An interventional study of SIPPV_VG_0_08 and SIPPV_VG_0_16 in Respiratory Distress Syndrome in Premature Infant, sponsored by Cambridge University Hospitals NHS Foundation Trust. Completed at 1 site in United Kingdom. Open to participants aged 1 Day to 2 Months. Per ClinicalTrials.gov, last updated 2019-05-20.
Sponsored by Cambridge University Hospitals NHS Foundation Trust · Not applicable, Interventional, and Treatment
Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →
During neonatal mechanical ventilation inflating pressures, tidal volumes, and inflation and expiration times need to be set and adjusted to optimise oxygenation and carbon dioxide removal. The flow of gas into the ventilator circuit has a big effect on ventilation but is usually set to a constant value (\~8 L/min) for all babies regardless of size or severity of illness, based on minimal research. High circuit flow may lead to lung damage and low flow to inadequate ventilation. The investigators recently developed a unique system to capture, record, analyse, and display ventilator data at high resolution over long periods. Using this the investigators will investigate, in within patient cross-over studies, how the level of gas flow affects ventilator parameters and ventilation, in two commonly used ventilation modes. The results will determine the lowest circuit flow that ventilates a baby safely and effectively. It will also provide preliminary data for a randomised trial.
About 1.5% of newborns require mechanical ventilation. Although mechanical ventilation can be life-saving for neonates with respiratory failure, it may cause lung injury due to excess airway pressure (barotrauma), delivery of high tidal volumes (volutrauma) and repetitive closing/re-opening of lung units (atelectotrauma)1. Very preterm neonates are especially vulnerable to ventilator induced lung injury. Ventilator associated lung injury is one of several factors contributing to the burden of chronic lung disease of infancy, also called bronchopulmonary dysplasia (BPD).
Ventilators have several different modes. The most frequently used mode is, time-cycled, pressure limited, where the doctor sets inspired oxygen level, the peak inflation pressure (PIP), rate, and inflation time of the ventilator. In this mode the inflations are usually synchronised with the baby's breathing. This is either Synchronised Intermittent Positive Pressure Ventilation (SIPPV), where the ventilator synchronises inflations with all the baby's breaths, or Synchronised Intermittent Mandatory Ventilation (SIMV), where the ventilator only synchronises with a set number of breaths. These can be delivered with or without targeting the tidal volume delivered to a set value by adjusting the peak inflating pressure. This mode is called volume guarantee (VG). A similar mode of ventilation is flow-cycled or Pressure Support Ventilation (PSV) and can be combined with tidal volume targeting 2. In this mode the inflation is terminated as soon as the flow rate falls to 15% of the maximum flow during inflation.
These modes have continuous gas flow through the ventilation circuit. Inflation starts and pressure rises when the expiratory valve is closed. The rate of circuit flow alters the ventilation waveforms. With time-cycled ventilation the higher the circuit flow the faster the pressure rises, the lung is distended and potentially injured, and the earlier the set PIP is achieved. A high flow, with a relatively long inflation time will result in a "pressure plateau", that is the PIP is sustained after the flow, into the baby, has stopped because the lung volume is maximal at the PIP used (Figure 1). In PSV inflation stops when the flow has decreased to \~15% of the peak inflation flow and so the inflation time is shorter. Increasing the device flow shortens the inflation time and thereby reduces the mean airway pressure.
Despite the effects on ventilation patterns and the speed of lung distension and injury, consideration has rarely been given to the circuit flow. By protocol, it is usually set at 7-10 L/min when the ventilator is turned on and it is not changed.
This is relatively high, and usually produces a "square" pressure waveform with a rapid distention of the lungs and a sustained pressure plateau. (see A in figure 1) With PSV a high flow results in a relatively short inflation time (\~0.2 sec; it needs to be at least 0.3 sec for adequate lung aeration). There is no evidence these high flows are best for optimum ventilation and minimum lung damage. In a preterm lamb model there were no adverse effects on gas exchange or cardiovascular parameters until the flow was reduced to 3 l/min 3. In animal studies ventilation with high flow resulted in histologic and molecular changes of lung injury 4. The effect of lowering the ventilator circuit flow rate has never been investigated in clinical studies.
The Dräger Babylog VN500 ventilator has an alternative to setting a gas flow: the user can set the slope time instead, that is time required to reach the set pressure. In Cambridge it is invariably set to 0.08 sec which results in a flow \~7 L/min. As we use an inflation time between 0.33 - 0.45 sec, there is a pressure plateau lasting at least 0.25 sec and sustained inflation with little or no flow.
The investigators are the first to develop a unique system for downloading and analysing data from the Dräger VN500 neonatal ventilator. Using DataGrabber software obtained from Dräger Medical, the investigators can retrieve ventilator parameters at 100Hz frequency over long periods (hours and even days). To analyse and visualise the large datasets the investigators developed a data analysis workflow using the Python programing language and its add-on packages (Figure 2). With this tool the investigators can now study details of each inflation and spontaneous breath. The Investigators have recorded detailed data from 30 babies and shown it is feasible, accurate, and the investigators have the expertise (Belteki et al, submitted).
In this application the investigators propose to investigate the effect of different slope times (and therefore different levels of circuit flow) on ventilation parameters and gas exchange in preterm infants. The investigators hypothesise that a longer slope time (= lower circuit flow) will distend the lungs more gently yet maintain ventilation and gas exchange.
The study is a within patient crossover design comparing short periods of ventilation with different slope times both in SIPPV-VG and PSV-VG modes with the following interventions:
A ventilator download tool and transcutaneous and expired CO2 monitors is attached to the ventilator and data download commences while the baby is ventilated with the parameters used by the clinical team. An arterial blood gas is performed. The ventilated parameters are changed as shown below. The order of these epochs is randomised. Another arterial blood gas is performed. Ventilator is changed back to the original parameters (or different as appropriate by the blood gas). Ventilator data and CO2 recording will be continued for another 30 minutes.
Interventions
Duration Ventilator More Slope time Inspiratory time[max] 15 min SIPPV-VG 0.08 0.40 15 min PSV-VG 0.08 [0.60] 15 min PSV-VG 0.16 [0.60] 15 min SIPPV-VG 0.16 0.40 15 min SIPPV-VG 0.24 0.40 15 min PSV-VG 0.24 [0.60] 15 min PSV-VG 0.32 [0.60] 15 min SIPPV-VG 0.32 0.40 15 min SIPPV-VG 0.40 0.40 115 min PSV-VG 0.40 [0.60]
Total study duration is 220 minutes. A researcher will be present continuously. FiO2 will be adjusted if needed to maintain saturations between 90-95%. If the FiO2 rises >15% or the end-tidal CO2 rises >1.5kPa over the pre-study level, that intervention will be abandoned.
Comparison:
At each slope time, the following parameters will be determined and compared with those recorded at 0.08 sec slope time: (1) peak inflating pressure, (2) inflation duration, (3) duration of inflation plateau, (4) duration of no gas flow, (5) expired tidal and minute volumes (mandatory/spontaneous, inspiratory/expiratory, (6) ventilator rate, (7) FiO2, (8) transcutaneous and/or end-tidal CO2 and, (9) interaction between ventilator inflations and baby's breaths. Values for SIPPV and PSV will also be compared.
33 studies on the registry are indexed under Respiratory Distress Syndrome In Premature Infants; 9 are open to participants now.
This study's enrollment of 12 is below the median of 43 across 28 interventional studies indexed under Respiratory Distress Syndrome In Premature Infants.
Browse Respiratory Distress Syndrome In Premature Infants studies →Cambridge University Hospitals NHS Foundation Trust is the lead sponsor of 171 studies on the registry; 38 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
SIPPV VG ventilation for 15 minutes slope time = 0.08 seconds inspiratory time = 0.40 seconds
Device: SIPPV_VG_0_08
PSV VG ventilation for 15 minutes slope time = 0.08 seconds maximum inspiratory time = 0.60 seconds
Device: PSV_VG_0_08
SIPPV VG ventilation for 15 minutes slope time = 0.16 seconds inspiratory time = 0.40 seconds
Device: SIPPV_VG_0_16
PSV VG ventilation for 15 minutes slope time = 0.16 seconds maximum inspiratory time = 0.60 seconds
Device: PSV_VG_0_16
SIPPV VG ventilation for 15 minutes slope time = 0.24 seconds inspiratory time = 0.40 seconds
Device: SIPPV_VG_0_24
PSV VG ventilation for 15 minutes slope time = 0.24 seconds maximum inspiratory time = 0.60 seconds
Device: PSV_VG_0_24
SIPPV VG ventilation for 15 minutes slope time = 0.32 seconds inspiratory time = 0.40 seconds
Device: SIPPV_VG_0_32
PSV VG ventilation for 15 minutes slope time = 0.32 seconds maximum inspiratory time = 0.60 seconds
Device: PSV_VG_0_32
SIPPV VG ventilation for 15 minutes slope time = 0.32 seconds inspiratory time = 0.40 seconds
Device: SIPPV_VG_0_40
PSV VG ventilation for 15 minutes slope time = 0.40 seconds maximum inspiratory time = 0.60 seconds
Device: PSV_VG_0_40
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.08 seconds, inspiratory time of 0.40 seconds for 15 minutes
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.16seconds, inspiratory time of 0.40 seconds for 15 minutes
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.24 seconds, inspiratory time of 0.40 seconds for 15 minutes
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.32 seconds, inspiratory time of 0.40 seconds for 15 minutes
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.40 seconds, inspiratory time of 0.40 seconds for 15 minutes
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.08 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.16 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.24 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.32 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.40 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
End tidal CO2 measurement
Primary outcome will be the difference in end-tidal CO2 concentration during the epochs with slope times of 0.40 and 0.08 sec.
Time frame: One year
Plan to share: Undecided
Ran from Jul 2017 to primary completion in Jun 2018, 11 months.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in May 2019. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Respiratory Distress Syndrome In Premature Infants →
Cambridge University Hospitals NHS Foundation Trust →