Automated Assessment of PVI Using a Novel EP Recording System

An interventional study of CathVision Cube® system in Paroxysmal Atrial Fibrillation and Persistent Atrial Fibrillation, sponsored by CathVision ApS. Completed at 4 sites in 2 countries. Open to participants aged 21 Years and older. Per ClinicalTrials.gov, last updated 2024-08-09.

Sponsored by CathVision ApS · Not applicable, Interventional, and Treatment

Phase
Not applicable
Study type
Interventional
Enrollment
101
Allocation
Not applicable
Ages
21 Years and older
Sex
All
01

Study summary

A prospective, multi-center study with the CathVision Cube® system and the PVI Analyzer software in radiofrequency (RF), cryo-balloon (CB), and Pulse Field Ablation (PFA) procedures.

Read the detailed description

Subjects with paroxysmal or persistent atrial fibrillation (AF) who are indicated to undergo first pulmonary vein isolation (PVI) procedure and meet all eligibility criteria will be enrolled in the Study. Intracardiac signals will be passively recorded using the investigational CathVision Cube® System in parallel with the conventional (CE marked) EP recording system. The investigational device will not be used for direct clinical care decisions or therapy. The validation of the automated algorithm for PVI will be performed offline

The primary objective is to validate the PVI Analyzer software with a novel EP recording system (CathVision Cube® System) for assisting assessment of isolation status following PVI ablation.

The secondary objective is to determine the feasibility of "real-time" assessment of PVI analysis and rhythm dependent performance using the PVI Analyzer and CathVision Cube® System.

02

Conditions studied

  • Paroxysmal Atrial Fibrillation
  • Persistent Atrial Fibrillation

Browse trials for

Keywords

  • atrial fibrillation
  • pulmonary vein isolation
  • radiofrequency
  • cryo balloon
  • ablation
  • software
  • pulse field ablation
03

In context

Atrial Fibrillation

3,957 studies on the registry are indexed under Atrial Fibrillation; 933 are open to participants now.

This study's enrollment of 101 is below the median of 145 across 2,437 interventional studies indexed under Atrial Fibrillation.

Browse Atrial Fibrillation studies →

Lead sponsor

CathVision ApS is the lead sponsor of 9 studies on the registry; 2 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
21 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Eligibility criteria

Eligible subjects will meet all of the following inclusion criteria:

  1. Subjects undergoing first-time pulmonary vein isolation indicated by investigator for the treatment of atrial fibrillation
  2. Male or non-pregnant female aged ≥21 years.
  3. Able and willing to provide written informed consent prior to any clinical investigation related procedure

Eligible subjects will not meet any of the following exclusion criteria:

  1. Pregnant or nursing subjects.
  2. Current participation in another investigational drug or device study that interferes with this Study.
  3. Subjects who, in the opinion of the investigator, are not candidates for this Study.
  4. Patients who have had a prior ablation procedure
  5. Presence of other anatomic or comorbid conditions, or other medical, social, or psychological conditions that, in the investigator's opinion, could limit the subject's ability to participate in the clinical investigation or to comply with follow-up requirements, or impact the scientific soundness of the clinical investigation results.
  6. Life expectancy less than 12 month
  7. Subjects who, in the opinion of the investigator, are considered part of any vulnerable population.
05

Study design

Phase
Not applicable
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
101 participants (actual)

Study arms

  • Experimental
    Single Group

    Index EP Procedure: Ablation of atrial fibrillation

    Device: CathVision Cube® system

Interventions

  • DeviceCathVision Cube® system

    Intracardiac signals will be passively recorded using the investigational CathVision Cube® System in parallel with the commercial (bearing CE mark) EP recording systems LabSystem Pro, Boston Scientific EP, or EP-Tracer, Schwarzer Cardiotek. The investigational device will not be used for direct clinical care decisions or therapy. The validation of the PVI Analyzer will be performed offline and retrospectively.

06

What researchers measure

Primary outcomes

  1. PVI Analyzer Sensitivity and Specificity, Measured as the Accuracy of Predictions for Non-isolated and Isolated Veins.

    The primary outcome is to evaluate the sensitivity and specificity of the PVI Analyzer in the classification of the isolation status of veins, following the medical procedure known as pulmonary vein isolation (PVI). The information used for this has been obtained from the investigational CathVision Cube System. The PVI Analyzer determines whether the veins are isolated or not after pulmonary vein ablation. The proportion of correct negative (non-isolated) predictions among non-isolated vein data points and the proportion of correct positive (isolated) data points from isolated veins, provide the sensitivity and specificity data, respectively.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

  2. PVI Analyzer Safety Profile, Measured as the Number of Adverse Events and/or Device Malfunctions.

    As part of the primary outcome, safety is also evaluated as the number of adverse events and device malfunctions reported by the study team at site, during the use of the CathVision Cube® System from the start of the procedure until discharge. The number of adverse events and device malfunctions is recorded in the site's source documents as well as in the case report form (CRF) and is duly informed to the Sponsor who also keeps the record.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

Secondary outcomes

  1. The Pulmonar Vein Isolation (PVI) Classification in Sinus Rhythm (SR) Considering the Electrophysiology Procedure Type.

    This outcome assesses the overall classification of pulmonary vein isolation (PVI) in sinus rhythm (SR) based on the type of electrophysiology procedure performed. The classification provides insights into the accuracy of the procedure in achieving SR. The overall classification was evaluated in terms of sensitivity and specificity. To calculate specificity from the total number of EGMS analyzed, all the correct negatives (non-isolated) / Correct negatives + False positives were considered. To calculate sensitivity, all the correct positives (isolated) / Correct positives + False negatives were considered. As the sample size was not adjusted to detect the differences for variables other than the primary endpoint, the p-values obtained from the analyses of secondary outcomes should be viewed as complementary to the primary outcome results. Their primary purpose was to generate hypotheses rather than draw definitive conclusions.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

  2. Feasibility of "Real-time" Assessment of Isolation, Measured by Retesting the Accuracy of the PVI Analyzer Classification of Isolation Status at Different Timepoints Before, During and After Ablation.

    The device's feasibility was evaluated by reassessing its accuracy. It refers to the consistency in the re-classification of pairs of non-overlapping samples throughout the different timepoints of ablation-prior to, during, and following the procedure. For this evaluation, the same EGMs of the primary endpoint were used, including only EGMs from T Baseline and T Final measurements from patients in sinus rhythm. The extraction of non-overlapping samples was selected by considering only those EGMs with a number of global isolation index values above 6. The global isolation index value is an indicator from the PVI analyzer that is used to assess the overall effectiveness of the ablation in isolating specific regions of the heart. The fourth and last global index returned by the PVI Analyzer were used as pairs for analyzing the outcome. The accuracy was calculated as the percentage of EGMs with agreement across timepoints with respect to the total number of EGMs performed.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

  3. PVI Isolation Classification in Non-sinus Rhythm (NSR) From EP Procedure

    The PVI isolation classification in non-sinus rhythm (NSR) from EP procedure was focused on the analysis of all EGM in non-sinus rhythm. The data analysis and processing were similar to those of the primary endpoint, but taking into consideration that subjects shall be in non-sinus rhythm during the duration of the EGM snippet. Apart from this, the processing was the same and the threshold was also defined as 50: values below 50 were classified as baseline, while values at or above 50 were considered as isolated veins.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

  4. Accuracy of PVI Analyzer Classification at the Time of Expert-defined Isolation

    This secondary endpoint aimed to determine the sensitivity of the PVI Analyzer at the time of isolation, which requires the EGMs at the T PVI. The T PVI EGMs were extracted by taking EGM snippets of 4 seconds around the T PVI event. All of those EGMs were expected to be from isolated veins (ground-truth is isolated). Then, they were processed by the PVI Analyzer, and the last valid global isolation index output was used for endpoint assessment.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

  5. Comparison of the PVI Analyzer Performance With CathVision Cube® System Data With Its Performance With Data From Other Conventional EP Systems Used

    The conventional EP system and CathVision Cube® System were running on two different PCs in the EP laboratory in parallel. The relevant EGM snippets were extracted from the conventional EP system using a time offset between the conventional EP system PC and the Cube PC. Since a very similar number of samples were used to compute the results, no statistical assessment of the sampling variation in the sensitivity and specificity was made. A paired McNemars test, with an alpha value of 0.05, was applied to analyze the differences in classification performance.

    Time frame: From EP procedure until study completion at discharge; an average of 24 hours.

07

Results

Posted Aug 9, 2024
Limitations and caveats
Only a limited amount of data (30 seconds) could be exported from the LS Pro system, which limited the correlation to only search a narrow region around the exported EGM data. The data obtained from the PFA procedures could not be assessed due to incompatibility with the PVI Analyzer algorithm.

Participant flow

The enrollment took place between September 15, 2021 and June 22, 2022 at four sites (hospitals) located in Bruges, Brussels, Ghent and Toulouse. At these hospitals, participants were assessed for eligibility to participate in the clinical investigation, and those who met the eligibility criteria, voluntarily gave their informed consent. Those who were successfully enrolled were monitored until their hospital discharge, and no further follow-up was deemed necessary.

Participant flow — Overall Study
MilestoneIndex EP Procedure: Ablation of Atrial Fibrillation
Started101
Completed97
Not completed4
Withdrew: Minimum required measurements could not be taken4

Outcome measures

PrimaryPVI Analyzer Sensitivity and Specificity, Measured as the Accuracy of Predictions for Non-isolated and Isolated Veins.

The primary outcome is to evaluate the sensitivity and specificity of the PVI Analyzer in the classification of the isolation status of veins, following the medical procedure known as pulmonary vein isolation (PVI). The information used for this has been obtained from the investigational CathVision Cube System. The PVI Analyzer determines whether the veins are isolated or not after pulmonary vein ablation. The proportion of correct negative (non-isolated) predictions among non-isolated vein data points and the proportion of correct positive (isolated) data points from isolated veins, provide the sensitivity and specificity data, respectively.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Count of units · Electrogram (EGM)
PVI Analyzer Sensitivity and Specificity, Measured as the Accuracy of Predictions for Non-isolated and Isolated Veins.
Electrogram (EGM)Single Group
Sensitivity216
Specificity215
Statistical analysis
  • Single Group · Z-test · p = 0.006 (The success of the PVI Analyzer output was assumed to follow a binominal distribution with a given true proportion of 87.50%. It was tested whether the specificity was superior to the 80% limit.) · Percentage (%): 87A one-tailed test with a significance level of 95% (α = 0.05) was applied.
  • Single Group · Z-test · p = 0.004 (The success of the PVI Analyzer output was assumed to follow a binominal distribution with a given true proportion of 87.50%. It was tested whether the sensitivity was superior to the 80% limit.) · Percentage: 86A one-tailed test with a significance level of 95% (α = 0.05) was applied.
PrimaryPVI Analyzer Safety Profile, Measured as the Number of Adverse Events and/or Device Malfunctions.

As part of the primary outcome, safety is also evaluated as the number of adverse events and device malfunctions reported by the study team at site, during the use of the CathVision Cube® System from the start of the procedure until discharge. The number of adverse events and device malfunctions is recorded in the site's source documents as well as in the case report form (CRF) and is duly informed to the Sponsor who also keeps the record.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Number · participants
PVI Analyzer Safety Profile, Measured as the Number of Adverse Events and/or Device Malfunctions.
participantsSingle Group
Mild0
Moderate1
Severe0
Device related0
Procedure related1
other0
Serious Adverse Events1
Serious adverse device effect (SADE) incidence0
Serious unanticipated serious adverse device effect (USADE) incidence0
Device deficiency incidence0
Statistical analysis
  • Single Group · Percentage: 0.99 · 95% CI 0.02 to 5.39Only one AE was reported for one subject ( which was considered moderate in severity and not related to the device, but related to the SOC procedure.
SecondaryThe Pulmonar Vein Isolation (PVI) Classification in Sinus Rhythm (SR) Considering the Electrophysiology Procedure Type.

This outcome assesses the overall classification of pulmonary vein isolation (PVI) in sinus rhythm (SR) based on the type of electrophysiology procedure performed. The classification provides insights into the accuracy of the procedure in achieving SR. The overall classification was evaluated in terms of sensitivity and specificity. To calculate specificity from the total number of EGMS analyzed, all the correct negatives (non-isolated) / Correct negatives + False positives were considered. To calculate sensitivity, all the correct positives (isolated) / Correct positives + False negatives were considered. As the sample size was not adjusted to detect the differences for variables other than the primary endpoint, the p-values obtained from the analyses of secondary outcomes should be viewed as complementary to the primary outcome results. Their primary purpose was to generate hypotheses rather than draw definitive conclusions.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Count of units · Electrogram (EGM)
The Pulmonar Vein Isolation (PVI) Classification in Sinus Rhythm (SR) Considering the Electrophysiology Procedure Type.
Electrogram (EGM)Radiofrequency ProcedureCryoballoon ProcedurePulsed Field Ablation (PFA) Procedure
Specificty120950
Sensitivity1071090
Statistical analysis
  • Radiofrequency Procedure · Z-test · p = 0.00002 (Computed p values are the result of running a one sample z-test with a null hypothesis that specificity is below 80%.) · Percentage (%): 94
  • Radiofrequency Procedure · Z-test · p = 0.32 (Computed p values are the result of running a one sample z-test with a null hypothesis that sensitivity is below 80% and alternative hypothesis that sensitivity is at or above 80%.) · Percentage (%): 82
  • Cryoballoon Procedure · Z-test · p = 0.66 (Computed p values are the result of running a one sample z-test with a null hypothesis that specificity is below 80%.) · Percentage (%): 79
  • Cryoballoon Procedure · Z-test · p = 0.0008 (Computed p values are the result of running a one sample z-test with a null hypothesis that sensitivity is below 80% and alternative hypothesis that sensitivity is at or above 80%.) · Percentage (%): 92
SecondaryFeasibility of "Real-time" Assessment of Isolation, Measured by Retesting the Accuracy of the PVI Analyzer Classification of Isolation Status at Different Timepoints Before, During and After Ablation.

The device's feasibility was evaluated by reassessing its accuracy. It refers to the consistency in the re-classification of pairs of non-overlapping samples throughout the different timepoints of ablation-prior to, during, and following the procedure. For this evaluation, the same EGMs of the primary endpoint were used, including only EGMs from T Baseline and T Final measurements from patients in sinus rhythm. The extraction of non-overlapping samples was selected by considering only those EGMs with a number of global isolation index values above 6. The global isolation index value is an indicator from the PVI analyzer that is used to assess the overall effectiveness of the ablation in isolating specific regions of the heart. The fourth and last global index returned by the PVI Analyzer were used as pairs for analyzing the outcome. The accuracy was calculated as the percentage of EGMs with agreement across timepoints with respect to the total number of EGMs performed.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Count of units · Electrogram (EGM)
Feasibility of "Real-time" Assessment of Isolation, Measured by Retesting the Accuracy of the PVI Analyzer Classification of Isolation Status at Different Timepoints Before, During and After Ablation.
Electrogram (EGM)Index EP Procedure: Ablation of Atrial Fibrillation
Feasibility of "Real-time" Assessment of Isolation, Measured by Retesting the Accuracy of the PVI Analyzer Classification of Isolation Status at Different Timepoints Before, During and After Ablation.475
Statistical analysis
  • Index EP Procedure: Ablation of Atrial Fibrillation · Z-test · p = 0.000005 (To compute the proportional agreement between two non-overlapping samples, a one-sample Z-test was performed with the alternative hypothesis that agreement was above 90%, which was considered sufficient for a real-time assessment.) · Percentage (%): 96
SecondaryPVI Isolation Classification in Non-sinus Rhythm (NSR) From EP Procedure

The PVI isolation classification in non-sinus rhythm (NSR) from EP procedure was focused on the analysis of all EGM in non-sinus rhythm. The data analysis and processing were similar to those of the primary endpoint, but taking into consideration that subjects shall be in non-sinus rhythm during the duration of the EGM snippet. Apart from this, the processing was the same and the threshold was also defined as 50: values below 50 were classified as baseline, while values at or above 50 were considered as isolated veins.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Count of units · Electrogram (EGM)
PVI Isolation Classification in Non-sinus Rhythm (NSR) From EP Procedure
Electrogram (EGM)Index EP Procedure: Ablation of Atrial Fibrillation
Specificity25
Sensitivity41
Statistical analysis
  • Index EP Procedure: Ablation of Atrial Fibrillation · Z-test · p = 1 (The p values calculated are the result of running a one sample z-test with a null hypothesis that the specificity are below 80%.) · Percentage (%): 54
  • Index EP Procedure: Ablation of Atrial Fibrillation · Z-test · p = 0.0007 (The p values calculated are the result of running a one sample z-test with a null hypothesis that the sensitivity are below 80% and an alternative hypothesis that sensitivity is at or above 80%.) · Percentage (%): 100
SecondaryAccuracy of PVI Analyzer Classification at the Time of Expert-defined Isolation

This secondary endpoint aimed to determine the sensitivity of the PVI Analyzer at the time of isolation, which requires the EGMs at the T PVI. The T PVI EGMs were extracted by taking EGM snippets of 4 seconds around the T PVI event. All of those EGMs were expected to be from isolated veins (ground-truth is isolated). Then, they were processed by the PVI Analyzer, and the last valid global isolation index output was used for endpoint assessment.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Count of units · Electrogram (EGM)
Accuracy of PVI Analyzer Classification at the Time of Expert-defined Isolation
Electrogram (EGM)Index EP Procedure: Ablation of Atrial Fibrillation
Accuracy of PVI Analyzer Classification at the Time of Expert-defined Isolation86
Statistical analysis
  • Index EP Procedure: Ablation of Atrial Fibrillation · Z-test · p = 0.04 · Percentage (%): 87The estimate was done using a one-sample Z-test with a null hypothesis that sensitivity is at or below 80% and a 95% significance level.
SecondaryComparison of the PVI Analyzer Performance With CathVision Cube® System Data With Its Performance With Data From Other Conventional EP Systems Used

The conventional EP system and CathVision Cube® System were running on two different PCs in the EP laboratory in parallel. The relevant EGM snippets were extracted from the conventional EP system using a time offset between the conventional EP system PC and the Cube PC. Since a very similar number of samples were used to compute the results, no statistical assessment of the sampling variation in the sensitivity and specificity was made. A paired McNemars test, with an alpha value of 0.05, was applied to analyze the differences in classification performance.

Time frame:
From EP procedure until study completion at discharge; an average of 24 hours.
Reported as:
Count of units · Electrogram (EGM)
Comparison of the PVI Analyzer Performance With CathVision Cube® System Data With Its Performance With Data From Other Conventional EP Systems Used
Electrogram (EGM)Conventional EP SystemCathVision Cube® System
Specificity118127
Sensitivity103124
Statistical analysis
  • Conventional EP System vs CathVision Cube® System · McNemars test · p = 0.0002 (McNemar test assumes as a null hypothesis that the two datasets cause the PVI Analyzer to have a similar proportion of errors.)A paired McNemars test with an alpha value of 0.05, was applied to analyze the differences in classification performance.

Adverse events

Collected over Adverse events data were collected from participants' electrophysiology procedure until study completion at discharge (hospitalization time approximately 1-2 days).. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Adult Patients With Cardiac Arrhythmia Undergoing Cryoballoon0/47 (0%)0/47 (0%)0/47 (0%)
Adult Patients With Cardiac Arrhythmia Undergoing Radiofrequency0/42 (0%)1/42 (2.4%)0/42 (0%)
Adult Patients With Cardiac Arrhythmia Undergoing Pulsed Field Ablation0/12 (0%)0/12 (0%)0/12 (0%)
Most frequent serious events
Most frequent serious events
EventAdult Patients With Cardiac Arrhythmia Undergoing CryoballoonAdult Patients With Cardiac Arrhythmia Undergoing RadiofrequencyAdult Patients With Cardiac Arrhythmia Undergoing Pulsed Field Ablation
Prolongation of an existing hospitalizationSurgical and medical procedures0/471/420/12

Baseline characteristics

Age, Continuous
Age, Continuous(years)Index EP Procedure: Ablation of Atrial Fibrillation
Mean62.76 ± 12.31
Sex: Female, Male
Sex: Female, Male(Participants)Index EP Procedure: Ablation of Atrial Fibrillation
Female37
Male64
Race and Ethnicity Not Collected
Race and Ethnicity Not Collected(Participants)Index EP Procedure: Ablation of Atrial Fibrillation
Height [cm]
Height [cm](cm)Index EP Procedure: Ablation of Atrial Fibrillation
Mean174.23 ± 9.83
Weight [kg]
Weight [kg](kg)Index EP Procedure: Ablation of Atrial Fibrillation
Mean82.04 ± 17.86
Systolic Blood Pressure [mmHg]
Systolic Blood Pressure [mmHg](mmHg)Index EP Procedure: Ablation of Atrial Fibrillation
Mean137.39 ± 20.04
Diastolic Blood Pressure [mmHg]
Diastolic Blood Pressure [mmHg](mmHg)Index EP Procedure: Ablation of Atrial Fibrillation
Mean79.44 ± 12.10
Heart Rate [BPM]
Heart Rate [BPM](BPM)Index EP Procedure: Ablation of Atrial Fibrillation
Mean72.69 ± 22.40

2 further baseline measures are reported on the registry.

08

Study locations

4 sites
  • AZ Sint-Jan Hospital
    Bruges, 8000, Belgium
  • Universitair Ziekenhuis Brussel
    Brussels, 1090, Belgium
  • Ghent University Hospital
    Ghent, 9000, Belgium
  • Clinique Pasteur Toulouse,
    Toulouse, 31300, France
09

References and documents

Publications

  • De Pooter J, Timmers L, Boveda S, Combes S, Knecht S, Almorad A, De Asmundis C, Duytschaever M. Validation of a machine learning algorithm to identify pulmonary vein isolation during ablation procedures for the treatment of atrial fibrillation: results of the PVISION study. Europace. 2024 May 2;26(5):euae116. doi: 10.1093/europace/euae116. PubMed 38682165 ↗

Study documents

  • Study protocol · Dec 20, 2021
  • Statistical analysis plan · Aug 30, 2022
  • Informed consent form · May 20, 2021

Documents are hosted by the registry — open the source record to download them.

Individual participant data

Plan to share: No

10

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Aug 9, 2024, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT05043883
Lead sponsor
CathVision ApS
Collaborators
AKRN Scientific Consulting, S.L.
Responsible party
Sponsor
First posted
Sep 14, 2021
Start date
Sep 15, 2021
Primary completion
Jun 22, 2022
Completion
Jun 22, 2022
Results posted
Aug 9, 2024
Last update
Aug 9, 2024

Study contacts

Mattias Duytschaever, Prof. MD PhD
principal investigator · AZ Sint-Jan Hospital, Ruddershove 10 8000 Brugge, Belgium

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Mar 2024. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion