Safety and Efficacy Study of the QuickFlex Micro Model 1258T Left Heart Pacing Lead

An interventional study of QuickFlex Micro Model 1258T Left Heart Lead in Heart Failure, sponsored by Abbott Medical Devices. Completed at 13 sites in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2019-02-19.

Sponsored by Abbott Medical Devices (part of Abbott) · Not applicable, Interventional, and Treatment

Phase
Not applicable
Study type
Interventional
Enrollment
86
Allocation
Non-randomized
Ages
18 Years and older
Sex
All
01

Study summary

This is a prospective, multi-center non-randomized clinical study to evaluate the safety and efficacy of the QuickFlex Micro Model 1258T left ventricular lead in a heart failure patient population. The study will be conducted at a maximum of 20 investigational centers located in the United States and a total of 68 patients will be enrolled.

Read the detailed description

The investigational left heart lead, QuickFlex Model 1258T, will be implanted with a legally marketed St. Jude Medical cardiac resynchronization therapy implantable cardioverter defibrillator (CRT-D) device.

Patients will be followed at Pre-Discharge, 1-Month, 3-Month,and then every 6 months until the end of the study.

02

Conditions studied

  • Heart Failure

Browse trials for

Keywords

  • Left heart pacing lead
  • Cardiac resynchronization therapy (CRT)
  • Heart failure
  • ICD
03

In context

Heart Failure

5,803 studies on the registry are indexed under Heart Failure; 1,222 are open to participants now.

This study's enrollment of 86 is above the median of 72 across 3,790 interventional studies indexed under Heart Failure.

Browse Heart Failure studies →

Lead sponsor

Abbott Medical Devices is the lead sponsor of 539 studies on the registry; 33 are open to participants now.

Of its 53 completed or terminated interventional studies of FDA-regulated products, 44 (83%) have results posted.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years and older
Sexes eligible
All
Accepts healthy volunteers
No

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Approved indication for CRT-D system

Exclusion criteria

Exclusion Criteria:

  • Previous cerebrovascular accident (CVA) or transient ischemic attack (TIA) within 3 months of enrollment
  • Previous left ventricular (LV) lead implant
  • Have had a heart attack,unstable angina or cardiac revascularization (percutaneous transluminal coronary angioplasty (PTCA),coronary artery bypass grafting (CABG), Stent) within 40 days prior to enrollment
05

Study design

Phase
Not applicable
Primary purpose
Treatment
Allocation
Non-randomized
Intervention model
Single group
Masking
None (open label)
Enrollment
86 participants (actual)

Study arms

  • Experimental
    QuickFlex micro 1258T left heart lead

    Device: QuickFlex Micro Model 1258T Left Heart Lead

Interventions

  • DeviceQuickFlex Micro Model 1258T Left Heart Lead

    Implantation of QuickFlex Micro 1258T left heart lead in combination with a CRT-D system.

    Also known as: Model 1258T left heart pacing lead

06

What researchers measure

Primary outcomes

  1. Freedom From Left Ventricular Lead-related Complications

    A Kaplan-Meier survival analysis was carried out for left ventricular lead related complications through 3 months. Percentage of patients who remained free from complications at 3 months was reported.

    Time frame: 3 months

  2. Percentage of Successful Left Ventricular Lead Implants

    Left ventricular lead implant success rate was calculated as the total number of patients who had a successful implant of the left ventricular QuickFlex Micro Model 1258T lead divided by the total number of patients who had an attempted implant. A successful implant was defined as the placement of the left ventricular lead in the coronary sinus for the purposes of pacing of the left ventricle with connection to the pulse generator.

    Time frame: 3 months

  3. Left Ventricular Bipolar Pacing Capture Threshold (Volts)

    The mean left ventricular capture threshold (amount of energy needed to pace the heart) is reported. An overall mean capture threshold of \< 3 volts is required to meet this endpoint.

    Time frame: 3 months

07

Results

Posted Nov 4, 2011

Participant flow

The study was conducted at 18 investigational centers located in the United States (US). The first patient was enrolled on October 30, 2008 and follow up was completed on May 31, 2009.

Participant flow — Overall Study
MilestoneQuickFlex Micro 1258T Left Heart Lead
Started86
Completed76
Not completed10
Withdrew: Unsuccessful implant5
Withdrew: Death4
Withdrew: Withdrawal by subject1

Outcome measures

PrimaryFreedom From Left Ventricular Lead-related Complications

A Kaplan-Meier survival analysis was carried out for left ventricular lead related complications through 3 months. Percentage of patients who remained free from complications at 3 months was reported.

Time frame:
3 months
Reported as:
Number · percent of participants
Freedom From Left Ventricular Lead-related Complications
percent of participantsQuickFlex Micro 1258T Left Heart Lead
Freedom From Left Ventricular Lead-related Complications100 (100 to 100)
Statistical analysis
  • QuickFlex Micro 1258T Left Heart Lead · Kaplan-Meier Survival Analysis · Objective performance criteria: 80
PrimaryPercentage of Successful Left Ventricular Lead Implants

Left ventricular lead implant success rate was calculated as the total number of patients who had a successful implant of the left ventricular QuickFlex Micro Model 1258T lead divided by the total number of patients who had an attempted implant. A successful implant was defined as the placement of the left ventricular lead in the coronary sinus for the purposes of pacing of the left ventricle with connection to the pulse generator.

Time frame:
3 months
Reported as:
Number · percentage of participants
Percentage of Successful Left Ventricular Lead Implants
percentage of participantsQuickFlex Micro 1258T Left Heart Lead
Percentage of Successful Left Ventricular Lead Implants94 (87 to 100)
Statistical analysis
  • QuickFlex Micro 1258T Left Heart Lead · Wilson score interval method · Objective performance criteria: 80
PrimaryLeft Ventricular Bipolar Pacing Capture Threshold (Volts)

The mean left ventricular capture threshold (amount of energy needed to pace the heart) is reported. An overall mean capture threshold of \< 3 volts is required to meet this endpoint.

Time frame:
3 months
Reported as:
Mean · volts
Left Ventricular Bipolar Pacing Capture Threshold (Volts)
voltsQuickFlex Micro 1258T Left Heart Leads
Left Ventricular Bipolar Pacing Capture Threshold (Volts)1.1 ± 0.3

Adverse events

Collected over Data was collected from the time the first patient was enrolled until follow up was completed; a total of 7 months.. Non-serious events are listed at a 0.1% frequency threshold.

Adverse event summary by group
GroupDeathsSeriousOther
QuickFlex Micro 1258T Left Heart Lead—4/86 (4.7%)20/86 (23.3%)
Most frequent serious events
Most frequent serious events
EventQuickFlex Micro 1258T Left Heart Lead
Cardiac vein thrombusCardiac disorders1/86
HematomaSkin and subcutaneous tissue disorders1/86
Right ventricular lead dislodgement/migrationSurgical and medical procedures1/86
Right ventricular lead cardiac perforationCardiac disorders1/86
Most frequent other events
Showing 10 of 11
Most frequent other events
EventQuickFlex Micro 1258T Left Heart Lead
Phrenic Nerve/Diaphragmatic Nerve stimulationSurgical and medical procedures10/86
HematomaSkin and subcutaneous tissue disorders4/86
High defibrillation thresholdCardiac disorders3/86
Therapy for non-ventricular rhythmCardiac disorders3/86
Cardiac vein thrombusCardiac disorders1/86
HypotensionCardiac disorders1/86
Left ventricular lead dislodgement/migrationSurgical and medical procedures1/86
Elevated left ventricular pacing thresholdCardiac disorders1/86
Right atrial lead undersensingCardiac disorders1/86
Device pocket stimulationSurgical and medical procedures1/86

Baseline characteristics

Age, Categorical
Age, Categorical(Participants)QuickFlex Micro 1258T Left Heart Lead
<=18 years0
Between 18 and 65 years21
>=65 years65
Age, Continuous
Age, Continuous(years)QuickFlex Micro 1258T Left Heart Lead
Mean71.0 ± 11.3
Sex: Female, Male
Sex: Female, Male(Participants)QuickFlex Micro 1258T Left Heart Lead
Female29
Male57
Region of Enrollment
Region of Enrollment(participants)QuickFlex Micro 1258T Left Heart Lead
United States86
08

Study locations

13 sites
  • University Hospital-University of Alabama at Birmingham
    Birmingham, Alabama 35233, United States
  • Arkansas Cardiology
    Little Rock, Arkansas 72205, United States
  • Arkansas Heart Hospital
    Little Rock, Arkansas 72211, United States
  • Glendale Memorial Hospital and Medical Center
    Glendale, California 91204, United States
  • Regional Cardiology Associates
    Sacramento, California 95819, United States
  • Midwest Heart Foundation
    Lombard, Illinois 60148, United States
  • Central Baptist Hospital
    Lexington, Kentucky 40503, United States
  • St. John Hospital and Medical Center
    Detroit, Michigan 48236, United States
  • Thoracic Cardiovascular Healthcare Foundation
    Lansing, Michigan 48910, United States
  • Deborah Heart and Lung Center
    Browns Mills, New Jersey 08015, United States
  • St. Francis Hospital
    Roslyn, New York 11576, United States
  • EMH Regional Medical Center
    Elyria, Ohio 44035, United States
  • St. Thomas Hospital
    Nashville, Tennessee 37205, United States
09

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Feb 19, 2019, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT00763698
Responsible party
Sponsor
First posted
Oct 1, 2008
Start date
Oct 2008
Primary completion
Jun 2009
Completion
May 2010
Results posted
Nov 4, 2011
Last update
Feb 19, 2019

Study contacts

Tamara Shipman
study director · Abbott Medical Devices

Oversight

Data monitoring committee
No
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Feb 2019. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion