CompletedNCT07086742SMV1: PPGUpdated Sep 17, 2026

Validation of the Smart Mask V1 System and Its Measurements of SpO2 and Pulse Rate

An interventional study of Smart Mask V1 System in Sleep-disordered Breathing (SDB), Sleep Apnea and Sleep Quality, sponsored by Pathway Medtech, LLC.. Completed at 1 site in Belgium. Open to participants aged 18 Years to 65 Years, including healthy volunteers. Per ClinicalTrials.gov, last updated 2026-09-17.

Sponsored by Pathway Medtech, LLC. · Not applicable, Interventional, and Other

Phase
Not applicable
Study type
Interventional
Enrollment
53
Allocation
Non-randomized
Ages
18 Years to 65 Years
Sex
All
01

Study summary

The goal of this clinical study is to investigate a new medical device, the Smart Mask V1 System (herein 'Smart Mask'), and particularly its measurements of blood oxygen levels (SpO2) and pulse rate (PR) in healthy adults aged 18 to 65.

More specifically, the study is directed at answering the following questions:

  • Can the Smart Mask accurately measure SpO2 levels compared to established reference devices (reference oximeters, blood gas) throughout the range of SpO2 \~ 70 - 100%?
  • Can the Smart Mask accurately measure PR during the same conditions?
  • Does skin pigmentation impact the accuracy of Smart Mask's measurements of SpO2 and PR.

The following reference devices will be used in the study:

  • An FDA-cleared fingertip pulse oximeter (Nellcor Portable SpO2 Patient Monitoring System, PM10N)
  • A laboratory-grade CO-oximeter (Radiometer ABL90 FLEX) that analyzes oxygen saturation from blood samples.

Study participants will:

  • Wear the Smart Mask on their face and a fingertip pulse oximeter while lying down
  • Breathe air with gradually reduced oxygen levels inside a specialized hypoxia room while being closely.
  • In a second phase of the study, have a catheter inserted into a wrist artery for blood sampling to directly measure oxygen levels with the CO-oximeter.
Read the detailed description

This clinical study evaluates the Smart Mask V1 System, a medical device designed to measure blood oxygen saturation (SpO2) and pulse rate (PR) using reflective photoplethysmography (PPG) sensors integrated into a face mask intended for use with positive airway pressure (PAP) therapy. The study aims to validate the accuracy of the Smart Mask's optical measurements under controlled, progressive hypoxia conditions in healthy adult volunteers.

The investigation is divided into two sequential phases; Non-Invasive (NI) and Invasive (IN) Study. All study sessions are conducted at the Complementary Medical Centre (CMC) in Genk, Belgium, within a certified normobaric hypoxia room. This room is sealed and climate-controlled, equipped with oxygen and carbon dioxide sensors to maintain a stable and safe hypoxic environment.

Hypoxia is induced gradually to reduce any discomfort or potential risk to participants. Comprehensive safety protocols are in place to protect volunteers. These include medical screening and obtaining informed consent prior to enrollment, as well as real-time monitoring by trained medical staff throughout the exposure. Immediate exit protocols are activated if a participant's SpO2 falls below 73% or if symptoms of acute mountain sickness are observed. Emergency oxygen and resuscitation equipment are readily available on-site. Additionally, participants are reminded that they may withdraw from the study at any time without penalty.

02

Conditions studied

  • Sleep-disordered Breathing (SDB)
  • Sleep Apnea
  • Sleep Quality

Keywords

  • Pulse Oximeters
  • SpO2
  • Sleep apnea
  • postive air pressure
03

In context

Sleep Apnea Syndromes

2,162 studies on the registry are indexed under Sleep Apnea Syndromes; 290 are open to participants now.

This study's enrollment of 53 is close to the median of 53 across 1,386 interventional studies indexed under Sleep Apnea Syndromes.

Browse Sleep Apnea Syndromes studies →

Lead sponsor

Pathway Medtech, LLC. is the lead sponsor of 2 studies on the registry; none are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years to 65 Years
Sexes eligible
All
Accepts healthy volunteers
Yes

Inclusion criteria

  • Healthy subjects with ASA health score of I or II
  • Subjects aged between ≥ 18 and ≤ 65 years.
  • Subjects who have provided informed consent and are willing to comply with the study procedures.

Exclusion criteria

Exclusion Criteria:

  • Heavy smokers or individuals exposed to high levels of carbon monoxide resulting in elevated carboxyhemoglobin levels.
  • Individuals with conditions that result in elevated methemoglobin levels.
  • Individuals with hypoxia (SpO2 \< 95% at 21% O2).
  • Severe claustrophobia.
  • Subjects known with altitude sickness.
  • Subjects who are obese (BMI ≥ 35 kg/m2).
  • Subjects with a known history of moderate to severe heart, lung, kidney or liver disease.
  • Subjects diagnosed with moderate to severe asthma.
  • Subjects with a hemoglobinopathy or history of anemia, e.g. sickle cell anemia, thalassemia, that in the investigator's opinion, would make them unsuitable for study participation.
  • Subjects with any other serious systemic illness.
  • Any injury, deformity, or abnormality at the sensor sites that, in the investigator's opinion, would interfere with the correct functioning of the test or reference devices.
  • Subjects with a history of fainting or vasovagal response.
  • Subjects with a history of sensitivity or allergy to local anesthetics or disinfectants if included in the IN study.
  • Subjects diagnosed with Raynaud's disease.
  • Subjects with unacceptable collateral circulation based on examination by the investigator.
  • Subjects who are pregnant, lactating or trying to become pregnant.
  • Subjects unable or unwilling to provide informed consent or comply with study procedures.
  • Subjects with any other condition that, in the investigator's opinion, would make them unsuitable for the study.
  • Subjects who refuse to remove nail polish from their index finger.
  • Volunteers not willing to remove facial make-up.
05

Study design

Phase
Not applicable
Primary purpose
Other
Allocation
Non-randomized
Intervention model
Sequential assignment
Masking
None (open label)
Enrollment
53 participants (actual)

Study arms

  • Experimental
    Non-Invasive (NI)

    The Non-Invasive (NI) study is designed to calibrate the Smart Mask's algorithms using data collected from healthy volunteers exposed to controlled, progressive hypoxia. During the study, participants lie in a supine position in a normobaric hypoxia room while breathing room air with gradually reduced oxygen levels, with the fraction of inspired oxygen (FiO2) decreasing from approximately 21% to as low as 9.5%. The Smart Mask derives SpO2 values from reflected light signals detected by its integrated sensors. Simultaneously, a benchmark pulse oximeter (Nellcor PM10N), worn on the finger, provides reference SpO2 and pulse rate (PR) values for algorithm calibration. Data from reference device are used to refine the conversion of raw photoplethysmography (PPG) signals into SpO2 and PR readings. This phase serves as a feasibility study and is a prerequisite for advancing to subsequent phase. 12 healthy subjects will be enrolled in this study.

    Device: Smart Mask V1 System

  • Experimental
    Invasive (IN)

    If the Non-Invasive (NI) phase demonstrates feasibility, the Invasive (IN) study will follow to validate the Smart Mask's measurements against a gold-standard method. This phase involves arterial blood gas analysis to directly measure arterial oxygen saturation (SaO2) using CO-oximetry. Participants undergo the same progressive hypoxia protocol within a controlled hypoxia room. An indwelling radial artery catheter is inserted to allow periodic arterial blood sampling at five predefined oxygen plateaus. The SaO2 values obtained from the CO-oximeter reference device (ABL90 FLEX) are then compared to the SpO2 values calculated by the Smart Mask. 24 healthy volunteers will be enrolled in this phase.

    Device: Smart Mask V1 System

Interventions

  • DeviceSmart Mask V1 System

    PPG-based sensors integrated into a PAP-compatible face mask to measure SpO2 and PR.

    Also known as: Smart Mask, SMV1

06

What researchers measure

Primary outcomes

  1. SpO2 Accuracy (Arms)

    Root mean square error (Arms) of SpO2 measured by the Smart Mask compared to SpO2 and SaO2 reference devices

    Time frame: During each hypoxia session (90-140 minutes)

07

Study locations

1 site
  • Complementair Medisch Centrum (CMC) Europe
    Genk, 3600, Belgium
08

References and documents

Individual participant data

Plan to share: No — Data collected during the study will be specifically used to develop Smart Mask, a device intended to be commercialized by the study sponsor. Thus, there does not appear to be any value or reason to share IPD.

No publications or documents are linked to this record.

09

Updates

Tracking since Sep 25, 2026
No changes since tracking began. The registry record was last updated on Sep 17, 2026, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT07086742
Lead sponsor
Pathway Medtech, LLC.
Collaborators
Haku Technology
Responsible party
Sponsor
First posted
Jul 25, 2025
Start date
Jul 20, 2025
Primary completion
Nov 23, 2025
Completion
Nov 23, 2025
Last update
Sep 17, 2026

Study contacts

Raf De Jongh, M.D.
principal investigator · Complementair Medisch Centrum (CMC) Europe

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Not currently enrolling

This study is completed, as verified in Jun 2025. You cannot join it, but the record below documents what was studied.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion