Validation of Perin Health Patch ECG Monitoring Device Performance and Placement Accuracy

An observational study in ECG, Electrocardiography and Monitoring, Physiologic, sponsored by Perin Health Devices. Completed at 1 site in United States. Open to participants aged 22 Years to 90 Years, including healthy volunteers. Per ClinicalTrials.gov, last updated 2025-08-29.

Sponsored by Perin Health Devices · Observational

Study type
Observational
Model
Cohort
Time perspective
Prospective
Enrollment
255
Ages
22 Years to 90 Years
Sex
All
01

Study summary

The goal of this clinical trial is to test how accurately and reliably a new chest-worn device called the Perin Health Patch measures heart rhythms (ECG) in adults aged 22 and older. The main questions it aims to answer are:

  • How accurately does the Perin Health Patch measure heart rhythm compared to standard hospital ECG monitoring equipment?
  • Does the patch still work properly if it's placed slightly off from the exact recommended position on the chest?
  • Does the patch maintain good performance when worn continuously for 15 days?

The study has three parts that test different aspects of the device. Participants will:

  • Wear the Perin Health Patch on their chest along with standard ECG monitoring equipment for comparison
  • Have ECG recordings taken while sitting quietly for 15 minutes
  • For the placement study: have the patch placed in several different positions to test if small placement errors affect performance
  • For the long-term study: wear the patch continuously for 15 days at home with regular check-ups on days 1, 7, 10, and 15

The study will include both healthy adults and people with heart or lung conditions like heart failure, COPD, high blood pressure, or diabetes to test how well the device works for different types of patients. Researchers will compare the heart rhythm measurements from the chest patch to standard medical monitoring equipment to determine if the patch is accurate enough for medical use. Some participants will complete just one part of the study, while others may participate in multiple parts depending on which aspect of the device is being tested.

02

Conditions studied

  • ECG
  • Electrocardiography
  • Monitoring, Physiologic

Keywords

  • ECG validation study
  • electrocardiogram accuracy
03

In context

Lead sponsor

Perin Health Devices is the lead sponsor of 6 studies on the registry; 2 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
22 Years to 90 Years
Sexes eligible
All
Accepts healthy volunteers
Yes
Sampling method
Non-probability sample

Study population

Participants will be recruited from multiple clinical sites including hospital settings, clinical research facilities, and home healthcare providers. The population includes both healthy adult volunteers and patients with chronic cardiovascular or pulmonary conditions such as COPD, congestive heart failure, hypertension, coronary artery disease, chronic kidney disease, and diabetes. Recruitment targets a demographically diverse sample representative of the intended use population for remote cardiac monitoring, with specific attention to age, sex, BMI, and racial/ethnic diversity to ensure robust validation across patient subgroups commonly encountered in clinical practice.

Inclusion criteria

  • Adults 22 to 90 years of age
  • Healthy or chronic cardiopulmonary conditions or recent procedures
  • Able to comply with study procedure

Exclusion criteria

Exclusion Criteria:

  • Pediatric or adolescent patients (age ≤21 years)
  • Pacemaker
  • A history of skin reactions to medical adhesives
  • Any psychosocial condition or circumstance that, in the opinion of the investigator, would interfere with the conduct of the study or the patient's care and/or safety
  • Pregnant
  • Current participation in any other clinical study
  • Refusal to participate
05

Study design

Observational model
Cohort
Time perspective
Prospective
Enrollment
255 participants (actual)
Patient registry
No

Interventions

  • DeviceChest-Worn ECG Monitoring Device

    The Perin Health Patch is a non-invasive, chest-worn wearable device designed to continuously monitor electrocardiogram (ECG) signals using dry conductive adhesive electrodes. In this validation study, the device will be evaluated for its ability to accurately measure ECG parameters including heart rate, QT interval, QRS duration, and R-R intervals compared to FDA-cleared Holter monitors. The study intervention involves three components: (1) simultaneous ECG recording with the patch and reference Holter monitor for 15 minutes while participants remain seated, (2) placement tolerance testing where the patch is positioned at various locations (±20mm from optimal placement) to assess performance under suboptimal positioning, and (3) long-term wear evaluation where participants wear the patch continuously for 15 days with periodic assessments. The intervention is limited to device application and ECG data collection under controlled conditions, with no therapeutic treatment provided.

06

What researchers measure

Primary outcomes

  1. ECG Timing Interval Equivalence Between Perin Health Patch and Reference Holter Monitor

    Equivalence of ECG measurements between the investigational chest patch and FDA-cleared Holter monitor using Two One-Sided Tests (TOST) statistical methodology. ECG recordings are taken simultaneously from both devices while participants sit quietly for comparison. Equivalence is demonstrated when confidence intervals for mean differences fall within pre-specified clinical margins for key ECG timing intervals including QT interval, QRS duration, and R-R interval. Signal morphology similarity is assessed using Pearson correlation coefficient (where 1.0 indicates perfect similarity) and normalized root mean square error (where 0% indicates identical signals).

    Time frame: During single 15-minute recording session and up to 15 days of continuous monitoring

Secondary outcomes

  1. Heart Rate Measurement Accuracy

    Accuracy of heart rate measurements from the investigational device compared to the reference Holter monitor. Heart rate is calculated from R-R intervals during simultaneous ECG recordings. Accuracy is assessed using mean absolute error between devices, where lower values indicate better agreement.

    Time frame: During single 15-minute recording session and up to 15 days of continuous monitoring

  2. ECG Signal Quality Equivalence

    Comparison of signal quality between the investigational device and reference monitor using signal-to-noise ratio analysis. Signal quality is measured in decibels (dB), where higher values indicate cleaner signals with less noise interference. Equivalence is assessed using statistical comparison methods to determine if signal quality differences fall within clinically acceptable ranges.

    Time frame: During single 15-minute recording session and up to 15 days of continuous monitoring

  3. Device Adhesion Success Rate During Extended Wear

    Percentage of participants who successfully complete the full monitoring period without device adhesion failure. Adhesion failure is defined as device detachment that prevents reliable ECG signal acquisition. Success is measured by the proportion of participants maintaining adequate device adherence throughout the intended wear duration.

    Time frame: Over 15-day continuous monitoring period

07

Study locations

1 site
  • Perin Health Devices
    Woodland Hills, California 91364, United States
08

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Aug 29, 2025, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
09

Registry details

Key details

Study ID
NCT07147920
Lead sponsor
Perin Health Devices
Responsible party
Sponsor
First posted
Aug 29, 2025
Start date
May 21, 2025
Primary completion
Jul 15, 2025
Completion
Jul 15, 2025
Last update
Aug 29, 2025

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Aug 2025. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion