An interventional study of Zip Surgical Skin Closure Device and Standard Suture Closure in Wound and Cardiac Arrhythmia, sponsored by University of California, San Diego. Completed at 1 site in United States. Open to participants aged 18 Years and older. Per ClinicalTrials.gov, last updated 2016-05-30.
Sponsored by University of California, San Diego · Not applicable and Interventional
Help build ClinicalTrials.gg — short surveys, remote interviews or testing, in your language. Take part →
ZipLine Medical, Inc. has developed a novel, non-invasive skin closure device called "Zip Surgical Skin Closure" to replace sutures, staples and glue for closure of the skin layer for surgical incisions or laceration repair. The study will be comparing the Zip Surgical Skin Closure to standard sutures in patients undergoing cardiovascular implantable electronic device (CIED) procedures on outcomes including closure time, cosmetic appearance of resulting scar, and overall cosmetic appearance of scar. The study will be following participants for 3 months following CIED procedure.
ZipLine Medical, Inc. has developed a novel, non-invasive skin closure device called "Zip Surgical Skin Closure" to replace sutures, staples and glue for closure of the skin layer for surgical incisions or laceration repair. The device is designed to provide closure speed superior to sutures, while resulting in a suture-like cosmetic outcome.The ZIPS study is a randomized, clinical trial comparing outcomes of procedures that use the Zip Surgical Skin Closure device with the outcomes of procedures that use stitches. The Zip Surgical Skin Closure is not an investigational device and is currently used in hospital across the United States. The Zip Surgical Skin Closure is a non-invasive (in other words, does not pierce the skin or enter into the wound) device to hold skin closed while healing can occur. It adheres to the skin close to the incision and uses adjustable straps to hold the incision closed. It is typically worn on the skin for 7-14 days and is removed by peeling from the skin. Patients will be randomized to either Zip Surgical Skin Closure or sutures, in addition to dissolvable sutures for the deeper layers, for surgical procedures for cardiovascular implantable electronic devices (CIED) i.e. pacemakers or defibrillators. Both of these closure techniques are considered standard of care at this facility and are not themselves experimental. However, the goal of this experimental study is to see if there are any significant differences between the two closure methods, in a controlled study environment.
Since the Zip Surgical Skin Closure device is relatively new to the market, there is a limited amount of information available describing the experience of patients and doctors using the device. In this study, the investigators will be looking to compare the Zip Surgical Skin Closure and stitches methods for the time necessary for the surgeon to close the incision, cosmetic appearance of the resulting scar, and satisfaction of you and your surgeon. The investigators will also be monitoring both study groups for occurrence of any adverse (unfavorable) events. Patients will be followed for a total of 3 months for this study. All study visits align with standard of care follow up for post-implantation of CIED. Non-identifiable pictures will be obtained of the scar during follow up and both patients and surgeons will complete questionnaires indicating their experience and satisfaction with the either skin closure method.
2,225 studies on the registry are indexed under Wounds and Injuries; 446 are open to participants now.
This study's enrollment of 40 is below the median of 70 across 1,428 interventional studies indexed under Wounds and Injuries.
Browse Wounds and Injuries studies →University of California, San Diego is the lead sponsor of 981 studies on the registry; 196 are open to participants now.
Of its 114 completed or terminated interventional studies of FDA-regulated products, 71 (62%) have results posted.
Counted across the registry records on this site, refreshed daily.
Go through the criteria one by one — yes, no or not sure — and take the list to your doctor or the study team. Kept in this browser only.
Exclusion Criteria:
Underlined terms in the criteria open these too. Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
The study doctor’s estimate of how long someone is likely to live. Criteria use it so that people can stay in the study long enough for it to answer its question.
An allergy, or another reaction of the immune system, to a substance. “Hypersensitivity to the study drug” means a known allergic reaction to it — not every side effect a medicine can have.
Life expectancy
The study doctor’s estimate of how long someone is likely to live. Criteria use it so that people can stay in the study long enough for it to answer its question.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
Hypersensitivity
An allergy, or another reaction of the immune system, to a substance. “Hypersensitivity to the study drug” means a known allergic reaction to it — not every side effect a medicine can have.
Explained by clinicaltrials.gg, not part of the registry record. The study team decides who can take part.
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Device: Zip Surgical Skin Closure Device
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Device: Standard Suture Closure
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Also known as: Wound closure device
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Overall Closure Time
Duration of time starting when suture needle (control) or Zip device touches the skin until final suture knot is cut or Zip device application is complete (e.g., top liner is removed.)
Time frame: 2 weeks
Wound Healing as Determined by the CVAS (Cosmetic Visual Analogue Scale)
Based on photographs taken of scars taken at 3 months following CIED procedure. The CVAS scale is measured from 0mm (representing best scar) to 100 mm (representing worst scar).
Time frame: 3 Months
Surgeon Wound Evaluation Scale (WES)
At 2 weeks post-procedure, the Wound Evaluation Scale was measured. The scale is based from 0 (representing normal skin) to 100 (representing a poor scar).
Time frame: 2 weeks
Surgeon Evaluation Based on the Wound Evaluation Scale (WES)
The surgeon will complete an assessment of the following: 1. Closure Method Satisfaction 2. Scar Satisfaction 3. Wound Healing as judged by Wound Evaluation Scale Scale rated on 1 (least favorable) to 6 (most favorable) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Time frame: 3 Months
Surgeon Satisfaction With Scar
At 3 months post-procedure, the surgeon evaluated their satisfaction on a scale from 1 to 5, 5 being the least favorable. Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Time frame: 3 months
Patient Comfort
At 2 weeks post-procedure, patient comfort was evaluated with a scale 1 (most favorable) to 5 (least favorable).
Time frame: 2 weeks
Patient Satisfaction With Scar
Patient will complete questionnaires that includes assessments of the following: 1. Pain 2. Closure Method Comfort 3. Closure Method Satisfaction 4. Scar Satisfaction Scar Satisfaction was measured on a scale from 1 (most satisfied) to 5 (least satisfied) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Time frame: 3 months
Patient Rating of Scar
At 3 months post-procedure, patients were asked to rate their scar. Scale was measured from 0 (best scar) - 10 (worst scar) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Time frame: 3 months
Patient Incision Pain
At the discharge visit, patients evaluated their level of pain at the incision site based on a scale 1 (least pain) through 10 (worst pain).
Time frame: 1 day (discharge)
Patient Incision Pain
At 2 week post-procedure, incision pain was evaluated on a scale 0 to 10 (worst pain).
Time frame: 2 weeks
Patient Incision Pain
Incision pain experienced by the patient was evaluated at 3 months post-procedure. This was evaluated on a scale of 0 (least pain) to 10 (worst pain).
Time frame: 3 months
| Milestone | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Started | 21 | 19 |
| Completed | 20 | 18 |
| Not completed | 1 | 1 |
| Withdrew: Death | 0 | 1 |
| Withdrew: Lost to follow-up | 1 | 0 |
Duration of time starting when suture needle (control) or Zip device touches the skin until final suture knot is cut or Zip device application is complete (e.g., top liner is removed.)
| seconds | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Overall Closure Time | 78 ± 30.36 | 216 ± 93.71 |
Based on photographs taken of scars taken at 3 months following CIED procedure. The CVAS scale is measured from 0mm (representing best scar) to 100 mm (representing worst scar).
| mm | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Wound Healing as Determined by the CVAS (Cosmetic Visual Analogue Scale) | 35.8 ± 40.6 | 40.6 ± 21.7 |
At 2 weeks post-procedure, the Wound Evaluation Scale was measured. The scale is based from 0 (representing normal skin) to 100 (representing a poor scar).
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Surgeon Wound Evaluation Scale (WES) | 5.82 ± 0.50 | 5.78 ± 0.55 |
The surgeon will complete an assessment of the following: 1. Closure Method Satisfaction 2. Scar Satisfaction 3. Wound Healing as judged by Wound Evaluation Scale Scale rated on 1 (least favorable) to 6 (most favorable) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Surgeon Evaluation Based on the Wound Evaluation Scale (WES) | 6 ± 0 | 6 ± 0 |
At 3 months post-procedure, the surgeon evaluated their satisfaction on a scale from 1 to 5, 5 being the least favorable. Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Surgeon Satisfaction With Scar | 1 ± 0 | 1.06 ± 0.24 |
At 2 weeks post-procedure, patient comfort was evaluated with a scale 1 (most favorable) to 5 (least favorable).
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Patient Comfort | 1.77 ± 1.15 | 1.94 ± 1.16 |
Patient will complete questionnaires that includes assessments of the following: 1. Pain 2. Closure Method Comfort 3. Closure Method Satisfaction 4. Scar Satisfaction Scar Satisfaction was measured on a scale from 1 (most satisfied) to 5 (least satisfied) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Patient Satisfaction With Scar | 1.32 ± 0.48 | 1.22 ± 0.43 |
At 3 months post-procedure, patients were asked to rate their scar. Scale was measured from 0 (best scar) - 10 (worst scar) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Patient Rating of Scar | 0.87 ± 0.76 | 1.04 ± 1.39 |
At the discharge visit, patients evaluated their level of pain at the incision site based on a scale 1 (least pain) through 10 (worst pain).
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Patient Incision Pain | 1.77 ± 2.48 | 2.15 ± 1.83 |
At 2 week post-procedure, incision pain was evaluated on a scale 0 to 10 (worst pain).
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Patient Incision Pain | 0.64 ± 0.92 | 0.71 ± 1.41 |
Incision pain experienced by the patient was evaluated at 3 months post-procedure. This was evaluated on a scale of 0 (least pain) to 10 (worst pain).
| units on a scale | Zip Surgical Skin Closure Device | Standard Suture Closure |
|---|---|---|
| Patient Incision Pain | 0.27 ± 0.41 | 0.27 ± 0.73 |
Non-serious events are listed at a 0% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| Zip Surgical Skin Closure Device | — | 0/21 (0%) | 0/21 (0%) |
| Standard Suture Closure | — | 0/19 (0%) | 0/19 (0%) |
| Age, Continuous(years) | Zip Surgical Skin Closure Device | Standard Suture Closure | Total |
|---|---|---|---|
| Mean | 71 (44 to 99) | 70 (41 to 100) | 70 (41 to 100) |
| Sex: Female, Male(Participants) | Zip Surgical Skin Closure Device | Standard Suture Closure | Total |
|---|---|---|---|
| Female | 8 | 7 | 15 |
| Male | 13 | 12 | 25 |
| Race (NIH/OMB)(Participants) | Zip Surgical Skin Closure Device | Standard Suture Closure | Total |
|---|---|---|---|
| American Indian or Alaska Native | 0 | 0 | 0 |
| Asian | 1 | 1 | 2 |
| Native Hawaiian or Other Pacific Islander | 0 | 0 | 0 |
| Black or African American | 1 | 1 | 2 |
| White | 19 | 17 | 36 |
| More than one race | 0 | 0 | 0 |
| Unknown or Not Reported | 0 | 0 | 0 |
| Region of Enrollment(participants) | Zip Surgical Skin Closure Device | Standard Suture Closure | Total |
|---|---|---|---|
| United States | 21 | 19 | 40 |
| Fitzpatrick Skin Scale(participants) | Zip Surgical Skin Closure Device | Standard Suture Closure | Total |
|---|---|---|---|
| 1 - Always burns, never tans | 2 | 0 | 2 |
| 2 - Usually burns, tans with difficulty | 1 | 1 | 2 |
| 3 - Somtimes mild burn, gradually tans | 7 | 6 | 13 |
| 4 - Rarely burns, tans with ease | 6 | 7 | 13 |
| 5 - Very rarely burns, tans very easily | 4 | 3 | 7 |
| 6 - Never burns, tans very easily | 1 | 2 | 3 |
The record gives a start of Aug 2014, a primary completion of Jun 2015 and a completion of Jun 2015. It does not say whether these dates are actual or estimated.
Results were first posted in May 2016, 11 months after primary completion.
How these figures are worked out, and which studies are left out of them.
This study is completed, as verified in May 2016. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
University of California, San Diego →