CTICU Device Study CRUISE

An observational study in Open Heart Surgery, sponsored by Hernando Gomez. Enrolling by invitation at 1 site in United States. Open to participants aged 18 Years to 105 Years. Per ClinicalTrials.gov, last updated 2026-03-06.

Sponsored by Hernando Gomez · Observational

How we know

Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →

Study type
Observational
Model
Other
Time perspective
Other
Enrollment
50
Ages
18 Years to 105 Years
Sex
All
01

Study summary

The goal of this study evaluate three novel, Food and Drug Administration (FDA)-approved devices currently used to monitor blood pressure and the amount of blood that the heart pumps during each beat.

These devices are the ClearSight (Edwards Life Sciences), CareTaker (CareTaker Medical) and CNAP (CNS Systems). These devices measure blood pressure, how much blood the heart is pumping around the body each minute and on a heartbeat to heartbeat basis and other parameters that can tell the doctor if the patient will benefit from specific treatments. The purpose of this study is to determine the accuracy of these devices. This knowledge will help us understand how investigators can use such non-invasive devices to care for patients at less risk than the devices currently used for such purposes.

The main question the study aims to answer is:

  • What monitor is the most accurate in capturing non-invasive patient data?
Read the detailed description

Investigators will compare predictors of preload responsiveness, including pulse pressure variation (PPV) and stroke volume variation (SVV), and arterial dynamic elastance (Eadyn) in all patients and estimates of cardiac output (CO) in patients with an indwelling pulmonary artery catheter. in 50 intensive care unit, intubated patients during the patient's normal course of treatment using their usually present invasive monitoring devices for the first 8 hours following cardiac surgery.

The primary goal of the study will be to compare PPV, SVV and Eadyn measured by the non-invasive monitors ClearSight, CareTaker and CNAP versus a reference standard that uses invasive arterial pressure waveform data, which in this case will be the LiDCO monitor \& FloTrac monitor. Investigators will calculate accuracy, precision and concordance of each of the parameters measured during steady state conditions and after specific interventions including the administration of crystalloid fluid boluses and changes in vasoactive or inotropic medications. As a secondary objective, investigators will compare measures of CO by the same non-invasive monitors with the pulmonary artery catheter as a reference standard during the same conditions as above (i.e., steady state conditions, and before/after interventions). CO will therefore be only compared in patients that have a pulmonary artery catheter in place.

02

Conditions studied

  • Open Heart Surgery

Keywords

  • Heart, Surgery, ICU, Monitor, Blood Pressure, Measurement
  • non-invasive
  • responsiveness
  • cardiac
03

In context

Lead sponsor

Hernando Gomez is the lead sponsor of 3 studies on the registry; 2 are open to participants now.

Counted across the registry records on this site, refreshed daily.

04

Who can participate

Ages eligible
18 Years to 105 Years
Sexes eligible
All
Accepts healthy volunteers
No
Sampling method
Non-probability sample

Study population

All patients with estimates of cardiac output (CO) in patients with an indwelling pulmonary artery catheter. in 50 intensive care unit, intubated patients during the patient's normal course of treatment using their usually present invasive monitoring devices for the first 8 hours following cardiac surgery.

Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.

Inclusion criteria

  • Adult patients (18 years and older)
  • Male and female patients following cardiac surgery admitted to the CTICU with or without a pulmonary artery catheter

Exclusion criteria

Exclusion Criteria:

  • Subjects whose sternum remains open post-surgery and/or the presence of a LV ventricular assist device.
  • Presence of a any ventricular assist device, ECMO, Impella, durable LVAD, or central L/RVAD.
  • Spontaneous breathing activity upon coming out of the operating room (i.e., patients extubated in the operating room).
  • Presenting chronic atrial fibrillation, acute atrial fibrillation to medical therapy or frequent extra-systoles.
  • Children will be excluded from this study. The hospital unit where all subjects will be recruited is for adults' (ages 18 and up) only. Gender, racial and ethnic sub groups will not be explicitly excluded.
Terms in these criteria, in plain language(1)

Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.

Relapsed and refractory

Relapsed: the illness came back after treatment had worked. Refractory: it did not respond to treatment, or stopped responding while still on it.

05

Study design

Observational model
Other
Time perspective
Other
Enrollment
50 participants (estimated)
Target follow-up
1 Day
Patient registry
Yes

Interventions

  • DeviceCRUISE CTICU CareTaker Vitalstream Device

    The CareTaker Vitalstream device is a patented PDA technology and artificial intelligence algorithms derive continuous hemodynamic parameters by analyzing reflective pulse wave analyzes the pulse pressure waveform at 500 times per second, isolates the individual central aortic pulses that give rise to the observed pulse shape, reads central aortic measurements therefore eliminating the effects of hydrostatic pressure, and tracks vascular tone.

  • DeviceCRUISE CTICU CNAP Device

    The CNAP Monitor provides advanced hemodynamic information from the non-invasive reusable CNAP finger sensor. It supports your Goal Directed Therapy whenever an arterial line is not indicated. Patient-friendly, easy-to-use, cost-efficient. CNAP Monitor offers a unique set of parameters to support your daily routine - all non-invasive \& continuous allowing for perioperative fluid management even in low and moderate-risk surgeries and patients.

  • DeviceCRUISE CTICU Edwards Clearsight Device

    The Edwards ClearSight system provides continuous blood pressure and advanced hemodynamic parameters from a noninvasive finger cuff. Continuous data offered by the ClearSight system enables you to proactively optimize perfusion through hemodynamic management.

06

What researchers measure

Primary outcomes

  1. Stroke Volume Variation

    The Stroke Volume Variation (SSV) will be collected continuously for 8 hours using all 3 devices (ClearSight, CareTaker and CNAP). The metric of performance will be the precision, accuracy and concordance assessed using Bland Altman analysis and Pearson's moment analysis comparing SVV from each of the non-invasive monitors with the reference standard obtained from the analysis of the arterial pressure waveform. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping), before each change in vasopressors or inotropes (within 5 minutes of administration), after each change in vasopressors or inotropes (within 5 minutes of stopping), before changing ventilatory mode.

    Time frame: 8 hours

  2. Pulse Pressure Variation (PPV)

    The Pulse Pressure Variation will be collected continuously for 8 hours using all 3 devices (ClearSight, CareTaker and CNAP). The metric of performance will be the precision, accuracy and concordance assessed using Bland Altman analysis and Pearson's moment analysis comparing PPV from each of the non-invasive monitors with the reference standard obtained from the analysis of the arterial pressure waveform. This metric of performance will be reported. The most reliable device will be based on the metric of performance - that is, the most precise, accurate and concordant. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping), before each change in vasopressors.

    Time frame: 8 hours

  3. Arterial Dynamic Elastance (Eadyn)

    The Arterial Dynamic Elastance (Eadyn) will be collected continuously for 8 hours using all 3 devices (ClearSight, CareTaker and CNAP). The metric of performance will be the precision, accuracy and concordance assessed using Bland Altman analysis and Pearson's moment analysis comparing Eadyn from each of the non-invasive monitors with the reference standard obtained from the analysis of the arterial pressure waveform. This metric of performance will be reported. The most reliable device will be based on the metric of performance - that is, the most precise, accurate and concordant. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping).

    Time frame: 8 hours

Secondary outcomes

  1. Cardiac Output

    Compare measures of CO by the ClearSight, CareTaker and CNAP non-invasive monitors with the pulmonary artery catheter as a reference standard during the same conditions as above (i.e., steady state conditions, and before/after interventions). CO will therefore be only compared in patients that have a pulmonary artery catheter in place. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping), before each change in vasopressors or inotropes (within 5 minutes of administration), after each change in vasopressors or inotropes (within 5 minutes of stopping), before changing ventilatory mode to pressure support ventilation, after 30+/-5 minutes of changing ventilatory.

    Time frame: 8 hours

  2. Volume Responsiveness Collection

    The performance of ClearSight, CareTaker and CNAP in predicting volume responsiveness as compared to LiDCO, using a change of 10% or more in CO in response to fluid administration measured with a PAC as the definition of volume responsiveness. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping), before each change in vasopressors or inotropes (within 5 minutes of administration), after each change in vasopressors or inotropes (within 5 minutes of stopping), before changing ventilatory mode to pressure support ventilation, after 30+/-5 minutes of changing ventilatory mode to pressure support ventilation, before extubation (within 5minutes), after extubation .

    Time frame: 8 hours

07

Study locations

1 site
08

References and documents

Publications

  • Michard F, Chemla D, Richard C, Wysocki M, Pinsky MR, Lecarpentier Y, Teboul JL. Clinical use of respiratory changes in arterial pulse pressure to monitor the hemodynamic effects of PEEP. Am J Respir Crit Care Med. 1999 Mar;159(3):935-9. doi: 10.1164/ajrccm.159.3.9805077. PubMed 10051276 ↗
  • Ramsey SD, Saint S, Sullivan SD, Dey L, Kelley K, Bowdle A. Clinical and economic effects of pulmonary artery catheterization in nonemergent coronary artery bypass graft surgery. J Cardiothorac Vasc Anesth. 2000 Apr;14(2):113-8. doi: 10.1016/s1053-0770(00)90001-6. PubMed 10794325 ↗

Individual participant data

Plan to share: No — We will not share any data outside our study team. The DOD requires all vendors to sign NDAs and commit to the rules of the study. Any party that may have access to the data will be part of the study team and committed thru NDAs \& subcontracts.

09

Timeline

Tracking since Sep 25, 2026
ElapsedEstimatedEstimated, not yet reached

Started Jun 2025 (actual). Primary completion is estimated for Jan 2027, 19 months after the start.

How these figures are worked out, and which studies are left out of them.

Registered
Jan 13, 2025
First posted
Feb 14, 2025
StartActual
Jun 3, 2025
Primary completionEst.
Jan 31, 2027
CompletionEst.
Jan 31, 2027
Last update
Mar 6, 2026
What these dates mean
Start.
When the first participant enrolled; while estimated, when the study expects to open to recruitment.
Primary completion.
When the last participant was examined or treated to collect data for the study's main question, its primary outcome. A study that stopped early gives the date its data collection stopped.
Completion.
When the last participant's last visit took place, for every outcome, secondary ones and side effects included: the end of the study. Often later than primary completion, sometimes the same day.
Actual and Est.
Actual means the registry says the date has happened. Est. (estimated) is a date the sponsor plans or expects. It can still move, and the sponsor is expected to change it to actual once it has passed.
No changes since tracking began. The registry record was last updated on Mar 6, 2026, before this site started recording changes on Sep 25, 2026. Its history is on ClinicalTrials.gov ↗
10

Registry details

Key details

Study ID
NCT06828432
Lead sponsor
Hernando Gomez
Collaborators
United States Department of Defense
Responsible party
Hernando Gomez (Associate Professor, University of Pittsburgh) — Sponsor-investigator
First posted
Feb 14, 2025
Start date
Jun 3, 2025
Primary completion
Jan 31, 2027 (estimated)
Completion
Jan 31, 2027 (estimated)
Last update
Mar 6, 2026

Study contacts

Hernando Gomez, MD
principal investigator · University of Pittsburgh

Oversight

Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
View the source record on ClinicalTrials.gov ↗

Interested in this study?

Eligibility is decided by the study team. Share this record with your doctor or contact the team directly.

No contact was published for this record. The registry link below has the sponsor’s details.

Follow this study

Get an email when the registry record changes — status, dates, results — or when someone posts here.

Sign in to follow

Discussion

Questions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.

Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.

Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.

Start the discussion