A Phase 4 interventional study of Crystalens AO and ReSTOR in Cataract, sponsored by Bausch & Lomb Incorporated. Completed at 1 site in United States. Open to participants aged 40 Years and older. Per ClinicalTrials.gov, last updated 2014-03-28.
Sponsored by Bausch & Lomb Incorporated · Phase 4, Interventional, and Treatment
The objective of this study is to evaluate the efficacy and subject satisfaction with three different FDA approved lenses for adults over 40 years of age who desire a reduction in spectacle dependence.
1,701 studies on the registry are indexed under Cataract; 230 are open to participants now.
This study's enrollment of 78 is close to the median of 80 across 1,155 interventional studies indexed under Cataract.
Browse Cataract studies →Bausch & Lomb Incorporated is the lead sponsor of 219 studies on the registry; 7 are open to participants now.
Of its 49 completed or terminated interventional studies of FDA-regulated products, 41 (84%) have results posted.
Counted across the registry records on this site, refreshed daily.
Exclusion Criteria:
Eligible subjects to undergo small incision cataract surgery and were implanted with the Crystalens AO bilaterally.
Device: Crystalens AO
Eligible subjects to undergo small incision cataract surgery and were implanted with the ReSTOR IOL bilaterally.
Device: ReSTOR
Eligible subjects to undergo small incision cataract surgery and were implanted with the Tecnis Multifocal IOL bilaterally.
Device: Tecnis Multifocal IOL
Crystalens AO surgically implanted bilaterally. Study observation up to 180 days.
ReSTOR surgically implanted bilaterally. Study observation up to 180 days.
Tecnis surgically implanted bilaterally. Study observation up to 180 days.
Mesopic Monocular Contrast Sensitivity Without Glare 1.5 and 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity Without Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity Without Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity Without Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity Without Glare 1.5 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 4 ( 4-6 months)
Mesopic Monocular Contrast Sensitivity Without Glare 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 4 ( 4-6 months)
Mesopic Monocular Contrast Sensitivity Without Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 4 ( 4-6 months)
Mesopic Monocular Contrast Sensitivity Without Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 4 ( 4-6 months)
Mesopic Monocular Contrast Sensitivity Without Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Postoperative visit 4 ( 4-6 months)
Binocular Mesopic Contrast Sensitivity Without Glare 1.5 & 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Binocular Mesopic Contrast Sensitivity Without Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Binocular Mesopic Contrast Sensitivity Without Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Binocular Mesopic Contrast Sensitivity Without Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Binocular Mesopic Contrast Sensitivity Without Glare 1.5, 3, 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Binocular Mesopic Contrast Sensitivity Without Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Binocular Mesopic Contrast Sensitivity Without Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Monocular Contrast Sensitivity With Glare 1.5 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity With Glare 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity With Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity With Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity With Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Monocular Contrast Sensitivity With Glare 1.5 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Monocular Contrast Sensitivity With Glare 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Monocular Contrast Sensitivity With Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Monocular Contrast Sensitivity With Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Monocular Contrast Sensitivity With Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Binocular Contrast Sensitivity With Glare 1.5, 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Binocular Contrast Sensitivity With Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Binocular Contrast Sensitivity With Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Binocular Contrast Sensitivity With Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 3 (2-3 months)
Mesopic Binocular Contrast Sensitivity With Glare 1.5, 3 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Binocular Contrast Sensitivity With Glare 6 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Binocular Contrast Sensitivity With Glare 12 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Mesopic Binocular Contrast Sensitivity With Glare 18 Cycles/Degree
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
Time frame: Visit 4 (4-6 months)
Optical Scatter
Objective scatter index (OSI) was assessed using the Optical Quality Analysis System(OQAS) and summarized as a continuous variable. For this assessment, an OSI score of 1 or less represents good quality vision, 1-2 indicates some degradation but otherwise normal vision, 2-3 significant light scatter possibly requiring treatment, and scores over 4 severely compromised vision requiring intervention.
Time frame: Visit 3 (2-3 months)
Optical Scatter
Objective scatter index (OSI) was assessed using the Optical Quality Analysis System(OQAS) and summarized as a continuous variable. For this assessment, an OSI score of 1 or less represents good quality vision, 1-2 indicates some degradation but otherwise normal vision, 2-3 significant light scatter possibly requiring treatment, and scores over 4 severely compromised vision requiring intervention.
Time frame: Visit 4 (4-6 months)
Monocular Halo and Starburst
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
Time frame: Visit 3 (2-3 months)
Monocular Halo and Starburst
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
Time frame: Visit 4 (4-6 months)
Binocular Halo and Starburst
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
Time frame: Visit 3 (2-3 months)
Binocular Halo and Starburst
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
Time frame: Visit 4 (4-6 months)
Manifest Refraction
Manifest refraction spherical equivalent (MRSE) was calculated as the value of the sphere plus one-half of the value of the cylinder.
Time frame: Visit 3 (2-3 Months)
Manifest Refraction
Manifest refraction spherical equivalent (MRSE) was calculated as the value of the sphere plus one-half of the value of the cylinder.
Time frame: Visit 4 (4-6 Months)
Uncorrected Distance Visual Acuity
High Contrast Uncorrected Distance (UCDVA) measured at 32 inches, Intermediate (UCIVA) at 32 inches, and Near Visual Acuity (UCNVA) at 16 inches
Time frame: Visit 3 (2-3 months) Visit 4 (4-6 months)
Best-corrected Distance Visual Acuity
High Contrast Distance-Corrected Distance,(BCDVA), Intermediate (DCIVA), and Near Visual Acuity (DCNVA)
Time frame: Visit 3 (2-3 months) Visit 4 (4-6 months)
This study took place at two investigative sites in the United States, with the first participant undergoing cataract surgery on 5/18/2010 and last participant completed the study on 6/27/2012.
| Milestone | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Started | 26 | 26 | 26 |
| Completed | 26 | 25 | 24 |
| Not completed | 0 | 1 | 2 |
| Withdrew: Withdrawal by subject | 0 | 1 | 2 |
| Milestone | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Started | 26 | 25 | 24 |
| Completed | 26 | 25 | 22 |
| Not completed | 0 | 0 | 2 |
| Withdrew: Central macular edema | 0 | 0 | 1 |
| Withdrew: Fellow eye explanted | 0 | 0 | 1 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| 1.5 cycles/degree | 1.494 ± 0.173 | 1.576 ± 0.209 | 1.410 ± 0.195 |
| 3 cycles/degree | 1.613 ± 0.171 | 1.679 ± 0.199 | 1.528 ± 0.175 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 6 Cycles/Degree | 1.563 ± 0.228 | 1.644 ± 0.253 | 1.481 ± 0.212 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 12 Cycles/Degree | 1.155 ± 0.238 | 1.335 ± 0.309 | 1.143 ± 0.189 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 18 Cycles/Degree | 0.943 ± 0.266 | 0.950 ± 0.297 | 0.697 ± 0.138 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 1.5 Cycles/Degree | 1.525 ± 0.203 | 1.588 ± 0.213 | 1.408 ± 0.172 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 3 Cycles/Degree | 1.638 ± 0.243 | 1.698 ± 0.221 | 1.519 ± 0.125 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 6 Cycles/Degree | 1.610 ± 0.241 | 1.622 ± 0.228 | 1.543 ± 0.261 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 12 Cycles/Degree | 1.229 ± 0.262 | 1.218 ± 0.238 | 1.217 ± 0.275 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity Without Glare 18 Cycles/Degree | 0.998 ± 0.297 | 0.789 ± 0.219 | 0.896 ± 0.275 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| 1.5 cycles/degree | 1.688 ± 0.221 | 1.749 ± 0.201 | 1.537 ± 0.129 |
| 3 cycles/degree | 1.775 ± 0.252 | 1.797 ± 0.213 | 1.667 ± 0.168 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Binocular Mesopic Contrast Sensitivity Without Glare 6 Cycles/Degree | 1.731 ± 0.213 | 1.677 ± 0.257 | 1.593 ± 0.225 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Binocular Mesopic Contrast Sensitivity Without Glare 12 Cycles/Degree | 1.189 ± 0.170 | 1.362 ± 0.323 | 1.153 ± 0.228 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Binocular Mesopic Contrast Sensitivity Without Glare 18 Cycles/Degree | 0.965 ± 0.258 | 0.994 ± 0.359 | 0.861 ± 0.241 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| 1.5 cycles/degree | 1.712 ± 0.212 | 1.743 ± 0.218 | 1.537 ± 0.147 |
| 3 cycles/degree | 1.785 ± 0.223 | 1.768 ± 0.243 | 1.594 ± 0.135 |
| 6 cycles/degree | 1.710 ± 0.234 | 1.682 ± 0.268 | 1.595 ± 0.265 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Binocular Mesopic Contrast Sensitivity Without Glare 12 Cycles/Degree | 1.261 ± 0.291 | 1.273 ± 0.269 | 1.235 ± 0.245 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) without glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Binocular Mesopic Contrast Sensitivity Without Glare 18 Cycles/Degree | 1.160 ± 0.356 | 1.008 ± 0.386 | 0.885 ± 0.287 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 1.5 Cycles/Degree | 1.329 ± 0.172 | 1.389 ± 0.219 | 1.233 ± 0.182 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 3 Cycles/Degree | 1.527 ± 0.190 | 1.518 ± 0.267 | 1.388 ± 0.245 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 6 Cycles/Degree | 1.514 ± 0.217 | 1.504 ± 0.313 | 1.397 ± 0.265 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 12 Cycles/Degree | 1.118 ± 0.225 | 1.192 ± 0.235 | 1.059 ± 0.208 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 18 Cycles/Degree | 0.883 ± 0.199 | 0.989 ± 0.336 | 0.903 ± 0 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 1.5 Cycles/Degree | 1.343 ± 0.254 | 1.367 ± 0.229 | 1.199 ± 0.145 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 3 Cycles/Degree | 1.536 ± 0.247 | 1.462 ± 0.212 | 1.367 ± 0.196 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 6 Cycles/Degree | 1.535 ± 0.274 | 1.475 ± 0.276 | 1.413 ± 0.265 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 12 Cycles/Degree | 1.341 ± 0.341 | 1.272 ± 0.358 | 1.190 ± 0.327 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Monocular Contrast Sensitivity With Glare 18 Cycles/Degree | 1.139 ± 0.315 | 1.192 ± 0.334 | 0.849 ± 0.353 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| 1.5 cycles/degree | 1.459 ± 0.239 | 1.439 ± 0.216 | 1.316 ± 0.210 |
| 3 cycles/degree | 1.609 ± 0.189 | 1.492 ± 0.264 | 1.405 ± 0.225 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Binocular Contrast Sensitivity With Glare 6 Cycles/Degree | 1.574 ± 0.183 | 1.496 ± 0.385 | 1.412 ± 0.261 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Binocular Contrast Sensitivity With Glare 12 Cycles/Degree | 1.199 ± 0.243 | 1.337 ± 0.273 | 1.141 ± 0.222 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Binocular Contrast Sensitivity With Glare 18 Cycles/Degree | 0.980 ± 0.295 | 1.009 ± 0.096 | 0.690 ± 0.125 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| 1.5 cycles/degree | 1.392 ± 0.222 | 1.404 ± 0.238 | 1.344 ± 0.151 |
| 3 cycles/degree | 1.573 ± 0.268 | 1.493 ± 0.304 | 1.410 ± 0.223 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Binocular Contrast Sensitivity With Glare 6 Cycles/Degree | 1.519 ± 0.296 | 1.515 ± 0.330 | 1.448 ± 0.271 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Binocular Contrast Sensitivity With Glare 12 Cycles/Degree | 1.321 ± 0.331 | 1.319 ± 0.450 | 1.337 ± 0.419 |
Logarithm of the mesopic (low light) contrast sensitivity (ability to detect detail with subtle gradations of grayness between test target and background) with glare at five spatial frequencies (number of light and dark bars per cycle). A higher mean indicates improved contrast sensitivity
| log contrast sensitivity units | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Mesopic Binocular Contrast Sensitivity With Glare 18 Cycles/Degree | 1.175 ± 0.258 | 1.101 ± 0.439 | 1.060 ± 0.648 |
Objective scatter index (OSI) was assessed using the Optical Quality Analysis System(OQAS) and summarized as a continuous variable. For this assessment, an OSI score of 1 or less represents good quality vision, 1-2 indicates some degradation but otherwise normal vision, 2-3 significant light scatter possibly requiring treatment, and scores over 4 severely compromised vision requiring intervention.
| units on a scale | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Optical Scatter | 1.1280 ± 0.7661 | 1.8080 ± 0.8685 | 1.9179 ± 0.9751 |
Objective scatter index (OSI) was assessed using the Optical Quality Analysis System(OQAS) and summarized as a continuous variable. For this assessment, an OSI score of 1 or less represents good quality vision, 1-2 indicates some degradation but otherwise normal vision, 2-3 significant light scatter possibly requiring treatment, and scores over 4 severely compromised vision requiring intervention.
| units on a scale | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Optical Scatter | 1.1308 ± 0.7337 | 1.9760 ± 1.1630 | 1.8318 ± 0.9840 |
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
| units on a scale | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Halo | 1.2 ± 0.9 | 1.3 ± 0.8 | 1.8 ± 0.9 |
| Starburst | 4.5 ± 1.1 | 4.7 ± 1.0 | 4.5 ± 1.2 |
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
| units on a scale | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Halo | 1.2 ± 0.6 | 1.5 ± 0.8 | 1.8 ± 0.7 |
| Starburst | 4.6 ± 1.0 | 4.7 ± 1.0 | 4.5 ± 1.2 |
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
| units on a scale | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Halo | 1.2 ± 1.1 | 1.4 ± 0.8 | 2.0 ± 1.0 |
| Starburst | 4.6 ± 1.1 | 4.7 ± 1.0 | 4.6 ± 0.9 |
Halo and starburst (both monocular and binocular) were measured using a glarometer on integer scales of 1-5, where the smaller the number the better the outcome.
| units on a scale | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Halo | 1.3 ± 0.7 | 1.8 ± 1.2 | 2.0 ± 0.9 |
| Starburst | 4.6 ± 1.0 | 4.9 ± 0.4 | 4.5 ± 0.9 |
Manifest refraction spherical equivalent (MRSE) was calculated as the value of the sphere plus one-half of the value of the cylinder.
| Diopters | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Manifest Refraction | -0.507 ± 0.433 | 0.038 ± 0.344 | -0.114 ± 0.362 |
Manifest refraction spherical equivalent (MRSE) was calculated as the value of the sphere plus one-half of the value of the cylinder.
| Diopters | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Manifest Refraction | -0.339 ± 0.455 | 0.145 ± 0.370 | -0.091 ± 0.339 |
High Contrast Uncorrected Distance (UCDVA) measured at 32 inches, Intermediate (UCIVA) at 32 inches, and Near Visual Acuity (UCNVA) at 16 inches
| logMAR | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| UCDVA Monocular Visit 3 | 0.122 ± 0.171 | 0.002 ± 0.115 | 0.072 ± 0.150 |
| UCDVA Monocular Visit 4 | 0.113 ± 0.173 | 0.016 ± 0.121 | 0.060 ± 0.148 |
| UCDVA Binocular Visit 3 | 0.005 ± 0.082 | -0.054 ± 0.102 | -0.048 ± 0.087 |
| UCDVA Binocular Visit 4 | -0.026 ± 0.110 | -0.061 ± 0.080 | -0.053 ± 0.104 |
| UCIVA Monocular Visit 3 | 0.055 ± 0.159 | 0.219 ± 0.150 | 0.237 ± 0.221 |
| UCIVA Monocular Visit 4 | 0.066 ± 0.168 | 0.244 ± 0.158 | 0.237 ± 0.197 |
| UCIVA Binocular Visit 3 | -0.012 ± 0.130 | 0.138 ± 0.140 | 0.126 ± 0.169 |
| UCIVA Binocular Visit 4 | -0.024 ± 0.111 | 0.169 ± 0.144 | 0.175 ± 0.146 |
| UCNVA Monocular Visit 3 | 0.192 ± 0.189 | 0.005 ± 0.108 | 0.164 ± 0.149 |
| UCNVA Monocular Visit 4 | 0.264 ± 0.195 | -0.008 ± 0.120 | 0.188 ± 0.185 |
| UCNVA Binocular Visit 3 | 0.082 ± 0.151 | -0.057 ± 0.074 | 0.041 ± 0.109 |
| UCNVA Binocular Visit 4 | 0.133 ± 0.148 | -0.061 ± 0.087 | 0.043 ± 0.130 |
High Contrast Distance-Corrected Distance,(BCDVA), Intermediate (DCIVA), and Near Visual Acuity (DCNVA)
| logMAR | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| BCDVA Monocular Visit 3 | -0.100 ± 0.070 | -0.094 ± 0.080 | -0.068 ± 0.098 |
| BCDVA Monocular Visit 4 | -0.100 ± 0.066 | -0.096 ± 0.106 | -0.075 ± 0.088 |
| BCDVA Binocular Visit 3 | -0.144 ± 0.049 | -0.133 ± 0.073 | 0.115 ± 0.057 |
| BCDVA Binocular Visit 4 | -0.136 ± 0.049 | -0.137 ± 0.058 | -0.117 ± 0.098 |
| DCIVA Monocular Visit 3 | 0.169 ± 0.144 | 0.292 ± 0.132 | 0.292 ± 0.162 |
| DCIVA Monocular Visit 4 | 0.186 ± 0.129 | 0.297 ± 0.119 | 0.328 ± 0.159 |
| DCIVA Binocular Visit 3 | 0.095 ± 0.146 | 0.222 ± 0.112 | 0.196 ± 0.112 |
| DCIVA Binocular Visit 4 | 0.084 ± 0.109 | 0.229 ± 0.108 | 0.229 ± 0.127 |
| DCNVA Monocular Visit 3 | 0.346 ± 0.130 | -0.004 ± 0.087 | 0.132 ± 0.143 |
| DCNVA Monocular Visit 4 | 0.360 ± 0.144 | -0.042 ± 0.132 | 0.136 ± 0.144 |
| DCNVA Binocular Visit 3 | 0.261 ± 0.109 | -0.079 ± 0.065 | -0.004 ± 0.100 |
| DCNVA Binocular Visit 4 | 0.247 ± 0.126 | -0.075 ± 0.096 | 0.025 ± 0.127 |
Collected over 6 months. Non-serious events are listed at a 5% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| Crystalens AO | — | 0/26 (0%) | 10/52 (19.2%) |
| ReSTOR | — | 1/25 (4%) | 13/50 (26%) |
| Tecnis Multifocal IOL | — | 3/23 (13%) | 4/46 (8.7%) |
| Event | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Vitreous hemorrhageEye disorders | 0/26 | 0/25 | 1/23 |
| Pulmonary embolismVascular disorders | 0/26 | 0/25 | 1/23 |
| PyelonephritisRenal and urinary disorders | 0/26 | 0/25 | 1/23 |
| Retinal vein occlusionEye disorders | 0/26 | 1/25 | 0/23 |
| Event | Crystalens AO | ReSTOR | Tecnis Multifocal IOL |
|---|---|---|---|
| Punctate KeratitisEye disorders | 8/52 | 1/50 | 2/46 |
| Increased Intraocular Pressure (IOP)Eye disorders | 0/52 | 7/50 | 0/46 |
| Corneal EdemaEye disorders | 2/52 | 5/50 | 2/46 |
| Age, Customized(participants) | Crystalens AO | ReSTOR | Tecnis Multifocal IOL | Total |
|---|---|---|---|---|
| <40 years | 0 | 0 | 0 | 0 |
| 40 - 49 years | 0 | 1 | 1 | 2 |
| 50-59 years | 10 | 5 | 8 | 23 |
| 60-69 years | 13 | 12 | 12 | 37 |
| 70-79 years | 3 | 8 | 4 | 15 |
| ≥ 80 | 0 | 0 | 1 | 1 |
| Sex: Female, Male(Participants) | Crystalens AO | ReSTOR | Tecnis Multifocal IOL | Total |
|---|---|---|---|---|
| Female | 17 | 15 | 19 | 51 |
| Male | 9 | 11 | 7 | 27 |
| Region of Enrollment(participants) | Crystalens AO | ReSTOR | Tecnis Multifocal IOL | Total |
|---|---|---|---|---|
| United States | 26 | 26 | 26 | 78 |
This study is completed, as verified in Feb 2014. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Bausch & Lomb Incorporated