An interventional study of Measurement of respiratory rate with DUT and comparison with reference in Respiratory Rate, sponsored by Biosency. Completed at 1 site in United States. Open to participants aged 18 Years and older, including healthy volunteers. Per ClinicalTrials.gov, last updated 2019-08-22.
Sponsored by Biosency · Not applicable, Interventional, and Diagnostic
Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →
The purpose of this study is to conduct a Respiratory Rate accuracy validation comparing the BORA BAND™ Wristband Monitoring System to the Reference. Respiratory Rate will be compared to an FDA cleared End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda). The testing will be conducted in two phases. The first phase will be an initial data collection to assist in the algorithm development of respiration rate. Upon a successful Phase one performance, the second phase will be the Respiratory Rate Validation conducted at a different period in time.
For phase 1, there will be 5 to 6 volunteer test subjects. For the final validation phase, there will be a minimum of 20 volunteer test subjects. All subjects will be 18 or older and selected to represent a range of body types including small, average, muscular, and large with a range of BMI's.
Each subject will be connected to a commonly used, FDA cleared ECG Impedance monitor (to be used for informational purposes) and an End Tidal Carbon Dioxide monitor (GE Datex-Ohmeda) and BiOSENCY BORA BAND™ wristband monitoring system. The end tidal CO2 monitor will be attached to the patient to determine performance of respiratory rate metrics (Reference). Each subject will be instrumented with a mouthpiece that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate (Reference) and tidal volume.
A range of stable respiratory rates will be elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40, 45 and 50 breaths per minute; with some natural variation from these exact numbers. Paced breathing app will be used. Not all subjects will be able to breathe at a constant level for every rate.
The respiratory rate will be measured simultaneously with EtCO2 (Reference) and the Devices Under Test.
For the data analysis, the EtCO2 monitor will be used to assess the stability of the data. If the respiratory rate varies by more than +3 bpm during a one-minute period of the target respiration rates, then the data for that period will be considered unstable and removed from analysis.
To 'Pass' this test the BiOSENCY BORA BAND™ wristband (Device Under Test) must demonstrate a Respiratory Rate of ≤3 when compared to the Reference End tidal carbon dioxide monitor.
Biosency is the lead sponsor of 5 studies on the registry; 2 are open to participants now.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
Subjects with known respiratory conditions such as:
Subjects with self-reported heart or cardiovascular conditions such as:
Self-reported health conditions as identified in the Health Assessment Form
Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.
You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.
Chronic obstructive pulmonary disease: long-term lung disease that makes it hard to breathe out, including emphysema and chronic bronchitis.
A “mini-stroke”: temporary stroke-like symptoms caused by reduced blood flow to part of the brain, spinal cord or eye, without signs of a new stroke on a scan.
Stroke symptoms need emergency care at once, even if they go away.
Informed consent
Before you join, the study team explains what taking part involves — its purpose, what happens, the risks and possible benefits — and answers your questions, so you can decide freely. Consent is usually recorded on a form (the informed consent form, ICF) signed by you or, where the law allows, by someone legally authorized to decide for you.
You can keep asking questions during the study, you are told of new information that matters, and you can choose to leave at any time.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
COPD
Chronic obstructive pulmonary disease: long-term lung disease that makes it hard to breathe out, including emphysema and chronic bronchitis.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
TIA (transient ischaemic attack)
A “mini-stroke”: temporary stroke-like symptoms caused by reduced blood flow to part of the brain, spinal cord or eye, without signs of a new stroke on a scan.
Stroke symptoms need emergency care at once, even if they go away.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Device: Measurement of respiratory rate with DUT and comparison with reference
The subject will follow the paced breathing app, once a stable Respiration Rate at the natural rate (without the breathing app) is captured. The breathing pattern on the app will be set to begin near the normal Respiration Rate of the subject and continue to go down by five intervals until reaching the Respiration Rate of 5 bpm. After that, we will pick back up at the rate above the subject's normal Respiration Rate and continue upwards until reaching 50 bpm or as high as the subject can tolerate up to 50 bpm. The respiratory rate will be monitored simultaneously with an EtCo2 (Reference) and the BORA BAND™ wristband monitoring system, Device Under Test. These data will be continuously recorded electronically. Additional study notes that describe conditions of the test as well as deviations, device issues and any adverse events will be recorded in written documentation.
Accuracy of BORA Band respiratory rate measurements versus reference end-tidal CO2
Accuracy of the BORA Band respiratory rate measurement as compared to reference end-tidal CO2. Results shall include: * Respiratory rate root mean square accuracy of BORA Band * Respiratory rate bias of BORA Band
Time frame: 5 days
Plan to share: No
No publications or documents are linked to this record.
Ran from Jun 2019 to primary completion in Jul 2019, 1 month.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in Aug 2019. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Biosency