An interventional study of OneTouch Select® Plus Glucose Test Strips in Blood Glucose Measurement, sponsored by Johnson & Johnson Medical, China. Withdrawn. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2019-11-15.
Sponsored by Johnson & Johnson Medical, China · Not applicable, Interventional, and Diagnostic
This study is to support the regulatory application of OneTouch Select® Plus test strips in China. The blood glucose test strip is regulated as an In Vitro Diagnostic (IVD) product and China Food and Drug Administration (CFDA) regulation of IVDs should be complied with. The result of this study will be used to support regulatory filing in China.
Investigational Product:
OneTouch Select® Plus test strips (test with OneTouch Select® Plus glucose meter)
Objectives:
Comparator(control) Product:
YSI 2300 glucose analyser
Acceptance Criteria:
User Performance of the OneTouch Select® Plus Blood Glucose Monitoring system(described as OneTouch Select Plus below) is met if the lay user and HCP results obtained with OneTouch Select Plus meet the requirement stated in section 7.4.1 of ISO () 15197:2003(E). Further to meet the requirements of ISO 15197:2013(E) , the lay user results obtained with OneTouch Select Plus should meet the requirements stated in section 8.2 of ISO 15197:2013(E). The effectiveness of instructions for use is met , if for each item on the Instructions for Use Questionnaire, the 90% lower confidence limit (90% LCL) for the correct response rate is equal to or greater than 70%.
Study Design and subject population:
This study is an open, non-randomized clinical evaluation and will be conducted at a minimum of 2 sites in China. At least 240 evaluable diabetic subjects who meet the enrolment criteria will participate in this evaluation.
Study procedures:
Johnson & Johnson Medical, China is the lead sponsor of 9 studies on the registry; none are open to participants now.
Counted across the registry records on this site, refreshed daily.
Exclusion Criteria:
The accuracy of glucose results obtained from OneTouch Select Plus BGMS
User Performance of the OneTouch Select® Plus Blood Glucose Monitoring system(described as OneTouch Select Plus below) is met if the lay user and HCP results obtained with OneTouch Select Plus meet the requirement stated in section 7.4.1 of ISO 15197:2003(E).Further to meet the requirements of ISO 15197:2013(E) , the lay user results obtained with OneTouch Select Plus should meet the requirements stated in section 8.2 of ISO 15197:2013(E).
Time frame: instantly after Glucose testing
The correct response rate of Instructions for Use Questionnaire
The effectiveness of instructions for use is met , if for each item on the Instructions for Use Questionnaire, the 90% lower confidence limit (90% LCL) for the correct response rate is equal to or greater than70%
Time frame: instantly after Glucose testing
No study locations are listed for this record.
This study is withdrawn, as verified in Nov 2019. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Johnson & Johnson Medical, China