial thickness; Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT.

The aim of the study is to establish the repeatability (i.e. the variability in measurements taken by a single examiner during a single visit) and reproducibility (i.e. the variability in measurements taken in the same conditions between two examiners) for each of the four devices for measuring corneal thickness and epithelial thickness.

02

Conditions studied

  • Corneal Thickness Measurement
  • Epithelial Thickness Measurement
  • LASIK Flap Thickness Measurement
  • SMILE Cap Thickness Measurement

Keywords

  • VHF digital ultrasound
  • Optical coherence tomography
  • Corneal thickness
  • Epithelial thickness
  • LASIK
  • SMILE
03

In context

Lead sponsor

London Vision Clinic is the lead sponsor of 6 studies on the registry; 2 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
21 Years to 70 Years
Sexes eligible
All
Accepts healthy volunteers
Yes

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

Only patients who are medically suitable for corneal laser refractive surgery can be included in the study.

As general inclusion criteria the following aspects are defined:

  • Subjects should be 21 years of age or older.
  • Contact lens wearers must stop wearing their contact lenses at least four weeks per decade of wear before baseline measurements in case of hard contact lenses and one week before baseline measurements in case of soft contact lenses.
  • Patient will be able to understand the patient information and willing to sign an .

For the normal eye populations:

  • Eyes with no ocular pathology
  • Eyes with no previous ocular or corneal surgery

For the post-op eye populations:

  • Eyes between 3 and 9 months after LASIK or SMILE

Exclusion criteria

Exclusion Criteria:

  • Previous ocular or corneal surgery (for normal eyes)
  • Patient not being able to tolerate local or topical anesthesia
  • Patient has active conjunctivitis or any active ocular infection or inflammation
  • Patient has a corneal ulcer
  • Patient has an open wound on the corneal epithelium
  • Patient has compromised epithelium (e.g. ≥Grade 3 superficial punctate keratitis)
  • Pregnancy
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorised to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

05

Study design

Phase
Not applicable
Primary purpose
Diagnostic
Allocation
Non-randomized
Intervention model
Parallel assignment
Masking
None (open label)
Enrollment
242 participants (estimated)

Study arms

  • Experimental
    Repeatability and Reproducibility - Normal eyes

    This arm will include 20 eyes of 20 patients with no previous ocular surgery. The research participant will be scanned a number of times using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). The scan sequence will be undertaken on a single day as follows: 1. 5 consecutive repeated scans of the cornea will be performed by the first operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set). 2. There will be a break of about 30 minutes. 3. 5 consecutive repeated scans of the cornea will be performed by the second operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set).

    Device: ArcScan Insight 100 very high-frequency digital ultrasound · Device: Carl Zeiss Meditec Cirrus HD OCT 5000 · Device: Optovue RTVue OCT · Device: CSO MS-39 OCT

  • Experimental
    Repeatability and Reproducibility - Post-op eyes

    This arm will include 20 eyes of 20 patients between 3 and 9 months after corneal laser refractive surgery. The research participant will be scanned a number of times using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). The scan sequence will be undertaken on a single day as follows: 1. 5 consecutive repeated scans of the cornea will be performed by the first operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set). 2. There will be a break of about 30 minutes. 3. 5 consecutive repeated scans of the cornea will be performed by the second operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set).

    Device: ArcScan Insight 100 very high-frequency digital ultrasound · Device: Carl Zeiss Meditec Cirrus HD OCT 5000 · Device: Optovue RTVue OCT · Device: CSO MS-39 OCT

  • Experimental
    Comparison between devices - Normal eyes

    This arm will include 101 eyes of 101 patients with no previous ocular surgery. The research participant will be scanned once using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). Expected total time approximately 20 minutes to complete all four scans.

    Device: ArcScan Insight 100 very high-frequency digital ultrasound · Device: Carl Zeiss Meditec Cirrus HD OCT 5000 · Device: Optovue RTVue OCT · Device: CSO MS-39 OCT

  • Experimental
    Comparison between devices - Post-op eyes

    This arm will include 101 eyes of 101 patients between 3 and 9 months after corneal laser refractive surgery. The research participant will be scanned once using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). Expected total time approximately 20 minutes to complete all four scans.

    Device: ArcScan Insight 100 very high-frequency digital ultrasound · Device: Carl Zeiss Meditec Cirrus HD OCT 5000 · Device: Optovue RTVue OCT · Device: CSO MS-39 OCT

Interventions

  • DeviceArcScan Insight 100 very high-frequency digital ultrasound

    The ArcScan Insight 100 VHF digital ultrasound scanner can measure individual layers within the cornea with very high precision, as well as being able to image the interior of the eye including the chamber between the iris and the cornea, the lens and other structures behind the iris. The ArcScan Insight 100 device is a digital ultrasound scanner, meaning that it uses ultrasound waves to measure parameters of the eye. The Insight examination involves the patient sitting in front of the ultrasound unit and looking through a disposable goggle-like eyepiece at a light source. Warm sterile balanced saline solution (like natural tears) are used to fill the space between the light source and the eye, to allow ultrasound transmission. There is no contact between the instrument and the eye other than being immersed in this water bath.

    Also known as: Artemis very high-frequency digital ultrasound

  • DeviceCarl Zeiss Meditec Cirrus HD OCT 5000

    The Cirrus HD OCT 5000 uses high definition spectral domain optical coherence tomography (OCT) to measure anatomical structures of the eye. OCT is non-contact and uses light, rather than ultrasound, to obtain cross-sectional images. An OCT scan involves the patient sitting in front of the device, placing their chin on a chin-rest and fixating on a light source. Nothing comes into contact with the eye and the patient is only aware of the instrument rotating in front of them.

  • DeviceOptovue RTVue OCT

    The Optovue RTVue OCT uses high definition Fourier/spectral domain optical coherence tomography (OCT) to measure anatomical structures of the eye. OCT is non-contact and uses light, rather than ultrasound, to obtain cross-sectional images. An OCT scan involves the patient sitting in front of the device, placing their chin on a chin-rest and fixating on a light source. Nothing comes into contact with the eye and the patient is only aware of the instrument rotating in front of them.

  • DeviceCSO MS-39 OCT

    The CSO MS-39 OCT uses high definition Fourier domain optical coherence tomography (OCT) to measure anatomical structures of the eye. OCT is non-contact and uses light, rather than ultrasound, to obtain cross-sectional images. The MS-39 also includes a Placido topographer within the device to simultaneously obtain a measurement of the shape (curvature) of the front surface of the cornea, which is then combined with the OCT thickness measurement to generate the shape of the back surface of the cornea. An OCT scan involves the patient sitting in front of the device, placing their chin on a chin-rest and fixating on a light source. Nothing comes into contact with the eye and the patient is only aware of the instrument rotating in front of them.

06

What researchers measure

Primary outcomes

  1. Repeatability and reproducibility of Cirrus HD OCT 5000 corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Time frame: 10 minutes (5 minutes per observer)

  2. Repeatability and reproducibility of RTVue OCT corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Time frame: 10 minutes (5 minutes per observer)

  3. Repeatability and reproducibility of MS-39 OCT corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Time frame: 10 minutes (5 minutes per observer)

  4. Repeatability and reproducibility of Insight 100 VHF digital ultrasound corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Time frame: 30 minutes (15 minutes per observer)

  5. Repeatability and reproducibility of Cirrus HD OCT 5000 corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Time frame: 10 minutes (5 minutes per observer)

  6. Repeatability and reproducibility of RTVue OCT corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Time frame: 10 minutes (5 minutes per observer)

  7. Repeatability and reproducibility of MS-39 OCT corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Time frame: 10 minutes (5 minutes per observer)

  8. Repeatability and reproducibility of Insight 100 VHF digital ultrasound corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Time frame: 30 minutes (15 minutes per observer)

  9. Difference in corneal thickness measurements between Cirrus HD OCT 5000, RTVue OCT, MS-39 OCT, and Insight 100 VHF digital ultrasound in normal eyes

    The difference between each pair of devices for measurements of corneal thickness and epithelial thickness in eyes with no previous ocular surgery.

    Time frame: 20 minutes for all four scans

  10. Difference in corneal thickness measurements between Cirrus HD OCT 5000, RTVue OCT, MS-39 OCT, and Insight 100 VHF digital ultrasound in post-op eyes

    The difference between each pair of devices for corneal thickness, epithelial thickness and interface thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Time frame: 20 minutes for all four scans

07

Study locations

1 site
  • London Vision Clinic
    London, W1G 7LA, United Kingdom
08

References and documents

Publications

  • Reinstein DZ, Archer TJ, Gobbe M, Silverman RH, Coleman DJ. Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner. J Refract Surg. 2010 Sep;26(9):646-59. doi: 10.3928/1081597X-20091105-01. Epub 2009 Nov 16. PubMed 19928698 ↗
  • Reinstein DZ, Yap TE, Archer TJ, Gobbe M, Silverman RH. Comparison of Corneal Epithelial Thickness Measurement Between Fourier-Domain OCT and Very High-Frequency Digital Ultrasound. J Refract Surg. 2015 Jul;31(7):438-45. doi: 10.3928/1081597X-20150623-01. PubMed 26186562 ↗
  • Yap TE, Archer TJ, Gobbe M, Reinstein DZ. Comparison of Central Corneal Thickness Between Fourier-Domain OCT, Very High-Frequency Digital Ultrasound, and Scheimpflug Imaging Systems. J Refract Surg. 2016 Feb;32(2):110-6. doi: 10.3928/1081597X-20151223-01. PubMed 26856428 ↗
  • Kiraly L, Stange J, Kunert KS, Sel S. Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices. J Ophthalmol. 2017;2017:6181405. doi: 10.1155/2017/6181405. Epub 2017 Mar 5. PubMed 28357136 ↗

Individual participant data

Plan to share: No — No data will be shared with other researchers

09

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reached

Start is estimated for Sep 2022; primary completion for Dec 2023, 15 months later.

How these figures are worked out, and which studies are left out of them.

Registered
Mar 8, 2018
First posted
Mar 22, 2018
StartEst.
Sep 1, 2022
Primary completionEst.
Dec 1, 2023
CompletionEst.
Dec 1, 2023
Last update
Apr 26, 2022
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Apr 26, 2022, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT03473847
Lead sponsor
London Vision Clinic
Responsible party
Sponsor
First posted
Mar 22, 2018
Start date
Sep 1, 2022 (estimated)
Primary completion
Dec 1, 2023 (estimated)
Completion
Dec 1, 2023 (estimated)
Last update
Apr 26, 2022

Study contacts

Dan Z Reinstein, MD, MA, FRCOphth
principal investigator · London Vision Clinic

Oversight

Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is suspended, as verified in Apr 2022. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow