An interventional study of Continuous Glucose Monitoring System in Diabetes, Diabetes Melletus, Type 1 and Diabetes Mellitus, Type 2, sponsored by Senseonics, Inc.. Recruiting at 4 sites in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2026-04-21.
Sponsored by Senseonics, Inc. · Not applicable, Interventional, and Diagnostic
The purpose of this clinical investigation is to demonstrate that the proposed physical changes to the ROME GS Sensor do not have any adverse impact on safety and performance when compared to the Eversense 365 Sensor.
A prospective, non-randomized, multi-center evaluation of accuracy and safety of the ROME GS system. The purpose of this clinical investigation is to demonstrate that the proposed physical changes to the ROME GS Sensor do not have any adverse impact on safety and performance when compared to the Eversense 365 Sensor. The investigation will also determine the accuracy of the ROME GS Sytem measurements compored to a reference glucose analyzer. Additionally, the study will demonstrate the safety of the ROME GS System post-sensor insertion or removal and follow-up by measuring the incidence of device-related and insertion/removal procedure-related serious adverse events during the investigation.
10,923 studies on the registry are indexed under Diabetes Mellitus; 1,318 are open to participants now.
This study's planned enrollment of 60 is below the median of 80 across 8,367 interventional studies indexed under Diabetes Mellitus.
Browse Diabetes Mellitus studies →Senseonics, Inc. is the lead sponsor of 11 studies on the registry; 3 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Subjects meeting all of the following inclusion criteria will be included in this study:
Exclusion Criteria:
Subjects meeting any of the following exclusion criteria at the time of screening will be excluded from this study:
Eversense 365 Continuous Glucose Monitoring System and ROME GS System
Device: Continuous Glucose Monitoring System
Eversense 365 Continuous Glucose Monitoring System and ROME GS System
Mean absolute relative difference (MARD) for paired sensor and reference measurements post-insertion.
To determine accuracy of the ROME GS System measurements compared to a reference glucose analyzer. The effectiveness measure will be mean absolute relative difference (MARD) for paired sensor and reference measurements post-insertion for reference glucose values from 40-400 mg/dL. Effectiveness measures will be evaluated descriptively. Neither inferential analysis nor hypothesis testing will be performed.
Time frame: 365 days post-insertion
Number of device-related or sensor insertion/removal procedure-related serious adverse events
Incidence of device-related or sensor insertion/removal procedure-related serious adverse events in clinic and during home use post-sensor insertion or removal and follow-up.
Time frame: 365 days post-insertion and follow-up
Plan to share: No
No publications or documents are linked to this record.
Eligibility is decided by the study team. Share this record with your doctor or contact the team directly.
Contact study teamGet an email when the registry record changes — status, dates, results — or when someone posts here.