An interventional study of Orotracheal intubation in Orotracheal Intubation, Videolaryngoscopes (VL) and 3D Printing, sponsored by Hospital General de México Dr. Eduardo Liceaga. Completed at 1 site in Mexico. Open to participants aged 18 Years and older, including healthy volunteers. Per ClinicalTrials.gov, last updated 2025-02-25.
Sponsored by Hospital General de México Dr. Eduardo Liceaga · Not applicable, Interventional, and Treatment
Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →
This study aims to compare the ORION video laryngoscope with the King Vision video laryngoscope for orotracheal intubation in patients without predictors of difficult airway scheduled for elective surgery under general anesthesia.
The video laryngoscope, compared to direct laryngoscopy, provides better visualization of the laryngeal and glottic structures, a higher success rate for intubation on the first attempt, and a lower incidence of complications such as mucosal injury, dental injury, and esophageal intubations. Despite the advantages offered by video laryngoscopy, its use is not widespread in developing countries due to high costs. For this reason, the ORION video laryngoscope was developed as a low-cost, reusable device that can undergo high-level disinfection. It is designed using additive manufacturing and meets the standards for biocompatibility, rigidity, and resistance. Additionally, it has undergone preclinical testing in simulation models where anesthesiologists have approved its use in real clinical scenarios. In this study, patients scheduled for elective surgery under general anesthesia without predictors of difficult airway will be included and randomized to be intubated with either the ORION video laryngoscope or the King Vision video laryngoscope. The study will compare the success of intubation on the first attempt, the time taken for intubation, the level of visualization of the glottic structures, and the associated complications between the two video laryngoscopes.
Hospital General de México Dr. Eduardo Liceaga is the lead sponsor of 14 studies on the registry; 3 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
The anaesthetists’ scale of health before surgery: I is healthy, II has a mild illness that is under control, III a severe one, IV a severe illness that is a constant threat to life. V and VI are used for the critically ill and for organ donors.
Weight in kilograms divided by height in metres squared. Example: 70 kg and 1.75 m gives 70 ÷ (1.75 × 1.75) ≈ 22.9.
What counts as overweight or obesity differs between countries (some use lower cut-offs), so go by the range the study states.
Chronic obstructive pulmonary disease: long-term lung disease that makes it hard to breathe out, including emphysema and chronic bronchitis.
Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.
You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.
ASA physical status
The anaesthetists’ scale of health before surgery: I is healthy, II has a mild illness that is under control, III a severe one, IV a severe illness that is a constant threat to life. V and VI are used for the critically ill and for organ donors.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
BMI (body mass index)
Weight in kilograms divided by height in metres squared. Example: 70 kg and 1.75 m gives 70 ÷ (1.75 × 1.75) ≈ 22.9.
What counts as overweight or obesity differs between countries (some use lower cut-offs), so go by the range the study states.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
COPD
Chronic obstructive pulmonary disease: long-term lung disease that makes it hard to breathe out, including emphysema and chronic bronchitis.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Informed consent
Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.
You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
The King Vision video laryngoscope used in this study has a 45° angulation, with a disposable No. 3 blade and no channel for the orotracheal tube.
Device: Orotracheal intubation
The ORION video laryngoscope is designed by additive manufacturing, has the same dimensions as a Macintosh No. 3 blade, with 45° angulation, is biocompatible and reusable when subjected to high-level disinfection with 2% Glutaraldehyde for 40 minutes. It has an endoscopic camera fully covered in the body of the videolaryngoscope of 8 mm diameter with 2 megapixels with a resolution of 1280x720, has a viewing angle of 70°, has 6 LEDS with adjustable intensity and the camera has an anti-fogging system and can be connected to any mobile device with Android operating system through a free application. (AN98 Version 2.2.0). It complies with ISO 7376:2020 standards for orotracheal intubation laryngoscopes because its blade withstands a stiffness of more than 60 Newtons and a resistance of more than 150 Newtons.
Device: Orotracheal intubation
Patients who meet the inclusion criteria and sign the informed consent form will be intubated with the ORION or King Vision video laryngoscope according to randomisation.
Successful first attempt
Measurement of orotracheal intubations obtained in the first attempt of videolaryngoscopy with the two videolaryngoscopes, successful orotracheal intubation will be corroborated by capnography. Measured in percentage (%).
Time frame: After the endotracheal intubation completed
Intubation time
Time elapsed from the time the video laryngoscope blade crosses the dental arch until the orotracheal tube is inserted into the trachea, this time will be measured for each video laryngoscope. Measured in seconds (s).
Time frame: After the endotracheal intubation completed
Laryngeal visualisation measured with the Cormack Lehane scale
Measurement of visualization of glottic structures using the Cormack Lehane scale in grade I, II, II and IV obtained with both video laryngoscopes. Measured in percentage (%).
Time frame: After the endotracheal intubation completed
Laryngeal visualisation measured with the POGO scale
Measurement of the visualization of the glottic structures using the POGO (Percentage of Glottic Opening) scale from 0% to 100% obtained with both video laryngoscopes. Measured in percentage (%).
Time frame: After the endotracheal intubation completed
Esophageal intubation
Measurement of the number of esophageal intubations obtained with both video laryngoscopes. Measured in percentage (%)
Time frame: After the endotracheal intubation completed
Complications
Measurement of complications (mucosal, lip, oral, pharyngeal and laryngeal injury, total or partial fracture of a tooth and Hypoxemia) that occur with the use of the ORION and King Vision Videolaryngoscopes. Measured in percent (%)
Time frame: After the endotracheal intubation completed
Number of video laryngoscopy attempts
Measurement of the number of video laryngoscopy attempts to successful intubation with both video laryngoscopes. Measurement in percentage (%)
Time frame: After the endotracheal intubation completed
Intubation failure
Three videolaryngoscopies with duration of 120 seconds each without achieving orotracheal intubation. The number of intubation failures of both videolaryngoscopes will be measured in percentage (%).
Time frame: After the endotracheal intubation completed
Plan to share: No
No publications or documents are linked to this record.
Ran from Oct 2024 to primary completion in Oct 2024, 0 months.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in Feb 2025. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.