An interventional study of Rhaeos Shunt Monitoring System in Hydrocephalus, sponsored by Rhaeos, Inc.. Recruiting at 1 site in United States. Per ClinicalTrials.gov, last updated 2026-03-18.
Sponsored by Rhaeos, Inc. · Not applicable, Interventional, and Diagnostic
This study is evaluating a new, noninvasive device designed to detect whether cerebrospinal fluid (CSF) is flowing through a surgically implanted shunt. CSF shunts are commonly used to treat hydrocephalus, but it can be difficult to tell whether a shunt is working properly without invasive testing or imaging that does not directly measure flow.
The study device is a small, wireless sensor placed on the skin over the shunt tubing, typically near the collarbone. It uses gentle, controlled warmth to measure temperature changes that indicate whether fluid is flowing inside the shunt. The device does not break the skin and does not change a participant's medical care.
This study will enroll children and adults who already have a CSF shunt and who do not have new or worsening symptoms of shunt malfunction. All participants will have a measurement taken while their shunt valve is at its usual prescribed setting. Some participants with programmable shunt valves will also have a second measurement taken after their valve is temporarily adjusted to a setting intended to stop or greatly reduce flow ("virtual off"). After this measurement, the valve will be returned to its prescribed setting.
The study device results will not be shown to participants or their medical providers and will not be used to make treatment decisions. The goal of the study is to determine how well the device can distinguish between shunts that are allowing flow and those that are not.
Participants will be monitored for a short period after device use and, if applicable, after any valve adjustment. The main risks of participation are mild skin irritation from the adhesive or temporary symptoms related to valve adjustment. Participants are not expected to receive a direct medical benefit, but the information gained may help improve future diagnosis of shunt function.
This is a prospective, multicenter, open-label, blinded, observational study designed to evaluate the performance of a noninvasive wireless thermal flow detection device for assessing cerebrospinal fluid (CSF) shunt flow in pediatric and adult patients with hydrocephalus.
The study enrolls participants with an existing implanted ventricular CSF shunt who do not have new or worsening symptoms suggestive of shunt malfunction. The investigational device is placed externally on intact skin overlying the shunt tubing, typically near the clavicle, and measures thermal patterns associated with the presence or absence of CSF flow within the shunt catheter. Measurements are obtained using a sponsor-provided tablet application, and results are encoded so that both participants and treating clinicians remain blinded.
All participants undergo at least one device measurement with the shunt valve set to the participant's prescribed flow-permissive ("valve normal") setting. A subset of participants with programmable valves capable of being adjusted noninvasively will also undergo an additional measurement after the valve is temporarily adjusted to a manufacturer-reported "virtual off" setting, defined as a valve opening pressure exceeding 400 mm H₂O. After this measurement, the valve is returned to the prescribed setting, with confirmation using standard valve programming tools.
Participants whose valves are already prescribed at a "virtual off" setting will undergo a single measurement at that setting. A small number of initial roll-in participants are included to refine procedural logistics; their data are not included in the primary endpoint analyses.
The study is non-significant risk. The investigational device does not alter standard-of-care treatment, and device data are not used to guide clinical decision-making. Valve adjustments performed as part of the study are routinely used in clinical care and are monitored closely. Participants are observed for adverse events during device use and for a defined follow-up period after device removal and, when applicable, after valve readjustment.
The primary objective of the study is to assess the ability of the device to distinguish between flow-permissive and non-flowing shunts based on known valve settings. Secondary and exploratory objectives include evaluation of device accuracy metrics and characterization of thermal flow patterns under different conditions, such as posture changes or valve transitions.
Adverse events, including device-related skin findings and any symptoms associated with valve adjustment, are recorded and reported according to protocol and regulatory requirements. Participants exit the study after completion of the measurement(s) and required observation period.
Inclusion Criteria:
Inclusion and Exclusion Criteria for Roll-In Subjects Inclusion criteria
Inclusion and Exclusion Criteria for Main Study Subjects
Inclusion criteria
Subjects with planned "virtual off" measurement:
-Existing ventricular CSF shunt with an available "virtual off" setting which can be programmed noninvasively
Subjects with planned adjustment to the "virtual off" setting only:
-Subject-reported or documented history of successful valve adjustment(s)
Exclusion Criteria:
Exclusion criteria
Exclusion Criteria for Main Study Subjects Exclusion criteria
Subjects with planned "valve normal" measurements:
Participants undergo noninvasive assessment of cerebrospinal fluid (CSF) shunt flow using a wireless thermal flow detection device placed on the skin overlying the shunt tubing. All participants have a measurement performed at their prescribed shunt valve setting. Participants with programmable valves capable of a manufacturer-reported "virtual off" setting may undergo an additional measurement after temporary valve adjustment to reduce or stop CSF flow, followed by return of the valve to the prescribed setting. Device results are encoded and blinded to participants and clinical staff and are not used for clinical decision-making. Participants are monitored for adverse events during and after device use.
Device: Rhaeos Shunt Monitoring System
characters, and fully aligned with your protocol: The intervention consists of use of a noninvasive, wireless thermal flow detection device to assess cerebrospinal fluid (CSF) flow in an existing implanted shunt. The device is externally applied to intact skin over the shunt tubing, typically near the clavicle, and delivers controlled, imperceptible thermal energy while measuring resulting temperature gradients associated with CSF flow. Measurements are performed using a sponsor-provided tablet application that encodes results to maintain blinding. In some participants with programmable shunt valves, measurements are obtained under both prescribed flow-permissive settings and a temporary manufacturer-reported "virtual off" valve setting, after which the valve is returned to the prescribed setting. Device data are collected for research purposes only and are not used to guide clinical care.
Sensitivity and Specificity
Evaluation of sensitivity and specificity using the main study data
Time frame: Day 1
Plan to share: No
No publications or documents are linked to this record.
Eligibility is decided by the study team. Share this record with your doctor or contact the team directly.
Contact study teamGet an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Rhaeos, Inc.