CompletedNCT04582877Updated Dec 2, 2024Results posted

Pressure Guidewire System Multi-center, Prospective, Self-Control, Clinical Trial

An interventional study of FFR measurement in Myocardial Ischemia, sponsored by Zurich Medical Inc.. Completed at 1 site in China. Open to participants aged 18 Years to 75 Years. Per ClinicalTrials.gov, last updated 2024-12-02.

Sponsored by Zurich Medical Inc. · Not applicable, Interventional, and Diagnostic

Phase
Not applicable
Study type
Interventional
Enrollment
300
Allocation
Not applicable
Ages
18 Years to 75 Years
Sex
All
01

Study summary

Patients with intermediate-grade coronary stenosis undergo measurement of fractional flow reserve using the test article (Zurich Pressure Guidewire System) and predicate article (Abbott PressureWire System). Both articles are used within the same participant and the FFR values are statistically compared with each other.

Read the detailed description

The objective of the 'Pressure guidewire system multi-center, prospective, single-subject design clinical trial' is to evaluate the effectiveness and safety of Zurich Medical's pressure guidewire system (including guidewire with high-fidelity sensors and a unique paired portable display unit), which is used to measure the coronary artery blood fractional flow reserve (FFR) to diagnose coronary artery disease, and to direct a catheter through a blood vessel. This clinical trial uses a multi-center, prospective, single-subject design trial method, using St. Jude Medical's pressure guidewire and Analyzer Express as a control device. The subjects who met the criteria for this study will be registered with the central registration system after enrollment. Participants with intermediate-grade coronary stenosis undergo measurement of fractional flow reserve using the test article (Zurich Pressure Guidewire System) and predicate article (Abbott PressureWire System). Both articles are used within the same participant and the FFR values are statistically compared with each other. The subject's FFR of coronary artery stenosis will first be measured using investigational device and recorded, and then measured using control device and recorded. The safety follow-up period is 48 hours.

02

Conditions studied

  • Myocardial Ischemia

Browse trials for

03

In context

Myocardial Ischemia

3,242 studies on the registry are indexed under Myocardial Ischemia; 390 are open to participants now.

This study's enrollment of 300 is above the median of 120 across 1,901 interventional studies indexed under Myocardial Ischemia.

Browse Myocardial Ischemia studies →

Lead sponsor

This is the only study on the registry with Zurich Medical Inc. as lead sponsor.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years to 75 Years
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Age 18 to 75 years old, gender-unrestricted, non-pregnant female;
  • Understand and be willing to sign an informed consent form;
  • Diagnosed with coronary heart disease;
  • Intrusive ICA and FFR measurement are needed;
  • Visual coronary angiography showed at least one moderate stenosis lesion (diameter stenosis of 30% - 70%) on the coronary artery with diameter ≥ 2.5mm

Exclusion criteria

Exclusion Criteria:

  • Patient who do not understand or are unwilling to sign an informed consent form;
  • Has a history of myocardial infarction;
  • Patient with other serious diseases are not suitable for clinical trials, such as a complex congenital heart disease history, severe heart failure(NYHA cardiac function level IV), long QT syndrome, severe hypertension, Severe asthma, severe chronic obstructive pulmonary disease, liver and kidney dysfunction and other serious infections and critical illnesses;
  • Coronary intervention surgery contraindications;
  • Patient with ATP contraindications (ATP contraindications: sinus syndrome, sinus insufficiency and the elderly with cautious use or no use);
  • The clinical manifestations of patients show acute instability, including acute chest pain (sudden appearance), cardiogenic shock, unstable blood pressure (systolic pressure less than 90mmHg), severe congestive heart failure or acute pulmonary edema;
  • The angiography shown or suspect of thrombosis;
  • The angiography shown or suspect of dissection;
  • Left main coronary artery disease, target blood vessels with severe curvature or calcification lesions, total occlusion;
  • There are any other factors that the investigator considers unsuitable for inclusion or completion of this study.
05

Study design

Phase
Not applicable
Primary purpose
Diagnostic
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
300 participants (actual)

Study arms

  • Other
    FFR Measurement in Intermediate-Grade Coronary Stenosis

    Participants with intermediate-grade coronary stenosis undergo measurement of fractional flow reserve using the test article (Zurich Pressure Guidewire System) and predicate article (Abbott PressureWire System).

    Device: FFR measurement

Interventions

  • DeviceFFR measurement

    FFR is measured sequentially with the test article and the predicate device.

06

What researchers measure

Primary outcomes

  1. Number of Participants With Diagnostic Consistency When Comparing FFR Measurement Between Control Device and Investigational Device

    Diagnostic consistency rate: When the measurement of control device is positive (FFRPW ≤ 0.80) or negative (FFRPW \> 0.80), investigational device is also positive (FFRWS ≤ 0.80) or negative (FFRWS \> 0.80). Each subject will be considered a research unit (for patients with multiple lesions, the method of converting the results of multiple lesions into patient level is defined in the statistical analysis plan), and the coronary artery FFRPW measured by control device will be taken as reference. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

Secondary outcomes

  1. The Sensitivity of the Trial Device Compared to the Control Device in Diagnosing Coronary Heart Disease

    After all subjects complete their procedures, using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.80 as a positive threshold, the sensitivity of diagnosing coronary heart disease using the trial device is calculated. Sensitivity = Number of subjects with FFRws (≤0.80) / Number of subjects with FFRpw (≤0.80). Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  2. The Specificity of the Trial Device Compared to the Control Device in Diagnosing Coronary Heart Disease

    After all subjects complete their procedures, using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.80 as a positive threshold, the specificity of diagnosing coronary heart disease using the trial device is calculated. Specificity = Number of subjects with FFRws (\>0.80) / Number of subjects with FFRpw (\>0.80). Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  3. Positive Predictive Value (PPV) of the Trial Device for Diagnosing Coronary Heart Disease

    using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.8 as a positive threshold, the positive predictive value of diagnosing coronary heart disease using the trial device is calculated. PPV = true positive / (true positive + false positive) Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  4. Negative Predictive Value (NPV) of the Trial Device for Diagnosing Coronary Heart Disease

    After all subjects complete their procedures, using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.8 as a positive threshold, the negative predictive value of diagnosing coronary heart disease using the trial device is calculated. NPV = true negative / (true negative + false negative) Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  5. Receiver Operating Characteristic (ROC) Curve of the Trial Device in Diagnosing Coronary Heart Disease

    On a two-dimensional coordinate graph, set the x-axis as the false positive rate (FPR), and the y-axis as the true positive rate (TPR). For the FFRws values measured by the trial device, a point pair of TPR and FPR can be obtained based on its performance on the test sample. Adjust the diagnostic threshold of FFRws to obtain a series of points, connect them in sequence (including the points (0,0) and (1,1)), which forms the ROC curve of FFRws. The value of Area Under Curve (AUC) is the size of the area under the ROC curve. The curve is plotted using MedCalc software. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  6. Passing-Bablok Regression Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease

    Using the vessel as the research unit, with the trial device's measured coronary artery FFRws value as the y-axis, and the control device's measured coronary FFRpw value as the x-axis, perform linear regression analysis on the FFR measurement results, and record the regression analysis equation, slope, and intercept. The analysis will be conducted at the lesion level. Regression analysis equation: Y=aX + b Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  7. Bland-Altman Bias Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease

    Bland-Altman analysis is a statistical method for evaluating the agreement between two measurement methods and for assessing bias. Using the vessel as the research unit, analyze the trial device's measured coronary artery FFRws value and the control device's measured coronary artery FFRpw value using Bland-Altman analysis, and plot the bias graph. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit. In the two-dimensional Cartesian coordinate of the bias graph, the x-axis represents the average value of the FFR measured by the two devices for each diseased vessel, and the y-axis represents the difference between the FFR values determined by the two devices for each diseased vessel. mean: (x-axis): (FFRws+FFRpw) / 2 absolute error (y-axis): FFRws - FFRpw The mean absolute error can then be calculated by taking the average of all absolute errors.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  8. Pearson Linear Correlation Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease

    After all subjects complete their procedures, using the vessel as the research unit, perform Pearson linear correlation analysis on the trial device's measured coronary artery FFRws value and the control device's measured coronary artery FFRpw value, and record the correlation coefficient. The correlation coefficient is a statistical measure that determines the strength of a linear relationship between two variables, ranging from -1 to 1. A value close to 1 represents a strong positive correlation between the x and y axis values (trail device FFRws and control device FFRpw). Because of the nature of this value, the measure type is "number". The coefficient is determined using the following formula: P = Cov(x,y) / OxOy where: Cov(x,y) = covariance of variables x and y Ox = standard deviation of x Oy = standard deviation of y

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

  9. Device Success Rate Analysis Results for Subjects

    After all subjects complete their procedures, using the vessel as the research unit, with valid FFRws and FFRpw readings as the standard for successful device use, compare the device success rates of the trial device and the control device. Success rate is determined as the number of units that were performed successfully on the patient. The case was deemed successful if the unit was able to read a valid FFRws and FFRpw reading. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

    Time frame: Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.

07

Results

Posted Dec 2, 2024

Participant flow

Participant flow — Overall Study
MilestoneFFR Measurement in Intermediate-Grade Coronary Stenosis
Started300
Completed292
Not completed8
Withdrew: Did not use device measurement5
Withdrew: No fractional flow reserve test results2
Withdrew: Protocol violation1

Outcome measures

PrimaryNumber of Participants With Diagnostic Consistency When Comparing FFR Measurement Between Control Device and Investigational Device

Diagnostic consistency rate: When the measurement of control device is positive (FFRPW ≤ 0.80) or negative (FFRPW \> 0.80), investigational device is also positive (FFRWS ≤ 0.80) or negative (FFRWS \> 0.80). Each subject will be considered a research unit (for patients with multiple lesions, the method of converting the results of multiple lesions into patient level is defined in the statistical analysis plan), and the coronary artery FFRPW measured by control device will be taken as reference. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Count of participants · Participants
Number of Participants With Diagnostic Consistency When Comparing FFR Measurement Between Control Device and Investigational Device
ParticipantsFFR Measurement in Intermediate-Grade Coronary Stenosis
Number of Participants With Diagnostic Consistency When Comparing FFR Measurement Between Control Device and Investigational Device274
SecondaryThe Sensitivity of the Trial Device Compared to the Control Device in Diagnosing Coronary Heart Disease

After all subjects complete their procedures, using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.80 as a positive threshold, the sensitivity of diagnosing coronary heart disease using the trial device is calculated. Sensitivity = Number of subjects with FFRws (≤0.80) / Number of subjects with FFRpw (≤0.80). Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · percentage of sensitivity
The Sensitivity of the Trial Device Compared to the Control Device in Diagnosing Coronary Heart Disease
percentage of sensitivityFFR Measurement in Intermediate-Grade Coronary Stenosis
asymptotic normal method89.2 (80.9 to 97.5)
exact probability method89.2 (79.1 to 95.6)
SecondaryThe Specificity of the Trial Device Compared to the Control Device in Diagnosing Coronary Heart Disease

After all subjects complete their procedures, using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.80 as a positive threshold, the specificity of diagnosing coronary heart disease using the trial device is calculated. Specificity = Number of subjects with FFRws (\>0.80) / Number of subjects with FFRpw (\>0.80). Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · Percentage of specificity
The Specificity of the Trial Device Compared to the Control Device in Diagnosing Coronary Heart Disease
Percentage of specificityFFR Measurement in Intermediate-Grade Coronary Stenosis
asymptotic normal method95.2 (92.1 to 98.2)
exact probability method95.2 (91.5 to 97.6)
SecondaryPositive Predictive Value (PPV) of the Trial Device for Diagnosing Coronary Heart Disease

using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.8 as a positive threshold, the positive predictive value of diagnosing coronary heart disease using the trial device is calculated. PPV = true positive / (true positive + false positive) Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · Percentage of PPV
Positive Predictive Value (PPV) of the Trial Device for Diagnosing Coronary Heart Disease
Percentage of PPVFFR Measurement in Intermediate-Grade Coronary Stenosis
asymptotic normal method84.1 (74.7 to 93.4)
exact probability method84.1 (73.3 to 91.8)
SecondaryNegative Predictive Value (NPV) of the Trial Device for Diagnosing Coronary Heart Disease

After all subjects complete their procedures, using the subjects as the research unit, with the control device's determined coronary artery FFRpw as a reference, and with FFRpw ≤ 0.8 as a positive threshold, the negative predictive value of diagnosing coronary heart disease using the trial device is calculated. NPV = true negative / (true negative + false negative) Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · Percentage of NPV
Negative Predictive Value (NPV) of the Trial Device for Diagnosing Coronary Heart Disease
Percentage of NPVFFR Measurement in Intermediate-Grade Coronary Stenosis
asymptotic normal method96.9 (94.3 to 99.4)
exact probability method96.9 (93.6 to 98.7)
SecondaryReceiver Operating Characteristic (ROC) Curve of the Trial Device in Diagnosing Coronary Heart Disease

On a two-dimensional coordinate graph, set the x-axis as the false positive rate (FPR), and the y-axis as the true positive rate (TPR). For the FFRws values measured by the trial device, a point pair of TPR and FPR can be obtained based on its performance on the test sample. Adjust the diagnostic threshold of FFRws to obtain a series of points, connect them in sequence (including the points (0,0) and (1,1)), which forms the ROC curve of FFRws. The value of Area Under Curve (AUC) is the size of the area under the ROC curve. The curve is plotted using MedCalc software. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · Proportion Probability
Receiver Operating Characteristic (ROC) Curve of the Trial Device in Diagnosing Coronary Heart Disease
Proportion ProbabilityFFR Measurement in Intermediate-Grade Coronary Stenosis
Receiver Operating Characteristic (ROC) Curve of the Trial Device in Diagnosing Coronary Heart Disease0.9511 (0.92 to 0.98)
SecondaryPassing-Bablok Regression Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease

Using the vessel as the research unit, with the trial device's measured coronary artery FFRws value as the y-axis, and the control device's measured coronary FFRpw value as the x-axis, perform linear regression analysis on the FFR measurement results, and record the regression analysis equation, slope, and intercept. The analysis will be conducted at the lesion level. Regression analysis equation: Y=aX + b Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · trial device FFRws/control device FFRpw
Passing-Bablok Regression Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease
trial device FFRws/control device FFRpwFFR Measurement in Intermediate-Grade Coronary Stenosis
Passing-Bablok Regression Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease1 (0.95 to 1)
SecondaryBland-Altman Bias Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease

Bland-Altman analysis is a statistical method for evaluating the agreement between two measurement methods and for assessing bias. Using the vessel as the research unit, analyze the trial device's measured coronary artery FFRws value and the control device's measured coronary artery FFRpw value using Bland-Altman analysis, and plot the bias graph. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit. In the two-dimensional Cartesian coordinate of the bias graph, the x-axis represents the average value of the FFR measured by the two devices for each diseased vessel, and the y-axis represents the difference between the FFR values determined by the two devices for each diseased vessel. mean: (x-axis): (FFRws+FFRpw) / 2 absolute error (y-axis): FFRws - FFRpw The mean absolute error can then be calculated by taking the average of all absolute errors.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Mean · mean absolute error
Bland-Altman Bias Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease
mean absolute errorFFR Measurement in Intermediate-Grade Coronary Stenosis
Bland-Altman Bias Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease-0.01 ± 0.03
SecondaryPearson Linear Correlation Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease

After all subjects complete their procedures, using the vessel as the research unit, perform Pearson linear correlation analysis on the trial device's measured coronary artery FFRws value and the control device's measured coronary artery FFRpw value, and record the correlation coefficient. The correlation coefficient is a statistical measure that determines the strength of a linear relationship between two variables, ranging from -1 to 1. A value close to 1 represents a strong positive correlation between the x and y axis values (trail device FFRws and control device FFRpw). Because of the nature of this value, the measure type is "number". The coefficient is determined using the following formula: P = Cov(x,y) / OxOy where: Cov(x,y) = covariance of variables x and y Ox = standard deviation of x Oy = standard deviation of y

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · Correlation Coefficient
Pearson Linear Correlation Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease
Correlation CoefficientFFR Measurement in Intermediate-Grade Coronary Stenosis
Pearson Linear Correlation Analysis Results for the Trial Device in Diagnosing Coronary Heart Disease0.9091
SecondaryDevice Success Rate Analysis Results for Subjects

After all subjects complete their procedures, using the vessel as the research unit, with valid FFRws and FFRpw readings as the standard for successful device use, compare the device success rates of the trial device and the control device. Success rate is determined as the number of units that were performed successfully on the patient. The case was deemed successful if the unit was able to read a valid FFRws and FFRpw reading. Measurements are taken on patients with both the trial device and control device, sequentially during the same visit.

Time frame:
Acute - all measurements are taken and assessed during each procedure. Each participant completes their procedure in one visit, with each procedure taking up to 30 minutes.
Reported as:
Number · Number of successful devices
Device Success Rate Analysis Results for Subjects
Number of successful devicesFFR Measurement in Intermediate-Grade Coronary Stenosis - Test ArticleFFR Measurement in Intermediate-Grade Coronary Stenosis - Predicate Article
Device Success Rate Analysis Results for Subjects292293

Adverse events

Collected over Acute - following an initial screening visit, each participant completes their procedure in one visit, with each procedure taking up to 30 minutes. Follow-up time is 48 hours post-operatively. Adverse event data can be collected during the day of the procedure, or up to 48 hours afterwards during the follow up period (Visit 3). During the follow-up period, laboratory tests are performed to check for adverse events, comparing blood and other biochemical changes before and after surgery.. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
FFR Measurement in Intermediate-Grade Coronary Stenosis - Subject Device0/295 (0%)0/295 (0%)1/295 (0.3%)
FFR Measurement in Intermediate-Grade Coronary Stenosis - Control Device0/295 (0%)0/295 (0%)0/295 (0%)
Most frequent other events
Most frequent other events
EventFFR Measurement in Intermediate-Grade Coronary Stenosis - Subject DeviceFFR Measurement in Intermediate-Grade Coronary Stenosis - Control Device
Myocardial injurySurgical and medical procedures1/2950/295

Baseline characteristics

The study started with 300 participants, however, 7 patients never completed the study. Additionally, 1 participant violated protocol, however, their baseline characteristics were still recorded.

Age, Continuous
Age, Continuous(Years)FFR Measurement in Intermediate-Grade Coronary Stenosis
Mean63.47 ± 7.6
Sex: Female, Male
Sex: Female, Male(Participants)FFR Measurement in Intermediate-Grade Coronary Stenosis
Female103
Male190
Race/Ethnicity, Customized
Race/Ethnicity, Customized(Participants)FFR Measurement in Intermediate-Grade Coronary Stenosis
Han292
Other1
Region of Enrollment
Region of Enrollment(participants)FFR Measurement in Intermediate-Grade Coronary Stenosis
China293
Weight
Weight(kg)FFR Measurement in Intermediate-Grade Coronary Stenosis
Mean68.79 ± 11.09
Body Mass Index (BMI)
Body Mass Index (BMI)(kg/m^2)FFR Measurement in Intermediate-Grade Coronary Stenosis
Mean25.04 ± 3.11
History of Allergies
History of Allergies(Participants)FFR Measurement in Intermediate-Grade Coronary Stenosis
No Allergies251
Yes Allergies42
08

Study locations

1 site
  • Zhongshan Hospital
    Shanghai, Shanghai Municipality, China
09

References and documents

Study documents

  • Protocol and statistical analysis plan · Jul 11, 2022

Documents are hosted by the registry — open the source record to download them.

Individual participant data

Plan to share: No

10

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Dec 2, 2024, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT04582877
Lead sponsor
Zurich Medical Inc.
Responsible party
Sponsor
First posted
Oct 12, 2020
Start date
Sep 23, 2020
Primary completion
Nov 15, 2023
Completion
Apr 30, 2024
Results posted
Dec 2, 2024
Last update
Dec 2, 2024

Study contacts

Junbo Ge, MD
principal investigator · Shanghai Zhongshan Hospital

Oversight

FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Nov 2024. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion