An interventional study of CardioWatch 287-2 in Hypoxia, sponsored by Corsano Health B.V.. Completed at 1 site in Belgium. Open to participants aged 18 Years to 65 Years, including healthy volunteers. Per ClinicalTrials.gov, last updated 2022-09-15.
Sponsored by Corsano Health B.V. · Not applicable, Interventional, and Screening
In an individual's SpO2 range from 100 to 73 %, this study calibrates and evaluates the accuracy of SpO2 measurement by CW2 in comparison to reference pulse oximeter (Nellcor PM10) or CO-oximetry, Pulse rate (PR) and respiratory rate (RR) were calibrated based on the same Nellcor equipment and the frequency of end-tidal CO2 respectively.
The study is designed in accordance with Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment ISO 80601-2-61:2017, Corrected version 2018-02 and ISO 80601-2-55.
SpO2: Testing to confirm the accuracy of the SpO2 monitoring according to ISO 80601-2-61 Clause 201.12.1.101 201.12.1.101
RR Testing: Testing to confirm the accuracy of the respiration rate monitoring according to ISO 80601-2-55. Volunteers respirated spontaneously, no efforts were made to include all ranges of respiration during the study.
PR Testing: Testing to confirm the accuracy of the PR monitoring according to ISO 80601-2-61 Clause 201.12.1.104. No efforts were made to influence PRs of volunteers.
An accelerometer and PPG filters are built in CW2 in order to form the basis to exclude non-eligible test results.
1,246 studies on the registry are indexed under Hypoxia; 239 are open to participants now.
This study's enrollment of 12 is below the median of 45 across 864 interventional studies indexed under Hypoxia.
Browse Hypoxia studies →Corsano Health B.V. is the lead sponsor of 9 studies on the registry; 5 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
Each participant wore one CW2 sensor attached to the back of the wrist with a wrist band. Additionally, each participant wore a reference pulse oximeter (Nellcor PM10) attached to the middle finger of the ipsilateral hand. This enabled the comparison of the paired results of the CW2 sensor and reference pulse oximeter equipment for SpO2 measurements at different oxygen saturation levels. Twelve subjects of the 24 participants received an arterial catheter in the contralateral radial artery for exvivo SaO2 determination by CO-oximetry and in order to define the final accuracy of CW2 for SpO2 measurement.
Device: CardioWatch 287-2
CardioWatch 287-2 (CW2 Test Device or CW2). This medical device is based on the precursor CardioWatch 287-1 approved by EU-MDR under number 20M00011CRT01, CW2 has two red, two infrared and two green LEDs, and two photosensors compared to its precursor.
Pulse Oximetry
SpO2
Time frame: 1 hour
Pulse rate
Heart rate in beats per minutes
Time frame: 1 hour
Respiratory rate
In breaths per minutes
Time frame: 1 hour
Plan to share: No
No publications or documents are linked to this record.
Ran from Jun 2022 to primary completion in Jun 2022, 0 months.
Of 32 industry-sponsored interventional studies with no phase under Hypoxia, half reached primary completion within 0 months of starting (the median), counting those still running; a quarter within 0 and three quarters within 5.
Counting every sponsor, half of 406 studies reached it within 13 months; a quarter within 4 and three quarters within 29.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in Sep 2022. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Corsano Health B.V.