TEE Image Quality Improvement With Our Devised Probe Cover

An interventional study of Suctioning orogastric tube which is attached to TEE probe cover in Image Quality, sponsored by Henry Ford Health System. Completed at 1 site in United States. Per ClinicalTrials.gov, last updated 2023-12-13.

Sponsored by Henry Ford Health System · Not applicable, Interventional, and Diagnostic

Phase
Not applicable
Study type
Interventional
Enrollment
53
Allocation
Not applicable
Sex
All
01

Study summary

With using suction tube attached TEE probe cover, we will assess its pinpoint suction capacity on image quality and surgical decision making.

Read the detailed description

Background: Transesophageal echocardiography (TEE) has become a standard intraoperative diagnosis technique for clinical management of patients during cardiac and non-cardiac surgery. Accurate intraoperative diagnosis by TEE improves as image quality improves. Although we recently reported enhanced image quality using a TEE probe with an attached orogastric tube, its clinical significance is still unknown. Also, we are concerned about potential clinical complications including damage to the upper gastrointestinal tract due to its rough surface. Therefore, we devised a new TEE probe cover equipped with a suction catheter and ultrasound gel containing pad that functions as a cushion for preventing surrounding tissue trauma. Our long-term goal is to improve TEE image quality, which will help surgical decision-making with precise assessment. The study population will be elective cardiac cases and liver transplant cases at Henry Ford Hospital. The objective of this grant is to assess the image quality change with our newly designed TEE probe, and its clinical utility on patient care in different types of procedures.

Aim 1. To assess TEE image quality before and after pin-point suctioning with our newly designed TEE probe. We hypothesize that decreasing stomach air between the TEE probe transducer and tissue wall by pin-point suction with our newly designed TEE probe will reduce ultrasound reflection and lead to better image quality.

Aim 2. To evaluate the clinical utility of our newly designed TEE probe cover. We hypothesize our newly designed TEE probe cover will facilitate decision-making for both anesthesiologists and surgeons and it will not increase the occurrence of TEE probe related trauma.

02

Conditions studied

  • Image Quality
03

In context

Lead sponsor

Henry Ford Health System is the lead sponsor of 303 studies on the registry; 50 are open to participants now.

Of its 37 completed or terminated interventional studies of FDA-regulated products, 24 (65%) have results posted.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
Child (0–17), Adult (18–64), Older adult (65+)
Sexes eligible
All
Accepts healthy volunteers
No

Eligibility criteria

Inclusion Criteria: Cardiac procedures and liver transplants which had intraoperative TEE -

Exclusion Criteria: general contraindication with TEE, apparent complications which are related to TEE.

-

05

Study design

Phase
Not applicable
Primary purpose
Diagnostic
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrollment
53 participants (actual)

Study arms

  • Other
    TEE image before and after suctioning orogastric tube

    for intraoperative TEE used cardiac or transplant cases, TEE images will be stored before and after suctioning orogastric tube which is attached to TEE probe cover. This Arm is TEE image BEFORE suctioning.

    Device: Suctioning orogastric tube which is attached to TEE probe cover

Interventions

  • DeviceSuctioning orogastric tube which is attached to TEE probe cover

    Suctioning orogastric tube which is atttached to TEE probe cover.

06

What researchers measure

Primary outcomes

  1. Image Quality Assessment Method#1: Difference in Likert Scale Before/After Suctioning

    Difference in image quality assessment by Likert scale before/after suctioning, by assessing quality of images stored before/after suctioning. Investigators categorized the quality of all acquired images on a numeric scale (the higher number means higher image quality) based on each investigator's impression (1: very poor, 2: poor, 3: acceptable, 4: good, and 5: very good). Image quality improvement was determined by increased number. Three investigators (A, B, C) assessed the quality of all TEE image sets (i.e. before and after suctioning) post hoc. The acquired numeric scales were combined, and compared before and after suctioning, and calculated how much percentage of patients had improved image quality (i.e. increase in numeric scale), same quality (i.e. same numeric scale), and worsened image quality (i.e. decrease in numeric scale)

    Time frame: TEE image sets were acquired after general anesthesia induction (before suctioning) and after 10minutes (after suctioning), and the outcome was the difference in image quality. In 6-8 months, investigator C did the same analysis on the same images.

Secondary outcomes

  1. Reproducibility of the LV FAC (Inter-observer)

    The reproducibility of the LV FAC was assessed, assuming that better image quality would yield better LV FAC reproducibility. Three investigators (A, B, and C) assessed the quality of all TEE image sets (i.e. before and after suctioning) post hoc. These image sets are combined and assessment was done for each group (i.e. before and after suctioning)

    Time frame: Images were acquired after general anesthesia induction (before suctioning) and after 10minutes (after suctioning images).

  2. Reproducibility of the LV FAC (Intra-observer)

    Investigator C analyzed all image sets again in 6-8 months to determine if there was intra-observer variability with the initial assessment of image quality. All investigators were blinded to which images were obtained before or after suctioning.

    Time frame: 6-8 months after initial images obtained during surgery.

  3. Reproducibility of the RV FAC (Inter-observer)

    The reproducibility of the RV FAC was assessed, assuming that better image quality would yield better RV FAC reproducibility. Three investigators (A, B, and C) assessed the quality of all TEE image sets (i.e. before and after suctioning) post hoc.

    Time frame: Images were acquired after general anesthesia induction (before suctioning) and after 10minutes (after suctioning).

  4. Reproducibility of the RV FAC (Intra-observer)

    Investigator C analyzed all image sets again in 6-8 months to determine if there was intra-observer variability with the initial assessment of image quality. All investigators were blinded to which images were obtained before or after suctioning.

    Time frame: 6-8 months after initial images obtained during surgery.

07

Results

Posted Dec 13, 2023

Participant flow

TEE Imaging Before Suctioning
Participant flow — TEE Imaging Before Suctioning
MilestoneTEE Image Before and After Suctioning Orogastric Tube
Started53
Completed50
Not completed3
Withdrew: Physician decision3
TEE Imaging After Suctioning
Participant flow — TEE Imaging After Suctioning
MilestoneTEE Image Before and After Suctioning Orogastric Tube
Started50
Completed50
Not completed0

Outcome measures

PrimaryImage Quality Assessment Method#1: Difference in Likert Scale Before/After Suctioning

Difference in image quality assessment by Likert scale before/after suctioning, by assessing quality of images stored before/after suctioning. Investigators categorized the quality of all acquired images on a numeric scale (the higher number means higher image quality) based on each investigator's impression (1: very poor, 2: poor, 3: acceptable, 4: good, and 5: very good). Image quality improvement was determined by increased number. Three investigators (A, B, C) assessed the quality of all TEE image sets (i.e. before and after suctioning) post hoc. The acquired numeric scales were combined, and compared before and after suctioning, and calculated how much percentage of patients had improved image quality (i.e. increase in numeric scale), same quality (i.e. same numeric scale), and worsened image quality (i.e. decrease in numeric scale)

Time frame:
TEE image sets were acquired after general anesthesia induction (before suctioning) and after 10minutes (after suctioning), and the outcome was the difference in image quality. In 6-8 months, investigator C did the same analysis on the same images.
Reported as:
Count of units · images
Image Quality Assessment Method#1: Difference in Likert Scale Before/After Suctioning
imagesTEE Image Before and After Suctioning Orogastric Tube
Improved after suction83
Same after suction65
Worse after suction2
SecondaryReproducibility of the LV FAC (Inter-observer)

The reproducibility of the LV FAC was assessed, assuming that better image quality would yield better LV FAC reproducibility. Three investigators (A, B, and C) assessed the quality of all TEE image sets (i.e. before and after suctioning) post hoc. These image sets are combined and assessment was done for each group (i.e. before and after suctioning)

Time frame:
Images were acquired after general anesthesia induction (before suctioning) and after 10minutes (after suctioning images).
Reported as:
Number · correlation coefficient
Reproducibility of the LV FAC (Inter-observer)
correlation coefficientTEE Image Before and After Suctioning Orogastric Tube
Intraclass correlation coefficients of LV FAC before suctioning0.94 (0.91 to 0.96)
Intraclass correlation coefficients of LV FAC after suctioning0.99 (0.98 to 0.99)
SecondaryReproducibility of the LV FAC (Intra-observer)

Investigator C analyzed all image sets again in 6-8 months to determine if there was intra-observer variability with the initial assessment of image quality. All investigators were blinded to which images were obtained before or after suctioning.

Time frame:
6-8 months after initial images obtained during surgery.
Reported as:
Number · correlation coefficient
Reproducibility of the LV FAC (Intra-observer)
correlation coefficientDifference in TEE Imaging Before/After Suctioning at 6 Months After the Surgery
Intraclass correlation coefficients of LV FAC before suctioning0.95 (0.91 to 0.99)
Intraclass correlation coefficients of LV FAC after suctioning1.00 (0.99 to 1.00)
SecondaryReproducibility of the RV FAC (Inter-observer)

The reproducibility of the RV FAC was assessed, assuming that better image quality would yield better RV FAC reproducibility. Three investigators (A, B, and C) assessed the quality of all TEE image sets (i.e. before and after suctioning) post hoc.

Time frame:
Images were acquired after general anesthesia induction (before suctioning) and after 10minutes (after suctioning).
Reported as:
Number · correlation coefficient
Reproducibility of the RV FAC (Inter-observer)
correlation coefficientDifference in TEE Imaging Before/After Suctioning at 6 Months After the Surgery
Intraclass correlation coefficients of RV FAC before suctioning0.89 (0.84 to 0.92)
Intraclass correlation coefficients of RV FAC after suctioning0.90 (0.86 to 0.94)
SecondaryReproducibility of the RV FAC (Intra-observer)

Investigator C analyzed all image sets again in 6-8 months to determine if there was intra-observer variability with the initial assessment of image quality. All investigators were blinded to which images were obtained before or after suctioning.

Time frame:
6-8 months after initial images obtained during surgery.
Reported as:
Number · correlation coefficient
Reproducibility of the RV FAC (Intra-observer)
correlation coefficientDifference in TEE Imaging Before/After Suctioning at 6 Months After the Surgery
Intraclass correlation coefficients of RV FAC before suctioning0.88 (0.81 to 0.92)
Intraclass correlation coefficients of RV FAC after suctioning0.85 (0.84 to 0.9)

Adverse events

Collected over 1month. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
TEE Image Before Suctioning Orogastric Tube0/50 (0%)0/50 (0%)0/50 (0%)
TEE Image After Suctioning Orogastric Tube0/50 (0%)0/50 (0%)0/50 (0%)

Baseline characteristics

Age, Categorical
Age, Categorical(Participants)TEE Image Before and After Suctioning Orogastric Tube
<=18 years0
Between 18 and 65 years50
>=65 years0
Age, Continuous
Age, Continuous(years)TEE Image Before and After Suctioning Orogastric Tube
Least squares mean61 (48 to 68)
Sex/Gender, Customized
Sex/Gender, Customized(Participants)TEE Image Before and After Suctioning Orogastric Tube
gender not known50
Race and Ethnicity Not Collected
Race and Ethnicity Not Collected(Participants)TEE Image Before and After Suctioning Orogastric Tube
Region of Enrollment
Region of Enrollment(participants)TEE Image Before and After Suctioning Orogastric Tube
United States50
08

Study locations

1 site
  • HFHS
    Detroit, Michigan 48202, United States
09

References and documents

Study documents

  • Protocol and statistical analysis plan · Jan 18, 2019
  • Informed consent form · Jan 18, 2019

Documents are hosted by the registry — open the source record to download them.

10

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reached

Ran from Jan 2019 to primary completion in Dec 2019, 11 months.

Results were first posted in Dec 2023, 3 years 11 months after primary completion.

How these figures are worked out, and which studies are left out of them.

Registered
Jan 18, 2019
First posted
Jan 23, 2019
StartActual
Jan 1, 2019
Primary completionActual
Dec 14, 2019
CompletionActual
Dec 31, 2019
Results first posted
Dec 13, 2023
Last update
Dec 13, 2023
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Dec 13, 2023, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT03812185
Collaborators
Henry Ford Hospital
Responsible party
Yoshihisa Morita (Principal INvestigator, Henry Ford Health System) — Principal investigator
First posted
Jan 23, 2019
Start date
Jan 1, 2019
Primary completion
Dec 14, 2019
Completion
Dec 31, 2019
Results posted
Dec 13, 2023
Last update
Dec 13, 2023

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Nov 2023. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion