An interventional study of Medtronic MiniMed 530G Insulin Pump in Diabetes, sponsored by Medtronic MiniMed, Inc.. Completed at 43 sites in United States. Open to participants aged 16 Years and older. Per ClinicalTrials.gov, last updated 2019-03-05.
Sponsored by Medtronic MiniMed, Inc. (part of Medtronic) · Not applicable, Interventional, and Treatment
The purpose of this study is to evaluate the Threshold Suspend (TS) feature of the Medtronic MiniMed® 530G insulin pump in patients 16 and older with insulin requiring diabetes.
Multi-center trial is initiated to observe the Threshold Suspend (TS) feature with a sensor-augmented insulin pump (Medtronic MiniMed® 530G insulin pump) in patients 16 and older with insulin requiring diabetes over a period of one year.
11,090 studies on the registry are indexed under Diabetes Mellitus; 1,319 are open to participants now.
This study's enrollment of 372 is above the median of 80 across 8,491 interventional studies indexed under Diabetes Mellitus.
Browse Diabetes Mellitus studies →Medtronic MiniMed, Inc. is the lead sponsor of 13 studies on the registry; 4 are open to participants now.
Of its 6 completed or terminated interventional studies of FDA-regulated products, 4 (67%) have results posted.
First submitted after Medtronic MiniMed, Inc. became part of Medtronic.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
All subjects received diabetes treatment using the Medtronic MiniMed 530G insulin pump.
Device: Medtronic MiniMed 530G Insulin Pump
Mean Change in A1C From Baseline to 1 Year
Comparison of A1C measurement from baseline to end of study in the CEP266 study population. The overall mean change in A1C from baseline will be estimated and compared by a non-inferiority test with an A1C margin of 0.4% and a significance level of 0.025 (one-sided) with the CEP 266 study population.
Time frame: 1 year
Mean Change in A1C From Baseline to 1 Year, Baseline A1c Below 7%
Mean Change in A1C From Baseline to 1 Year, for subjects with baseline A1c below 7%.
Time frame: 1 year
Mean Change in A1C From Baseline to 1 Year, Baseline A1c of 7% to 9%
Mean Change in A1C From Baseline to 1 Year, for subjects with Baseline A1c of 7% to 9%
Time frame: 1 year
Mean Change in A1C From Baseline to 1 Year, Baseline A1c > 9%
Mean Change in A1C From Baseline to 1 year, for subjects with Baseline A1c \> 9%
Time frame: 1 year
| Milestone | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Started | 372 |
| Completed | 305 |
| Not completed | 67 |
| Withdrew: Death | 3 |
| Withdrew: Lost to follow-up | 12 |
| Withdrew: Withdrawal by subject | 52 |
Comparison of A1C measurement from baseline to end of study in the CEP266 study population. The overall mean change in A1C from baseline will be estimated and compared by a non-inferiority test with an A1C margin of 0.4% and a significance level of 0.025 (one-sided) with the CEP 266 study population.
| percentage | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Mean Change in A1C From Baseline to 1 Year | 0.034 (-0.03 to 0.098) |
Mean Change in A1C From Baseline to 1 Year, for subjects with baseline A1c below 7%.
| percentage | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Mean Change in A1C From Baseline to 1 Year, Baseline A1c Below 7% | 0.32 ± 0.53 |
Mean Change in A1C From Baseline to 1 Year, for subjects with Baseline A1c of 7% to 9%
| percentage | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Mean Change in A1C From Baseline to 1 Year, Baseline A1c of 7% to 9% | -0.04 ± 0.67 |
Mean Change in A1C From Baseline to 1 year, for subjects with Baseline A1c \> 9%
| percentage | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Mean Change in A1C From Baseline to 1 Year, Baseline A1c > 9% | -0.39 ± 0.96 |
Collected over 1 year. Non-serious events are listed at a 0% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| Medtronic MiniMed 530G Insulin Pump | 3/372 (0.8%) | 19/372 (5.1%) | 37/372 (9.9%) |
| Event | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Diabetic ketoacidosisMetabolism and nutrition disorders | 5/372 |
| HyperglycaemiaMetabolism and nutrition disorders | 4/372 |
| Acute myocardial infarctionCardiac disorders | 1/372 |
| Altered state of consciousnessNervous system disorders | 1/372 |
| Bile duct obstructionHepatobiliary disorders | 1/372 |
| Cellulitis staphylococcalInfections and infestations | 1/372 |
| Chest painGeneral disorders | 1/372 |
| DeathGeneral disorders | 1/372 |
| Gastric ulcerGastrointestinal disorders | 1/372 |
| HypoglycaemiaMetabolism and nutrition disorders | 1/372 |
| Event | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| HyperglycaemiaMetabolism and nutrition disorders | 10/372 |
| Puncture site haemorrhageGeneral disorders | 9/372 |
| HypoglycaemiaMetabolism and nutrition disorders | 4/372 |
| ContusionInjury, poisoning and procedural complications | 2/372 |
| Skin infectionInfections and infestations | 2/372 |
| Skin irritationSkin and subcutaneous tissue disorders | 2/372 |
| Abscess limbInfections and infestations | 1/372 |
| CellulitisInfections and infestations | 1/372 |
| ConvulsionNervous system disorders | 1/372 |
| DermatophytosisInfections and infestations | 1/372 |
| Age, Continuous(Years) | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Mean | 45.5 ± 15.9 |
| Sex: Female, Male(Participants) | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Female | 201 |
| Male | 171 |
| Ethnicity (NIH/OMB)(Participants) | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| Hispanic or Latino | 15 |
| Not Hispanic or Latino | 351 |
| Unknown or Not Reported | 6 |
| Region of Enrollment(participants) | Medtronic MiniMed 530G Insulin Pump |
|---|---|
| United States | 372 |
Documents are hosted by the registry — open the source record to download them.
Plan to share: No
Ran from Nov 2013 to primary completion in Feb 2018, 50 months.
Results were first posted in Mar 2019, 1 year after primary completion.
Of 192 industry-sponsored interventional studies with no phase under Diabetes Mellitus, half reached primary completion within 14 months of starting (the median), counting those still running; a quarter within 6 and three quarters within 29. This study's 50 months is above that.
Counting every sponsor, half of 2,388 studies reached it within 20 months; a quarter within 9 and three quarters within 37.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in Feb 2019. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.