ClinicalTrials.gg
CompletedNCT03416946Updated Oct 4, 2021Results posted

Custom Cutting Block Instrument vs Regular Instrumentation Total Knee Replacement (TKR)

An interventional study of Visionaire and Traditional in Osteoarthritis, Knee, sponsored by Orthopaedic Innovation Centre. Completed at 1 site in Canada. Open to participants aged 18 Years and older, including healthy volunteers. Per ClinicalTrials.gov, last updated 2021-10-04.

Sponsored by Orthopaedic Innovation Centre · Not applicable, Interventional, and Treatment

How we know

From the registry’s dates

  • Registered 6 years after the study started (first participant enrolled Dec 2011, registered Dec 2017).

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Phase
Not applicable
Study type
Interventional
Enrollment
54
Allocation
Randomized
Ages
18 Years and older
Sex
All
01

Where this study stands

Counted across the records on this site · 2026-11-27

Among 171 studies in Osteoarthritis, Knee with a site in Canada — 26 open to participants now, 39 under way, 6 of them phase 3. Osteoarthritis, Knee · Canada.

Enrollment
87th of 160
54 · median 63
Started
2011
with 7 others
No phase
100 of 171
19 open now
Registered
87th of 171
by date

Among the 171 studies in Osteoarthritis, Knee with a site in Canada, this one is the 87th largest by enrollment of the 160 interventional ones that give a count, is one of 8 that started in 2011, is one of 100 with no phase, and was the 87th registered.

Osteoarthritis, Knee

3,303 studies on the registry are indexed under Osteoarthritis, Knee; 627 are open to participants now.

Browse Osteoarthritis, Knee studies →

Open to participants now

26 of the 171 studies in Osteoarthritis, Knee with a site in Canada are open now. Recruiting first, the newest first:

21 more — sign in →

All 26 open studies →

A study can be open while its sites in Canada have not opened yet or have stopped. Only those recruiting at a site in Canada →

Who these studies take

Of the 386 Osteoarthritis, Knee studies recruiting now, 179 (46%) set no upper age limit and 114 (30%) stop at 75 or younger; 19 take children under 18; 39 take healthy volunteers.

This study: 18 Years and older · all sexes · healthy volunteers too.

The full table on the hub →

This tab as a report to print or save as PDF → With an account.

Was this useful?
Get alerts for this landscapeDownload CSVNarrow it in search →With an account: a weekly mail of what changed among 171 studies in Osteoarthritis, Knee with a site in Canada, or all of them as a file.
02

Study summary

To determine if the use of patient-specific custom cutting blocks for implantation of total knee components results in improved limb alignment and component positioning compared to regular instrumentation.

Read the detailed description

This is a randomized controlled trial. Patients will be randomized to have their knee implanted using either the patient-specific custom cutting blocks or the regular instrumentation system. Functional and radiographic outcomes will be assessed in a blinded fashion. The Smith and Nephew Legion Primary® total knee system will be used in both groups; this is an implant with a good long term track record based on the Genesis II design. All surgeons have used and are familiar with the regular instrumentation. They will be instructed in the use of the patient-specific custom cutting blocks prior to study initiation.

03

Conditions studied

  • Osteoarthritis, Knee

Keywords

  • Total knee replacement
04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
Yes

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  1. Patient is male or female ages 18 and over.
  2. Patient is having primary total knee replacement
  3. Patient is willing to sign the and to come for all study visits.

Exclusion criteria

Exclusion Criteria:

  1. Deformity of the femur preventing use of the intra-medullary guide utilized in the regular instrumentation set.
  2. Necessity for the use of constrained implants. These types of implants have intra-medullary stems, therefore all bone cuts need to be referenced off intramedullary guides, making image guided bone cuts inappropriate.
  3. Patients undergoing knee replacement revision surgery. Theses types of implants also have stems, making the use of image guided bone cuts inappropriate for the same reasons.
  4. Patients scheduled for bilateral knee surgery (simultaneous or staged)
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Informed consent

Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.

You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.

05

Study design

Phase
Not applicable
Primary purpose
Treatment
Allocation
Randomized
Intervention model
Parallel assignment
Masking
Single (Participant)
Enrollment
54 participants (actual)

Study arms

  • Active comparator
    Custom Block Instrumentation

    Patients-specific custom cutting blocks using the Smith and Nephew Visionaire system

    Device: Visionaire

  • Active comparator
    Traditional Instrumentation

    Traditional cutting methods for Total Knee Replacement

    Other: Traditional

Interventions

  • DeviceVisionaire

    Custom cutting block using MRI to create patient specific instrumentations

  • OtherTraditional

    Traditional cutting method

06

What researchers measure

Primary outcomes

  1. Radiographic Alignment

    Using 3 foot standing film

    Time frame: 8 Weeks

  2. Number of HKA Angle Outliers

    The angle formed between the mechanical axes of the femur and tibia create the hip-knee-ankle (HKA) angle.

    Time frame: 8 Weeks

Secondary outcomes

  1. EuroQol (EQ) Five Dimension (5D) Survey

    Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the EQ-5D (patient reported). The digits for the five dimensions can be combined into a 5-digit number that describes the patient's health state, this health state is range from 0 to 1.0. Higher scores indicate improved quality of life within each dimension.

    Time frame: Preoperative, 1 and 2 years

  2. Oxford Knee Score

    Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the Oxford 12 scores (patient reported). The score ranges from 0 to 48 with higher scores indicating improved function and less pain.

    Time frame: Preoperative, 1 and 2 years

  3. Pain Catastrophizing Scale (PCS) Score

    Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the PCS (patient reported). The total score ranges from 0 to 52 with higher scores indicating greater pain catastrophizing. Sub-scales will not be used in this study.

    Time frame: Preoperative, 1 and 2 years

  4. Hip Pain

    Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the Pain Visual Analog Scale (VAS). The scale is from 0 to 100, 0=no pain and at 100= worst pain. (patient reported).

    Time frame: Preoperative, 1 and 2 years

  5. University of California Los Angeles (UCLA) Activity Score

    Evaluate change from preoperative baseline to 1 and 2 year post operative in activity level assessments using the UCLA Activity score (patient reported). The scale ranges from 1-10 with higher values associated with more intense activities.

    Time frame: Preoperative, 1 and 2 years

  6. Patient Satisfaction

    Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the Satisfaction Visual Analog Scale (VAS). The scale is from 0 to 100, 0=Not Satisfied and at 100=Satisfied. (patient reported).

    Time frame: Preoperative, 1 and 2 years

Other outcomes

  1. Other Measure: Time

    Surgical and tourniquet time

    Time frame: Intra-operative

  2. Blood Loss

    Blood loss during surgery \>100ml (n)

    Time frame: Intra-operative

  3. Length of Stay (LOS)

    Length of hospital stay

    Time frame: Peri-operative

  4. Other Measures: Adverse Events

    Peri-operative complications such as deep venous thrombosis, pulmonary emboli, wound complications and cardiac events. Post-operative complication such as infection, loosening, etc.

    Time frame: Peri-operative, intra-operative and post-operative

07

Results

Posted Oct 4, 2021

Participant flow

Participant flow — Overall Study
MilestoneCustom Block InstrumentationTraditional Instrumentation
Started2529
Completed2529
Not completed00

Outcome measures

PrimaryRadiographic Alignment

Using 3 foot standing film

Time frame:
8 Weeks
Reported as:
Mean · degree
Radiographic Alignment
degreeCustom Block InstrumentationTraditional Instrumentation
Post-op HKA angle-1.6 ± 3.2-1.8 ± 2.7
Femoral coronal angle89.4 ± 1.989.3 ± 2.5
Tibial coronal angle89.2 ± 1.589.1 ± 1.6
Femoral sagittal angle87.4 ± 2.287.3 ± 2.7
Tibial sagittal angle3.8 ± 3.07.7 ± 3.6
Statistical analysis
  • Custom Block Instrumentation vs Traditional Instrumentation · t-test, 2 sided · p = 0.790 (Post-op HKA angle)
PrimaryNumber of HKA Angle Outliers

The angle formed between the mechanical axes of the femur and tibia create the hip-knee-ankle (HKA) angle.

Time frame:
8 Weeks
Reported as:
Number · outliers
Number of HKA Angle Outliers
outliersCustom Block InstrumentationTraditional Instrumentation
Number of HKA Angle Outliers610
SecondaryEuroQol (EQ) Five Dimension (5D) Survey

Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the EQ-5D (patient reported). The digits for the five dimensions can be combined into a 5-digit number that describes the patient's health state, this health state is range from 0 to 1.0. Higher scores indicate improved quality of life within each dimension.

Time frame:
Preoperative, 1 and 2 years
Reported as:
Mean · units on a scale
EuroQol (EQ) Five Dimension (5D) Survey
units on a scaleCustom Block InstrumentationTraditional Instrumentation
Pre-op0.760 ± 0.0580.744 ± 0.061
1 Year0.101 ± 0.0900.110 ± 0.086
2 Years0.112 ± 0.1100.120 ± 0.098
SecondaryOxford Knee Score

Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the Oxford 12 scores (patient reported). The score ranges from 0 to 48 with higher scores indicating improved function and less pain.

Time frame:
Preoperative, 1 and 2 years
Reported as:
Mean · units on a scale
Oxford Knee Score
units on a scaleCustom Block InstrumentationTraditional Instrumentation
Pre-op22.8 ± 7.819.6 ± 6.8
Improvement at 1 Year10.7 ± 11.415.3 ± 7.7
Improvement at 2 Year13.3 ± 12.412.8 ± 8.9
SecondaryPain Catastrophizing Scale (PCS) Score

Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the PCS (patient reported). The total score ranges from 0 to 52 with higher scores indicating greater pain catastrophizing. Sub-scales will not be used in this study.

Time frame:
Preoperative, 1 and 2 years
Reported as:
Mean · units on a scale
Pain Catastrophizing Scale (PCS) Score
units on a scaleCustom Block InstrumentationTraditional Instrumentation
Pre-op14.0 ± 12.317.4 ± 11.9
Improvement at 1 Year9.1 ± 12.310.5 ± 14.4
Improvement at 2 Year8.0 ± 13.76.4 ± 14.9
SecondaryHip Pain

Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the Pain Visual Analog Scale (VAS). The scale is from 0 to 100, 0=no pain and at 100= worst pain. (patient reported).

Time frame:
Preoperative, 1 and 2 years
Reported as:
Mean · units on a scale
Hip Pain
units on a scaleCustom Block InstrumentationTraditional Instrumentation
Pre-op61.6 ± 23.467.4 ± 20.6
Improvement at 1 Year-38.4 ± 34.4-40.1 ± 32.9
Improvement at 2 Year-40.8 ± 40.1-46.3 ± 24.2
SecondaryUniversity of California Los Angeles (UCLA) Activity Score

Evaluate change from preoperative baseline to 1 and 2 year post operative in activity level assessments using the UCLA Activity score (patient reported). The scale ranges from 1-10 with higher values associated with more intense activities.

Time frame:
Preoperative, 1 and 2 years
Reported as:
Mean · units on a scale
University of California Los Angeles (UCLA) Activity Score
units on a scaleCustom Block InstrumentationTraditional Instrumentation
Pre-op4.4 ± 1.83.4 ± 1.1
Improvement at 1 Year0.8 ± 1.921.6 ± 1.6
Improvement at 2 Year1.0 ± 2.01.6 ± 1.7
SecondaryPatient Satisfaction

Evaluate change from preoperative baseline to 1 and 2 year post operative in functional outcomes assessments using the Satisfaction Visual Analog Scale (VAS). The scale is from 0 to 100, 0=Not Satisfied and at 100=Satisfied. (patient reported).

Time frame:
Preoperative, 1 and 2 years
Reported as:
Mean · units on a scale
Patient Satisfaction
units on a scaleCustom Block InstrumentationTraditional Instrumentation
Pre-op23.0 ± 23.525.6 ± 26.2
Improvement at 1 Year59.1 ± 31.657.4 ± 28.5
Improvement at 2 Year62.0 ± 31.356.0 ± 29.6
Other pre-specifiedOther Measure: Time

Surgical and tourniquet time

Time frame:
Intra-operative
Reported as:
Mean · Minutes
Other Measure: Time
MinutesCustom Block InstrumentationTraditional Instrumentation
Other Measure: Time64.2 ± 16.557 ± 14
Other pre-specifiedBlood Loss

Blood loss during surgery \>100ml (n)

Time frame:
Intra-operative
Reported as:
Count of participants · Participants
Blood Loss
ParticipantsCustom Block InstrumentationTraditional Instrumentation
Blood Loss510
Other pre-specifiedLength of Stay (LOS)

Length of hospital stay

Time frame:
Peri-operative
Reported as:
Mean · Days
Length of Stay (LOS)
DaysCustom Block InstrumentationTraditional Instrumentation
Length of Stay (LOS)4.0 ± 1.54.2 ± 2.6
Other pre-specifiedOther Measures: Adverse Events

Peri-operative complications such as deep venous thrombosis, pulmonary emboli, wound complications and cardiac events. Post-operative complication such as infection, loosening, etc.

Time frame:
Peri-operative, intra-operative and post-operative
Reported as:
Number · participants
Other Measures: Adverse Events
participantsCustom Block InstrumentationTraditional Instrumentation
Other Measures: Adverse Events41

Adverse events

Collected over 2 Years. Non-serious events are listed at a 0% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
Custom Block Instrumentation0/25 (0%)4/25 (16%)0/25 (0%)
Traditional Instrumentation0/29 (0%)1/29 (3.4%)0/29 (0%)
Most frequent serious events
Most frequent serious events
EventCustom Block InstrumentationTraditional Instrumentation
RevisionInfections and infestations2/250/29
RevisionSurgical and medical procedures1/250/29
ManipulationSurgical and medical procedures1/250/29
RevisionSurgical and medical procedures0/251/29

Baseline characteristics

Age, Continuous
Age, Continuous(years)Custom Block InstrumentationTraditional InstrumentationTotal
Mean63.8 ± 9.265.7 ± 9.264.8 ± 9.1
Sex: Female, Male
Sex: Female, Male(Participants)Custom Block InstrumentationTraditional InstrumentationTotal
Female182240
Male7714
Race and Ethnicity Not Collected
Race and Ethnicity Not Collected(Participants)Custom Block InstrumentationTraditional InstrumentationTotal
Count of participants——0
Region of Enrollment
Region of Enrollment(Participants)Custom Block InstrumentationTraditional InstrumentationTotal
Canada252954
Body Mass Index
Body Mass Index(kg/m2)Custom Block InstrumentationTraditional InstrumentationTotal
Mean33.6 ± 7.232.0 ± 6.032.76 ± 6.6
Knee
Knee(Participants)Custom Block InstrumentationTraditional InstrumentationTotal
Right131225
Left121729
Pre-op HKA angle
Pre-op HKA angle(degree)Custom Block InstrumentationTraditional InstrumentationTotal
Mean-5.2 ± 5.9-6.7 ± 7.5-6.0 ± 6.8
08

Study locations

1 site
  • Concordia Hospital
    Winnipeg, Manitoba R2K 3S8, Canada
09

References and documents

Publications

  • Turgeon TR, Cameron B, Burnell CD, Hedden DR, Bohm ER. A double-blind randomized controlled trial of total knee replacement using patient-specific cutting block instrumentation versus standard instrumentation. Can J Surg. 2019 Dec 1;62(6):460-467. doi: 10.1503/cjs.018318. PubMed 31782643 ↗

Study documents

PDFs hosted by the registry, ClinicalTrials.gov.

Individual participant data

Plan to share: No

10

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reachedSimilar studies: half by the dot; the band is the middle halfHalf of similar studies reached it by the dot; the band is the middle half, from a quarter to three quarters

Ran from Dec 2011 to primary completion in May 2017, 65 months.

Results were first posted in Oct 2021, 4 years 5 months after primary completion.

Of 1,118 non-industry interventional studies with no phase under Osteoarthritis, Knee, half reached primary completion within 18 months of starting (the median), counting those still running; a quarter within 8 and three quarters within 36. This study's 65 months is above that.

Counting every sponsor, half of 1,258 studies reached it within 18 months; a quarter within 8 and three quarters within 36.

How these figures are worked out, and which studies are left out of them.

Registered
Dec 13, 2017
First posted
Jan 31, 2018
StartActual
Dec 1, 2011
Primary completionActual
May 1, 2017
CompletionActual
May 1, 2017
Results first posted
Oct 4, 2021
Last update
Oct 4, 2021
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Oct 4, 2021, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
11

Registry details

Key details

Study ID
NCT03416946
Lead sponsor
Orthopaedic Innovation Centre
Responsible party
Dr. Thomas Turgeon (Orthopaedic Surgeon, Orthopaedic Innovation Centre) — Principal investigator
First posted
Jan 31, 2018
Start date
Dec 1, 2011
Primary completion
May 1, 2017
Completion
May 1, 2017
Results posted
Oct 4, 2021
Last update
Oct 4, 2021

Study contacts

Thomas Turgeon, MD MPH FRCSC
principal investigator · Orthopaedic Innovation Centre

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Sep 2021. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion