An interventional study of MiniMed 670G system in Type1diabetes, Hypoglycemia Unawareness and Nocturnal Hypoglycemia, sponsored by University of Pennsylvania. Completed at 3 sites in United States. Open to participants aged 25 Years to 70 Years. Per ClinicalTrials.gov, last updated 2024-10-24.
Sponsored by University of Pennsylvania · Not applicable, Interventional, and Treatment
Can a type 1 diabetic adult avoid low glucoses and regain hypoglycemia awareness using a hybrid closed loop insulin delivery system? Involvement is 22 months (13 visits) and includes a 4-week Screening Phase and an 18-month Intervention Phase. Participants will undergo 3 Hyperinsulinemic Clamps done at: Baseline (before starting the device and after completing the screening), 6 months (after using the device 6 months), and after using the device for 18 months. This metabolic testing will allow us to measure improvement in hypoglycemia awareness.
The purpose of this small, proof-of-concept, mechanistic study is to determine whether hybrid closed-loop insulin delivery can achieve sufficient hypoglycemia avoidance in patients with long standing type 1 diabetes experiencing hypoglycemia and symptom unawareness despite receiving intensive insulin therapy standard-of-care to improve glucose counterregulation against insulin-induced hypoglycemia. As it remains critically important to understand the potential physiologic benefits of hypoglycemia avoidance by means of a hybrid closed-loop system in this population, studying each well characterized subject as his/her own control by a within subject design should serve the study purpose, and has been the standard approach to assessing the effects of various approaches to hypoglycemia avoidance in patients with unawareness. The data generated will be available to power future randomized clinical trials to determine the comparative efficacy of emerging artificial pancreas and β-cell replacement approaches to achieve target glycemic control with amelioration of problematic hypoglycemia in type 1 diabetes.
Screening Phase At the screening visit, the study details and procedures will be discussed with a research coordinator and at least one of the PI or the research nurse practitioner. The potential participant is given adequate time to ask questions and review the informed consent document. Once satisfied that all questions have been answered, the potential participant will either decline to participate or sign the informed consent document. This may occur at a subsequent visit if the potential participant desires, in order to think further about what participation means and/or to consult with family, friends and/or a personal physician. The consent form is signed in the presence of a witness (research coordinator +/- family member). All participants must read, sign, and date a consent form before entering the study, undergoing physical examination or undergoing any testing. The informed consent form will be revised whenever important new safety information is available, whenever the protocol is amended, and/or whenever any new information becomes available that may affect participation in the studies.Eligibility will be confirmed through the performance of a history and physical examination by the PI or the research nurse practitioner, EKG, urine pregnancy test (if applicable), serum chemistries, TSH, cell counts, HbA1c and C-peptide, completion of glycemic lability and hypoglycemia awareness and hypoglycemia severity questionnaires, placement of a 7 day blinded CGM (iPro 2, Medtronic Diabetes, Northridge, CA) unless on CGM available for downloading, and 7 day accelerometry (WGT3X-BT, Actigraph LLC,) to define the nocturnal period.
Only after all eligibility criteria (inclusion and exclusion) are met, will a potential subject be enrolled. Repeated clinical testing throughout the study will ensure the continued safety and minimization of risk for the enrolled participants.
Study Intervention Phase Eligible subjects will complete a baseline assessment of glucose counterregulation by stepped- hyperinsulinemic hypoglycemic clamp prior to starting intervention with the hybrid closed-loop system (MiniMed 670G system, Medtronic Diabetes, Northridge, CA). Unless this system becomes available to the subject via their insurance carrier, one will be provided for them. Subjects will also receive a study glucometer (Contour Next Link 2.4, Bayer, Indianapolis, IN) that communicates with the MiniMed 670G insulin pump for bolus dosing calculation and glucose sensor calibration. Subjects who cannot maintain > 80% (or 6/7 day) compliance with the sensor component as assessed at each study visit may be dropped since less compliance has not been associated with any benefit of CGM to glycemic control and limits the potential for benefit from LGS on hypoglycemia avoidance. Accuracy of the sensor will be assessed at each visit through devise download and interpretation. Study visits will occur weekly for the first month, then monthly until month 6, and then every 3 months until month 18. This schedule will allow for determination of possible benefit from hybrid closed- loop insulin delivery on glucose counterregulation after 6 months of intensive provider support, and then for assessment of the durability or potential further gains in beneficial effects after another 12 months of more typical provider interaction occurring every 3 months. Weekly visits may be performed via telephone with uploading devise data to Care Link for review and interpretation. Uploaded or downloaded insulin delivery, blood and sensor glucose monitoring, insulin dose settings and CGM calibration accuracy, alert settings, time spent in auto and manual modes, and LGS threshold and activity will be assessed at each visit, targeting > 80% CGM and LGS compliance, adjusting basal and bolus insulin dosing in order to minimize glycemic excursions while maximizing hypoglycemia (\< 60 mg/dl) avoidance, with adjustment of alarms set to alert the subject to rapidly increasing or decreasing glucose and predict the occurrence of elevated or low blood glucose. During manual mode, target glucose ranges will be 90 - 140 mg/dl before meals, \< 180 mg/dl after meals, and 120 - 160 mg/dl at bedtime, with correction dosing to no lower than 100 mg/dl during the day, and 120 mg/dl overnight. Alarm settings may be individualized to target these ranges, but the hypoglycemia alarm for LGS will not be set lower than 70 mg/dl. During auto mode, the automated interprandial basal insulin delivery will adjust according to the closed-loop algorithm to target a sensor glucose of 120 mg/dl, which may be temporality increased to 150 mg/dl if needed to further minimize exposure to hypoglycemia during exercise or overnight. Prior to each 3 monthly visit, subjects will wear an actigraph monitor ( WGT3X-BT, Actigraph LLC,) for three weeks in order to define the nocturnal period. Every 6 months measures of hypoglycemia awareness (Clarke score) and severity (HYPO score), and the glycemic lability index (LI) will be calculated from questionnaires, event diaries, and device downloads, respectively. At 6 months and at 18 months, subjects will again undergo assessment of glucose counterregulation by stepped-hyperinsulinemic hypoglycemic clamp testing.
3,522 studies on the registry are indexed under Diabetes Mellitus, Type 1; 577 are open to participants now.
This study's enrollment of 10 is below the median of 40 across 2,649 interventional studies indexed under Diabetes Mellitus, Type 1.
Browse Diabetes Mellitus, Type 1 studies →University of Pennsylvania is the lead sponsor of 1,635 studies on the registry; 239 are open to participants now.
Of its 154 completed or terminated interventional studies of FDA-regulated products, 104 (68%) have results posted.
Counted across the registry records on this site, refreshed daily.
Exclusion Criteria:
Eligible subjects with type 1 diabetes will initiate hybrid closed-loop insulin delivery based on interstitial glucose monitoring via the MiniMed 670G system according to Medtronic's labeling. This system combines subject-delivered pre-meal boluses with automatic interprandial insulin delivery that includes automated functions for both predictive and threshold suspension of insulin delivery intended to minimize exposure to glucose levels \< 70 mg/dl.
Device: MiniMed 670G system
Eligible subjects with type 1 diabetes will initiate hybrid closed-loop insulin delivery based on interstitial glucose monitoring via the MiniMed 670G system according to Medtronic's labeling. This system combines subject-delivered pre-meal boluses with automatic interprandial insulin delivery that includes automated functions for both predictive and threshold suspension of insulin delivery intended to minimize exposure to glucose levels \< 70 mg/dl.
Endogenous Glucose Production
The primary outcome measure will be endogenous glucose production in response to insulin-induced hypoglycemia.
Time frame: After 6 months of hybrid closed-loop insulin delivery.
Endogenous Glucose Production
The secondary outcome measure will be endogenous glucose production in response to insulin-induced hypoglycemia.
Time frame: After 18 months of hybrid closed-loop insulin delivery.
| Milestone | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Started | 10 |
| Completed | 10 |
| Not completed | 0 |
The primary outcome measure will be endogenous glucose production in response to insulin-induced hypoglycemia.
| [mg/(kgxmin)] | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Endogenous Glucose Production | .66 ± .13 |
The secondary outcome measure will be endogenous glucose production in response to insulin-induced hypoglycemia.
| [mg/(kgxmin)] | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Endogenous Glucose Production | .69 ± .11 |
Collected over 18 months. Non-serious events are listed at a 0% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| Subjects With Type 1 Diabetes Using MiniMed 670G System | 0/10 (0%) | 0/10 (0%) | 8/10 (80%) |
| Event | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| HypoglycemiaEndocrine disorders | 4/10 |
| URIInfections and infestations | 2/10 |
| Bacterial InfectionsInfections and infestations | 2/10 |
18 participants were consented. Of the 18, 6 did not meet inclusion and 2 were dropped.10 were enrolled and completed the study.
| Age, Categorical(Participants) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| <=18 years | 0 |
| Between 18 and 65 years | 9 |
| >=65 years | 1 |
| Age, Continuous(years) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Mean | 49 ± 16 |
| Sex: Female, Male(Participants) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Female | 7 |
| Male | 3 |
| Ethnicity (NIH/OMB)(Participants) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Hispanic or Latino | 0 |
| Not Hispanic or Latino | 10 |
| Unknown or Not Reported | 0 |
| Race (NIH/OMB)(Participants) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| American Indian or Alaska Native | 0 |
| Asian | 0 |
| Native Hawaiian or Other Pacific Islander | 0 |
| Black or African American | 0 |
| White | 10 |
| More than one race | 0 |
| Unknown or Not Reported | 0 |
| Region of Enrollment(participants) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| United States | 10 |
| Endogenous Glucose Produciton([mg/(kgxmin)]) | Subjects With Type 1 Diabetes Using MiniMed 670G System |
|---|---|
| Mean | .73 ± .23 |
Documents are hosted by the registry — open the source record to download them.
Plan to share: Yes — It is not yet known if there will be any further plan to make IPD available besides the Informed Consent.
Supporting information: Icf
This study is completed, as verified in Oct 2024. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.