An interventional study of Enlite sensors in Insulin Requiring Diabetes, sponsored by Medtronic Diabetes. Completed at 11 sites in United States. Open to participants aged 2 Years to 18 Years. Per ClinicalTrials.gov, last updated 2018-09-27.
Sponsored by Medtronic Diabetes · Not applicable, Interventional, and Treatment
The purpose of this study is to evaluate the Enlite Sensor™ and Enlite 3 Sensor performance in pediatric patients with insulin requiring diabetes at the beginning, middle and end of sensor wear.
This study is a prospective, single arm, multi-center, in-clinic study.. There will be two consecutive phases of the study:
A total of up to 260 subjects (2-18 years old) will be enrolled at up to 15 investigational centers
Subjects will come for clinic visit for YSI reference glucose testing and for Enlite Sensor™ and Enlite 3 Sensor™ testing. SMBG (Self-Monitoring of Blood Glucose) reference measurement is for subjects 2-6 years of age and cannot tolerate YSI.
Study has a run-in period and a study period with a total of 10 visits
Medtronic Diabetes is the lead sponsor of 61 studies on the registry; none are open to participants now.
Of its 22 completed or terminated interventional studies of FDA-regulated products, 21 (95%) have results posted.
Counted across the registry records on this site, refreshed daily.
Exclusion Criteria:
Each subject will first wear enlite sensors and will come for clinic visit for YSI (Yellow Spring Instruments). Then each subject will wear enlite 3 sensors and will come for clinic visit for YSI (Yellow Spring Instruments) or SMBG (Self-Monitoring of Blood Glucose) testing. SMBG testings are for subjects 2-6 years of age and cannot tolerate YSI. In addition. In addition, subjects aged 2-6 years old will not participate in Enlite phase. They only participate in Enlite 3 phase.
Device: Enlite sensors
Subject wears enlite sensor first and followed by enlite 3 sensor.
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump
Enlite Sensor accuracy relative to the YSI (Yellow Spring Instruments) reference. Each subject wore two 530G pumps. The results for the lower serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the primary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Time frame: 6 days
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System
Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG (Self-Monitoring of Blood Glucose) reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 days
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the GST3C Transmitter
Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 days
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump
Enlite Sensor accuracy relative to the YSI reference. Each subject wore two 530G pumps. The results for the higher serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the secondary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Time frame: 6 Days
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump, Using One Additional Calibration During FST
Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Time frame: 6 Days
Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump, Using One Additional Calibration During FST
Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
Time frame: 6 Days
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System, Using One Additional Calibration During FST
Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 Days
Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With GST3C Transmitter, Using One Additional Calibration During FST
Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
Time frame: 7 Days
| Milestone | 2-18 Years Old Wearing Sensors |
|---|---|
| Started | 161 |
| Completed | 141 |
| Not completed | 20 |
| Withdrew: Withdrawal by subject | 18 |
| Withdrew: Adverse event | 1 |
| Withdrew: Inclusion/exclusion not met | 1 |
| Milestone | 2-18 Years Old Wearing Sensors |
|---|---|
| Started | 158 |
| Completed | 146 |
| Not completed | 12 |
| Withdrew: Lost to follow-up | 1 |
| Withdrew: Physician decision | 2 |
| Withdrew: Physician decision | 8 |
| Withdrew: Due to multiple sensor issues | 1 |
Enlite Sensor accuracy relative to the YSI (Yellow Spring Instruments) reference. Each subject wore two 530G pumps. The results for the lower serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the primary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
| percentage of difference | Enlite Sensor Accuracy MARD, Lower Serial Number |
|---|---|
| Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump | 15.02 ± 13.37 |
Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG (Self-Monitoring of Blood Glucose) reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
| percentage of difference | Enlite 3 Sensor Accuracy MARD, Guardian Mobile |
|---|---|
| Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System | 10.88 ± 10.65 |
Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
| percentage of difference | Enlite 3 Sensor Accuracy MARD,GST 3C |
|---|---|
| Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the GST3C Transmitter | 11.08 ± 10.58 |
Enlite Sensor accuracy relative to the YSI reference. Each subject wore two 530G pumps. The results for the higher serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the secondary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
| percentage of difference | Enlite Sensor Accuracy MARD, Higher Serial Number |
|---|---|
| Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump | 17.50 ± 19.70 |
Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
| percentage of difference | Enlite Sensor Accuracy MARD, Lower Serial Number |
|---|---|
| Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump, Using One Additional Calibration During FST | 14.37 ± 12.99 |
Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.
| percentage of difference | Enlite Sensor Accuracy MARD, Higher Serial Number |
|---|---|
| Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump, Using One Additional Calibration During FST | 16.94 ± 18.74 |
Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
| percentage of difference | Enlite 3 MARD, Guardian Mobile, One Additional Fingerstick |
|---|---|
| Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System, Using One Additional Calibration During FST | 10.05 ± 9.26 |
Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) \* 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.
| percentage of difference | Enlite 3 MARD,GST 3C, One Additional Fingerstick |
|---|---|
| Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With GST3C Transmitter, Using One Additional Calibration During FST | 10.85 ± 10.35 |
Non-serious events are listed at a 0% frequency threshold.
| Group | Deaths | Serious | Other |
|---|---|---|---|
| 2-18 Year Olds Wearing Enlite Sensors | 0/161 (0%) | 2/161 (1.2%) | 25/161 (15.5%) |
| 2-18 Year Olds Wearing Enlite3 Sensors | 0/158 (0%) | 1/158 (0.6%) | 23/158 (14.6%) |
| Event | 2-18 Year Olds Wearing Enlite Sensors | 2-18 Year Olds Wearing Enlite3 Sensors |
|---|---|---|
| Influenzavirus BInvestigations | 0/161 | 1/158 |
| Diabetic KetoacidosisMetabolism and nutrition disorders | 1/161 | 0/158 |
| Pilonidal AbscessInfections and infestations | 1/161 | 0/158 |
| Event | 2-18 Year Olds Wearing Enlite Sensors | 2-18 Year Olds Wearing Enlite3 Sensors |
|---|---|---|
| Upper Resiratory InfectionInfections and infestations | 3/161 | 4/158 |
| severe hyperglycemiaMetabolism and nutrition disorders | 2/161 | 3/158 |
| coughRespiratory, thoracic and mediastinal disorders | 1/161 | 2/158 |
| GastroenteritisInfections and infestations | 0/161 | 2/158 |
| common coldInfections and infestations | 2/161 | 0/158 |
| HeadacheNervous system disorders | 1/161 | 1/158 |
| Nasal CongestionRespiratory, thoracic and mediastinal disorders | 1/161 | 1/158 |
| Sinus InfectionInfections and infestations | 1/161 | 1/158 |
| Abrasion on right arm and right legInjury, poisoning and procedural complications | 0/161 | 1/158 |
| Cold soreInfections and infestations | 0/161 | 1/158 |
a total of 186 subjects participated the study
| Age, Continuous(years) | 2-18 Year Olds Wearing Enlite Sensor |
|---|---|
| Mean | 13.2 ± 3.79 |
| Sex: Female, Male(Participants) | 2-18 Year Olds Wearing Enlite Sensor |
|---|---|
| Female | 83 |
| Male | 103 |
| Region of Enrollment(participants) | 2-18 Year Olds Wearing Enlite Sensor |
|---|---|
| United States | 186 |
This study is completed, as verified in Aug 2018. You cannot join it, but the record below documents what was studied.
Get an email when the registry record changes — status, dates, results — or when someone posts here.
Sign in to followQuestions and observations about this study, from anyone following it. Not medical advice, and not a channel to the study team — their contact details are on the registry record.
Sign in to join the discussion. Reading takes no account; posting does. You choose a display name, and a pseudonym is the default.
Nothing here yet. If you are running this trial, taking part in it, or weighing whether to, this is the place to say so.
Medtronic Diabetes